← Product Code [MWI](/productcode/MWI) · K172221

# VTRUST Vital Signs Monitor, Model: TD-2300; FORA Vital Signs Monitor, Model: VSM100 (K172221)

_Taidoc Technology Corporation · MWI · Jun 6, 2018 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K172221

## Device Facts

- **Applicant:** Taidoc Technology Corporation
- **Product Code:** [MWI](/productcode/MWI.md)
- **Decision Date:** Jun 6, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Vital Signs Monitor is intended for the monitoring of parameters such as ECG, non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, body temperature, blood glucose and oxygen saturation of arterial hemoglobin (SpO2) of adult patient. The monitor is intended to be used by clinicians and medically qualified personnel and trained users.

## Device Story

Multi-parameter vital signs monitor; inputs include ECG signals, NIBP, pulse rate, body temperature, blood glucose, and SpO2; device processes physiological signals to provide real-time monitoring data; used by clinicians and trained personnel in clinical settings; output displayed on monitor for assessment of patient status; assists healthcare providers in clinical decision-making; benefits patient through continuous or spot-check monitoring of critical physiological parameters.

## Clinical Evidence

Bench testing and clinical evaluations performed. Bench testing verified performance against specifications for ECG, SpO2, NIBP, blood glucose, and temperature modules. Clinical evaluation conducted to ensure device performance complies with design specifications. No specific sensitivity, specificity, or AUC metrics provided; results indicate performance is equivalent to predicate devices.

## Technological Characteristics

Multi-parameter monitor; includes ECG, NIBP, SpO2, temperature, and blood glucose modules. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-27, IEC 60601-2-49, IEC 80601-2-30, ISO 80601-2-61, ISO 10993, and ISO 14971. Software developed per IEC 62304. Connectivity includes data transmission capabilities.

## Regulatory Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

## Predicate Devices

- V-TRUST Model TD-8002 Multi-Parameter Spot Check Monitor ([K101259](/device/K101259.md))
- V-TRUST TD-2202 Portable ECG Recorder ([K101569](/device/K101569.md))
- TD-4268 Blood Glucose Monitoring System ([K120042](/device/K120042.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 6, 2018

TaiDoc Technology Corporation Anne Kuo Regulatory Affairs Specialist 6F, No. 127, Wugong 2nd Rd., Wugu District New Taipei City, Taiwan 24888

Re: K172221

Trade/Device Name: VTRUST Vital Signs Monitor, Model: TD-2300; FORA Vital Signs Monitor, Model: VSM100 Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI, DQA, DXN, FLL, NBW, LFR, DPS Dated: May 1, 2018 Received: May 4, 2018

Dear Anne Kuo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

M.A. Wilhelm

for Bram D. Zuckerman, M.D.
 Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K172221

#### Device Name

VTRUST Vital Signs Monitor, Model: TD-2300 FORA Vital Signs Monitor, Model: VSM100

#### Indications for Use (Describe)

The Vital Signs Monitor is intended for the monitoring of parameters such as ECG, non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, blood glucose and oxygen saturation of arterial hemoglobin (SpO2) of adult patient. The monitor is intended to be used by clinicians and medically qualified personnel and trained users.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for TaiDoc. The logo consists of the word "TaiDoc" in a gray sans-serif font, followed by three red hearts. Below the hearts is a gray rectangle with the text "www.taidoc.com" in white.

Tel : +886-2-6625-8188 Fax : +886-2-6625-0288

# 510(k) Summary

This summary of 510(k) information is being submitted in accordance with the requirements of 21 CFR 807.92.

