← Product Code [FOZ](/productcode/FOZ) · K172204

# BD Intima II PLUS Closed IV Catheter System (K172204)

_Becton, Dickinson and Company · FOZ · Aug 17, 2017 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K172204

## Device Facts

- **Applicant:** Becton, Dickinson and Company
- **Product Code:** [FOZ](/productcode/FOZ.md)
- **Decision Date:** Aug 17, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5200
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The BD Intima II PLUS Closed IV Catheter System is inserted into a patient's vascular system for short-term use to monitor blood pressure or administer fluids intravascularly. These devices may be used for any patient with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy.

## Device Story

Closed IV catheter system; designed to contain blood during insertion. Components: radiopaque Vialon polyurethane catheter, notched needle for flashback visualization (BD Instaflash technology), septum, pinch clamp, extension tubing, Luer connector. Used in clinical settings by healthcare providers for short-term vascular access. Modifications from predicate include non-PVC extension tubing/shrink wrap, removal of slide clamp, addition of dual-port Y-connection configurations, and 0.56" catheter length. Device maintains sterility via EO sterilization. Healthcare providers use the system to establish venous access; the closed design reduces blood exposure risk during insertion. Benefits include reliable fluid delivery and pressure monitoring with improved safety features.

## Clinical Evidence

Bench testing only. Performance testing included pull force, joint leakage pressure, clamp seal pressure, and package integrity. Biocompatibility testing conducted per ISO 10993-1:2009 included cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, haemocompatibility, and pyrogenicity. Particulate analysis of surface and fluid path also performed. All tests met predetermined acceptance criteria.

## Technological Characteristics

Polyurethane (Vialon) catheter; stainless steel needle/wedge; silicone lubricant; non-PVC extension tubing (thermoplastic polyurethane); polyisoprene septum; polypropylene Luer; EO sterilized (SAL 10^-6). Integrated extension set with single or dual-port configurations. BD Instaflash technology for flashback visualization.

## Regulatory Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

## Predicate Devices

- BD Intima II Closed IV Catheter System ([K143610](/device/K143610.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 17, 2017

Becton, Dickinson and Company % Mark Job Responsible Third Party Official Regulatory Technology Services. LLC 1394 25th Street, NW Buffalo, Minnesota 55313

Re: K172204

Trade/Device Name: BD Intima II PLUS Closed IV Catheter System Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular catheter Regulatory Class: Class II Product Code: FOZ Dated: July 19, 2017 Received: July 21, 2017

Dear Mark Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

# Tara A. Ryan -S

for

Lori Wiggins Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

Device Name

BD Intima II PLUS Closed IV Catheter System

Indications for Use (Describe)

The BD Intima II PLUS Closed IV Catheter System is inserted into a patient's vascular system for short-term use to monitor blood pressure or administer fluids intravascularly. These devices may be used for any patient with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy.

| Type of Use (Select one or both, as applicable)                                                 |                                               |
|-------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over The Counter Use (21 CFR 801 Subpart C) |

|X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for BD, a medical technology company. The logo consists of two parts: a symbol on the left and the letters "BD" on the right. The symbol is a blue circle with a white figure inside, topped by a red sunburst. Below the logo is the company's slogan, "Helping all people live healthy lives," written in red.

