K172111 · Vevazz, LLC · OLI · Nov 22, 2017 · General, Plastic Surgery
Device Facts
Record ID
K172111
Device Name
Vevazz LED
Applicant
Vevazz, LLC
Product Code
OLI · General, Plastic Surgery
Decision Date
Nov 22, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5400
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K172111 · Nov 22, 2017
Vevazz LED
Vevazz, LLC
Retrospective review of existing treatment records from multiple private practice locations
A retrospective analysis of 77 patients was conducted to determine the circumferential reduction of the waist, hips, and thighs following a single 28-minute treatment session with the Vevazz LED, and to compare these results with the predicate device (Photonica Professional).
Retrospective analysis; Circumferential reduction; Body contouring; Real-world clinical records
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
A Retrospective Analysis of the Effects of Red Light Therapy on Body Contouring; Retrospective analysis of existing treatment records; Follow-up/Duration: Single 28-minute treatment session
77 patients (64 female, 13 male) at multiple private practice locations; age range 18-77 years.; Sample Size: 77; Number of Sites: 4
Photonica Professional (K160880) clinical trial data
Combined circumferential reduction of the waist, hips, and thighs
Indications for Use
The Vevazz LED can be used to provide the non-invasive dermatological aesthetic treatment for the reduction of circumference of hips, waist, and thighs.
Device Story
Vevazz LED is a non-invasive red-light system for aesthetic body contouring. Device consists of a base unit supporting up to 16 large and 4 small LED paddles. Each paddle contains 28 LEDs emitting 650 nm (+/- 25 nm) light at 50 mW/cm2. Operated by clinicians in private practice settings. Paddles are placed directly on patient skin over target areas (hips, waist, thighs). Device delivers a uniform bath of red light to induce lipolysis in subcutaneous fat. Treatment is preset to 7-minute intervals per area, totaling 28 minutes. Output is controlled via a power switch. Clinical benefit includes immediate circumferential reduction of treated areas without surgery, diet, or exercise. Device does not prevent future weight gain.
Clinical Evidence
Retrospective multi-site analysis of 77 patients (64 female, 13 male; ages 18-77). Primary endpoint: combined circumferential reduction of waist, hips, and thighs after a single 28-minute treatment session. Results showed an average reduction of 2.1 inches (std dev 1.5). 100% of patients achieved at least 1.0 inch loss. No device-related adverse events reported. Study compared results to historical data from the Photonica Professional predicate device.
Technological Characteristics
Non-invasive red-light system. Base unit (120/240 VAC) powers up to 16 LED paddles. Each paddle features 28 LEDs (650 nm +/- 25 nm). Power density: 50 mW/cm2. Compliance with IEC 60601-1 (electrical safety), IEC 60601-2-57 (non-laser light source), IEC 62471 (photobiological safety), and IEC 60601-1-2 (EMC). Constant waveform. No software component.
Indications for Use
Indicated for non-invasive dermatological aesthetic treatment for the reduction of circumference of hips, waist, and thighs in adult patients.
Regulatory Classification
Identification
A Low Level Laser System for Aesthetic Use is a device using low level laser energy for the disruption of adipocyte cells within the fat layer for the release of fat and lipids from these cells for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Industry and Food and Drug Administration Staff; Class II Special Controls Guidance Document: Low Level Laser System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 22, 2017 Vevazz, LLC Jamie Fettig Owner 3839 McKinney Ave. #251 Dallas, Texas 75204
Re: K172111
Trade/Device Name: Vevazz LED Regulation Number: 21 CFR 878.5400 Regulation Name: Low Level Laser System For Aesthetic Use Regulatory Class: Class II Product Code: OLI Dated: July 10, 2017 Received: July 13, 2017
Dear Jamie Fettig:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820):
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@Ida.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S.
Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172111
Device Name Vevazz LED
Indications for Use (Describe)
The Vevazz LED can be used to provide the non-invasive dermatological aesthetic treatment for the reduction of circumference of hips, waist, and thighs.
| Type of Use (Select one or both, as applicable) |
|--------------------------------------------------|
| ☑ Prescription Use (Part 21 CFR 201.2 Subpart D) |
| ☐ Over-The-Counter Use (21 CFR 201.2 Subpart C) |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510-(k) Summary
#### 1. Submitters Information
- Name: Vevazz, LLC 3839 McKinney Ave. #251 DALLAS, TX 75204 Tel: 773-665-4005 Contact: Jamie Fettig, Owner 773-620-9500 jamie@healthyhuman.org
#### 2. Date Prepared:
July 10, 2017
#### 3. Device Name and Classification
| Device | Vevazz LED |
|------------------------|------------------------------------------|
| Trade Name | Vevazz LED |
| Regulation Description | Low Level Laser System for aesthetic use |
| Regulation Number | 21 CFR 878.5400 |
| Device Class | II |
| Review Panel | General and Plastic Surgery |
| Product Code | OLI |
#### 4. Intended Use
The Vevazz LED can be used to provide the non-invasive dermatological aesthetic treatment for the reduction of circumference of hips, waist, and thighs.
#### 5. Device Description
The Vevazz is a non-invasive red-light system. The device is a moderately complex device that consists of a base unit that plugs into 120 VAC or 240 VAC main power with Ground (earth) and a 2 amp 250 VAC Fuse. The base unit can support up to 16 large LED paddles and 4 small LED paddles. The connections on the base are female phono jacks that are low voltage having a maximum electrical output of 16 VDC @ .5 amps for a total of 8 Watts per port. The large paddles consist of 28 LEDs in a series/parallel electrical connection that has a cord with a male phono jack. Small paddles consist of 4 LEDs and have matching electrical characteristics as the large paddles. Each LED produces 50 mw/cm2 of light energy with a wavelength of 650 nm; +/- 25 nm uniformly over the treatment area without producing hot spots. Treatment time is preset to 7 minutes for non-invasive dermatological aesthetic treatment for the reduction of circumference of hips, waist, and thighs.
#### Predicate Devices 6.
The Vevazz LED (K172111) is substantially equivalent to the following predicate device:
- Photonica Professional (K160880) manufactured by Ward Photonics, LLC. ●
#### 7. Substantial Equivalency Discussion
The intended use for this type of device (OLI) is equivalent to the predicate device, Photonica Professional by Ward Photonics. Both devices, the Vevazz (K172111) and Photonica Professional
{4}------------------------------------------------
(K160880) cause lipolysis. which reduces the size of subcutaneous fat accumulations as a result of exposure to 650 nm red light. The Vevazz uses 650 nm ± 25 nm (visible red-light spectrum) which is the same red-light wavelength used by Photonica Professional.
The Vevazz is substantially equivalent to the predicate device and does not raise any new risks or hazards.
The principles of operation and base elements of the Vevazz LED device are very similar to the Photonica Professional predicate device in that the Vevazz LED and predicate device is fully compliant with USFDA recognized standards for electrical safety and emit a substantially equivalent amount of power and energy. Both the Vevazz LED and Photonica Professional devices achieve treatment area coverage through a soft uniform bath of red light. While the Photonica Professional uses an array of 150 diodes and a treatment time of 8 minutes, the Vevazz LED device uses an array of 8 or 16 LED light paddles placed about the treatment areas of the patient, each paddle consisting of 28 LED diodes for a treatment time of 7 minutes. As dosage is a mechanism of energy, coverage and time, the Vevazz LED has matched the effectiveness of the predicate with similar, more efficient method of treatment.
There are no differences between the subject device and the Photonica Professional with respect to indications and intended use. Any other differences where they occur are minor and do not represent any risk or hazard.
| Performance<br>Characteristic | Submission Device<br>Vevazz | Predicate Device<br>Photonica<br>Professional |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510K | K172111 | K160880 |
| Manufacturer | Vevazz LLC | Ward Photonics LLC |
| Date of FDA<br>Clearance | TBD | September 23, 2016 |
| Intended Use | The Vevazz LED can<br>be used to provide the<br>non-invasive<br>dermatological<br>aesthetic treatment for<br>the reduction of<br>circumference of hips,<br>waist, and thighs. | Photonica<br>Professional is<br>indicated for use as a<br>non-invasive<br>dermatological<br>aesthetic treatment for<br>the reduction of<br>circumference of hips,<br>waist, and thighs. |
| Output (nominal) | 120/240 VAC<br>16 VDC | 120 VAC |
| Power | 50 mW/cm2 | 105 mW/cm2 |
| Wavelength | 650 nm +/- 25nm | 635 nm +/-2nm |
| Waveform | Constant | Constant |
| Energy Source | Multi LED diode<br>dispersed over<br>treatment area | Multi LED diode<br>dispersed over<br>treatment area |
| Energy Delivery | Handheld Treatment<br>Probes (1-16 paddles) | LED array mounted on<br>an articulated arm |
| Timer range | 0-20 minutes | 8 or 20 minutes<br>treatment cycles |
| Treatment Time | 7 minutes (default) | 8 minutes or 20<br>minutes |
| Software | None | None |
| Patient override<br>control | Power off switch | Power off switch |
{5}------------------------------------------------
## 8. Non-Clinical Performance
The Vevazz is in compliance with the following standards listed below. The Vevazz has been tested for all designated tests, as applicable, given in the Guidance document: Class II Special Controls Guidance Document: Low Level Laser System for Aesthetic Use (Document issued on: April 14, 2011). A detailed review of the performance and safety testing is given in Section: Performance Testing - Bench.
