Shoocin Introducer Kit
K172099 · Lepu Medical Technology (Beijing) Co., Ltd. · DYB · Oct 5, 2017 · Cardiovascular
Device Facts
| Record ID | K172099 |
| Device Name | Shoocin Introducer Kit |
| Applicant | Lepu Medical Technology (Beijing) Co., Ltd. |
| Product Code | DYB · Cardiovascular |
| Decision Date | Oct 5, 2017 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.1340 |
| Device Class | Class 2 |
Intended Use
The ShoocinTM Introducer Kits are intended for use to facilitate the introduction of guide wires, catheters and other accessory medical devices through the skin into a vein or artery and minimize blood loss associated with such introduction.
Device Story
Shoocin™ Introducer Kit facilitates vascular access for interventional procedures. Components include puncture needle, stainless-steel guide wire with collimator, sheath introducer, and dilator. Puncture needle provides initial vascular access; guide wire acts as a track for sheath insertion. Sheath introducer, containing Barium sulfate (BaSO4) for fluoroscopic visibility, provides a conduit for interventional devices; features hemostasis valve and side port with 3-way stopcock to minimize blood loss. Dilator provides stability during deployment. Used in clinical settings by physicians; device provides physical conduit for catheterization. Benefits include reliable vascular access and reduced blood loss during device exchange.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included mechanical integrity (tensile force, leakage, fracture, flexing), radio-detectability, and chemical/particulate analysis. Biocompatibility testing per ISO 10993-1 confirmed non-cytotoxic, non-sensitizing, non-pyrogenic, and thromboresistant properties.
Technological Characteristics
Components: stainless-steel guide wire, Barium sulfate (BaSO4) radiopaque sheath, dilator, puncture needle. Sterilization: Ethylene Oxide (EO) per AAMI/ANSI/ISO 11135:2014 (SAL 10^-6). Biocompatibility: ISO 10993-1 (External Communicating Device, blood contact <24 hours). Mechanical testing: ASTM F756 (hemolysis).
Indications for Use
Indicated for patients requiring vascular access to facilitate the introduction of guide wires, catheters, and other accessory medical devices into a vein or artery while minimizing blood loss.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Predicate Devices
- Shoocin™ Introducer Kit (K123475)
Related Devices
- K123475 — SHOOCIN INTRODUCER KIT · Lepu Medical Technology (Beijing) Co., Ltd. · Feb 3, 2014
- K172331 — Type I, Type II, Type III, Type IV · Lepu Medical Technology (Beijing) Co., Ltd. · Mar 2, 2018
- K140768 — TYPE I BRILLIANT INTRODUCER KIT, TYPE II BRILLIANT INTRODUCER KIT, TYPE III BRILLIANT INTRODUCER KIT, TYPE IV BRILLIANT · Lepu Medical Technology (Bejing) Co., Ltd. · Nov 25, 2014
- K110051 — RADIALSOURCE TRANSRADIAL ACCESS INTRODUCER KIT · Great Batch Medical · Feb 2, 2012
- K052557 — SUPER SHEATH INTRODUCER SHEATH · Togo Medikit Co., Ltd. · Jan 17, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other to form a single, abstract figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 5, 2017
Lepu Medical Technology (Beijing) Co., Ltd. % Arthur Goddard President FDA Regulatory and Quality Systems Consultant 31853 Cedar Road Mayfield Heights, Ohio 44124-4445
Re: K172099
Trade/Device Name: Shoocin™ Introducer Kit Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: August 4, 2017 Received: August 9, 2017
Dear Arthur Goddard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K172099
Device Name ShoocinTM Introducer Kit
Indications for Use (Describe)
The ShoocinTM Introducer Kits are intended for use to facilitate the introduction of guide wires, catheters and other accessory medical devices through the skin into a vein or artery and minimize blood loss associated with such introduction.
