Etest Ceftazidime/Avibactam (CZA) (0.016 - 256 ug/mL)

K172150 · bioMerieux, Inc. · JWY · Sep 15, 2017 · Microbiology

Device Facts

Record IDK172150
Device NameEtest Ceftazidime/Avibactam (CZA) (0.016 - 256 ug/mL)
ApplicantbioMerieux, Inc.
Product CodeJWY · Microbiology
Decision DateSep 15, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K172150 · Sep 15, 2017Etest Ceftazidime/Avibactam (CZA) (0.016 - 256 ug/mL)bioMerieux, Inc.Clinical isolates (fresh and stock)Clinical isolates were used in external evaluations to compare the performance of the Etest CZA against the CLSI broth microdilution reference method.Clinical isolates; External evaluation; Antimicrobial susceptibility testing

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
External evaluations of Etest Ceftazidime/Avibactam; Comparative performance evaluation against reference methodGram-negative aerobic bacteria (Enterobacteriaceae and Pseudomonas aeruginosa)CLSI broth microdilution reference methodEssential Agreement (EA) and Category Agreement (CA)

Indications for Use

Etest is a quantitative technique for determination of antimicrobial susceptibility of both non-fastidious Gram-negative and Gram-positive aerobic bacteria such as Enterobacteriaceae, Pseudomonas, Staphylococcus, and Enterococcus species and fastidious bacteria, such as anaerobes, N. gonorrhoeae, S. pneumoniae, Streptococcus and Haemophilus species. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC), in µg/mL, of different antimicrobial agents against microorganisms as tested on agar media using overnight incubation. Ceftazidime/avibactam has been shown to be active against the Gram negative aerobic microorganisms listed below, according to the FDA label for this antimicrobial agent. Etest CZA can be used to determine the MIC of ceftazidime/avibactam against the microorganisms listed below: Active both in vitro and in clinical infections: Citrobacter freundii, Enterobacter cloacae, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus mirabilis, Pseudomonas aeruginosa. The following in vitro data are available, but clinical significance is unknown: Citrobacter koseri, Enterobacter aerogenes, Morganella morganii, Providencia rettgeri, Providencia stuartii, Serratia marcescens.

Device Story

Etest is a thin, inert, non-porous plastic strip with a predefined antibiotic gradient on one side and an MIC reading scale on the other. Used in clinical microbiology laboratories; applied to inoculated agar surfaces. Antibiotic gradient transfers into agar, forming a stable, continuous concentration gradient. Following overnight incubation, bacterial growth forms a symmetrical inhibition ellipse. The MIC value is determined by the intersection of the ellipse with the strip scale. Provides quantitative susceptibility data to guide antimicrobial therapy decisions for Gram-negative infections.

Clinical Evidence

Bench testing only. Performance evaluated using 1,153 clinical isolates and 86 challenge isolates across three sites. Comparison against CLSI broth microdilution reference method showed 99.1% Essential Agreement (EA) and 99.6% Category Agreement (CA) for Enterobacteriaceae, and 99.3% EA and 99.3% CA for P. aeruginosa. A trend toward higher MIC readings was observed for Enterobacteriaceae. Very major error rate for Enterobacteriaceae was 4.4% (adjusted to 0% due to EA). Limitations noted for M. morganii and C. koseri due to lack of resistant strains or specific interpretive discrepancies.

Technological Characteristics

Thin, inert, non-porous plastic strip; predefined antibiotic gradient; manual reading of MIC scale; agar diffusion principle; overnight incubation.

Indications for Use

Indicated for determination of antimicrobial susceptibility of Gram-negative aerobic bacteria (Enterobacteriaceae, Pseudomonas aeruginosa) to Ceftazidime/Avibactam. Used for clinical isolates and challenge strains.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Predicate Devices

Submission Summary (Full Text)