. IEC 60601-1:2005 +C1:2006, +C2:2007, AM1:2012, Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
IEC 60601-2-57:2011. Medical electrical equipment – Part 2-57: Particular requirements for basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
IEC 62471 Ed. 1, International Standard for Photobiological Safety for Lamps and Lamp Systems - General Requirements
IEC 60601-1-2:2007, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests.
### 9. Clinical Performance Data:
A multi-site retrospective analysis has been performed on the records of 77 patients, which demonstrates the efficacy of the device using the Vevazz LED protocol with a 7-minute exposure to the front of the body. 7 minutes to the left side. 7 minutes to the back, and 7 minutes to the right side. with a total of 28 minutes exposure time for each treatment session.
The objective of this study is to retrospectively review existing treatment records at multiple private practice locations in the USA to determine the combined circumferential reduction of the waits, hips and thighs with a single 28 minute treatment session using the subject Vevazz LED product, without benefit of dieting or exercise, in order to establish a baseline of expected outcomes for patients who undertake this procedure and to compare the procedure's effectiveness with non-invasive fat reduction using the Photonica Professional fat reducing low level laser system by Ward Photonics LLC.
The studies differ in that the Ward Photonics LLC protocol was performed four 8-minute treatments.
In order to assure a fair comparison of the measurements with the predicate's measurement schema, the measurements in this study were taken in exact accordance with the Photonica Male Measurement Guide. Female Measurement Guide. Male Measurement Form, and Female Measurement Forms as were used in the clinical trials upon which the predication was approved. Those results were published in A Retrospective Analysis of the Effects of Red Light Therapy on Body Contouring.
The same self-tensioning tape measures were used for both studies. As detailed in the Photonica measurement quides and forms, multiple circumferential measurements were recorded for each patient. The measurement reference points are described in the Photonica Male Measurement Guide as quoted here for five measurements:
| Back | Circumference at level of nipples |
|---------------|-----------------------------------------|
| Upper-Abdomen | Distance above umbilicus |
| Mid-Abdomen | 2-3 inches above the umbilicus |
| Flanks | Circumference around the "love handles" |
| Lower-Abdomen | Distance from umbilicus |
{6}------------------------------------------------
The measurement reference points are described in the Photonica Female Measurement Guide as quoted here for six measurements:
| Back | Circumference where fat is under bra and note<br>distance in inches above the umbilicus |
|-------------------------|-----------------------------------------------------------------------------------------|
| Waist | 2-3 inches above umbilicus (note distance) |
| Mid-Abdomen | 2-3 inches below the umbilicus (note distance) |
| Hips | Note distance below umbilicus |
| Thighs (right and left) | Greatest circumference, then note distance from<br>top of kneecap. |
The results of the retrospective analysis are summarized in the table below:
| Title of Study: | A Retrospective Analysis of the Effects of Red Light Therapy on Body Contouring. |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Study Objective | The objective of this study is to retrospectively review existing treatment records at multiple private practice locations in the USA to determine the combined circumferential reduction of the waist, hips, and thighs with a single 28 minute treatment session using the subject Vevazz LED product, without benefit of dieting or exercise, in order to establish a baseline of expected outcomes for patients who undertake this procedure and to compare the procedure's effectiveness with non-invasive fat reduction using Photonica Professional by Ward Photonics LLC |
| Number of Sites: | 4 |
| Enrollment Size: | 77 |
| Patient Population | For the study, 64 of the patients are female (83%) and 13 are male (17%). Females ranged in age from 18 to 77, with an average age of 41.7 years. Males ranged in age from 35 to 68, with an average age of 43.3 years. Overall, patients ranged in age from 18 to 77, with an average age of 42 years. |
