K172114 · Zimmer, Inc. · HSB · Aug 10, 2017 · Orthopedic
Device Facts
Record ID
K172114
Device Name
Zimmer Natural Nail System
Applicant
Zimmer, Inc.
Product Code
HSB · Orthopedic
Decision Date
Aug 10, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Zimmer Natural Nail System is intended for temporary fracture fixation and stabilization of the bone. Indications for use of the Greater Trochanter and Piriformis Fossa Antegrade Femoral nails in the femur include: - Compound and simple shaft fractures - Proximal, metaphyseal and distal shaft fractures - Segmental fractures - Comminuted fractures - Fractures involving osteopenic and osteoporotic bone - Pathological fractures - Fractures with bone loss - Pseudoarthrosis, non-union, mal-union and delayed union - Periprosthetic fractures - Surgically created defects such as osteotomies - Intertrochanteric and subtrochanteric fractures. The Zimmer Natural Nail System is intended for temporary fracture fixation and stabilization of the bone. Indications for use of the retrograde femoral nails in the femur include: - Compound and simple shaft fractures - Proximal, metaphyseal, and distal shaft fractures - Segmental fractures - Closed supracondylar fractures - Severely comminuted supracondylar fractures with articular involvement - Fractures involving femoral condyles - Comminuted fractures - Fractures involving osteopenic and osteoporotic bone - Pathological fractures - Fractures with bone loss - Pseudoarthrosis, non-union, and mal-union - Periprosthetic fractures - Poly trauma patients
Device Story
Zimmer Natural Nail System is a family of intramedullary nails for temporary femoral fracture fixation and stabilization. System includes nails of various lengths/diameters, secured by 6.0mm fixed angle screws passing through proximal/distal nail holes. Components include nail caps, cortical washers, and cortical nuts. Used by orthopedic surgeons in clinical settings to stabilize femoral fractures. Device provides mechanical support to bone segments during healing. Output is the physical stabilization of the fracture site, enabling patient recovery.
Clinical Evidence
No clinical data. Bench testing performed including screw three-point bend test, proximal nail fatigue construct test, distal nail fatigue construct test, and non-pyrogenicity testing.
Technological Characteristics
Intramedullary fixation rod system. Components made of Ti-6Al-4V alloy; cortical nut contains UHMWPE. Fixed angle screw design. Mechanical fixation via intramedullary placement. Sterilization method same as predicate.
Indications for Use
Indicated for temporary fracture fixation and stabilization of the femur in patients with compound/simple shaft fractures, proximal/metaphyseal/distal shaft fractures, segmental fractures, comminuted fractures, osteopenic/osteoporotic bone fractures, pathological fractures, bone loss, pseudoarthrosis, non-union, mal-union, delayed union, periprosthetic fractures, surgically created defects (osteotomies), intertrochanteric/subtrochanteric fractures, closed supracondylar fractures, severely comminuted supracondylar fractures with articular involvement, femoral condyle fractures, and poly trauma.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Predicate Devices
Zimmer Natural Nail System Piriformis Fossa and Greater Trochanter Antegrade Femoral Nails (K083497)
Zimmer Natural System Retrograde Femoral Nails (K101622)
Submission Summary (Full Text)
{1}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Zimmer Inc. Dhaval Saraiya Regulatory Affairs Senior Specialist 345 East Main Street Warsaw, Indiana 46580
Re: K172114
Trade/Device Name: Zimmer Natural Nail System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: July 10, 2017 Received: July 13, 2017
Dear Mr. Saraiya:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-
Image /page/0/Picture/10 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is written in a simple, sans-serif font.
August 10, 2017
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
### Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
2301V932 ИАМИН ОНА НТЈАЭН ЗО ТИЭМТЯАЧЭО Food and Drug Administration
udications for Use
### See PRA Statement below Expiration Date: January 31, 2017 Form Approved: OMBBIolo. 0900-0120
### 510(k) Number (if know
77777717
Device Name Zimmer Natural Nail System
## Indications for Use (Describe)
ANTEGRADE FEMORAL NAILS:
:ebuloui uumst edi ui sliku asso zimrofi.iff bas returner of sent and the sur institution and and to noitezilidate bus noitezit entroques not bebrient a imeter likke kunska sell
FOR CONTINUOUS MONITORING AND
PERIODIC INSPECTIONS OF THE VISUAL
CONDITIONS FOR THE INTEGRITY OF
CONTAINMENT AND ASSOCIATED
COMPONENTS
- Proximal, metaphyseal and distal shaft fractures Componing and simple shaft fractures
- Segmental fractures
- Comminuted fractures
- Fractures involving osteopenic and osteoporotic bon
- Pathological fractures
- Fractures with bone loss
- and partirosis, pur noinn-lem (university of the desires and
- Periprosthetic fractures
- Surgically created defects such as osteotomies
- Interfrochanteric and subtrochanteric fracture
## RETROGRADE FEMORAL NAILS:
on one in formations but unders in the firsten from the first list. Inches of t Innications for use of the retroges ade femore in the femur include:
- Compound and simple shaft fractures
- Proximal, metaphyseal and distal shaft fractures
- Segmental fractures
- Closed supracondylar fractures
- Severy commininted supracondy lar fractures with articular involvement
- Fractures involving femoral condyles
-
- Comminuted fractures
- Fractures involving osteopenic and osteoporotic bon
-
- Pathological fractures
-
- Fractures with bone loss
-
- Pseudoarthrosis, non-union, and mal-unio
-
-
-
-
-
- Periprosthefic fractures
-
-
-
-
- Poly frauna patients
-
FORM FDA 3888 (8814
Page 1 of 2
SC Publishing Services (301) 413-6740
8 I
-
-
-
-
-
-
-
-
-
-
-
-
-
-
Type of Use (Select one or both, as applicable)
A Prescription Use (Part 21 CFF 801 Suppart D
O
Clover The-Counter Use (11 CFF 801 Support C
-
-
-
-
-
-
-
-
-
-
-
{4}------------------------------------------------
# . SEET To ISA noiloubes yourses and to stremeriuper of your seilede noiloses sidT
* WO START YOU COMPLETED FORM O THE PRATE ADDRESS BLOWN
:of ,uebrud sid) ഉന്നാսber 101 году 2017 г. подредног подредног подільні від 10 to and the vir and the new ind sint published streming on the 2 . noils mornity to noticellos and works in the əri birinci ilə sənədlər və qalan və bir sənəyə olunan mərkəzi və və ilə və ilə və ilə və bir nəti məbul əri
Dogs PARSITAR@fdla.hhs.gov Office of Chief Information Officer Pagerwork Reduction Act (PRA) State Food and Drug Administration
าง แต่เวอร์เวอร์
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Zimmer Biomet. The logo features a large, stylized blue letter "Z" enclosed in a blue circle. Below the circle, the word "zimmer" is written in lowercase blue letters, using a modern sans-serif font.