## K172204 510(k) Summary

21 CFR §807.92

### BD Intima II PLUS™ Closed IV Catheter System

| Submitter<br>Information                                      | Submitter Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Becton Dickinson Infusion Therapy Systems Inc.                                                       |  |
|---------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|--|
|                                                               | Submitter Address:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 9450 South State Street<br>Sandy, UT 84070                                                           |  |
|                                                               | Contact Person:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Henry Boland<br>Staff Regulatory Affairs Specialist<br>henry.boland@bd.com<br>(801) 565-2550 (phone) |  |
|                                                               | Date of Preparation:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | June 30, 2017                                                                                        |  |
| Subject Device                                                | Trade Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | BD Intima II PLUS™ Closed IV Catheter System                                                         |  |
|                                                               | Common Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Peripheral Intravascular or IV Catheter                                                              |  |
|                                                               | Regulation Number:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 21 CFR §880.5200                                                                                     |  |
|                                                               | Regulation Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Intravascular Catheter                                                                               |  |
|                                                               | Regulatory Class:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | II                                                                                                   |  |
|                                                               | Product Code:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | FOZ                                                                                                  |  |
|                                                               | Classification Panel:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | General Hospital                                                                                     |  |
| Predicate<br>Device                                           | Trade Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | BD Intima II™ Closed IV Catheter System                                                              |  |
|                                                               | 510(k) Reference:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | K143610                                                                                              |  |
|                                                               | Common Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Peripheral Intravascular or IV Catheter                                                              |  |
|                                                               | Regulation Number:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 21 CFR §880.5200                                                                                     |  |
|                                                               | Regulation Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Intravascular Catheter                                                                               |  |
|                                                               | Regulatory Class:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | II                                                                                                   |  |
|                                                               | Product Code:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | FOZ                                                                                                  |  |
|                                                               | Classification Panel:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | General Hospital                                                                                     |  |
| Reason for<br>Submission                                      | The reason for his submission is to notify CDRH of modifications to the BD Intima II™ Closed IV Catheter System, including: 1) a change to extension tubing and PRN shrink wrap band from PVC to non-PVC materials; 2) the removal of the slide clamp option (only a pinch clamp option will be available); 3) the addition of new Y connection (dual port) with dual PRN adapter configurations; 4) the addition 0.56" length catheter configurations.                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                      |  |
| Device<br>Description                                         | The BD Intima II PLUS™ Closed IV Catheter System is a closed system IV catheter designed to keep blood contained within the device throughout the insertion process. The system consists of a radiopaque Vialon™ material catheter, a notched needle for flashback visualization, a septum to remove visible blood from the needle tubing, a pinch clamp, extension tubing and Luer connector. The system incorporates an integrated extension set which is available in three configurations: 1) Y Connection (dual port) with PRN adapter and end cap; 2) Y Connection (dual port) with two PRN adapters; and 3) Straight Connection (single port) with PRN adapter. The Luer connector is color-coded to indicate catheter gauge size. The system is provided EO sterilized to achieve a Sterility Assurance Level (SAL) of 10-6. The package system is able to maintain |                                                                                                      |  |
|                                                               | SUBJECT DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | PREDICATE DEVICE (K143610)                                                                           |  |
| Indications for<br>Use<br>(21 CFR<br>§807.92(a)(5))           | The subject device Indications for Use is the same as the predicate BD Intima II<br>Closed IV Catheter System (K143610), with the exception of the device trade<br>name. As such, the differences between the subject and predicate device<br>Indications for Use do not change the intended use of the predicate devices or<br>raise different questions of safety and effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                      |  |
| product sterility during its specified shelf life of 3 years. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                      |  |

{4}------------------------------------------------

| SUBJECT DEVICE                                                                                                                                                                                                                                                                                                                                                                                    | PREDICATE DEVICE (K143610)                                                                                                                                                                                                                                                                                                                                                                   |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The BD Intima II PLUS™ Closed IV<br>Catheter System is inserted into a patient's<br>vascular system for short-term use to<br>monitor blood pressure or administer fluids<br>intravascularly. These devices may be used<br>for any patient population with consideration<br>given to adequacy of vascular anatomy,<br>procedure being performed, fluids being<br>infused, and duration of therapy. | The BD Intima II™ Closed IV Catheter<br>System is inserted into a patient's vascular<br>system for short-term use to monitor blood<br>pressure or administer fluids intravascularly.<br>These devices may be used for any patient<br>population with consideration given to<br>adequacy of vascular anatomy, procedure<br>being performed, fluids being infused, and<br>duration of therapy. |

Technological Technological characteristics of the subject device are substantially equivalent to Characteristics the predicate device. The subject BD Intima II PLUS™ Closed IV Cather System achieves its intended use based on the same technology and principles of operation as the predicate BD Intima IITM Closed IV Cather System. The subject device has been modified from the predicate device as listed below. The results of design verification demonstrate that these changes are substantially equivalent to the predicate device. All other aspects of the subject device are identical to those of the predicate device.

- · Change to Extension Tubing and PRN Shrink Wrap Band materials from PVC to non-PVC materials
- · Removal of slide clamp component (only pinch clamp provided)
- · Qualification of new product configurations:
	- o Y Connection (dual port) with two pre-attached PRN adapters
		- o 0.56" length catheter configuration

| Attribute | SUBJECT DEVICE                                                                                                                                                                                                                | PREDICATE DEVICE (K143610)                                                                                                                                                                                                    |
|-----------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design    | A single-winged, polyurethane IV catheter with an integrated extension set incorporating either a single port or Y (dual)-port injection site. Incorporates BD Instaflash™ technology to assist with flashback visualization. | A single-winged, polyurethane IV catheter with an integrated extension set incorporating either a single port or Y (dual)-port injection site. Incorporates BD Instaflash™ technology to assist with flashback visualization. |
| Materials | <b>Component</b>                                                                                                                                                                                                              | <b>PREDICATE DEVICE</b>                                                                                                                                                                                                       |
|           | Catheter Tubing                                                                                                                                                                                                               | BD Vialon™ Polyurethane                                                                                                                                                                                                       |
|           | Catheter Tipping Lubricant                                                                                                                                                                                                    | Silicone                                                                                                                                                                                                                      |
|           | Catheter Lubricant                                                                                                                                                                                                            | Silicone                                                                                                                                                                                                                      |
|           | Metal Wedge                                                                                                                                                                                                                   | Stainless Steel                                                                                                                                                                                                               |
|           | Y-Adapter                                                                                                                                                                                                                     | Propionate                                                                                                                                                                                                                    |
|           | Septum/Sleeve Stopper                                                                                                                                                                                                         | Polyisoprene                                                                                                                                                                                                                  |
|           | Needle                                                                                                                                                                                                                        | Stainless Steel                                                                                                                                                                                                               |

