The LIASYS 600 Electrolyte Measurement System is intended for the quantitative determination of sodium, chloride, and potassium in human serum. Sodium measurements are used in the diagnosis and treatment of aldosteronism (excessive secretion of hormone aldosterone), diabetes insipidus, adrenal hypertension, Addison’s disease, dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance. Measurements of potassium are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis. The LIASYS 600 creatinine reagent is intended for the quantitative determination of creatinine in human serum. Packaged for the ease of use with LIASYS 600. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. The LIASYS 600 is a random access, computer controlled, standalone clinical analyzer intended for the in-vitro determination of analytes in human serum.
Device Story
LIASYS 600 is a random-access, computer-controlled, standalone clinical chemistry analyzer. It processes human serum samples loaded onto dual carousels (100-sample capacity). Sodium, potassium, and chloride are measured via an Ion Selective Electrode (ISE) module; creatinine is measured via a photometric Jaffe reaction (picric acid/sodium hydroxide). The system translates electrical potential (Nernst equation) or absorbance (510 nm) into analyte concentrations. Operated by laboratory personnel in clinical settings, the device provides quantitative results to assist physicians in diagnosing electrolyte imbalances, renal disease, and metabolic disorders. It supports barcode specimen identification and requires external quality control materials and specific calibrators.
Clinical Evidence
Bench testing only. Performance established via precision studies (CLSI EP05-A3), linearity (CLSI EP06), detection limits (CLSI EP17-A2), and interference testing (CLSI EP07-A2). Method comparison study (N=109-117 samples) against legally marketed comparators showed strong correlation (R=0.990-0.998) and acceptable Deming regression slopes (0.956-1.056). No clinical trials performed.
Technological Characteristics
Random-access clinical chemistry analyzer. ISE module for Na/K/Cl; photometric/colorimetric (Jaffe) for creatinine. Reagents: 10 mM picric acid, 250 mM NaOH. Throughput: 600 tests/hr (absorbance), 120 tests/hr (ISE). Connectivity: Standalone, barcode-enabled, no wireless transmission. Traceable to NIST SRM 919b, 918c, 956d, 914a.
Indications for Use
Indicated for quantitative determination of sodium, potassium, chloride, and creatinine in human serum for patients requiring diagnosis or monitoring of electrolyte imbalances, metabolic disorders (e.g., cystic fibrosis, diabetic acidosis), or renal diseases. No specific age or gender contraindications provided, though icteric samples are contraindicated for sodium and creatinine testing.
Regulatory Classification
Identification
A sodium test system is a device intended to measure sodium in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k172232
B. Purpose for Submission:
New device
C. Measurand:
Sodium, potassium, chloride, and creatinine
D. Type of Test:
Quantitative, potentiometric, ion selective electrodes for sodium, potassium, and chloride
Quantitative, photometric/colorimetric method for creatinine
E. Applicant:
AMS Diagnostics
F. Proprietary and Established Names:
LIASYS 600 Electrolyte Measurement System, LIASYS 600 Creatinine reagent, LIASYS 600
G. Regulatory Information:
| Regulation | Classification | Product Code | Panel |
| --- | --- | --- | --- |
| 21 CFR 862.1665
Sodium Test System | Class II | JGS | Clinical Chemistry (75) |
| 21 CFR 862.1600
Potassium Test System | Class II | CEM | Clinical Chemistry (75) |
| 21 CFR 862.1170
Chloride Test System | Class II | CGZ | Clinical Chemistry (75) |
| 21CFR 862.1225
Creatinine Test System | Class II | CGX | Clinical Chemistry (75) |
| 21 CFR 862.2160
Discrete photometric chemistry analyzer for clinical use | Class I | JJE | Clinical Chemistry (75) |
{1}
H. Intended Use:
1. Intended use:
See Indications for use.
