The Pantheris System is intended to remove plaque (atherectomy) from partially occluded vessels in the peripheral vasculature with a reference diameter of 3.0 mm to 7.0 mm, using OCT-assisted orientation and imaging. The system is as an adjunct to fluoroscopy by providing images of vessel lumen, wall structures and vessel morphologies. The Pantheris System is NOT intended for use in the iliac, coronary, cerebral, renal or carotid vasculature.
Device Story
Pantheris System is an OCT-guided directional atherectomy device for peripheral vasculature. System components: Pantheris catheter (7Fr/8Fr), Lightbox Sled, and Lightbox HS Imaging Console. Catheter features optical fiber for real-time OCT imaging and a cutter for plaque excision. Physician advances catheter through sheath to lesion site; uses OCT images to identify vessel lumen, wall structures, and morphologies; performs directional atherectomy; excised tissue packs into catheter nosecone. Lightbox console processes optical signals via PC-based system and displays images on monitors. Device used in cardiac catheter labs or hospitals. Software patch attenuates laser when Sled drive is not rotating. Provides real-time visualization to assist physician in plaque removal, potentially improving procedural outcomes.
Clinical Evidence
No clinical testing was performed for the modifications. Substantial equivalence is supported by bench testing, including mechanical testing (sheath insertion, balloon burst, torque, tensile strength, life-cycle, simulated use, heat generation, rotational speed, plaque removal efficiency) and software verification/validation. Biocompatibility testing was performed on modified accessories.
Technological Characteristics
Peripheral atherectomy catheter (7Fr/8Fr, 110cm length, 0.014" guidewire compatible). Imaging: OCT (near-infrared light, <30mW output, Class 1M laser). Materials: Biocompatible, stainless steel accessories. Connectivity: PC-based Lightbox console. Electrical: Class I, Type CF, defibrillation proof (IEC 60601-1). EMC: IEC 60601-1-2. Laser safety: 21 CFR 1040, IEC 60825. Software: Moderate level of concern. Sterilization: Sterile, single-use.
Indications for Use
Indicated for patients with partially occluded peripheral vasculature vessels (3.0-7.0 mm diameter) requiring plaque removal (atherectomy). Contraindicated for use in iliac, coronary, cerebral, renal, or carotid vasculature.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
October 30, 2017
Avinger, Inc. Thomas Lawson Director, Regulatory and Clinical Affairs 400 Chesapeake Drive Redwood City, California 94063
#### Re: K172236
Trade/Device Name: Pantheris System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW, NQQ Dated: September 29, 2017 Received: October 2, 2017
Dear Mr. Lawson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Kenneth J. Cavanaugh - S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172236
Device Name Pantheris System
Indications for Use (Describe)
The Pantheris System is intended to remove plaque (atherectomy) from partially occluded vessels in the peripheral vasculature with a reference diameter of 3.0 mm, using OCT-assisted orientation and imaging. The system is an adjunct to fluoroscopy by providing images of vessel lumen, wall structures, and vessel morphologies. The Pantheris System is NOT intended for use in the iliac, coronary, cerebral, renal or carotid vasculature.
Type of Use (Select one or both, as applicable)
| <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------|
| <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### General Information
| Submitter | Avinger, Inc. |
|-------------------------|-------------------------------------------------------------------------------|
| Address | 400 Chesapeake Drive<br>Redwood City, CA 94063 |
| FDA Registration Number | 3007498664 |
| Correspondence Person | Thomas Lawson, PhD<br>Director, Clinical & Regulatory Affairs<br>Avinger Inc. |
| Contact Information | Email: tlawson@avinger.com<br>Phone: 510-206-1794 |
| Date Prepared | 29 September 2017 |
### Proposed Device
| Trade Name | Pantheris System |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Common Name | Avinger Pantheris Catheter |
| Regulation Number and<br>Classification Name | 21 CFR§870.4875, Intraluminal Artery Stripper<br>21 CFR§892.1560, Ultrasonic Pulsed Echo Imaging<br>System |
| Product Code | MCW, NQQ |
| Regulatory Class | II |
# Predicate Device
| Trade Name | Pantheris System |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Common Name | Avinger Pantheris Catheter |
| Premarket Notification | K162326 |
| Regulation Number and<br>Classification Name | 21 CFR§870.4875, Intraluminal Artery Stripper<br>21 CFR§892.1560, Ultrasonic Pulsed Echo Imaging<br>System |
| Product Code | MCW, NQQ |
| Regulatory Class | II<br>Note: This predicate device has not been subject to a design-related recall. |
# Device Description and Proposed Modifications
The Pantheris System (cleared initially under K152275) combines the use of Avinger's optical coherence tomography (OCT) technology with peripheral vascular atherectomy
{4}------------------------------------------------
capabilities. The Pantheris System consists of the Pantheris catheter, Lightbox Sled with integrated Umbilical (referred to as "Sled") and the Lightbox HS Imaging Console (referred to as "Lightbox").
