Treatment of dentinal hypersensitivity. For use in adults over the age of 21.
Device Story
Riva Star is a two-component dental desensitizing system; consists of silver diamine fluoride liquid (Step 1) and aqueous potassium iodide liquid (Step 2). Applied by clinician to teeth using applicator brush. Principle of operation: chemical reaction between components forms immediate silver iodide precipitate; precipitate physically occludes open dentinal tubules to reduce fluid flow and sensitivity. Used in dental clinic setting. Output is clinical reduction of tooth sensitivity. Benefits patient by alleviating pain associated with dentinal hypersensitivity.
Clinical Evidence
Double-blind randomized clinical trial comparing Riva Star to Super Seal in patients with bilateral tooth sensitivity. Riva Star applied to one side, Super Seal to the other. Primary endpoint: reduction in sensitivity measured via Visual Analogue Scale (VAS) at 1 week. Results showed Riva Star was more effective than Super Seal in reducing dentine hypersensitivity after 7 days.
Technological Characteristics
Two-component liquid system: silver diamine fluoride and aqueous potassium iodide. Tubule occlusion via silver iodide precipitate. Biocompatibility tested per ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-11, and ISO 7405. Shelf life 25 months at 2ºC-8ºC.
Indications for Use
Indicated for treatment of dentinal hypersensitivity in adults over age 21.
Regulatory Classification
Identification
Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.
Special Controls
*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
SDI Limited Rav Cahill Chief Quality and Compliance Officer 3-13 Brunsdon Street Bayswater, Victoria 3153, Australia
Re: K172047
Trade/Device Name: Riva Star Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity Varnish Regulatory Class: Class II Product Code: PHR Dated: March 2, 2018 Received: March 7, 2018
Dear Ray Cahill:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
April 11, 2018
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mary-S. Runner -S
forTina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172047
Device Name Riva Star
Indications for Use (Describe) Treatment of dentinal hypersensitivity. For use in adults over the age of 21.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary K172047
This 510(k) summary is submitted in accordance with the requirements of 21 CFR 807.92.
### 1. Submitters information:
SDI Limited
3-13 Brunsdon Street, Bayswater, Victoria 3153, Australia
### 2. Contact information:
Rav Cahill Chief Quality and Compliance Officer Phone: +61 3 8727 7111 Fax: +61 3 8727 7222 Email:ray.cahill@sdi.com.au
### 3. Device Information:
| Proprietary Name: | Riva Star |
|----------------------|--------------------------|
| Common name: | Tooth desensitiser |
| Classification Name: | Cavity varnish |
| Regulation Number: | 21 CFR 872.3260 |
| Product Code: | PHR |
| Classification: | Medical device, Class II |
#### 4. Date summary prepared: 2 March 2018
### 5. Primary Predicate device:
Riva Star is substantially equivalent to predicate device Advantage Arrest (K102973) by Elevate Oral Care, product code PHR.
### 6. Reference predicate devices:
Super Seal (K120109), D/Sense (K120176), Remesense (K082594), with Product Code LBH.
#### 7. Indications for Use:
Treatment of dentinal hypersensitivity. For use in adults over the age of 21.
#### 8. Purpose and description Device Description and technological characteristics:
Riva Star is a two-component device made up of Riva Star Step 1, containing silver diamine fluoride liquid, and Riva Star Step 2, containing aqueous potassium iodide liquid.
Riva Star Step 1 silver diamine fluoride solution is applied to the tooth, immediately followed by Riva Star Step 2 aqueous potassium iodide, which forms an immediate reaction precipitate of silver iodide, thereby reducing tooth sensitivity by physically occluding the open dentinal tubules.
{4}------------------------------------------------
Riva Star as well as predicates are liquids or gels that are applied to the teeth using an applicator brush or similar application. With D/Sense, however, the product is in contact with the teeth by foam strips impregnated with the solution.
### 9. Comparison of technological characteristics:
The fundamental principle and mode of action of Riva Star and the predicate devices in reducing dentinal hypersensitivity is the occlusion of open dentin tubules by the formation of a precipitate.
