NAJA™ Ligament Correction System

K172206 · Cousin Biotech Sas · OWI · Apr 10, 2018 · Orthopedic

Device Facts

Record IDK172206
Device NameNAJA™ Ligament Correction System
ApplicantCousin Biotech Sas
Product CodeOWI · Orthopedic
Decision DateApr 10, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3010
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The NAJA System is a temporary implant for use in orthopedic surgery. The system is intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures. The indications for use include the following applications: 1. Spinal trauma surgery, used in sublaminar or facet wiring techniques. 2. Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as idiopathic and neuromuscular scoliosis in patients 8 years of age and older, adult scoliosis, kyphosis and spondylolisthesis. 3. Spinal degenerative surgery, as an adjunct to spinal fusions. The NAJA System may also be used in conjunction with other medical grade implants made of titanium or cobalt chrome alloy whenever "wiring" may help secure the attachment of the other implants.

Device Story

NAJA Ligament Correction System; temporary orthopedic implant for bone stabilization/fracture repair. Comprises titanium alloy connector and polyethylene terephthalate (PET) woven band. Used in spinal surgery (trauma, reconstruction, degenerative) as alternative to sublaminar wires/hooks; secures spine to rod constructs without bone screws. Operated by surgeons in clinical/OR settings. Provides temporary stabilization during bony fusion; functions as bone anchor. Can be used with other titanium or cobalt chrome implants to secure attachments. Benefits patient by providing stable fixation during healing process.

Clinical Evidence

Bench testing only. Mechanical testing included static and dynamic band tension, static axial and torsional grip (ASTM F1798-13), and dynamic axial compression (ASTM F1717-15). Biocompatibility verified per ISO 10993-1:2009. Packaging validated per ISO 11607-1/2:2006. Sterilization validated per ISO 11137-1/2:2006/2013.

Technological Characteristics

Materials: Titanium alloy connector, polyethylene terephthalate (PET) woven band. Mechanical testing per ASTM F1798-13 and ASTM F1717-15. Sterilization: Gamma radiation (SAL 10^-6) per ISO 11137-1/2. Packaging: ISO 11607-1/2. Biocompatibility: ISO 10993-1.

Indications for Use

Indicated for patients 8 years of age and older requiring spinal stabilization for trauma, reconstructive surgery (scoliosis, kyphosis, spondylolisthesis), or degenerative spinal fusion. Used as a bone anchor or adjunct to other titanium/cobalt chrome implants.