| Study Design | Vevazz LED paddles were positioned directly over the skin with four 7-minute exposures. For males, the paddles were positioned:<br><b>Front:</b> Over the chest and abdomen, extending 23" toward suprapubic area. <b>Back:</b> Starting at the top of buttocks, extending 23 inches toward the upper back. <b>Left:</b> From the left hip area, extending 23 inches toward the left underarm area. <b>Right:</b> From the right hip area, extending 23 inches toward the right underarm area. For females, the paddles were positioned: <b>Front:</b> From below the bra and extending 23 inches to include the top of the thighs. <b>Back:</b> Covering the “back” measurement area and extending 23 inches to include the top of the thighs. <b>Left:</b> The paddles are placed to cover the left "love handle", the outside of the left thigh, and the inside of the right thigh |
{7}------------------------------------------------
| • Right: The paddles are placed to cover the right "love handle", the outside of the right thigh, and the inside of the left thigh, along with the lower abdomen | | | | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|------|-----------|-----------|
| Measurements and Timing | Measurements were taken immediately before and after the 28-minute treatment using a self-tensioning tape measure, measurement forms, and instructions (Measurement Guides) exactly as used in the clinical trials for the predicate Photonica Professional.<br><br>Since measurements were taken immediately before and after the treatment, the observed results are therefore immediate. | | | | |
| Treatment | The treatments consist of four 7-minute exposures with the 650nm light positioned directly over the skin.<br><br>Vevazz LED uses large paddles consisting of 28 LEDs in a series/parallel electrical connection. Each LED produces 50 mw/cm2 of light energy with a wavelength of 650 nm; +/- 25 nm uniformly over the treatment area. | | | | |
| Results | Inch loss by gender is: | | | | |
| | Gender | Patients | Age | Inch Loss | Std. Dev. |
| | Males | 13 | 43.3 | 1 5/8 | 1 2/8 |
| | Females | 64 | 41.7 | 2 2/5 | 1 6/8 |
| | Total | 77 | 42.0 | 2 1/80 | 1 1/2 |
| | Inch loss by treatment site is: | | | | |
| | Site | Patients | Age | Inch Loss | Std. Dev. |
| | A | 8 | 49.1 | 3 4/7 | 1 2/8 |
| | B | 16 | 48.7 | 1 2/3 | 1 5/8 |
| | C | 23 | 40.7 | 2 1/2 | 2 |
| | D | 30 | 37.2 | 2 1/6 | 1 6/8 |
| | Total | 77 | 43.9 | 2 1/2 | 1 5/8 |
| | Patients lost a minimum of 1". The maximum lost is 7 1/2". | | | | |
| | For males, inch loss is computed by adding the inch loss measurements for back, upper-abdomen, mid-abdomen, flanks and lower abdomen. | | | | |
| | For females, inch loss is computed by adding the inch loss measurements for back, waist, mid-abdomen, hips and thighs. | | | | |
| Adverse Events | There were no device-related adverse events | | | | |
| Substantially Equivalent to Predicate | Vevazz LED achieved an average combined circumferential reduction of the waits, hips, and thighs of 2.5" at one visit using the same measurements schema as in the Photonica Professional study which found an average loss of 3.5" with one 32-minute treatment | | | | |
{8}------------------------------------------------
Further, 100% of patients had clinically significant results at one visit with Vevazz LED, considering that all patients lost at least 1.0" with one treatment. By comparison, traditional liposuction is limited to removal of no more than 500 ml of fat at one surgery, due to the attendant trauma and potential sideeffects.
NOTE: While treatment with Vevazz LED immediately reduces the size of subcutaneous fat accumulations exposed to the lights, treatments do not prevent the patients from subsequently gaining weight or increasing the size of the fat accumulations.
The durability of the treatment has not been studied or established in follow-up clinical trials with a sufficiently statistically significant sample.
### 10. Conclusion
The Vevazz is substantially equivalent to the Photonica Professional manufactured by Ward Photonics (K160880).
The results of non-clinical performance data confirm that the Vevazz LED delivers substantially equivalent Total Treatment Dose of 650 nm ±25 light to the subcutaneous fat cells in 28 minutes at one visit rather than in 32 minutes at one visit.
The results of clinical testing confirm that the Vevazz LED achieves circumferential reduction substantially equivalent to the Photonica Professional, but in 28 minutes rather than 32 minutes.
The information provided in this submission supports the substantial equivalence to the predicate device and that the system is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.