K172060 · Boston Scientific Corporation · PWI · Nov 3, 2017 · Obstetrics/Gynecology
Device Facts
Record ID
K172060
Device Name
Capio SLIM Suture Capturing Device
Applicant
Boston Scientific Corporation
Product Code
PWI · Obstetrics/Gynecology
Decision Date
Nov 3, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4910
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Capio™ SLIM Suture Capturing Device is intended for use in general suturing applications during surgery to assist in the placement of suture material in tissues at the operative site. The Capio™ SLIM Suture Capturing Device is to be used with sensory and/or direct visual control. The Capio SLIM Suture Capturing Device is also intended for use as an aid in the insertion, placement, fixation, or anchoring of surgical mesh during urogynecologic procedures.
Device Story
Capio™ SLIM Suture Capturing Device is a sterile, single-use surgical instrument designed for use in open surgical procedures to facilitate suture placement in difficult-to-access locations. The device features a proximal one-hand-activated plunger, a tubular shaft, and a distal curved needle driver. Operation involves depressing the plunger to actuate the carrier, which drives a suture/dart through tissue; the needle is then automatically captured by a catch mechanism for retrieval or suture tying. The device is operated by a physician to assist in general suturing or the anchoring of urogynecologic surgical mesh. By automating the throw, catch, and retrieval of sutures in a single step, the device simplifies suture management, potentially reducing operative time and improving precision in tissue fixation.
Clinical Evidence
Bench testing only. Performance testing included sterilization validation, package integrity (dye penetration, seal strength), dimensional analysis, functional testing, mechanical stress testing (compression, tip tensile, side/top load deflection, needle retention, catch pull, actuation), distribution challenge, biocompatibility (cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity), and shelf-life testing. Results demonstrate the device is biocompatible and meets mechanical performance requirements.
Technological Characteristics
Materials: Plastic and stainless steel. Design: Tubular shaft with proximal plunger and distal curved needle driver. Energy: Manual actuation. Connectivity: None. Sterilization: Ethylene Oxide (EO). Shelf-life: 3 years. Single-use.
Indications for Use
Indicated for general surgical suturing and as an aid in the insertion, placement, fixation, or anchoring of surgical mesh during urogynecologic procedures. Used with sensory and/or direct visual control.
Regulatory Classification
Identification
Specialized surgical instrumentation for use with urogynecologic surgical mesh is a prescription device specifically intended for use as an aid in the insertion, placement, fixation, or anchoring of surgical mesh during urogynecologic procedures. These procedures include transvaginal pelvic organ prolapse repair, sacrocolpopexy (transabdominal pelvic organ prolapse repair), and treatment of female stress urinary incontinence. Examples of specialized surgical instrumentation include needle passers and trocars, needle guides, fixation tools, and tissue anchors. This device is not a manual gastroenterology-urology surgical instrument and accessories (§ 876.4730) or a manual surgical instrument for general use (§ 878.4800).
Special Controls
*Classification.* Class II (special controls). The special controls for specialized surgical instrumentation for use with urogynecologic surgical mesh are:(1) The device must be demonstrated to be biocompatible;
(2) The device must be demonstrated to be sterile and, if reusable, it must be demonstrated that the device can be adequately reprocessed;
(3) Performance data must support the shelf life of the device by demonstrating package integrity and device functionality over the requested shelf life;
(4) Non-clinical performance testing must demonstrate that the device meets all design specifications and performance requirements, and that the device performs as intended under anticipated conditions of use; and
(5) Labeling must include:
(i) Information regarding the mesh design that may be used with the device;
(ii) Detailed summary of the clinical evaluations pertinent to use of the device;
(iii) Expiration date; and
(iv) Where components are intended to be sterilized by the user prior to initial use and/or are reusable, validated methods and instructions for sterilization and/or reprocessing of any reusable components.
Predicate Devices
Pinnacle Lite Pelvic Floor Repair Kit, Posterior Uphold Lite Vaginal Support System (K122459)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 3, 2017
Boston Scientific Corporation Michelle Becker Regulatory Affairs Specialist Urology and Pelvic Health 100 Boston Scientific Way Marlborough, Massachusetts 01752
Re: K172060
Trade/Device Name: Capio™ SLIM Suture Capturing Device Regulation Number: 21 CFR 884.4910 Regulation Name: Specialized surgical instrumentation for use with urogynecologic surgical mesh Regulatory Class: Class II Product Code: PWI Dated: September 29, 2017 Received: October 2, 2017
Dear Michelle Becker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172060
Device Name Capio™ SLIM Suture Capturing Device
#### Indications for Use (Describe)
The Capio™ SLIM Suture Capturing Device is intended for use in general suturing applications during surgery to assist in the placement of suture material in tissues at the operative site. The Capio™ SLIM Suture Capturing Device is to be used with sensory and/or direct visual control.