The Pantheris Catheter comes in both a 7Fr and an 8Fr size. Both sizes have a working length of 110 cm and are sterile, single-use devices that are compatible with 7 and 8F sheaths (respectively) and 0.014" guidewires. The Pantheris Catheter incorporates an Optical Fiber that allows for real-time OCT guided directional atherectomy during the procedure.
The Pantheris catheter is connected to the Lightbox via the Sled. The Lightbox is an optical transceiver, transmitting light to the intraluminal environment through the optical fiber on the Pantheris Catheter and receiving and interpreting the signal from the tissue using a PCbased processing system. The Lightbox provides a visualization platform for the real-time OCT-assisted directional atherectomy. The Lightbox consists of a cart with two monitors, a PC-based processing system, software, an isolation transformer, and an OCT system.
The Pantheris catheter is to be used in a healthcare facility, such as a cardiac catheter lab or a hospital. It is to be used and in contact with patient tissue for less than 24 hours and is made of materials that are biocompatible.
The submission of K162326 for the Pantheris system was focused solely on a change in the system software that permits the use of OCT images to identify vessel lumen, wall structures, and vessel morphologies.
This Special 510(k) details modifications to improve manufacturing compatibility, modify accessories that have no direct patient contact, and describes a patch of the system software that allows the attenuation of the laser whenever the drive in the Sled is not rotating.
#### Indications for Use
The indications for use for the Pantheris System cleared under K163264 is not altered due to the software and other modifications disclosed in this submission:
The Pantheris System is intended to remove plaque (atherectomy) from partially occluded vessels in the peripheral vasculature with a reference diameter of 3.0mm to 7.0mm, using OCT-assisted orientation and imaging. The system is as an adjunct to fluoroscopy by providing images of vessel lumen, wall structures and vessel morphologies. The Pantheris System is NOT intended for use in the iliac, coronary, cerebral, renal or carotid vasculature.
{5}------------------------------------------------
Both the subject device and the predicate device have the same intended use for the removal of plaque from partially occluded vessels in the peripheral vasculature.
### Comparison of Technological Characteristics with the Predicate Device
Excision of plaque from peripheral vessels through atherectomy is the technological principle for both the subject and predicate devices.
At a high level, the subject and predicate devices are based upon the same technological elements:
- . Device is advanced to the point of treatment through an indwelling sheath;
- Measurement and display of vessel lumen, wall structures, and vessel morphology ● via the OCT-imaging component located directly behind the tissue cutter component;
- Rotation of the OCT-imaging and cutter components by drives in the Sled ● component of the system;
- Display of OCT-generated images on monitors attached to the Lightbox component ● of the system;
- Excision of targeted tissue by a cutter component; and .
- Packing of excised tissue into the nosecone component of the catheter. ●
The only technological difference is that the patch release of the software in the subject device allows the attenuation of the laser whenever the drive in the Sled is not rotating. In the predicate device, the laser is not stopped when the drive is not rotating.