The Assigned 510(k) number is: K172221_

## 1. Submitter Information

| Company Name:   | TaiDoc Technology Corporation                                          |
|-----------------|------------------------------------------------------------------------|
| Address:        | 6F, No. 127, Wugong 2nd Rd., Wugu Dist., New Taipei City, Taiwan 24888 |
| Contact Person: | Anne Kuo                                                               |
| Title:          | Regulatory Affairs Specialist                                          |
| Phone:          | +886-2-6625-8188 #1195                                                 |
| Fax:            | +886-2-6625-0608                                                       |
| E-mail:         | Anne.Kuo@taidoc.com.tw                                                 |

## 2. Proposed Device Information:

| Trade/Device Name:           | VTRUST Vital Signs Monitor, Model: TD-2300<br>FORA Vital Signs Monitor, Model: VSM100 |
|------------------------------|---------------------------------------------------------------------------------------|
| Regulation Name:             | Monitor, physiological, patient (without arrhythmia detection or alarms)              |
| Regulatory Class:            | Class II                                                                              |
| Regulation Number:           | 21 CFR §870.2300                                                                      |
| Review Panel:                | Cardiovascular                                                                        |
| Product Code:                | MWI                                                                                   |
| Subsequent Product<br>Codes: | DPS Electrocardiograph                                                                |
|                              | DQA Oximeter                                                                          |
|                              | DXN System, measurement, blood-pressure, non-invasive                                 |
|                              | FLL Thermometer, Electronic, Clinical                                                 |
|                              | NBW Blood Glucose Test System, Over-the-Counter                                       |
|                              | LFR Glucose Dehydrogenase                                                             |

{4}------------------------------------------------

|                                  | 泰博科技股份有限公司<br>TaiDoc Technology Corp. |                                               | 新北市24888五股區五工二路127號6樓<br>6F., No.127, Wugong 2nd Rd., Wugu Dist.,<br>New Taipei City 24888, Taiwan | Tel: +886-2-6625-8188<br>Fax: +886-2-6625-0288 |
|----------------------------------|---------------------------------------|-----------------------------------------------|----------------------------------------------------------------------------------------------------|------------------------------------------------|
|                                  | www.taidoc.com                        |                                               |                                                                                                    |                                                |
| Subsequent Regulation<br>Number: | 21 CFR §870.2340                      | Electrocardiograph                            |                                                                                                    |                                                |
|                                  | 21 CFR §870.2700                      | Oximeter                                      |                                                                                                    |                                                |
|                                  | 21 CFR §870.1130                      | Noninvasive blood pressure measurement system |                                                                                                    |                                                |
|                                  | 21 CFR §880.2910                      | Clinical electronic thermometer               |                                                                                                    |                                                |
|                                  | 21 CFR §862.1345                      | Glucose test system                           |                                                                                                    |                                                |

## 3. Intended Use:

The Vital Signs Monitor is intended for the monitoring of parameters such as ECG, non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, body temperature, blood glucose and oxygen saturation of arterial hemoglobin (SpO2) of adult patient. The monitor is intended to be used by clinicians and medically qualified personnel and trained users.

#### 4. Device Description:

The Vital Signs Monitor is intended for the monitoring of parameters such as ECG, non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, body temperature, blood glucose and oxygen saturation of arterial hemoglobin (SpO2) of adult patient. The monitor is intended to be used by clinicians and medically qualified personnel and trained users.

The body temperature and blood glucose measurements are cleared by FDA under k101259 and k120042 respectively.

VTRUST Vital Signs Monitor and FORA Vital Signs Monitor are exactly identical in design technical and physical appearance. The only difference is the brand name and model name.

| Primary predicate device                       | V-TRUST Model TD-8002 Multi-Parameter Spot Check Monitor<br>k101259                                 |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Subsequent Predicate Device<br>(ECG)           | V-TRUST TD-2202 Portable ECG Recorder<br>k101569                                                    |
| Subsequent Predicate Device<br>(Blood Glucose) | TD-4268 Blood Glucose Monitoring System<br>TD-4268 Multi Blood Glucose Monitoring System<br>k120042 |
| Similarities                                   | ECG technology                                                                                      |

