K172040 · W.O.M. World of Medicine GmbH · HIG · Feb 5, 2018 · Obstetrics/Gynecology
Device Facts
Record ID
K172040
Device Name
Aquilex Fluid Control System AQL-100S
Applicant
W.O.M. World of Medicine GmbH
Product Code
HIG · Obstetrics/Gynecology
Decision Date
Feb 5, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.1700
Device Class
Class 2
Indications for Use
The Aquilex® Fluid Control System AOL-100S is intended to provide fluid distension of the uterus during diagnostic and operative hysteroscopy and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus.
Device Story
Aquilex AQL-100S is a microprocessor-controlled fluid management system for hysteroscopy; comprises irrigation pump unit (AQL-110P) and fluid monitoring unit (AQL-100CBS) on a roller stand. System uses peristaltic pump to deliver irrigation fluid; monitors fluid deficit by measuring inflow and outflow volumes via integrated bag and container scales. Features include fluid bag/container change detection and bag deflector. Suction pump provides up to 225 mmHg pressure for fluid removal. Operated by clinicians in clinical settings. Output includes real-time fluid volume monitoring; assists clinicians in maintaining fluid balance during procedures to prevent complications like fluid overload. System requires sterile single-use irrigation/outflow tube sets and non-sterile vacuum tube sets.
Clinical Evidence
Bench testing only. Testing demonstrated accuracy of fluid deficit determination using bag and container scales compared to secondary predicate. Electrical safety, EMC, biocompatibility, and sterilization validation performed per recognized standards.
Technological Characteristics
Microprocessor-controlled peristaltic pump; fluid monitoring via load cells/scales. Components: irrigation pump, fluid monitoring unit, tube sets. Connectivity: standalone. Sterilization: Ethylene Oxide (SAL ≤ 10^-6). Standards: IEC 60601-1, IEC 60601-1-2, ISO 10993 series, ISO 11135-1, ISO 14937, ISO 11607-1, ASTM F1980.
Indications for Use
Indicated for patients undergoing diagnostic and operative hysteroscopy requiring uterine fluid distension and fluid deficit monitoring.
Regulatory Classification
Identification
A hysteroscopic insufflator is a device designed to distend the uterus by filling the uterine cavity with a liquid or gas to facilitate viewing with a hysteroscope.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the agency's acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 5, 2018
W.O.M. World of Medicine GmbH % Susanne Raab Regulatory Consultant Susanne Raab Guenthersburgallee 75 60389 Frankfurt GERMANY
K172040 Re:
Trade/Device Name: Aquilex® Fluid Control System AOL-100S Regulation Number: 21 CFR 884.1700 Regulation Name: Hysteroscopic insufflator Regulatory Class: Class II Product Code: HIG Dated: January 3, 2018 Received: January 10, 2018
Dear Susanne Raab:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172040
Device Name Aquilex® Fluid Control System AQL-100S
#### Indications for Use (Describe)
The Aquilex® Fluid Control System AQL-100S is intended to provide fluid distension of the uterus during diagnostic and operative hysteroscopy and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | □ |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with a circular shape at the top and the letters 'WOM' at the bottom. Inside the circle, there is a stylized design that resembles a figure eight or an infinity symbol. A small, red triangle is positioned at the bottom of the circular design, adding a pop of color to the logo. The overall design is simple and modern, with a focus on clean lines and geometric shapes.
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
### General Information:
| Submitter: | W.O.M. WORLD OF MEDICINE GmbH<br>Salzufer 8<br>10587 Berlin<br>Germany<br>Phone: +49 30 39981 594<br>Fax: +49 30 39981 593 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Registration Number: | 3001556604 |
| Contact Person: | Susanne Raab<br>Regulatory Consultant<br>Guenthersburgallee 75<br>60389 Frankfurt<br>Phone: 0049 69 28606161<br>e-mail: sbraab@comcast.net |
| Date Prepared: | January 30, 2018 |
| Proposed Device: | |
| Trade Name: | Aquilex® Fluid Control System AQL-100S |
| Common Name: | Hysteroscopic Insufflator, Fluid Monitoring<br>System and Tube Sets |
| Classification Name: | Hysteroscopic Insufflator and Accessories |
| Regulation Number: | 21 CFR 884.1700 |
| Regulation Name: | Hysteroscopic Insufflator |
| Regulatory Class: | II |
| Product Code: | HIG |
| Predicate Devices: | |
| Primary Predicate Device: | |
| Trade Name: | Aquilex Fluid Control System H112 |
| 510(k) Number: | K112642 |
| Applicant: | W.O.M. WORLD OF MEDICINE GmbH |
| Regulation Number: | 21 CFR 884.1700 |
| Regulatory Class: | II |
| Product Code: | HIG |
{4}------------------------------------------------
#### Secondary Predicate Device:
| Trade Name: | Hysteroscopy Pump HM6 |
|--------------------|-------------------------------|
| 510(k) Number: | K123732 |
| Applicant: | W.O.M. WORLD OF MEDICINE GmbH |
| Regulation Number: | 21 CFR 884.1700 |
| Regulatory Class: | II |
| Product Code: | HIG |
The predicate devices were not subject to a design-related recall.