The Capio SLIM Suture Capturing Device is also intended for use as an aid in the insertion, placement, fixation, or anchoring of surgical mesh during urogynecologic procedures.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span style="font-size:20px">✖</span> </div> |
|----------------------------------------------|----------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary for Capio™ SLIM Suture Capturing Device (K172060)
## A. Date Prepared
November 2, 2017
#### B. Sponsor
Boston Scientific Corporation Urology and Pelvic Health Division 100 Boston Scientific Way Marlborough, MA 01752
## C. Contact
Rachel Nankervis Specialist I, Regulatory Affairs Phone: 508-683-4755 Fax: 508-683-5041 Rachel.Nankervis@bsci.com
Or
Michelle Becker Principal Specialist, Regulatory Affairs Phone: 508-210-6988 Fax: 508-683-5041 Michelle.Becker@bsci.com
## D. Device Name
Trade Name: Capio™ SLIM Suture Capturing Device Common/usual Name: Suture Placement Device Regulation Number: 21 CFR 884.4910 Regulation Name: Specialized surgical instrumentation for use with urogynecologic surgical mesh Classification: Class II Product Code: PWI
{4}------------------------------------------------
## E. Predicate Device(s)
Trade Name: Pinnacle Lite Pelvic Floor Repair Kit, Posterior Uphold Lite Vaginal Support System Common/usual Name: Surgical Mesh for transvaginal pelvic organ prolapse repair Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Classification: II Product Code: OTP Premarket Notification: K122459
## F. Device Description
The Capio™ SLIM Suture Capturing Device is urogynecologic mesh instrumentation used with urogynecologic surgical mesh. The subject device is also intended for use in general suturing applications during surgery to assist in placement of suture material in tissues, and assist in insertion or anchoring of surgical mesh for transabdominal pelvic organ prolapse repair.
The Capio™ SLIM Suture Capturing Device is a sterile, single use device and is designed to be used by the physician to facilitate the placement of sutures in difficult to access locations during open surgical procedures. The Capio™ SLIM device is also designed to throw, catch and retrieve sutures in one step. The device consists of a one-handactivated plunger at the proximal end, a tubular shaft and a curved needle driver at the distal end. Depressing the plunger actuates the carrier and drives the suture/dart through the tissue. The needle is automatically caught by the device's needle catch mechanism for easy suture tying, or the device may be re-loaded for additional suture throws.
#### G. Comparison of Intended Use with Predicate Device
| | K172060 (Subject Device) | K122459 (Predicate Device) |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The Capio™ SLIM Suture Capturing<br>Device is intended for use in general<br>suturing applications during surgery to<br>assist in the placement of suture<br>material in tissues at the operative site.<br>The Capio™ SLIM Suture Capturing<br>Device is to be used with sensory<br>and/or direct visual control.<br><br>The Capio SLIM Suture Capturing<br>Device is also intended for use as an | The Pinnacle Lite Pelvic Floor Repair<br>Kit is indicated for tissue reinforcement<br>in women with pelvic organ prolapse, for<br>transvaginal repair of posterior vaginal<br>vault prolapse.<br><br>The Uphold Lite Vaginal Support System<br>is indicated for tissue reinforcement in<br>women with pelvic organ prolapse, for<br>transvaginal repair of anterior and apical<br>vaginal vault prolapse |
{5}------------------------------------------------
| aid in the insertion, placement, | |
|-----------------------------------------|--|
| fixation, or anchoring of surgical mesh | |
| during urogynecologic procedures. | |
The indications for use statements for subject device and the predicate device are not identical, as the predicate device is a surgical mesh that is indicated pelvic to treat organ prolapse. The intended use of the subject and predicate device is the same, as the subject device is an accessory to a surgical mesh.
| Device(s): | K172060 (Subject Device) | K122459 (Predicate Device) |
|--------------------------------|---------------------------------------------|---------------------------------------------------------------------|
| General Device Characteristics | | |
| Patient<br>contact | < 24 hours (tissue/bone) | Permanent (tissue/bone) |
| Operating<br>Principle | Aid in placement of surgical<br>mesh/suture | Transvaginal repair of anterior<br>and apical vaginal wall prolapse |
| Device<br>Design | Surgical mesh<br>instrumentation | Surgical mesh |
| Materials | Plastic<br>Stainless steel | Polypropylene |
| Shelf-life | 3 years | 3 years |
| Sterility | EO | EO |
## H. Comparison of Technological Characteristics
As discussed above. the subject device has different technological characteristics from the predicate device. The predicate device is a surgical mesh whereas the subject device is used for the placement of surgical mesh. Briefly, the predicate device differs from the subject device as follows:
- 1. Operating principle
- 2. Device Design
- 3. Materials
The differences between the subject device and the predicate device can raise different types of safety and effectiveness questions, as we are comparing an accessory and parent device. The differences in technological characteristics were evaluated through completion of special controls (performance testing, biocompatibility, shelflife/reprocessing, labeling, and sterilization) published in the final order reclassifying urogynecologic surgical mesh instrumentation from class I to class II published on January 6, 2017.
{6}------------------------------------------------
#### I. Non-Clinical Performance Test Summary
The following tests were performed to demonstrate that the proposed Capio Slim Device met the applicable design and performance requirements and support a determination of substantial equivalence under the specified testing parameters. These tests include:
- Sterilization Validation
- . Package integrity testing following real-time aging
- o Dve Penetration
- o Seal Strength
- o Visual Inspection
- Dimensional Test ●
- Functional Test ●
- Compression Test
- Tip Tensile Test ●
- Side Load Deflection
- Top Load Deflection ●
- Needle Retention Test ●
- Catch Pull Test
- Actuation Test
- Distribution Challenge of Packaging
- Biocompatibility ●
- o Cytotoxicity
- o Sensitization
- Irritation or Intracutaneous Reactivity O
- o Acute systemic toxicity
- o Material-Mediated Pyrogenicity
- Shelf-life Testing ●
The results of performance testing demonstrate the Capio SLIM Suture Capturing Device is biocompatible, has sufficient mechanical performance for its intended use, and supports a three-year shelf life.
## J. Conclusion
The Capio SLIM is substantially equivalent and complies with the special controls outlined in 21 CFR 884.4910, Specialized surgical instrumentation for use with urogynecologic surgical mesh.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.