2. Indications for use:
The LIASYS 600 Electrolyte Measurement System is intended for the quantitative determination of sodium, chloride, and potassium in human serum. Sodium measurements are used in the diagnosis and treatment of aldosteronism (excessive secretion of hormone aldosterone), diabetes insipidus, adrenal hypertension, Addison’s disease, dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance. Measurements of potassium are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
The LIASYS 600 creatinine reagent is intended for the quantitative determination of creatinine in human serum. Packaged for the ease of use with LIASYS 600. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
The LIASYS 600 is a random access, computer controlled, standalone clinical analyzer intended for the in-vitro determination of analytes in human serum.
3. Special conditions for use statements:
For prescription use only.
4. Special instrument requirements:
LIASYS 600
I. Device Description:
The LIASYS 600 sodium, potassium, and chloride assays are supplied through an ISE (ion specific electrode) module installed onto the LIASYS 600 instrument. Internal solution composition: sodium 140 mmol/L, potassium 4.0 mmol/L, and chloride 125 mmol/L.
The LIASYS 600 creatinine assay is supplied as reagent kit in a single vial with ready to use liquid. Reagent concentrations: 10 mM picric acid, 250 mM sodium hydroxide 250 mM; pH 12.8 to 13.2.
The LIASYS 600 is a fully automated, random access, analyzer for clinical chemistry assays.
{2}
J. Substantial Equivalence Information:
1. Predicate device names:
LIASYS Sodium, potassium, and chloride - k080468
Roche Reagent for Creatinine - k941837
LIASYS 450 SAT 450 - k113131
2. Predicate 510(k) numbers:
See predicate device names above.
3. Comparison with predicate:
| Similarities and Differences for Na/K/Cl | | |
| --- | --- | --- |
| Item | Candidate device k172232 | Predicate device k080468 |
| Intended use | For the quantitative determination of the sodium, potassium, and chloride concentration in serum. | Same |
| Method | Ion Selective Electrode | Same |
| Specimen types | Serum | Same |
| Measuring Range | Na: 53 - 200 mmol/L
K: 1.8 - 10.1 mmol/L
Cl: 50 - 200 mmol/L | Na: 27 - 200 mmol/L
K: 0.4 - 20 mmol/L
Cl: 27 - 200 mmol/L |
| Similarities and Differences for Creatinine | | |
| --- | --- | --- |
| Item | Subject device k172232 | Predicate device k941837 |
| Intended use | For the quantitative determination of the creatinine concentration in serum. | Same |
| Method | Colorimetric – Jaffe reaction. | Same |
| Specimen types | Serum | Serum, plasma, and urine |
| Measuring Range | 0.2 - 20 mg/dL | 0 - 30 mg/dL serum
0 - 200 mg/dL urine |
{3}
| Similarities and Differences for LIASYS 600 | | |
| --- | --- | --- |
| Item | Candidate device k172232 | Predicate device k113131 |
| Intended use | A random access, computer controlled, standalone clinical analyzer for clinical chemistry intended for the in-vitro determination of analytes in human serum. | Same |
| Measurement Modes | Absorbance and ion selective electrodes. | Same |
| Throughput | Absorbance: 600 tests per hour
Ion selective electrodes: Serum 120 tests/hr | Absorbance: 450 tests per hour
Ion selective electrodes: Serum 120 tests/hr Urine 60 tests/hr |
K. Standard/Guidance Document Referenced:
CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition.
CLSI EP07-A2 Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition.
CLSI EP17-A2 Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition.
L. Test Principle:
The sodium, potassium, and chloride assays are based on ion selective electrode technologies that are specific for each ion of interest in the sample. An electrical potential is developed according to the Nernst Equation for a specific ion. When compared to the internal solution, this electrical potential is translated into voltage and then into the ion concentration of the sample.
The creatinine assay is a photometric assay based on Jaffe reaction, in which creatinine reacts with picric acid under alkaline conditions to form a colored complex with absorbance at 510 nm.
{4}
# M. Performance Characteristics:
# 1. Analytical performance:
# a. Precision/Reproducibility:
The precision performance for sodium, potassium, chloride, and creatinine was established by a within-laboratory study following the recommendations in CLSI EP05-A3 guideline. In the study, serum based control material samples were tested in duplicate per run with 2 runs per day over 20 days for a total of 80 measurements. The results are summarized below.