#### Device Description:
The Aquilex® Fluid Control System AQL-100S is a modified version of the primary predicate device, Aquilex Fluid Control System H112 (K112642). The proposed device is a microprocessor-controlled device that consists of the following two main components: (1) an irrigation pump unit including suction pumps (AQL-110P) and (2) and fluid monitoring unit (AQL-100CBS) that are to be placed on a roller stand. The irrigation pump unit (AQL-110P) of the Aquilex® Fluid Control System AQL-100S is a microprocessor-controlled device that functions according to the peristaltic principle and consists of the following components: (1) a casing. (2) a power switch, (3) a power supply, (4) mains cable, (5) a roller wheel, (6) a pump head, (7) suction pumps, (8) various setting keys and (9) display elements. The irrigation pump unit (AQL-110P) is to be used with specially designed single use irrigation and outflow tube sets that are delivered sterile (AQL-110 and AQL-111). In addition, the suction pumps of the irrigation pump unit are to be used with specially designed non-sterile vacuum tube sets (AQL-114). The fluid monitoring unit (AQL-100CBS) consists of the following main components: (1) two scale units, (2), a bag holder, (4) a bag deflector, (5) a container holder, and (6) a roller wheel base. The irrigation pump unit of the proposed device is only operational in conjunction with the fluid monitoring unit.
#### Intended Use / Indication for Use:
The Aquilex® Fluid Control System AOL-100S is intended to provide fluid distension of the uterus during diagnostic and operative hysteroscopy and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a logo for WOM. The logo consists of a circle with a stylized figure inside. The figure appears to be a person standing with their arms raised, and the lower part of the figure is colored in pink. The text "WOM" is written below the logo.
### Substantial Equivalence and Comparison of Technological Characteristics:
The proposed device, Aquilex® Fluid Control System AQL-100S, is substantially equivalent to the predicate devices, Aquilex Fluid Control System H112 (K112642) and Hysteroscopy Pump HM6 (K123732). As stated above the proposed device, Aquilex® Fluid Control System AQL-100S, is a modified version of the primary predicate device, Aquilex Fluid Control System H112.
The proposed device, Aquilex® Fluid Control System AQL-100S, and the predicate devices, Aquilex Fluid Control System H112 and Hysteroscopy Pump HM6, are all intended to provide fluid distension of the uterus during diagnostic and operative hysteroscopy and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus. The proposed device and the predicate devices have the same intended use.
Furthermore, tThe proposed device, Aquilex® Fluid Control System AQL-100S, and the predicate device, Aquilex Fluid Control System H112 and Hysteroscopy Pump HM6, use the same basic operating principles and incorporate the same basic design. All three devices incorporate the following main components: (1) irrigation pump, (2) fluid monitoring unit, (3) irrigation and outflow tube sets.
The modification to the Aquilex® Fluid Control System AQL-100S in comparison to the primary predicate device, Aquilex Fluid Control System H112, do not raise different questions of safety and effectiveness. The differences between both devices are the following:
- Incorporation of a separate bag scale into the fluid monitoring unit of the . Aquilex® Fluid Control System AQL-100S to calculate the inflow volume, whereas the Aquilex Fluid Control System H112 determines the inflow volume based on the number of rotation of the roller wheel.
- . Change in design of the cart to allow for a mounting of the bag scale and addition of bag deflector.
- . Incorporation of a fluid bag and container change detection feature into the Aquilex® Fluid Control System AQL-100S.
- . Increase in the suction pressure of the low suction pump from 125 to 225 mmHg.
- . The inflow tubing AQL-110 of the primary predicate device has been slightly modified.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a logo for WOM. The logo consists of a gray circle with a stylized symbol inside. The symbol resembles an infinity sign above a pink triangle. Below the circle, the letters "WOM" are written in a simple, sans-serif font.
- · The filter of the reusable vacuum tubing is manufactured by a different supplier.