Both Riva Star and primary predicate Advantage Arrest employ silver diamine fluoride as the main chemical to form the precipitate to occlude the open dentinal tubules. Riva Star, however, also uses a second component, potassium iodide solution, which allows for an immediate reaction precipitate of silver iodide to be formed to occlude the open dentinal tubules. Advantage Arrest, a one component device containing silver diamine fluoride, reacts with the hydroxyapatite of the tooth component. Upon reaction with the silver diamine fluoride, the calcium and phosphate of the hydroxyapatite are converted into calcium fluoride and silver phosphate, respectively to occlude the open dentinal tubules.
With the reference predicate devices Super Seal, D/Sense and Remesense, the oxalate reacts with the dentin calcium to form a deposit of calcium oxalate crystals to occlude the open dentinal tubules and thus reduce tooth sensitivity.
### 10. Comparison of Indications for Use:
This device is indicated for the treatment of dentinal hypersensitivity, for use in adults over the age of 21.
Its Indications for Use are identical to predicate device Advantage Arrest.
Although Riva Star's indications for use are not identical to the reference predicate devices. the differences in phrasing does not change the overall intended purpose as a tooth desensitiser.
#### 11. Comparison to the Predicate Devices:
A comparison of Riva Star with the predicate devices is as follows:
| | Subject device | Primary<br>predicate device | Reference<br>predicate device | Reference<br>predicate<br>device<br>D/Sense | Reference<br>predicate<br>device<br>Remesense |
|--------------------------------|-----------------------|--------------------------------------------------------------------------------|-------------------------------|---------------------------------------------|-----------------------------------------------|
| | RIVA STAR | ADVANTAGE<br>ARREST | Super Seal | | |
| 510(k) number | K172047 | K102973 | K120109 | K120176 | K082594 |
| Company | SDI Limited | Elevate Oral Care<br>(named as ADP<br>Silver Dental<br>Arrest, LLC in<br>510k) | Phoenix Dental,<br>Inc | Centrix, Inc. | Remedent |
| Classification<br>name | Cavity varnish | Cavity varnish | Cavity varnish | Cavity varnish | Cavity varnish |
| Common name | Tooth<br>desensitiser | Tooth<br>desensitiser | Tooth<br>desensitiser | Tooth<br>desensitiser | Tooth<br>desensitiser |
| Regulation<br>number | 872.3260 | 872.3260 | 872.3260 | 872.3260 | 872.3260 |
| Classification<br>Product Code | PHR | PHR | LBH | LBH | LBH |
| Indications for | Treatment of | Treatment of | Assists in the | Dual-action | Relief of tooth |
{5}------------------------------------------------
| Use | dentinal<br>hypersensitivity.<br>For use in adults<br>over the age of<br>21 | dentinal<br>hypersensitivity.<br>For use in adults<br>over the age of 21 | removal of the<br>smear layer, seals<br>the tubules,<br>desensitizes and<br>functions as a re-<br>wetting agent. | dentin<br>desensitizer<br>and cavity<br>liner. | sensitivity |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technical<br>method /<br>characteristics: | Silver diamine<br>fluoride and<br>potassium<br>iodide form a<br>precipitate of<br>silver iodide to<br>block open<br>dentinal tubules. | Silver diamine<br>fluoride forms<br>precipitates with<br>calcium or<br>phosphate in the<br>dentinal tubules to<br>block open<br>dentinal tubules | Demineralises the<br>organic and<br>mineral debris of<br>the smear layer<br>and the outer<br>most ring of<br>peritubular dentin<br>(the very hard<br>mineralized dentin<br>of each tubule<br>complex) and<br>restructures the<br>demineralized<br>material as a<br>calcium oxalate<br>precipitate. It<br>creates an acid<br>resistant layer<br>bound both to the<br>surface as well as<br>into the dentinal<br>tubules. | Reacts with<br>the smear later<br>to precipitate<br>micro crystals<br>of calcium<br>oxalate and<br>potassium<br>nitrate. These<br>crystals<br>penetrate<br>deeply into the<br>tubules, and<br>seal the entire<br>dentinal<br>surface with a<br>continuous,<br>acid-resistant<br>complex. | Potassium<br>oxalate breaks<br>down into<br>potassium and<br>oxalic acid. The<br>oxalic acid<br>reacts with the<br>calcium ions to<br>form calcium-<br>oxalate crystals.<br>These crystals<br>block the dentin<br>tubules, thereby<br>alleviating<br>dental<br>sensitivity. |