Summary of Technological Characteristics of the Subject Device Compared to Those of the Predicate Device
| Technological<br>Characteristic | Pantheris Catheter<br>Predicate Device<br>K162326 | Pantheris Catheter<br>Subject Device of this<br>Submission |
|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| Indication for Use | The Pantheris System is<br>intended to remove plaque<br>(atherectomy) from partially<br>occluded vessels in the<br>peripheral vasculature with a<br>reference diameter of 3.0 mm<br>to 7.0 mm, using OCT-<br>assisted orientation and<br>imaging. The system is as an<br>adjunct to fluoroscopy by<br>providing images of vessel | Same |
| | lumen, wall structures and vessel morphologies.<br>The Pantheris System is NOT intended for use in the iliac, coronary, cerebral, renal or carotid vasculature." | Same |
| Anatomical Site of Use | Peripheral Vasculature | Same |
| Components of the System | Lightbox console<br>Sled<br>Pantheris Catheter | Same |
| Imaging Modality | Optical Coherence Tomography | Same |
| Imaging Energy Type | Near-infrared light | Same |
| Optical Output Power | < 30 mW<br>(Class 1M laser output) | Same |
| Optical sensitivity (signal:noise ratio) | 90 db minimum | Same |
| Attenuation of the laser when the Sled driver is not rotating | No | Yes |
| Imaging Capabilities | OCT-assisted orientation and imaging of vessel lumen and wall structures in the peripheral vasculature to facilitate atherectomy.<br>Identify clinically relevant morphologies and assess complex lesions in the peripheral vasculature.<br>Differentiate and classify plaque | Same |
| Electrical Safety | Class I, Type CF, defibrillation proof IEC 60601-1 | Same |
| Electromagnetic compatibility | IEC 60601-1-2 | Same |
| Laser Safety | 21 CFR Part 1040<br>IEC 60825 | Same |
| Software Level of Concern | Moderate | Same |
| Sheath compatibility for<br>the catheter | 7 Fr & 8 Fr | Same |
| Working length of the<br>catheter | 110 cm | Same |
| Guidewire compatibility<br>of the catheter | 0.014 in | Same |
| Distal tip OD of the<br>catheter | 0.100 (7 Fr)<br>0.110 (8 Fr) | Same |
| Provided Sterile | Yes | Yes |
| Single-use catheter | Yes | Yes |
{6}------------------------------------------------
{7}------------------------------------------------
While the technique of plaque excision between the subject and predicate devices are the same, there have been design changes to the subject device to improve manufacturability and usability of the catheter, changes to the tweezers and flush fixture accessories, modification of the Sled component of the system, updates to the instructions for use, and a patch release of the system's software. The design changes are summarized below.
Summary of design and other changes of the subject device covered in this 510(k) submission:
Manufacturing
- . Modify balloon inflation test during assembly
- Shorten the torque shaft hypotube length ●
- Add strain relief in the handle to reduce tension pulling on fibers from the lens ●
- Modify the edge of the cutter ●
- Modify the bushing that aligns the cutter
Sled
- Modify the integrated connector of the umbilical cable
- Modify the FORJ, which increased the length of the Sled slightly
Accessories
- . Change tweezers material from nylon to stainless steel and change the shape of its tip
- Length the flush fixture increased .
Labeling
- . Revise the IFU in terms of volume of CO2 used to inflate the balloons
- Revise the IFU to instruct user to test balloon inflation prior to insertion of the ● catheter
{8}------------------------------------------------
Software
- Update of software to allow attenuation of the laser whenever the drive in the Sled is not rotating
Testing was performed to demonstrate that these changes do not impact the safety and effectiveness of the final device.
#### Performance Data
The performance testing conducted establishes that these minor modifications do not raise new questions of the safety and effectiveness for the Pantheris System cleared under K163264.
#### Biocompatibility testing
There are no changes in patient-contacting materials in the subject device when compared to the predicate device. The only material change is to an accessory provided with the device, a pair of tweezers, which were nylon in the predicate device and in the subject device are stainless steel. The stainless steel tweezers do not have direct patient contact, but were subjected nonetheless to the following testing and passed all tests:
- o Cytotoxicity:
- Sensitization: O
- Irritation: and O
- o Systemic toxicity.
#### Electrical safety and electromagnetic compatibility (EMC)
The predicate and subject devices comply with IEC 60601-1 standard for safety and the IEC 60601-1-2 standard for EMC.
#### Software Verification and Validation Testing
Software verification and validation testing, as well as regression testing, were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The software for this device is considered as a "moderate" level of concern.
## Mechanical Testing
The mechanical testing of the subject device included:
- Sheath insertion cycle test; ●
- Destructive balloon burst test; ●
{9}------------------------------------------------
- Torque capacity; ●
- Leak evaluation; ●
- Bond joint tensile strength; ●
- . Life-cycle tests:
- 8-month accelerated shelf life study;
- Simulated use testing;
- Corrosion testing*;
- Heat generation testing;
- Rotational speed testing*; ●
- Dimensional verification*;
- Balloon inflation/deflation cycle test; ●
- Plaque removal efficiency; and ●
- Flushing tool capacity. ●
*device historical data/information was provided to support the test conclusions
## Animal Testing
No preclinical testing of the modifications of the subject device was necessary.
## Clinical Studies
No clinical testing of the modifications of the subject device was necessary.
## Conclusion
The information submitted in this premarket notification confirms that the minor modifications to the Pantheris catheter, components, and system software raise no new questions of safety and effectiveness and that the device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.