Ins Hilden Humeral Arzzt, Ins Hilden Cannulated Tibial Arzzt, Ins Hilden Femoral Arzzt & Ins Hilden Retrograde Femoral Arzzt
Applicant
Truemed Group, LLC
Product Code
HSB · Orthopedic
Decision Date
Jan 16, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Ins Hilden Arzzt Nail Systems, consisting of intramedullary solid and cannulated nails, end cap and locking screws, are intended for fixation of fractures of different types: of the shaft, open and closed shaft fractures; and malunion and non-unions of the Femur, Tibia, and Humerus.
Device Story
The Ins Hilden Arzzt Nail Systems are orthopedic intramedullary fixation devices used for stabilizing long bone fractures. The system comprises solid or cannulated titanium alloy (Ti6Al4V) nails, end caps, and locking screws. During surgery, the nail is inserted into the medullary canal of the femur, tibia, or humerus to immobilize the fracture site. The nails feature multiple orifices for proximal and distal locking screws to secure the implant within the bone. The device is intended for use in hospital settings by surgeons. By providing mechanical stability to the bone shaft, the system facilitates fracture healing and addresses malunions or non-unions. The design varies in geometry, curvature, and length to accommodate specific anatomical requirements of the target long bones.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, specifically engineering analyses comparing static bending and static torsional yield of the subject devices against predicate devices. The devices met requirements of ASTM F-136-13, ASTM F-983-86, and ISO 5832-3:1996.
Technological Characteristics
Constructed from Ti6Al4V titanium alloy (ASTM F-136-13, ISO 5832-3:1996). Intramedullary fixation rod system with solid/cannulated nail designs, end caps, and self-tapping locking screws. Features include distal and proximal locking orifices. Permanent marking per ASTM F-983-86. Single-use, sterile-capable implants.
Indications for Use
Indicated for adult men and women requiring fixation of open and closed shaft fractures, malunions, and non-unions of the femur, tibia, and humerus.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
ARZZT 3.5/4.5 Small and Large Fragment System (K162507)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of human services. To the right of the symbol is the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG" in a larger, bold blue font, and the word "ADMINISTRATION" in a smaller, regular blue font.
Truemed Group LLC Nina Galeana Rodriguez Coordinator 2002 Timberloch Place Suite 200 The Woodlands, Texas 77380
January 16, 2018
Re: K172189
Trade/Device Name: Ins Hilden Humeral Arzzt, Ins Hilden Cannulated Tibial Arzzt, Ins Hilden Femoral Arzzt & Ins Hilden Retrograde Femoral Arzzt Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: HSB Dated: December 12, 2017 Received: December 14, 2017
Dear Ms. Nina Galeana Rodriguez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image contains the logo for TrueMed Group. The logo consists of a teal-colored abstract symbol on the left, resembling a stylized person within a circle. To the right of the symbol, the word "TRUEMED" is written in a bold, dark gray font, with "GROUP" written in teal below it. A horizontal line extends from the word "GROUP" to the right.
2002 Timberloch Pl Suite 200 The Woodlands, Texas 77380
## Indications for Use
510(k) Number (if known): K172189
## Device Name:
Ins Hilden Humeral Arzzt, Ins Hilden Cannulated Tibial Arzzt, Ins Hilden Femoral Arzzt & Ins Hilden Retrograde Femoral Arzzt.
## Indications for use:
Ins Hilden Arzzt Nail Systems, consisting of intramedullary solid and cannulated nails, end cap and locking screws, are intended for fixation of fractures of different types: of the shaft, open and closed shaft fractures; and malunion and non-unions of the Femur, Tibia, and Humerus.