Sodium
| Mean, mmol/L | Repeatability %CV | Between-Run %CV | Between-Day %CV | Within-laboratory %CV |
| --- | --- | --- | --- | --- |
| 161 | 0.6% | 0.6% | 0.5% | 1.0% |
| 141 | 0.6% | 0.6% | 0.7% | 1.1% |
| 122 | 0.6% | 0.3% | 0.7% | 1.0% |
Potassium
| Mean, mmol/L | Repeatability %CV | Between-Run %CV | Between-Day %CV | Within-laboratory %CV |
| --- | --- | --- | --- | --- |
| 6.13 | 0.9% | 1.3% | 1.3% | 2.0% |
| 4.48 | 1.1% | 1.1% | 1.1% | 1.9% |
| 2.57 | 1.1% | 0.9% | 2.1% | 2.5% |
Chloride
| Mean, mmol/L | Repeatability %CV | Between-Run %CV | Between-Day %CV | Within-laboratory %CV |
| --- | --- | --- | --- | --- |
| 121 | 0.8% | 0.2% | 0.7% | 1.1% |
| 105 | 0.6% | 0.7% | 0.2% | 0.9% |
| 91.9 | 0.7% | 0.5% | 0.9% | 1.2% |
Creatinine
| Mean, mg/dL | Repeatability %CV | Between-Run %CV | Between-Day %CV | Within-laboratory %CV |
| --- | --- | --- | --- | --- |
| 0.48 | 5.7% | 2.3% | 5.8% | 8.5% |
| 1.11 | 1.7% | 1.4% | 4.0% | 4.6% |
| 4.55 | 1.7% | 2.2% | 2.2% | 3.6% |
| 12.48 | 0.9% | 1.8% | 1.0% | 2.3% |
| 16.73 | 0.8% | 0.8% | 0.9% | 1.5% |
{5}
# b. Linearity/assay reportable range:
The linearity performance of the LIASYS 600 for sodium, potassium, chloride, and creatinine was established in a study by testing six samples across the measurement range in replicates of 2. The samples were commercially available standards. The expected values were plotted against the observed values and a linear regression line was fitted with the following regression equations:
| Analyte | Slope | Intercept | Correlation coefficient, R |
| --- | --- | --- | --- |
| Sodium | 0.989 | 0.8 | 0.978 |
| Potassium | 1.000 | 0.14 | 0.999 |
| Chloride | 0.994 | 1.6 | 0.988 |
| Creatinine | 0.990 | 0.13 | 0.980 |
The linearity studies support the following claimed measurement ranges:
| Analyte | Analytical Measurement Range | Sample Range Tested |
| --- | --- | --- |
| Sodium | 53 – 200 mmol/L | 26 – 204 mmol/L |
| Potassium | 1.8 – 10.1 mmol/L | 0.6 – 10.1 mmol/L |
| Chloride | 50 – 200 mmol/L | 26 – 200 mmol/L |
| Creatinine | 0.2 – 20.0 mg/dL | 0.1 – 19.7 mg/dL |
# c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The sodium, potassium, chloride, and creatinine assays on the LIASYS 600 are traceable to commercially available standards whose concentrations were verified against NIST Standard Reference Material (SRM) 919b for sodium, SRM 918c for potassium, SRM 956d for chloride, and SRM 914a for creatinine.
# d. Detection limit:
The detection limit of the LIASYS 600 creatinine assay was assessed in a study measuring the limit of quantitation (LoQ) following the recommendation in CLSI EP17-A2 guideline. The low level creatinine samples were measured in 10 replicates per run with 3 runs per day for 3 days. The results support the claimed LoQ for creatinine of $0.19\mathrm{mg / dL}$ using the sponsor definition of LoQ as the lowest concentration with $< 20\%$ CV.
The lower limit of the measurement range, supported by linearity studies above in M.1.b defines the detection limit for the LIASYS 600 sodium, potassium, and chloride assays.