{1} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K172150 B. Purpose for Submission: To obtain a substantial equivalence determination for ceftazidime/avibactam at concentrations of 0.016 – 256 µg/mL for susceptibility testing of gram negative aerobic microorganisms with Etest. C. Measurand: Ceftazidime/avibactam 0.016 – 256 µg/mL. The avibactam concentration is fixed at 4 µg/mL in this combination. D. Type of Test: Quantitative AST growth-based detection E. Applicant: bioMérieux, Inc. F. Proprietary and Established Names: Etest Ceftazidime/Avibactam (CZA) (0.016 – 256 µg/mL) G. Regulatory Information: 1. Regulation section: 866.1640 Antimicrobial Susceptibility Test Powder 2. Classification: II 3. Product code: JWY – Manual Antimicrobial Test Systems {2} 4. Panel: 83 – Microbiology H. Intended Use: 1. Intended use(s): Etest is a quantitative technique for determination of antimicrobial susceptibility of both non-fastidious Gram-negative and Gram-positive aerobic bacteria such as Enterobacteriaceae, Pseudomonas, Staphylococcus, and Enterococcus species and fastidious bacteria, such as anaerobes, N. gonorrhoeae, S. pneumoniae, Streptococcus and Haemophilus species. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC), in µg/mL, of different antimicrobial agents against microorganisms as tested on agar media using overnight incubation. Ceftazidime/avibactam has been shown to be active against the Gram negative aerobic microorganisms listed below, according to the FDA label for this antimicrobial agent. Etest CZA can be used to determine the MIC of ceftazidime/avibactam against the microorganisms listed below: Active both in vitro and in clinical infections: - Citrobacter freundii - Enterobacter cloacae - Escherichia coli - Klebsiella oxytoca - Klebsiella pneumoniae - Proteus mirabilis - Pseudomonas aeruginosa The following in vitro data are available, but clinical significance is unknown: - Citrobacter koseri - Enterobacter aerogenes - Morganella morganii - Providencia rettgeri - Providencia stuartii - Serratia marcescens 2. Indication(s) for use: Same as the Intended Use 3. Special conditions for use statement(s): For prescription use only 2 {3} 3 Limitations: Due to the lack of an intermediate interpretive category for ceftazidime/avibactam, testing of Morganella morganii with this drug has resulted in very major discrepancies for isolates that are otherwise within essential agreement of the reference method. Testing should be repeated using an alternative testing/reference method prior to reporting results for M. morganii when the Etest MIC is 8 µg/mL. The ability of Etest to detect resistance with the following microorganisms is unknown because resistant strains were not available at the time of comparative testing. If such a strain is observed, it should be submitted to a reference laboratory for further testing Ceftazidime/Avibactam: Citrobacter koseri 4. Special instrument requirements: N/A I. Device Description: Etest consists of a thin, inert and non-porous plastic strip 5mm wide and 60 mm long. One side of the strip carries a two letter code designating the identity of the antibiotic and is calibrated with MIC values in terms of µg/mL. On the reverse, a predefined exponential gradient of the dried and stabilized antibiotic covers a continuous concentration range across 15 two-fold dilutions of a conventional MIC method. J. Substantial Equivalence Information: 1. Predicate device name(s): Etest Ceftolozane/Tazobactam 2. Predicate 510(k) number(s): K170670 3. Comparison with predicate: {4} Table 1. Comparison with the Predicate Device | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | | K172150 Etest Ceftazidime/Avibactam | K170670 Etest Ceftolozane/Tazobactam | | Intended Use | Etest is a quantitative technique for determination of antimicrobial susceptibility of both non-fastidious Gram-negative and Gram-positive aerobic bacteria such as Enterobacteriaceae, Pseudomonas, Staphylococcus, and Enterococcus species and fastidious bacteria, such as anaerobes, N. gonorrhoeae, S. pneumoniae, Streptococcus and Haemophilus species. