# Medrobotics Flex Robotic System (K172036)

_Medrobotics Corporation · FDF · Aug 3, 2017 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K172036

## Device Facts

- **Applicant:** Medrobotics Corporation
- **Product Code:** [FDF](/productcode/FDF.md)
- **Decision Date:** Aug 3, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Medrobotics Flex Robotic System is intended to provide robot-assisted control of the Flex Colorectal Drive during visualization of and surgical site access to the anus, rectum and distal colon. The Flex Robotic System is intended for use in adults (≥22 years of age). The Flex Colorectal Drive is intended for robot-assisted visualization of and surgical site access to the anus, rectum, and distal colon in adults (≥22 years of age). The Flex Colorectal Drive also provides accessory channels for compatible flexible instruments used in surgery.

## Device Story

The Medrobotics Flex Robotic System is an operator-controlled flexible endoscopic platform combining rigid scope benefits with computer-assisted control. It facilitates transanal access to the anus, rectum, and distal colon. The system utilizes a user-selectable 2D or 3D HD camera at the distal end for visualization and includes accessory channels for flexible surgical instruments. Operated by a physician, the system provides robot-assisted control via a user interface to navigate the anatomy. The output is high-definition visual feedback, enabling the surgeon to perform procedures with enhanced access and visualization. The device is intended for use in a clinical/surgical setting to assist in surgical site access, potentially improving procedural outcomes for adult patients.

## Clinical Evidence

Bench testing only. The system was tested for function, performance, and safety per IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-18. Biocompatibility and toxicity testing were performed per ISO 10993-1. Shelf life and shipping validation were conducted per FDA-recognized standards. Sterility was validated for ETO and Steam to a SAL of 10^-6.

## Technological Characteristics

Flexible robotic endoscope system with 2D/3D HD camera. Materials tested per ISO 10993-1. Sterilization via ETO and Steam (SAL 10^-6). Complies with IEC 60601-1 (General safety), IEC 60601-1-2 (EMC), and IEC 60601-2-18 (Endoscopic equipment).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Medrobotics Flex Robotic System ([K162330](/device/K162330.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, arranged in a stacked formation.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 3, 2017

Medrobotics Corporation John D. Bonasera Vice President of Clinical. Regulatory and Ouality Affairs 475 Paramount Drive Raynham, MA 02767

Re: K172036

> Trade/Device Name: Medrobotics Flex Robotic System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDF Dated: July 5, 2017 Received: July 5, 2017

Dear John D. Bonasera:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Charles Viviano -S

For

Beniamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K172036

Device Name Medrobotics Flex Robotic System

#### Indications for Use (Describe)

The Medrobotics Flex Robotic System is intended to provide robot-assisted control of the Flex Colorectal Drive during visualization of and surgical site access to the anus, recturn and distal colon. The Flex Robotic System is intended for use in adults (≥22 years of age).

The Flex Colorectal Drive is intended for robot-assisted visualization of and surgical site access to the anus, rectum, and distal colon in adults (≥22 years of age). The Flex Colorectal Drive also provides accessory channels for compatible flexible instruments used in surgery.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary Medrobotics Flex Robotic System

| Submitter's Name:        | Medrobotics Corporation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Address:     | 475 Paramount Drive<br>Raynham, MA 02767                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Contact Person:          | John D. Bonasera                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Phone Number:            | 508-692-6460                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Fax Number:              | 508-823-1703                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Date Prepared:           | July 5, 2017                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Device Trade Name:       | Medrobotics Flex® Robotic System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Device Common Name:      | Endoscope/Endoscope and Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Product Code:            | FDF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Classification:          | The Medrobotics Flex Robotic System is classified as Class II per 21 CFR §<br>876.1500                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Predicate Device:        | Medrobotics Flex Robotic System (K162330)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Device Description:      | The Medrobotics Flex Robotic System is an operator controlled flexible scope<br>that include the benefits of both a rigid scope and a computer assisted<br>controller.  This allows for the Flex Colorectal Drive to be introduced via an<br>operator controlled user interface, easily providing transanal access to the<br>anus, rectum and distal colon.  Visualization is provided by a user selectable<br>2D or 3D HD camera incorporated in distal end of the scope. The Flex<br>Robotic System's scope also provides accessory channels for the use of varied<br>flexible surgical instruments. |
| Intended Use:            | The Medrobotics Flex Robotic System is intended to provide robot-assisted<br>control of the Flex Colorectal Drive during visualization of and surgical site<br>access to the anus, rectum and distal colon. The Flex Robotic System is<br>intended for use in adults (≥22 years of age).<br>The Flex Colorectal Drive is intended for robot-assisted visualization of and<br>surgical site access to the anus, rectum, and distal colon in adults (≥22 years<br>of age). The Flex Colorectal Drive also provides accessory channels for<br>compatible flexible instruments used in surgery.            |
| Performance Data:        | The Flex Robotic system has been successfully tested for function,<br>performance, and safety as per FDA recognized Standards: IEC 60601-1,<br>IEC 60601-1-2, IEC 60601-2-18, and Biocompatibility and Toxicity testing of<br>the patient contacting materials to ISO 10993-1. It has been tested and met<br>acceptance criteria per FDA recognized standards for the establishment of<br>labeled shelf life and shipping, and validated for sterility by ETO and Steam<br>to a SAL of 10-6.                                                                                                           |
| Substantial Equivalence: | The Medrobotics Flex Robotic System is substantially equivalent to and has<br>the intended use as the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |

---

**Source:** [https://fda.innolitics.com/device/K172036](https://fda.innolitics.com/device/K172036)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).