Prescription Use: ____________________________________________________________________________________________________________________________________________________________
AND/OR
Over-The-Counter Use: ________________________________________________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
## Premarket Notification 510(k) Summary
| 1. | Submitter's Name: | Trumed Group LLC |
|-----|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. | Contact Person: | Jorge Trujillo Zavala |
| | | 2002 Timberloch Pl Suite 200 |
| | | The Woodlands, Texas 77380 |
| | | Phone 832 442 2310 |
| 3. | Date Prepared: | June 5, 2017 |
| 4. | Device Name: | Ins Hilden Humeral Arzzt, Ins Hilden Cannulated Tibial Arzzt, Ins Hilden<br>Femoral Arzzt & Ins Hilden Retrograde Femoral Arzzt. |
| 5. | Common Name: | Rod, Fixation, Intramedullary and Accessories |
| 6. | Classification Name: | Rod, Fixation, Intramedullary and Accessories per 21 CFR<br>section 888.3020 |
| 7. | Product Codes: | HSB |
| 8. | Devices Classification: | Class II |
| 9. | Regulation Numbers: | 21 CFR 888.3020 |
| 10. | Predicate Devices: | Ins Hilden Tibial Arzzt K133166<br>DEPUY ACE Antegrade Retrograde Humeral Nail System K033878<br>DEPUY ACE VERSANAIL TIBIAL NAIL K032097<br>T2 Femoral Nail (K010801)<br>ARZZT 3.5/4.5 Small and Large Fragment System (K162507)<br>Devices Description:<br>I. The Ins Hilden Humeral Arzzt is a single use system that consist the<br>following components: |
- i. Nail, is solid and cannulated, with five orifices, two for distal locking and three for the proximal locking. One of the orifices on the proximal locking is for static locking and one for dynamic
{4}------------------------------------------------
locking. Also the nail is milled on the distal end and flat at the proximal end with a threaded insertion/extraction hole.
- ii. End caps.
- iii. Locking Screws with conical head and self-tapping on the head and blunt.
The system is manufactured from titanium alloy Ti6Al4V, a medical-grade material that is recognized as generally accepted for the manufacture of implants for their level of biocompatibility with surrounding tissues. It is intended to be used in adult men and women, and implanted during surgeries performed in hospitals.
II. The Ins Hilden Femoral Arzzt is a single use system that consist the following components:
- i. Nail, is solid and cannulated with four orifices, two for distal locking and two for the proximal locking. One of the orifices on the proximal locking is for static locking and one for dynamic locking. Also the nail is milled on the distal end and flat at the proximal end with a threaded insertion/extractionhole.
- End caps. ii.
- iii. Locking Screws with conical head and self-tapping on the head and blunt.
The system is manufactured from titanium alloy Ti6Al4V, a medicalgrade material that is recognized as generally accepted for the manufacture of implants for their level of biocompatibility with surrounding tissues. It is intended to be used in adult men and women, and implanted during surgeries performed in hospitals.
III. The Ins Hilden Femoral Retrograde Arzzt is a single use system that consist the following components:
{5}------------------------------------------------
- i. Nail, is solid and cannulated, with five orifices, two for distal locking and three for the proximal locking. All three proximal orifices for static locking. Also the nail is milled on the distal end and flat at the proximal end with a threaded insertion/extraction hole.
- ii. End caps.
- Locking Screws with conical head and self-tapping on the head iii. and blunt.
IV. The Ins Hilden Cannulated Tibial Arzzt is a single use system that consist the following components:
- i. Nail, is cannulated, with four orifices, two for distal locking and four for the proximal locking. All three proximal orifices for static locking. Also the nail is milled on the distal end and flat at the proximal end with a threaded insertion/extraction hole.
- End caps. ii.
- iii. Locking Screws with conical head and self-tapping on the head and blunt.
The system is manufactured from titanium alloy Ti6Al4V, a medicalgrade material that is recognized as generally accepted for the manufacture of implants for their level of biocompatibility with surrounding tissues. It is intended to be used in adult men and women, and implanted during surgeries performed in hospitals.
- 11. Intended Uses: Ins Hilden Arzzt Nail Systems, consisting of intramedullary solid and cannulated nails, end cap and locking screws, are intended for fixation of fractures of different types: of the shaft; open and closed shaft fractures; and malunion and non-unions of the Femur, Tibia, and Humerus.