{6}
# e. Analytical specificity:
Interference testing on the LIASYS 600 for sodium, potassium, chloride, and creatinine was evaluated follow the recommendations in CLSI EP07-A2 guideline. The study tested for possible interference from bilirubin, lipemia, and hemolysis. Testing was conducted using samples with the following analyte concentrations: sodium = 138 mmol/L, potassium = 4 mmol/L, chloride = 93 mmol/L, creatinine = 1.1 mg/dL. The samples were divided into two aliquots: control sample with no added interferent and test sample with interferent added. The test samples were serially diluted into a series and measured in duplicates to identify the highest concentration without significant bias (i.e., <10% difference between test and control). The results of the study are given below.
| Sodium – Substances tested | Highest concentration without significant interference |
| --- | --- |
| Hemoglobin | 1000 mg/dL |
| Bilirubin conjugate, ditaurate | 10.7 mg/dL |
| Intralipid | 1130 mg/dL |
| Potassium – Substances tested | Highest concentration without significant interference |
| --- | --- |
| Hemoglobin | 900 mg/dL |
| Bilirubin conjugate, ditaurate | 17.3 mg/dL |
| Intralipid | 1130 mg/dL |
| Chloride – Substances tested | Highest concentration without significant interference |
| --- | --- |
| Hemoglobin | 1000 mg/dL |
| Bilirubin conjugate, ditaurate | 17.3 mg/dL |
| Intralipid | 1130 mg/dL |
| Creatinine – Substances tested | Highest concentration without significant interference |
| --- | --- |
| Hemoglobin | 500 mg/dL |
| Bilirubin conjugate, ditaurate | 7.9 mg/dL |
| Intralipid | 538 mg/dL |
The following limitations are identified in the product labeling:
Do not test for sodium in icteric samples.
Do not test for creatinine in icteric samples.
{7}
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
The accuracy performance of the LIASYS 600 was evaluated in a method comparison study by comparing sodium, potassium, chloride, and creatinine measurements for agreement to legally marketed comparator methods. For the study, serum specimens were tested in singlicate on the subject device and in duplicate on the comparator methods. To cover the measurement range, no more than 10% of the samples were diluted and/or spiked. The results of the method comparison study are summarized as follows.
Deming regression analysis
| Analyte | Slope | Intercept | R | N | Range Tested |
| --- | --- | --- | --- | --- | --- |
| Sodium | 1.019 | -3.2 | 0.990 | 110 | 53 - 197 mmol/L |
| Potassium | 1.056 | -0.3 | 0.998 | 111 | 1.80 - 9.70 mmol/L |
| Chloride | 0.956 | 6.8 | 0.990 | 117 | 50 - 192 mmol/L |
| Creatinine | 1.031 | 0.08 | 0.998 | 109 | 0.2 to 19.8 mg/dL |
b. Matrix comparison:
Not applicable, only human serum samples are indicated.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data:
Not applicable.
4. Clinical cut-off:
Not applicable.
{8}
9
5. Expected values/Reference range:
The product labeling identifies the following reference ranges from the literature:
Sodium¹ 136 – 145 mmol/L
Potassium¹ 3.5 – 5.1 mmol/L
Chloride¹ 98 – 107 mmol/L
Creatinine²
men 0.72 – 1.18 mg/dL
women 0.55 – 1.02 mg/dL
Sources:
1. Tietz, textbook of Clinical Chemistry 2nd ed, Philadelphia, W.B. Saunders, (1994).
2. Henry RJ (Ed). Clin. Chem. Principles and Techniques (2nd Ed). Harper and Row. 1974, 548-551.
N. Instrument Name:
LIASYS 600
O. System Descriptions:
1. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ X or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ or No ☐ X
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
3. Specimen Identification:
Barcode option for specimen identification.
4. Specimen Sampling and Handling:
Sample specimens are loaded onto the instrument’s two independent carousels which can host a total of 100 samples.
{9}
5. Calibration:
The product labeling recommends for calibration of sodium, potassium, and chloride the use of an appropriate ISE calibrator, and for the creatinine assay the use of AMS Chemistry Calibrator.
6. Quality Control:
The product labeling recommends for quality control the use of two levels of external quality control materials.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Not applicable.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.