With regards to the incorporation of an additional bag scale, the proposed device, Aquilex® Fluid Control System AQL-100S, is substantially equivalent to the secondary predicate device, Hystersocopy Pump HM6, that also utilizes a separate scale for determination of the inflow volume.
With regards to the incorporation of a separate bag scale into the fluid monitoring unit to determine the inflow volume and change in design of the cart to allow for a mounting of the bag scale, the proposed device, Aquilex® Fluid Control System AQL-100S, is substantial equivalent to the secondary predicate device, Hysteroscopy Pump HM6. Like the proposed device, Aquilex® Fluid Control System AQL-100S, the secondary predicate device, Hysteroscopy Pump HM6, incorporates a separate scale to determine the inflow volume for calculation of the fluid deficit.
With regards to the fluid bag change detection feature is substantially equivalent to the secondary predicate device, Hysteroscopy Pump HM6, that incorporates the same feature. The bag deflector was added to the fluid monitoring unit of the proposed device. Aquilex® Fluid Control System AOL-100S. to protect the bag scale from fluid.
The increase in the suction pressure of the suction pump from 125 to 225 mmHg, that was implemented due to end user demand to ensure fluid suction even if the tubing is draped over the patient's leg.
The modifications that were made on the inflow tubing AQL-110 are minor and have not changed the material, suppliers and manufacturing method.
# Performance Data:
#### Electrical Safety and Electromagnetic Compatibility Testing:
Electrical safety and electromagnetic compatibility testing was performed by independent laboratories and the test results demonstrate that the proposed device conforms to the below standards:
- IEC 60601-1 2005/(R)2012 and A1: 2012, C1:2009/(R)2012 andA2: ● 2010/(R)2012 (consolidated text); and
- IEC 60601-1-2 Fourth Edition: 2014.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo for WOM. The logo consists of a circle with a stylized figure inside, resembling a person with arms raised. The figure is positioned above the letters "WOM" in a sans-serif font. The figure is colored in pink, while the circle and the letters are in gray.
### Software:
The software was considered as major level of concern in accordance with the FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". Software verification and validation testing of the Aquilex® Fluid Control System AQL-100S demonstrates that the device performs as intended. It was developed, tested and verified in accordance with the above FDA guidance document and in accordance with the following standard:
- IEC 62304 First Edition 2006-05. .
# Biocompatibility:
Biocompatibility testing was performed in accordance with:
- AAMI/ANSI/ISO 10993-1 Fourth Edition 2009-10-15 Biological . evaluation of medical devices- Evaluation and testing within a risk management system;
- . ISO 10993-5: 2009 (R) 2014 - Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity;
- . ISO 10993-10 Third Edition 2010-08-01 - Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization; and
- . ISO 10993-11 Second Edition 2006-08-15- Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity;
- . ISO 10993-12:2007 Biological evaluation of medical devices -- Part 12: Sample preparation and reference materials.
# Sterilization and Shelf Life:
Sterilization validation was performed in accordance with:
- . ISO 10993-7 Second Edition 2008-10-15- Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals;
- . AAMI TIR 28 2009/(R)2013. Product adoption and process equivalency for ethylene oxide sterilization;
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows a logo for WOM. The logo consists of a circle with a stylized symbol inside, resembling a person with arms raised. The symbol is partially colored in pink. Below the symbol, the letters "WOM" are written in a sans-serif font.
- ISO 11135-1:2007 Sterilization of health care products Ethylene oxide - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices;
- . ISO 14937 Second Edition 2009-10-15- Sterilization of health care products - General criteria for characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical devices.
Residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) data shows that the limit of EO < 4 mg and ECH < 5 mg after 4 days of aeration (gas release) that remain on the tube set will not be exceeded. As a result of the sterilization validation, a sterility assurance level (SAL) ≤ 106 in accordance with ANSI/AAMI/ISO 11135-1 was achieved. Package and product integrity of the tube sets were tested in accordance with ISO11607-1 - Packaging for terminally sterilized medical devices and ASTM-F1980:2002 - Standard for accelerated aging of sterile medical device packages.
# Bench Testing:
Bench testing was performed to demonstrate that the fluid deficit determination of the proposed device, Aquilex® Fluid Control System AQL-100S, using a bag and container scale, is substantially equivalent to that of the secondary predicate device, Hysteroscopy Pump HM6, in terms of accuracy.
# Conclusion:
The Aquilex® Fluid Control System AQL-100S is substantially equivalent to the predicate devices: Aquilex Fluid Control System H112 and Hysteroscopy Pump HM6.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.