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC), in μg/mL, of different antimicrobial agents against microorganism as tested on agar media using overnight incubation. | Same | | Antimicrobial Concentration Range | 0.016 – 256 μg/mL (Avibactam: fixed at 4 μg/mL) | Same (Tazobactam: fixed at 4 μg/mL) | | Test Design | A predefined exponential gradient of the dried and stabilized antibiotic covers a continuous concentration range across 15 two-fold dilutions of a conventional MIC method. | Same | | Inoculation | Isolated colonies from culture | Same | | Incubation | 35°±2°C for 16–20 hours | Same | | Result | MIC | Same | {5} | Differences | | --- | | Item | Device | Predicate | | Claimed organisms | Citrobacter freundii | Citrobacter koseri | | Citrobacter koseri | Enterobacter cloacae | | Enterobacter aerogenes | Escherichia coli | | Enterobacter cloacae | Klebsiella oxytoca | | Escherichia coli | Klebsiella pneumoniae | | Klebsiella oxytoca | Morganella morganii | | Klebsiella pneumoniae | Proteus mirabilis | | Morganella morganii | Proteus vulgaris | | Proteus mirabilis | Providencia rettgeri | | Providencia rettgeri | Providencia stuartii | | Providencia stuartii | Pseudomonas aeruginosa | | Pseudomonas aeruginosa | Serratia liquifaciens | | Serratia marcescens | Serratia marcescens | | Antimicrobial Agent | Ceftazidime/Avibactam | Ceftolozane/Tazobactam | ## K. Standard/Guidance Document Referenced (if applicable): Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA CLSI M07-A10, Method for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard, January 2015. CLSI M100-S26, Performance Standards for Antimicrobial Susceptibility Testing; Volume 36, No. 1, January 2016. ## L. Test Principle: The Etest consists of a thin, inert, nonporous plastic strip that is used to determine the antimicrobial susceptibility of bacteria. One side of the strip carries the minimum inhibitory concentration (MIC) reading scale expressed in $\mu \mathrm{g} / \mathrm{mL}$. The other side of the strip contains a predefined continuous gradient of antibiotic concentrations. When the strip is applied to an inoculated agar surface, the preformed antibiotic gradient immediately transfers into the agar matrix, then forming a stable, continuous and exponential gradient of antibiotic concentrations directly underneath the strip. Bacteria growth becomes visible during incubation, and a symmetrical inhibition ellipse centered along the strip appears. After incubation, the MIC value is read from the scale in terms of $\mu \mathrm{g} / \mathrm{mL}$ at complete inhibition of bacterial growth, where the pointed end of the ellipse intersects the strip. Since Etest generates MIC values which fall between two-fold dilutions for interpretation, the MIC value read must be recorded to the next two-fold dilution. The MIC gradient on Etest Ceftazidime/Avibactam ranges from 0.016 to $256~\mu \mathrm{g / mL}$. The {6} avibactam concentration is fixed at $4\ \mu\mathrm{g/mL}$ in this combination. ## M. Performance Characteristics: ### 1. Analytical performance: #### a. Precision/Reproducibility: A reproducibility study was conducted at three external sites using 25 isolates of gram negative bacilli that were consistent with the intended use. The isolates tested included *Citrobacter freundii* (two isolates), *Citrobacter koseri* (two isolates), *Enterobacter aerogenes* (two isolates), *Enterobacter cloacae* (two isolates), *Escherichia coli* (five isolates), *Klebsiella oxytoca* (three isolates), *Klebsiella pneumoniae* (three isolates) and *Pseudomonas aeruginosa* (six isolates). All results were within $\pm 1$ doubling dilution of the mode MIC value for ceftazidime/avibactam. Eleven results were on scale. The reproducibility was acceptable at $100.0\%$. #### b. Linearity/assay reportable range: Not applicable #### c. Traceability, Stability, Expected values (controls, calibrators, or methods): **Inoculum Density Check.** Inoculum density checks were performed for all quality control and reproducibility organism suspensions and for $10\%$ of the suspensions prepared for susceptibility testing of the fresh clinical isolates. The mean inoculum density was acceptable at $4.40 \times 10^{5}$ CFU/mL. **Purity Check.** All clinical, challenge and reproducibility test suspensions were subcultured to assure purity. **Growth or Device Failure.** There were no growth or device failures during the course of the study. **Quality Control Testing.