CapSure® PS System

K172175 · Spine Wave, Inc. · NKB · Aug 16, 2017 · Orthopedic

Device Facts

Record IDK172175
Device NameCapSure® PS System
ApplicantSpine Wave, Inc.
Product CodeNKB · Orthopedic
Decision DateAug 16, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The CapSure® PS System is a non-cervical spinal fixation system intended for posterior pedicle screw fixation (T1-S2/ilium) in skeletally mature patients. The CapSure® PS System is indicated for degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis). The CapSure® PS System is also indicated for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebral joint, having fusions with autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3-S2/ilium), and for whom the device is intended to be removed after solid fusion is attained.

Device Story

CapSure® PS System is a non-sterile, single-use spinal fixation system; components include titanium alloy screws, connectors, and titanium/cobalt-chrome rods. System creates rigid construct to stabilize non-cervical spine during vertebral fusion; intended for removal post-fusion. Used by surgeons in clinical settings. Subject submission adds 4.75 mm diameter cobalt-chrome straight rods and dual-diameter titanium rods to existing system. Device provides mechanical stabilization to facilitate bone fusion in skeletally mature patients.

Clinical Evidence

Bench testing only. Dynamic axial compression bend testing was performed on worst-case samples of the modified implants in accordance with ASTM F1717. Results demonstrated that the modified implants do not represent a new worst case for the system.

Technological Characteristics

Non-sterile, single-use spinal fixation system. Materials: titanium alloy and cobalt-chrome. Components: pedicle screws, connectors, and rods (including 4.75 mm cobalt-chrome straight and dual-diameter titanium variants). Mechanical stabilization principle. Testing performed per ASTM F1717.

Indications for Use

Indicated for skeletally mature patients requiring posterior pedicle screw fixation (T1-S2/ilium) for degenerative spondylolisthesis with neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, or pseudoarthrosis. Also indicated for severe spondylolisthesis (Grades 3-4) at L5-S1 using autogenous bone graft, with fixation at L3-S2/ilium, intended for removal after fusion.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Spine Wave, Inc. Amy Noccioli Regulatory Affairs Specialist 3 Enterprise Drive. Suite 210 Shelton, Connecticut 06484 August 16, 2017 Re: K172175 Trade/Device Name: CapSure® PS System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbar Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: July 18, 2017 Received: July 19, 2017 Dear Amy Noccioli: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device {1}------------------------------------------------ related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) #### K172175 Device Name CapSure® PS System Indications for Use (Describe) The CapSure® PS System is a non-cervical spinal fixation system intended for posterior pedicle screw fixation (T1-S2/ilium) in skeletally mature patients. The CapSure® PS System is indicated for degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis). The CapSure® PS System is also indicated for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebral joint, having fusions with autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3-S2/ilium), and for whom the device is intended to be removed after solid fusion is attained. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the words "Spine Wave" in a bold, sans-serif font. The words are arranged horizontally, with "Spine" on the left and "Wave" on the right. A blue wave design is located under the word "Wave" and extends to the left. # 510(k) Summary CapSure® PS System #### 1. Submitter Information | Submitter: | Spine Wave, Inc. | |----------------|----------------------------------------------------------| | Address: | Three Enterprise Drive<br>Suite 210<br>Shelton, CT 06484 | | Telephone: | 203-712-1842 | | Telefax: | 203-944-9493 | | Contact: | Amy Noccioli | | Date Prepared: | August 11, 2017 | ## 2. Device Information | Trade Name: | CapSure® PS System | |----------------------|-----------------------------------------| | Common Name: | Pedicle Screw Spinal System | | Classification: | Class II per 21 CFR 888.3070 | | Classification Name: | Thoracolumbosacral Pedicle Screw System | | Product Code: | NKB | #### 3. Purpose of Submission The purpose of this submission is to gain clearance for the 4.75 mm diameter cobalt-chrome straight rod and dual-diameter titanium rod additions to the previously cleared CapSure® PS System. #### 4. Predicate Device Information The CapSure® PS System described in this submission is substantially equivalent to the following predicate: | Predicate Device | Manufacturer | 510(k) No. | |--------------------|------------------|------------| | CapSure® PS System | Spine Wave, Inc. | K132154 | #### 5. Device Description The CapSure® PS System consists of a selection of non-sterile, single-use, titanium alloy screw and connector components and titanium alloy and cobaltchrome rod components that are assembled to create a rigid spinal construct. The components of the CapSure® PS System are attached to the non-cervical spine of skeletally mature patients to stabilize the spine during fusion of vertebral bodies and are intended to be removed after spinal fusion is achieved. {4}------------------------------------------------ #### 6. Indications for Use The CapSure® PS System is a non-cervical spinal fixation system intended for posterior pedicle screw fixation (T1-S2/ilium) in skeletally mature patients. The CapSure® PS System is indicated for degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis). The CapSure® PS System is also indicated for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebral joint, having fusions with autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3-S2/ilium), and for whom the device is intended to be removed after solid fusion is attained. ## 7. Comparison of Technological Characteristics The subject CapSure® PS System has technological characteristics similar to the predicate device, including intended use and indications for use, performance, design, and material composition. The only difference between the subject device and the predicate device is the addition of 4.75 mm diameter cobalt-chrome straight rods and dual-diameter titanium rods. # 8. Performance Data Spine Wave, Inc. performed dynamic axial compression bend testing on the worst-case samples of the modified implants in accordance with ASTM F1717. The results of this testing show that the modified implants do not represent a new worst case for the system and are therefore substantially equivalent to the predicate device. #### 9. Conclusion The indications for use, technological characteristics, and comparison to the predicate show that the subject CapSure® PS System is substantially equivalent to the predicate device identified in this submission, and does not present any new issues of safety or effectiveness.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...