The Hysteroscopy System is indicated for rapid transcervical aspiration of the uterine cavity during the first trimester of pregnancy. It has an additional feature of visualization.
Device Story
Device facilitates vacuum aspiration of uterine cavity during first trimester pregnancy; provides visualization. System comprises EO-sterilized single-use cannula and reusable image processing unit. Cannula features internal channel for vacuum/negative pressure (up to 500 mmHg) and distal optical system (CMOS sensor, lens, 40 mW LED). Lens cover protects electronics from fluids. Used in clinical settings by physicians. Physician views uterine cavity via image processing unit to guide aspiration procedure. Visualization aids procedure accuracy; vacuum aspiration enables pregnancy termination or uterine evacuation. Patient benefits from combined visualization and aspiration in single procedure.
Clinical Evidence
Bench testing only. Validated via ISO 11135-1:2007 (sterilization), ISO 10993-5/10 (biocompatibility), IEC 60601-1-2:2014 (EMC), AAMI/ANSI ES60601-1 and IEC 60601-2-18 (electrical safety), and ISO 8600 series (image quality). Mechanical testing included vacuum integrity, tensile strength, and leak testing. No clinical data provided.
Technological Characteristics
Materials: medical grade stainless steel, polycarbonate, medical glue, medical nylon. Sensing: CMOS image sensor with 40 mW LED illumination. Connectivity: electrical/data connection and vacuum connection. Sterilization: Ethylene oxide (ISO 11135-1:2007). Form factor: single-use cannula in five sizes. Standards: ISO 10993, IEC 60601-1-2, AAMI/ANSI ES60601-1, IEC 60601-2-18, ISO 8600.
Indications for Use
Indicated for rapid transcervical aspiration of the uterine cavity in patients during the first trimester of pregnancy.
Regulatory Classification
Identification
A vacuum abortion system is a device designed to aspirate transcervically the products of conception or menstruation from the uterus by using a cannula connected to a suction source. This device is used for pregnancy termination or menstrual regulation. This type of device may include aspiration cannula, vacuum source, and vacuum controller.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is the Department of Health and Human Services logo, and on the right, there is the FDA logo. The FDA logo is in blue and consists of the letters "FDA" followed by the words "U.S. Food & Drug Administration".
April 9, 2018
Chongqing Jinshan Science & Technology (Group) Co., Ltd. Qing Xu Regulatory Manager Yubei District No. 18 Ningshang Avenue Jinshan International Industrial City Chongqing, 400000 China
Re: K172151 Trade/Device Name: Hysteroscopy System Regulation Number: 21 CFR§ 884.5070 Regulation Name: Vacuum Abortion System Regulatory Class: II Product Code: HGH Dated: March 6, 2018 Received: March 9, 2018
Dear Oing Xu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows the name "Benjamin R. Fisher -S" in a large, bold font. The text is black and appears to be on a white background. The letters are clearly legible and well-defined. The overall impression is clean and professional.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K172151
Device Name Hysteroscopy System
Indications for Use (Describe)
The Hysteroscopy System is indicated for rapid transcervical aspiration of the uterine cavity during the first trimester of pregnancy. It has an additional feature of visualization.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image contains the logo for JINSHAN. The logo consists of a red circular shape with a white triangle inside, resembling a mountain. To the right of the logo is the word "JINSHAN" in black, bold letters. The overall design is simple and corporate.