** Organisms recommended by both FDA and the CLSI were tested with ceftazidime/avibactam at three sites. According to the CLSI M100 S27 document, *K. pneumoniae* ATCC 700603 must be used for routine QC testing of ceftazidime/avibactam. The QC organisms tested included: *E. coli* ATCC 25922, *E. coli* ATCC 35218, *K. pneumoniae* ATCC 700603 and *P. aeruginosa* ATCC 27853. These QC strains were tested a minimum of 20 times per site by both the Etest and the reference method. One quality control strain (*Staphylococcus aureus* ATCC 29213) was tested with the reference method only to provide additional control for this method; $100\%$ of the results obtained with *S. aureus* ATCC 29213 were in the acceptable range. The results demonstrate that the ceftazidime/avibactam Etest can 6 {7} produce quality control results in the recommended range $&gt;95\%$ of the time. See Table 2 below for a summary of the QC results. Per the recommendations in the CLSI document M100, additional testing was performed using the reference method with $K$ pneumoniae 700603 with ceftazidime alone and $K$ pneumoniae ATCC 700603 and $E$ coli ATCC 35218 with ampicillin to assure the integrity of the $\beta$ -lactamase encoding plasmid. The testing demonstrated that the QC strains maintained their resistance profiles throughout the testing as evidenced by obtaining MIC values of $16 - 64~\mu \mathrm{g / mL}$ with ceftazidime for $K$ pneumoniae 700603 and obtaining MIC values of $&gt;128~\mu \mathrm{g / mL}$ with ampicillin for the same organism. MIC values of $&gt;32~\mu \mathrm{g / mL}$ were obtained with ampicillin for $E$ coli ATCC 35218. The sponsor included the following comment in the device labeling: K. pneumoniae ATCC 700603 should be used for routine QC of ceftazidime/avibactam. This QC strain should also be tested against ceftazidime alone to ensure that the plasmid encoding the beta-lactamase has not been lost in this strain. Table 2. Quality Control Data for Ceftazidime/Avibactam | QC organism | Ceftazidime/ Avibactam expected range (μg/mL) | Concentration (μg/mL) | Reference | Etest | | --- | --- | --- | --- | --- | | E. coliATCC 25922 | 0.06 – 0.5 | <0.06 | - | - | | | | 0.06 | 1 | - | | | | 0.125 | 74 | 71 | | | | 0.25 | 10 | 15 | | | | 0.5 | 1 | - | | | | >0.5 | - | - | | | | | | | | P. aeruginosaATCC 27853 | 0.5 - 4 | <0.5 | - | - | | | | 0.5 | - | - | | | | 1 | 14 | 46 | | | | 2 | 71 | 40 | | | | 4 | 1 | - | | | | >4 | - | - | | | | | | | | E. coliATCC 35218 | 0.03 – 0.12 | <0.03 | - | - | | | | 0.03 | 1 | - | | | | 0.06 | 77 | 76 | | | | 0.125 | 8 | 10 | | | | >0.125 | - | - | | | | | | | | K. pneumoniaeATCC 700603 | 0.25 - 2 | <0.25 | - | - | | | | 0.25 | - | - | | | | 0.5 | 50 | 1 | | | | 1 | 35 | 79 | {8} | QC organism | Ceftazidime/Avibactam expected range (μg/mL) | Concentration (μg/mL) | Reference | Etest | | --- | --- | --- | --- | --- | | | | 2 | 1 | 6 | | | | >2 | - | - | d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Results obtained with Etest Ceftazidime/Avibactam were compared to results obtained with the CLSI broth microdilution reference panel. The CLSI panel was prepared and interpreted according to CLSI recommendations outlined in the CLSI Standard: CLSI Document M07-A10, Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard – Tenth Edition, Vol. 35, No. 2; January 2015. The testing conditions for the reference method consisted of the following: - Medium – Cation-adjusted Mueller Hinton Broth with appropriate dilutions of antimicrobial solution added - Inoculum – Direct colony suspension to achieve a suspension equivalent to a 0.5 McFarland standard suspension - Incubation 35 °C in ambient air; 16-20 hours for all organisms Clinical testing was performed at three sites using a total of 1153 clinical isolates (1033 isolates of Enterobacteriaceae and 120 isolates of P. aeruginosa). Of the clinical isolates, 699 were fresh isolates (60.6 %) and 454 were stock isolates (39.4%). Clinical isolates were tested using both Etest and the reference method. A total of 86 challenge isolates were tested at a single site using Etest and the reference method (54 isolates of Enterobacteriaceae and 32 isolates of P. aeruginosa). {9} Combined clinical and challenge isolates included: C. freundii (93 isolates), C. koseri (91 isolates), E. aerogenes (96 isolates), E. cloacae (99 isolates), E. coli (109 isolates), K. oxytoca (102 isolates), K. pneumoniae (107 isolates), M. morganii (65 isolates), P. mirabilis (95 isolates), P. vulgaris (45 isolates), P. rettgeri (57 isolates), P. stuartii (59 isolates), S. marcescens (69 isolates) and P. aeruginosa (152 isolates). For P. aeruginosa, the combined results from clinical and challenge testing demonstrated a combined EA of 99.3% and CA of 99.3% (Table 3). A total of 145 isolates were determined to have evaluable results; the EA of evaluable results was 99.3%. For the Enterobacteriaceae, the combined results from clinical and challenge testing demonstrated a combined EA of 99.1% and CA of 99.6% (Table 3). A total of 1060 isolates were determined to have evaluable results; the EA of evaluable results was 99.2%. Table 3. Performance of Clinical and Challenge Isolates, Ceftazidime/Avibactam Etest. | | Tot | No. EA | EA % | Eval Tot | Eval EA | EA Eval % | No. CA | CA % | No. R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | P. aeruginosa | | | | | | | | | | | | | | Clinical | 120 | 119 | 99.2 | 114 | 113 | 99.1 | 119 | 99.2 | 16 | NA^{a} | 1 | 0 | | Challenge | 32 | 32 | 100.0 | 31 | 31 | 100.0 | 32 | 100.0 | 18 | NA^{a} | 0 | 0 | | Combined | 152 | 151 | 99.3 | 145 | 144 | 99.3 | 151 | 99.3 | 34 | NA^{a} | 1 | 0 | | | | | | | | | | | | | | | | Enterobacteriaceae^{b} | | | | | | | | | | | | | | Clinical | 1033 | 1023 | 99.0 | 1015 | 1006 | 99.1 | 1029 | 99.6 | 33 | NA^{a} | 2 | 2 | | Challenge | 54 | 54 | 100.0 | 45 | 45 | 100.0 | 54 | 100.0 | 12 | NA^{a} | 0 | 0 | | Combined | 1087 | 1077 | 99.1 | 1060 | 1051 | 99.2 | 1083 | 99.6 | 45 | NA^{a} | 2^{c} | 2^{c,d} | a NA, Not Applicable due to a lack of an intermediate breakpoint for ceftazidime/avibactam b Enterobacteriaceae included: Citrobacter freundii, Enterobacter cloacae, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus mirabilis, Citrobacter koseri, Enterobacter aerogenes, Morganella morganii, Providencia rettgeri, Providencia stuartii, and Serratia marcescens c Etest results are within essential agreement of the reference method Due to the lack of an intermediate interpretive category for ceftazidime/avibactam, overall very major error rate for Enterobacteriaceae of 4.4% (2/45) was adjusted to 0% when considering that all error results were within EA of the reference method EA - Essential Agreement (+/- 1 dilution) CA - Category Agreement EVAL - Evaluable isolates R - Resistant isolates min - minor discrepancies maj - major discrepancies vmj - very major discrepancies Essential Agreement (EA) occurs when the result of the reference method and that of Etest Ceftazidime/Avibactam are within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both Etest Ceftazidime/Avibactam and the reference method. Category Agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of Etest Ceftazidime/Avibactam. {10} A total of 79 resistant isolates were evaluated; however, no resistant strains of Citrobacter koseri were evaluated. The sponsor added the following limitation to the device labeling: The ability of Etest to detect resistance with the following microorganisms is unknown because resistant strains were not available at the time of comparative testing. If such a strain is observed, it should be submitted to a reference laboratory for further testing. Ceftazidime/Avibactam: Citrobacter koseri Three clinical isolates (one isolate each of P. stuartii, E. aerogenes and P. aeruginosa) were determined to be susceptible by the reference method but resistant with ceftazidime/avibactam Etest. Both the P. stuartii and E. aerogenes MIC results were within essential agreement of the results obtained with the reference method. For the single isolate of P. aeruginosa Etest results were not within essential agreement with the reference method. However, the major error rates for Enterobacteriaceae and for P. aeruginosa were determined to be acceptable at 0.2% and 0.8%, respectively. Two isolates of M. morganii were determined to be resistant by the reference method but susceptible with Etest Ceftazidime/Avibactam resulting in an overall very major error rate for Enterobacteriaceae of 4.4%. Because there is no intermediate breakpoint for ceftazidime/avibactam and because the Etest MIC values obtained with these two isolates were within essential agreement of the reference method, the overall very major error rate for Enterobacteriaceae was adjusted to 0.0% and