| Mode of action | Tubule<br>occlusion | Tubule occlusion | Tubule occlusion | Tubule<br>occlusion | Tubule<br>occlusion |
| Material<br>composition: | Silver<br>diamine<br>fluoride,<br>Aqueous<br>Potassium<br>lodide | Silver<br>diamine<br>fluoride | Oxalic<br>acid,<br>potassium salt | Potassium<br>binoxalate,<br>nitric<br>acid,<br>water. | Water, Glycerin,<br>Dipotassium<br>Oxalate, Aroma,<br>EDTA, BHT,<br>Sodium<br>Hydroxide |
| Application: | liquid | liquid | gel | liquid | Tray/kit/gel |
| Rx / OTC | Rx | Rx | Rx | Rx | Rx |
#### 12. Non-clinical performance testing:
A Hydraulic Conductance study of Riva Star and Super Seal tooth desensitisers was conducted. The data demonstrated that Riva Star is an effective agent to reduce fluid flow through dentine, and is substantially equivalent to Super Seal in treating dentinal hypersensitivity. Scanning electron microscope (SEM) images showed both Riva Star and Super Seal formed a precipitate to occlude the open dentinal tubules.
A comparative 7-day in vitro fluoride and silver ion release study was conducted on Riva Star and the primary predicate device Advantage Arrest. The data showed that Riva Star released less fluoride and silver ions compared to Advantage Arrest, demonstrating that Riva Star is as safe as Advantage Arrest.
Biocompatibility testing was conducted according to ISO 10993-1:2009 - Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process, ISO 7405:2008 – Dentistry – Evaluation of biocompatibility of medical devices used in dentistry, and FDA General Guidance on the Use of International Standard ISO 10993-1 (2016).
The tests carried out for Riva Star and standards applied for each test are as follows: - Cytotoxicity (ISO Agarose, ISO Direct contact) - ISO 10993-5; ISO 7405
{6}------------------------------------------------
- Sensitisation (ISO Guinea Pig Maximization Sensitization) - ISO 10993-10
- Oral Mucosal Irritation (Hamsters, collar method, 7 day) - ISO 10993-10
- Oral Acute Toxicity (Mice - 7 day) - ISO 10993-11.
Riva Star has limited contact with dentine, being an externally communicating device.
The test carried out for predicate device Advantage Arrest and standards applied is as follows: - Cytotoxicity (ISO Direct contact) - ISO 10993-5; ISO 7405
Riva Star and predicate device Advantage Arrest cytotoxicity testing produced equivalent results. Riva Star testing for irritation, sensitisation and acute oral toxicity provides evidence of biocompatibility.
Stability testing of Riva Star was conducted by evaluating the physical properties of the device to confirm a shelf life of 25 months when stored between 2ºC (35ºF) to 8ºC (45ºF).
#### 13. Clinical performance testing:
A double blind randomised clinical trial on patients with recognisable tooth sensitivity on both sides of their upper arch was performed, with Riva Star applied to the sensitive tooth on one arch, and predicate device Super Seal applied to the sensitive tooth on the other arch, and patients responses were recorded initially and following 1 week, on a visual analogue scale (VAS). The study showed that Riva Star was more effective than Super Seal in reducing dentine hypersensitivity after 7 days.
#### 14. Conclusion:
The information presented in this submission, including compositions for use, mode of action and technological characteristics, biocompatibility, non-clinical and clinical performance testing, establishes that Riva Star is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.