### 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Zimmer Natural Nail (ZNN) – 6.0mm Fixed Angle Screws 510(k) premarket notification.
| Sponsor: | Zimmer, Inc.<br>345 E Main St<br>Warsaw, IN 46580<br>Establishment Registration Number: 1822565 | | |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|-------------|
| Contact Person: | Dhaval Saraiya<br>Regulatory Affairs Sr. Specialist<br>Telephone: (305-269-6386)<br>Fax: (305-269-6400) | | |
| Date: | August 9, 2017 | | |
| Subject Device: | Trade Name: Zimmer Natural Nail System<br>Common Name: Intramedullary Fixation Road<br>Classification Name:<br>• HSB- Rod, Fixation, Intramedullary and Accessories<br>(21 CFR 888.3020) | | |
| Predicate Device(s): | K083497 | Zimmer Natural Nail System Piriformis<br>Fossa and Greater<br>Trochanter Antegrade<br>Femoral Nails | Zimmer Inc. |
| | K101622 | Zimmer Natural<br>System Retrograde<br>Femoral Nails | Zimmer Inc. |
| Purpose and Device<br>Description: | The intent of this premarket notification is to provide new<br>mechanical testing data for the Zimmer Natural Nail<br>System - 6.0mm Fixed Angle Screws which is different<br>than the performance specifications submitted in the<br>original premarket notifications mentioned above. The<br>reason for two predicate devices being identified in this<br>submission is because these screws are used with both<br>Zimmer Natural Nail System - Piriformis Fossa and Greater | | |
{6}------------------------------------------------
Trochanter Antegrade Femoral Nails and Retrograde Femoral Nails and different sizes were cleared in both submissions. Since these screws are used for various nails in the same product family, we think that this still qualifies as a special 510(k) submission.
The Zimmer Natural Nail System is a family of temporary fixation intramedullary nails designed for fracture fixation and stabilization of the femur. The nails are available in a variety of lengths and diameters to meet the assorted anatomical needs. Each of the intramedullary nails is secured by a series of screws that pass through holes manufactured into the proximal and distal sections of each nail. Nail caps, cortical washers and cortical nuts are available for use with the system. All components are made of Ti-6Al-4V alloy. The cortical nut also contains UHMWPE.
### Intended Use and Indications for Use:
### ANTEGRADE FEMORAL NAILS:
The Zimmer Natural Nail System is intended for temporary fracture fixation and stabilization of the bone. Indications for use of the Greater Trochanter and Piriformis Fossa Antegrade Femoral nails in the femur include:
- Compound and simple shaft fractures
- · Proximal, metaphyseal and distal shaft fractures
- Segmental fractures
- · Comminuted fractures
- · Fractures involving osteopenic and osteoporotic bone
- Pathological fractures
- Fractures with bone loss
- · Pseudoarthrosis, non-union, mal-union and delayed union
- Periprosthetic fractures
- · Surgically created defects such as osteotomies
- · Intertrochanteric and subtrochanteric fractures
### RETROGRADE FEMORAL NAILS:
The Zimmer Natural Nail System is intended for temporary fracture fixation and stabilization of the bone. Indications for use of the retrograde femoral nails in the femur include:
- · Compound and simple shaft fractures
- Proximal, metaphyseal, and distal shaft fractures
- Segmental fractures
- Closed supracondylar fractures
{7}------------------------------------------------
- Severely comminuted supracondylar fractures with articular involvement
- · Fractures involving femoral condyles
- · Comminuted fractures
- · Fractures involving osteopenic and osteoporotic bone
- Pathological fractures
- Fractures with bone loss
- · Pseudoarthrosis, non-union, and mal-union
- · Periprosthetic fractures
- Poly trauma patients
### Summary of Technological Characteristics:
The rationale for substantial equivalence is based on consideration of the following characteristics:
- Intended Use: Same as the predicate device
- Indications for Use: Same as the predicate device
- Materials: Same as the predicate device ●
- Design Features: Same as the predicate device ●
- . Sterilization: Same as the predicate device
### Summary of Performance Data (Nonclinical and/or Clinical)
### ● Non-Clinical Tests:
- Screw Three Point Bend Test O
- Proximal Nail Fatigue Construct Test o
- Distal Nail Fatigue Construct Test O
- Testing has been performed to establish product o non-pyrogenicity
- Clinical Tests: ●
- o N/A
### Substantial Equivalence Conclusion
The information provided within this premarket notification demonstrates that the Zimmer Natural Nail System Screws are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.