Type of Use (Select one or both, as applicable)
| <span style="font-size:120%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------|-----------------------------------------------|
|-------------------------------------------------------------------------------------|-----------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
The information on the Shoocial™ Introducer Kit is being submitted in accordences of 21 C.F. R. §807.92 and The results of future studies and or tests may require alterations of the conclusions or recommendations set forth.
| Applicant: | Lepu Medical Technology (Beijing) Co., Ltd.<br>No. 37 Chaoqian Road<br>Changping District, Beijing 102200<br>P.R. China |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone: | +86-10-80123510 |
| Contact: | Xiangdan Kim |
| Date: | June 30, 2017 |
| Name: | ShoocinTM Introducer Kit |
| Classification Name: | Catheter Introducer, 870.1340 |
| Product Code: | DYB |
| Predicate: | ShoocinTM Introducer Kit, Lepu Medical Technology (Beijing) Co.,<br>Ltd., K123475 with market clearance dates of February 3, 2014. |
| Modification to<br>Predicate K123475 | 1. Addition of two 7Fr models to the Radial Sheath Introducer<br>(RS070718 and RS071118).<br>2. Change of the radiocontrast agent in the Introducer Sheath from<br>Bismuth to Barium sulfate (BaSO4).<br>3. Removal of the radiocontrast agent Bismuth in the Dilator |
| Description: | The ShoocinTM Introducer Kit consists of a puncture needle, a guide<br>wire with a guide wire collimator, a sheath introducer, and a dilator.<br>The Puncture Needle incorporates a lumen, which provides a conduit<br>for the insertion of the Guide Wire into the vascular system. The<br>flexible stainless-steel guide wire is utilized as a guiding mechanism<br>for the insertion of the Introducer Sheath into the vascular system. The<br>Guide Wire contains a wire collimator, which assists in funneling the<br>wire through the lumen of the Puncture Needle. The Guide Wire is<br>radio-detective under fluoroscopy. The Sheath Introducer contains<br>Barium sulfate (BaSO4), making the device visible under fluoroscopy<br>and provides a conduit for introducing other interventional devices,<br>including guide wires and interventional catheters, into the vascular<br>system. The main components of this assembly are a non-hydrophilic<br>coated Sheath Introducer, hemostasis valve housing, and a side port<br>tubing with a 3-way stopcock/valve. The Dilator is used to provide<br>support and stability to the Sheath Introducer during deployment into<br>the vascular system. The proximal end of the Dilator includes a luer<br>port and has a tapered, atraumatic distal tip. There is no radiocontrast<br>agent on the dilator. |
| Indication For Use: | The ShoocinTM Introducer Kits are intended for use to facilitate the<br>introduction of guide wires, catheters and other accessory medical<br>devices through the skin into a vein or artery and minimize blood loss<br>associated with such introduction. |
{4}------------------------------------------------
| Predicate Device | Lepu Medical Technology (Beijing) Co., Ltd. added two models to the |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison: | Radial Sheath Introducer, the sheath introducer radiocontrast agent<br>was changed from Bismuth to Barium sulfate (BaSO4), and the<br>removal of a radiocontrast agent on the dilator to the currently<br>marketed ShoocinTM Introducer Kit. The subject device is<br>substantially equivalent to the predicate device in terms of intended<br>use, indication for use, operational characteristics, and fundamental<br>design and technology characteristics. |
Biocompatibility: The Shoocin™ Introducer Kit produced by Lepu Medical was assessed against the International Standard ISO 10993-1, "Biological evaluation of medical devices. Part 1. Guidance on selection of tests." The Shoocin ™ Introducer Kit would be classified as an External Communicating Device in contact with the Circulating Blood for a Limited Duration (<24 hours). The following test per ISO 10993 and results are:
| Test | Standard | Results | |
|--------------------------------------|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity | ISO 10993-5 | Under the conditions of this study, the MEM test extracts would be considered no cytotoxicity potential. The negative controls, blank controls, and the positive controls performed as anticipated. | |