Regulatory Classification

Identification

A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font in blue. April 10, 2018 Cousin Biotech S.A.S. Mr. Franck Pelletier Regulatory Affairs Director 8 Rue De L'abbe Bonpain Wervicq-Sud 59117 FRANCE Re: K172206 Trade/Device Name: NAJATM Ligament Correction System Regulation Number: 21 CFR 888.3010 Regulation Name: Bone fixation cerclage Regulatory Class: Class II Product Code: OWI Dated: March 6, 2018 Received: March 9, 2018 Dear Mr. Pelletier: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {2}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Ronald P. Jean -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K172206 Device Name NAJATM Ligament Correction System #### Indications for Use (Describe) The NAJA System is a temporary implant for use in orthopedic surgery. The system is intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures. The indications for use include the following applications: 1. Spinal trauma surgery, used in sublaminar or facet wiring techniques. 2. Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as idiopathic and neuromuscular scoliosis in patients 8 years of age and older, adult scoliosis, kyphosis and spondylolisthesis. 3. Spinal degenerative surgery, as an adjunct to spinal fusions. The NAJA System may also be used in conjunction with other medical grade implants made of titanium or cobalt chrome alloy whenever "wiring" may help secure the attachment of the other implants. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------| | <div> <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 510(k) Summary provided in accordance with 21 CFR §807.92(C) | Date Prepared: | April 9, 2018 | |-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | Cousin Biotech S.A.S.<br>8 Rue De L'abbe Bonpain<br>59117 Wervicq-Sud, France | | Submitter Contact: | Ms. Mathilde Collet<br>Regulatory Affairs Engineer<br>Phone: +011 33 (3) 20144130<br>Email: m.collet@cousin-biotech.com | | Application<br>Correspondent: | Mr. Franck Pelletier<br>Regulatory Affairs Director<br>Phone: +011 33 (3) 20144119<br>Email: f.pelletier@cousin-biotech.com | | Manufacturing Site: | Cousin Biotech S.A.S.<br>Allée des roses<br>F-59117 Wervicq-Sud, France | | Trade Name: | NAJATM Ligament Correction System | | Common Name: | Bone Fixation Cerclage, Sublaminar | | Classification<br>Regulation: | 21 CFR §888.3010 - Bone Fixation Cerclage, Class II | | Product Code: | OWI | | Predicate Devices: | K110348 - Zimmer Spine - Universal Clamp Spinal Fixation System (Primary)<br>K922952 - Pioneer Surgical Technology - Songer Cable System (Additional) | | Device Description: | The NAJA™ Ligament Correction System (NAJA System) is a<br>temporary implant intended to provide temporary stabilization as a<br>bone anchor during the development of solid bony fusion and aid in<br>the repair of bone fractures. The NAJA implant consists of a titanium<br>alloy connector and a polyethylene terephthalate (PET) woven band.<br>The NAJA System allows the spine to be secured to a rod construct<br>without the use of a bone screw, and acts as an alternative to<br>sublaminar wires and hooks. | | Intended Use: | The NAJA System is a temporary implant for use in orthopedic<br>surgery. The system is intended to provide temporary stabilization as a<br>bone anchor during the development of solid bony fusion and aid in the<br>repair of bone fractures. The indications for use include the following<br>applications: | | | 1. Spinal trauma surgery, used in sublaminar or facet wiring techniques. | | | 2. Spinal reconstructive surgery, incorporated into constructs for the<br>purpose of correction of spinal deformities such as idiopathic and<br>neuromuscular scoliosis in patients 8 years of age and older, adult<br>scoliosis, kyphosis and spondylolisthesis. | | | 3. Spinal degenerative surgery, as an adjunct to spinal fusions. | | | The NAJA System may also be used in conjunction with other<br>medical grade implants made of titanium or cobalt chrome alloy<br>whenever "wiring" may help secure the attachment of the other<br>implants. | | Technology<br>Comparison: | The NAJA System is substantially equivalent to the predicates<br>identified in terms of intended use/indications for use, technological<br>characteristics, device materials and function. | | Shelf Life | The packaging for the NAJA System was validated in accordance with the following standards:<br>ISO 11607-1: 2006 Packaging for terminally sterilized medical devices – Part 1: requirements for materials, sterile barrier systems and packaging systems; and ISO 11607-2: 2006 Packaging for terminally sterilized medical devices – Part 2: Validation requirements for forming, sealing and assembly processes. Validation results indicate that the packaging for the NAJA System complies with the standards. | | Biocompatibility | NAJA System patient contact materials were verified in accordance with the following standard:<br>ISO 10993-1: 2009, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process. Verification results indicated that the NAJA System patient contact materials comply with the standard. | | Performance Testing – Bench | The following mechanical tests were performed on the NAJA System:<br>Static and dynamic band tension testing Static axial grip and static torsional grip testing per ASTM F1798-13 Dynamic axial compression testing per ASTM F1717-15 Test results demonstrate that the NAJA System has substantially equivalent mechanical performance as compared to the predicates listed. | | Conclusion | Verification and validation activities were conducted to establish the performance of NAJA System. The results of these activities demonstrate that NAJA System is as safe and as effective as the predicate devices.<br>Therefore, the NAJA System is considered substantially equivalent to the predicate devices. | {5}------------------------------------------------ #### Summary of Performance Testing: #### The NAJA System is gamma radiation sterilized and was validated to a Sterilization sterility assurance level of 10-6 in accordance with the following standards: - ISO 11137-1: 2006, Am1: 2013, Sterilization of health care products . – Requirements for validation and routine control – Radiation sterilization; and - ISO 11137-2:2013, Sterilization of health care products Radiation . – Part 2: Establishing the sterilization dose. Validation results indicate that the NAJA System complies with the standards. {6}------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...