considered to be acceptable. The following footnote was added to the performance table in the device labeling: The overall very major error rate for Enterobacteriaceae was 4.4%. The two very major errors were one dilution apart from the reference method and as such fall within essential agreement. Based on the essential agreement and the lack of an intermediate breakpoint, the adjusted very major error rate is 0.0% In addition, because both of the very major errors occurred with M. morganii (for a very major error rate of 100% for this species) the following limitation was added to the device labeling to address the possibility of very major errors when testing M. morganii: Due to the lack of an intermediate interpretive category for ceftazidime/avibactam, testing of Morganella morganii with this drug has resulted in very major discrepancies for isolates that are otherwise within essential agreement of the reference method. Testing should be repeated using an alternative testing/reference method prior to reporting results for M. morganii when the Etest MIC is 8 µg/mL. 10 {11} Enzyme Group Molecular Characterization: Isolates of Enterobacteriaceae and *P. aeruginosa* harboring various molecular mechanisms of resistance noted in the FDA drug label were tested with ceftazidime/avibactam. The following resistance mechanisms were evaluated: - Enterobacteriaceae – ESBL, not specified, CTX-M, TEM, SHV, KPC - *P. aeruginosa* – chromosomal AmpC, OprD loss, KPC, IMP, VIM, other metallo-β-lactamase Enterobacteriaceae isolates harboring AmpC and OXA were not evaluated. The sponsor included the following footnote to the performance table in the device labeling: Enzyme characterization was not available for the following organisms at the time of comparative testing, and therefore the performance of Etest Ceftazidime/Avibactam is unknown for Pseudomonas aeruginosa (up-regulation of MexXY and MexAB) and Enterobacteriaceae (OXA and AmpC). The sponsor also included the following footnote regarding additional resistance mechanisms: Ceftazidime/avibactam is not active against bacteria that produce metallo-beta-lactamase enzymes and may not have activity against gram-negative bacteria that overexpress efflux pumps or have porin mutations. MIC Trending An analysis of trending was conducted using the combined clinical and challenge data for *P. aeruginosa* and for Enterobacteriaceae. This trending calculation takes into account MIC values that are determined to be one or more doubling dilution lower or higher compared to the reference method irrespective of whether the device MIC values are on-scale or not. The data for 152 *P. aeruginosa* results determined to be evaluable for trending analysis showed no evidence of trending. The data for 1061 Enterobacteriaceae results determined to be evaluable for trending analysis is presented in Table 4. 11 {12} 12 Table 4. Trending in Enterobacteriaceae, Clinical and Challenge Isolates | Organism | Difference in MIC as Compared to the CLSI Reference Method | | | | | --- | --- | --- | --- | --- | | | No. results evaluable for trending analysis | No. results ≥1 dilution lower | No. results Exact | No. results ≥1 dilution higher | | Enterobacteriaceae | 1061 | 80 (7.54%)* | 628 (59.19%) | 353 (33.27%)* | *Difference: 25.73%, 95% CI (22.46 – 28.96%) A trend towards higher MIC readings was observed in the overall performance of Enterobacteriaceae compared to the CLSI broth microdilution method which raises concerns for the potential occurrence of major errors. The sponsor included the following footnote to the performance table to address the trending observed for ceftazidime/avibactam: Etest Ceftazidime/Avibactam MIC values tended to be in exact agreement or at least one doubling dilution higher when testing Enterobacteriaceae compared to the reference broth microdilution method. b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): No applicable 4. Clinical cut-off: Not applicable {13} 5. Expected values/Reference range: Table 5. Breakpoints and Interpretive Categories for Ceftazidime/Avibactam (FDA Drug Label) | Organism | FDA Interpretive Criteria for Ceftazidime/Avibactam MIC (μg/mL) | | | | --- | --- | --- | --- | | | S | I | R | | Enterobacteriaceae and P. aeruginosa | ≤ 8/4 | NA | ≥ 16/4 | N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...