| ISO Intracutaneous study | ISO 10993-10 | Under the conditions of this study, the test article met the requirements of the test since the difference between each test extract overall mean score and corresponding control overall mean score was 0.0 and 0.0 for the SC and CSO test extracts, respectively. | |
| ISO Guinea Pig Maximum Sensitization | ISO 10993-10 | Under the conditions of this study, the test article extracts showed no evidence of causing delayed dermal contract sensitization in the Guinea pig. The test article was not considered a sensitizer in the Guinea pig maximization test. | |
| Biocompatibility continue: | Test | Standard | Results |
| | ISO Systemic Toxicity | ISO 10993-11 | Under the conditions of this study, there was no mortality or evidence of systemic toxicity from the extracts injected into mice. Each test article extract met the requirements of the study. |
| | Complement Activation Assay | ISO 10993-4 | Under the test condition, reading of C3a or Sc5b-9 of test article were significantly different (P<0.05) than those of negative control article. |
| | ASTM Hemolysis | ISO 10993-4<br>ASTM F756 | Under the conditions of this study, the Hemolytic Index for the test article in direct contact with blood was 0.9% and that for the test article extract was 0.4%. Both the test article in direct contact with blood and test article were non-hemolytic. |
| | USP Pyrogen Study | ISO 10993-11<br>USP <151> | Under the conditions of this study, the maximum rise of each rabbit temperatures did not show a rise of 0.5° C or more above its baseline temperature during 3 hour observation period. The test article was judged as nonpyrogenic. |
| | Partial Thromboplastin Time | ISO 10993-4 | Under the conditions of this study, the % negative control is 68.69%, the test article would be considered mild thrombogenicity response. |
| | In Vivo Thromboresistance | ISO 10993-4 | Under the conditions of this study, the test article showed no thrombosis which was similar with the control article. The test article would be considered as thromboresistant. |
{5}------------------------------------------------
{6}------------------------------------------------
Performance Testing: The Shoocin ™ Introducer Kit modifications and new models were subjected to the following test methods (bench test) to demonstrate that these devices comply with the performance data from the predicate device.
| ShoocinTM Introducer Kit Performance Testing | |
|----------------------------------------------|-------------------------------------------------|
| Sheath<br>Introducer | Appearance |
| | Size designation |
| | Radio-detectability |
| | Freedom from leakage from sheath introducer |
| | Freedom from leakage through haemostasis valve |
| | Hub |
| | Peak tensile force |
| | Dilator accessibility |
| 3-way<br>stopcock/<br>valve | Hub |
| Dilator | Appearance |
| | Size designation |
| | Hub |
| | Strength of union between hub and dilator |
| | Guidewire accessibility |
| Puncture<br>Needle | Appearance |
| | Radio-detectability |
| | Corrosion resistance |
| | Size Designation |
| | Needle point |
| | Hub |
| | Strength of union of needle tube and needle hub |
| | Guidewire accessibility |
| Guidewire | Appearance |
| | Size designation |
| | Fracture test |
| | Flexing test |
| | Peak tensile force of guidewire |
| | Corrosion resistance |
| | Radio-detectability |
| Particulate evaluation | |
| | Reduced substances |
| Chemical<br>performance | Heavy metal |
| | pH variation value |
| | Nonvolatile matter |
| | Residual EO and ECH |
| | |
| Biological | Sterile |
| | Bacterial endotoxin |
{7}------------------------------------------------
Sterilization: The method used is based on practices recommended by AAMI / ANSI / ISO 11135:2014 and provides a Sterility Assurance Level (SAL) of 10-6.
- Conclusion: The information provided in this submission and comparing intended use, principle of operation and overall technological characteristics (i.e. puncture needle, guide wire, dilator, and sheath introducer to obtain access to the vascular system), the Shoocin™ Introducer Kit supports a determination of substantially equivalent to existing legally marketed predicate device Shoocin™ Introducer Kit (K123475).