- 12. Analysis Performed: We performed engineering analyses comparing the static bending and
{6}------------------------------------------------
static torsional yield of the predicate devices with the subject devices confirming substantial equivalence within them.
| 13. Standards: | · ASTM F-136-13 Standard Specification for Wrought Titanium-<br>6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical<br>Implant Applications.<br>• ASTM F-983-86 Standard Practice for Permanent Marking of<br>Orthopedic Implant Components<br>• ISO 5832-3:1996. Implants for surgery - Metallic materials - Part 3:<br>Wrought titanium 6-aluminium 4-vanadium alloy<br>The Ins Hilden Nails Arzzt Systems met the requirements of the above<br>standards. |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 15. Substantial Equivalence: | Comparing the INS HILDEN HUMERAL ARZZT, INS HILDEN CANNULATED<br>TIBIAL ARZZT, INS HILDEN FEMORAL ARZZT, and the INS HILDEN<br>FEMORAL RETROGRADE ARZZT systems with the PREDICATE DEVICES<br>(K133166, K033878, K032097) which have the device names in order<br>shown INS HILDEN TIBIAL ARZZT, DEPUY ACE Antegrade Retrograde<br>Humeral Nail System, T2 Femoral Nail (K010801) and ARZZT 3.5/4.5<br>Small and Large Fragment System (K162507) we found the following<br>conclusion: |
## SUBSTANTIAL EQUIVALENCE CONCLUSION
## SIMILARITIES
Stabilization of fractures, osteotomies and non-unions of long bones. Intended Uses:
Once the nail is inside the bone, the fixation of the fracture is the same, since the limb is Fixation: immobilized.
Types of fractures: Closed and open fractures type A, B and C from AO classification.
- O Target Population: Men and Women in adulthood.
- O Material: Same Titanium alloy Ti6Al4V
- Implant technique: All nails use the same principle of attaching to a O
{7}------------------------------------------------
modular assembly to insert the nail.
#### PERFORMANCE
All systems are intended for fixation of fractures, of long bones, closed and open fractures type A, B and C from AO classification, with delay or without consolidation and bones osteotomies and non-unions.
With similar clinical, the patients had no complications with either system during the healing process, after the reduction and fixation of the fracture with the subject devices, we can conclude substantial equivalence between the predicate K133166 Ins Hilden Tibial Arzzt and the subject devices.
According to the engineering analysis, the subject devices are stronger when applying the same force than the predicate devices K033878, K010801 which has the device name in the same order as shown, DEPUY ACE Antegrade Retrograde Humeral Nail System, DEPUY ACE VERSANAIL TIBIAL NAIL and T2 Femoral Nail. For this reason, when patients are recovering, the nail is strong enough to support them as the predicate devices do.
#### BIOCOMPATIBILITY
The system is made of TAV (Ti6Al4V), titanium alloy implant medical grade, which is considered as a biomaterial and is according to the ISO 5832- 03: Implants for surgery - Metallic materials - wrought titanium 6- aluminum 4-vanadium alloy.
There is no potential impact of the manufacturing processes when using titanium alloy Ti6Al4V on the biocompatibility of the final device according FDA submission K133166. It provides scientific information to demonstrate that the formulation change does not alter the chemical or physical properties of the medical device in its final finished form, and therefore, results from the Ins Hilden Tibial Arzzt PREDICATE DEVICE (K133166) can be applied to the medical devices in its final finished form.
There is no potential impact of the manufacturing processes when using titanium alloy Stainless Steel AISI 304 on the biocompatibility of the final device according FDA submission K162507. It provides scientific information to demonstrate that the formulation change does not alter the chemical or physical properties of the medical device in its final finished form, and therefore, results from the ARZZT 3.5/4.5 Small and Large Fragment System can be applied to the medical device in its final finished form, since both predicate and subject devices manufactured from the same material with the same manufacturing process with similar body contact and duration to each of the patient contacting instruments and the titanium screw.
{8}------------------------------------------------
## DIFFERENCES
Design: Each design has slightly different geometry, with slight differences in angles, curvatures, and different lengths to fit the different long bones. These differences do not affect the final result, since each nail fits a specific bone, helping the stabilization and healing of the fracture.
## REASONS WHY EACH DIFFERENCE DOES OR DOES NOT ADD NEW OR INCREASED RISKS AND COMPLICATIONS
## Design
The design differences do not increase risks and complications. Since the systems designed to fit specific bones, so as the predicate devices are designed to fit a specific bone. The performance of all, is according.
## POTENTIAL BENEFITS
Each system is designed to fit the specific type of bone.
## FINAL CONCLUSION
Based on the technological properties of the subject devices as compared to the predicate devices, we understand that no new questions of safety and effectiveness have been raised, and that the subject devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.