{5}------------------------------------------------

|                                            | Needle Cover                                                                                                                                                                 | Same as predicate                        | Polyethylene                                                                                                          |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
|                                            | Extension Tubing                                                                                                                                                             | Thermoplastic Polyurethane               | Polyvinyl Chloride                                                                                                    |
|                                            | Pinch Clamp                                                                                                                                                                  | Same as predicate                        | Polyoxymethylene                                                                                                      |
|                                            | Slide Clamp                                                                                                                                                                  | No slide clamp option                    | Polystyrene                                                                                                           |
|                                            | Luer Connection Site                                                                                                                                                         | Same as predicate                        | Polypropylene                                                                                                         |
|                                            | PRN Body                                                                                                                                                                     | Same as predicate                        | Polycarbonate                                                                                                         |
|                                            | PRN Injection Port                                                                                                                                                           | Same as predicate                        | Polyisoprene                                                                                                          |
|                                            | PRN Shrink Wrap Band                                                                                                                                                         | Polyethylene Terephthalate               | Polyvinyl Chloride                                                                                                    |
|                                            | Cap                                                                                                                                                                          | Same as predicate                        | Acrylonitrile Butadiene Styrene                                                                                       |
|                                            | Extension Tubing Adhesive                                                                                                                                                    | Same as predicate                        | UV-Cured Epoxy                                                                                                        |
|                                            | Needle Adhesive                                                                                                                                                              | Same as predicate                        | UV-Cured Epoxy                                                                                                        |
|                                            | Septum Adhesive                                                                                                                                                              | Same as predicate                        | UV-Cured Epoxy                                                                                                        |
|                                            | Paddle Hub                                                                                                                                                                   | Same as predicate                        | Polystyrene                                                                                                           |
| Physical /<br>Mechanical<br>Specifications | SUBJECT DEVICE                                                                                                                                                               |                                          | PREDICATE DEVICE                                                                                                      |
|                                            | Catheter Diameters<br>18G, 20G, 22G, 24G<br>1.16"                                                                                                                            | Catheter Lengths<br>0.56", 0.75", 1.00", | Catheter Diameters<br>Catheter Lengths<br>18G, 20G, 22G, 24G<br>0.75", 1.00", 1.16"                                   |
| Product<br>Configurations                  | • Y Connection (dual port) with PRN<br>adapter and end cap<br>• Y Connection (dual port) with two PRN<br>adapters<br>• Straight Connection (single port) with<br>PRN adapter |                                          | • Y Connection (dual port) with PRN adapter<br>and end cap<br>• Straight Connection (single port) with PRN<br>adapter |

Summary of Performance tests completed on the subject device were limited to those tests Performance required to support a determination of substantial equivalence to the predicate devices. A risk analysis was conducted to assess the impact of the proposed Tests modifications to the subject devices. When technological characteristics between the subject and predicate devices were found to be identical, results of performance testing conducted on the predicate devices were applied to the subject device. The performance tests listed below were identified to ensure that specified design requirements were met:

- Extension Tubing / Catheter Adapter Pull Force .
- . Extension Tubing / Catheter Adapter Joint Leakage Pressure
- Extension Tubing Pull Out Force .
- Clamp Seal Pressure ●
- Package Integrity .

In addition, the following biocompatibility testing was conducted in accordance with ISO 10993-1:2009, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process.

- . Cytotoxicity
- Sensitization .
- . Intracutaneous Reactivity
- Systemic Toxicity (Acute) .
- . Haemocompatibility

{6}------------------------------------------------

|  |  | Pyrogenicity (Material-Mediated Rabbit Pyrogen) |  |
|--|--|-------------------------------------------------|--|
|--|--|-------------------------------------------------|--|

Finally, particulate analysis of the surface and fluid path was conducted.

Per the design control requirements specified in 21 CFR §820.30, the subject device met all predetermined acceptance criteria for the above-listed performance tests, demonstrating substantial equivalence to the predicate devices.

| Summary of<br>Substantial<br>Equivalence | Based on the indications for use, technological characteristics, and performance<br>testing, the subject BD Intima II PLUS™ Closed IV Catheter meets all pre-<br>determined requirements in accordance with 21 CFR §820.30, Design Controls,<br>and demonstrates that the subject device is substantially equivalent to the<br>predicate BD Intima II™ Closed IV Catheter System. |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

---

**Source:** [https://fda.innolitics.com/device/K172204](https://fda.innolitics.com/device/K172204)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