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
(As Required by 21 CFR 807.92)
# 1. Date Prepared [21 CFR 807.92(a)(1)] 04/06/2018
### 2. Submitter's Information [21 CFR 807.92(a)(1)]
| Company Name: | Chongqing Jinshan Science & Technology (Group) Co., Ltd. |
|------------------|--------------------------------------------------------------------------------------------|
| Company Address: | Yubei District No.18 Ningshang Avenue Jinshan<br>International Industrial City, Chongqing, |
| Contact Person: | Qing Xu |
| Phone: | 023 - 86098111 |
| Fax: | 023 - 86098777 |
| Email: | xuq@jinshangroup.net |
#### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
| Trade Name: | Hysteroscopy System |
|--------------------|---------------------------------|
| Common Name: | Vacuum Abortion System |
| Product Code: | HGH (cannula, suction, uterine) |
| Regulation Name: | Vacuum Abortion System |
| Regulation Number: | 21CFR 884.5070 |
| Device Class: | II |
#### 4. Identification of Predicate Devices(s) [21 CFR 807.92(a)(3)[
The identification of predicates within this submission is as follow: Manufacturer: DONGGUAN MICROVIEW MEDICAL TECHNOLOGY CO., LTD. Trade Name: Disposable Endoscopic Cannula K140213 FDA 510(k) #:
The predicate device has not been subject to a design related recall.
### 5. Description of the Device [21 CFR 807.92(a)(4)]
{4}------------------------------------------------
# Chongqing Jinshan Science & Technology (Group) Co., Ltd.
Image /page/4/Picture/2 description: The image contains the logo for Jinshan. The logo consists of a red circle with a white triangle inside, followed by the word "JINSHAN" in black, sans-serif font. The triangle inside the circle is pointing upwards.
The Hysteroscopy System is designed to visualize the uterus and to facilitate physicians in the procedure of vacuum aspiration of the uterine cavity during the first trimester of pregnancy. Components of the Hysteroscopy System include an EOsterilized single-use cannula and reusable image processing unit. The Hysteroscopy System is available in five sizes that differ in length and width.
Each cannula includes an internal channel to draw vacuum or negative pressure up to 500 mmHg. The optical system, which includes a CMOS image sensor, lens, and 40 mW LED, is located at the distal end of the cannula. A lens cover shields the electronics from bodily fluids. The Hysteroscopy System has two connectors: one is an electrical/data connection and the second connection is for vacuum. The cannula is not indicated for use with any other accessories. The patient-contacting materials are medical grade stainless steel, polycarbonate, medical glue, and medical nylon.
# 6. Indications for Use [21 CFR 807.92(a)(5)]
The Hysteroscopy System is indicated for rapid transcervical aspiration of the uterine cavity during the first trimester of pregnancy. It has an additional feature of visualization.
The subject and predicate device have the same intended use.
# 7. Technological Characteristics [21 CFR 807.92(a)(6)]
The Hysteroscopy System has different technological characteristics in comparison to the named predicate, primarily with respect to the optics and image quality, internal device construction, and materials used for device construction. However, these differences in technological characteristics do not raise different questions of safety and effectiveness.
# 8. Summary of Performance Data [21 CFR 807.92(b)]
The following performance data were provided to support substantial equivalence:
- Ethylene oxide sterilization validation per ISO 11135-1:2007
- Shelf Life including packaging integrity following simulated shipping and mechanical performance testing as described below
- . Biocompatibility including cytotoxicity per ISO 10993-5:2009. sensitization per ISO 10993-10:2010 and irritation per ISO 10993-10:2010
- Software documentation per the FDA Software Guidance
- Electromagnetic compatibility testing per IEC 60601-1-2:2014 ●
{5}------------------------------------------------
### Chongqing Jinshan Science & Technology (Group) Co., Ltd.
Image /page/5/Picture/2 description: The image contains the logo for Jinshan. The logo consists of a red circle with a white triangle inside, followed by the word "JINSHAN" in black, sans-serif font. The triangle inside the circle is pointing upwards and has a small notch at the top.
- Electrical safety testing per AAMI / ANSI ES60601-1:2005/(R)2012 and ● C1:2009/(R)2012 and A2:2010/(R)2012 and IEC 60601-2-18: 2009
- Mechanical performance testing including dimensional analysis, tensile ● strength, vacuum integrity, vacuum performance, and leak testing
- . Image quality testing including field of view, direction of view, illumination, and resolution per the ISO 8600 series of standards
The protocol and results of all performance testing were acceptable.
# 9. Conclusion [21 CFR 807.92(b)(3)]
The performance data demonstrate that the Hysteroscopy System is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.