#### 5. Substantial Equivalence Information:

{5}------------------------------------------------

|  | 泰博科技股份有限公司<br>TaiDoc Technology Corp. | 新北市24888五股區五工二路127號6樓<br>6F., No.127, Wugong 2nd Rd., Wugu Dist.,<br>New Taipei City 24888, Taiwan | Tel : +886-2-6625-8188<br>Fax : +886-2-6625-0288 |
|--|---------------------------------------|----------------------------------------------------------------------------------------------------|--------------------------------------------------|
|  | www.taidoc.com                        |                                                                                                    |                                                  |
|  | NIBP technology                       |                                                                                                    |                                                  |
|  | SpO2 technology                       |                                                                                                    |                                                  |
|  | Blood glucose technology              |                                                                                                    |                                                  |
|  | Temperature technology                |                                                                                                    |                                                  |
|  | Data transmission                     |                                                                                                    |                                                  |
|  | Power source                          |                                                                                                    |                                                  |

#### 6. Test Principle:

The Vital Signs Monitor applies the same test principle as those of the predicate devices. No changes were made in the peripheral blood glucose meter and thermometer.

## 7. Safety and Effectiveness Characteristics:

The evaluation of safety, electromagnetic compatibility, biocompatibility, and software validation of the Vital Signs Monitor were conducted based on the following standards:

| Standard       | Title                                                                                                                                                                                                                                                              |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1    | Medical Electrical Equipment, Part 1: General requirements for basic safety and<br>essential performance                                                                                                                                                           |
| IEC 60601-1-2  | Medical electrical equipment - Part 1-2: General requirements for basic safety<br>and essential performance - Collateral standard: Electromagnetic compatibility -<br>Requirements and tests                                                                       |
| IEC 60601-1-8  | Medical electrical equipment - Part 1-8: General requirements for basic safety<br>and essential performance - Collateral Standard: General requirements, tests and<br>guidance for alarm systems in medical electrical equipment and medical<br>electrical systems |
| IEC 60601-2-27 | Medical electrical equipment - Part 2-27: Particular requirements for the basic<br>safety and essential performance of electrocardiographic monitoring equipment                                                                                                   |
| IEC 60601-2-49 | Medical electrical equipment - Part 2-49: Particular requirements for the basic<br>safety and essential performance of multifunction patient monitoring equipment                                                                                                  |
| IEC 62304      | Medical device software - Software life-cycle processes                                                                                                                                                                                                            |
| IEC 80601-2-30 | Medical electrical equipment - Part 2-30: Particular requirements for the basic<br>safety and essential performance of automated non-invasive<br>sphygmomanometers                                                                                                 |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for TaiDoc, a company that specializes in medical devices. The logo consists of the word "TaiDoc" in gray letters, followed by three red hearts. Below the logo is the website address "www.taidoc.com".

泰博科技股份有限公司 TaiDoc Technology Corp.

新北市24888五股區五工二路127號6樓 6F., No.127, Wugong 2nd Rd., Wugu Dist. New Taipei City 24888, Taiwan

Tel : +886-2-6625-8188 Fax : +886-2-6625-0288

| ISO 80601-2-61   | Medical electrical equipment -- Part 2-61: Particular requirements for basic<br>safety and essential performance of pulse oximeter equipment |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 10993 series | Biological evaluation of medical devices                                                                                                     |
| ISO 14971        | Medical device - Application of Risk Management to Medical Devices                                                                           |

## 8. Performance Bench Testing:

The laboratory and clinical evaluations for the performance of Vital Signs Monitor demonstrate that the modifications do not impact on the effectiveness. The performance of this system meets its intended use and is equivalent to the predicated devices.

| Description                       | Test Objective                                                                            |
|-----------------------------------|-------------------------------------------------------------------------------------------|
| ECG performance testing           | To ensure the ECG performance meets the specification<br>cleared under k101569.           |
| SpO2 performance testing          | To ensure the SpO2 performance meets the specification<br>cleared under k101259.          |
| NIBP performance testing          | To ensure the NIBP performance meets the specification<br>cleared under k101259.          |
| Blood Glucose performance testing | To ensure the blood glucose performance meets the<br>specification cleared under k120042. |
| Thermometer performance testing   | To ensure the thermometer performance meets the<br>specification cleared under k101259.   |
| Clinical Evaluation               | To ensure the device performance complies with the designed<br>specification.             |

#### 9. Conclusion:

Based on the information provided in this submission, the Vital Signs Monitor is believed to be substantially equivalent with the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K172221](https://fda.innolitics.com/device/K172221)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
