Gebauer’s Pain Ease Topical Anesthetic Skin Refrigerant (Mist Spray and Medium Spray)

K172028 · Gebauer Company · MLY · Oct 27, 2017 · PM

Device Facts

Record IDK172028
Device NameGebauer’s Pain Ease Topical Anesthetic Skin Refrigerant (Mist Spray and Medium Spray)
ApplicantGebauer Company
Product CodeMLY · PM
Decision DateOct 27, 2017
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic, Pediatric

Indications for Use

Gebauer's Pain Ease Topical Anesthetic Skin Refrigerant (Mist Spray and Medium Spray): a vapocoolant (skin refrigerant) intended for topical application to skin, mucous membranes and minor open wounds. Gebauer's Pain Ease controls pain associated with minor surgical procedures (such as lancing boils, incisions, drainage of small abscesses, and sutures), injections (venipuncture, IV starts, cosmetic procedures) and the temporary relief of minor sports injuries (sprains, bruising, cuts and abrasions). The Medium Spray is also intended for the treatment of myofascial pain caused by trigger points, restricted motion and muscle tension.

Device Story

Prescription vapocoolant spray; delivers HFC 245fa and HFC 134a mixture via pressurized aerosol container. Cooling effect achieved through rapid evaporation of volatile propellants on skin. Applied directly via spray or via saturated cotton ball/swab/gauze. Used in clinical settings for pre-injection anesthesia, minor surgery, and sports injury relief; Medium Spray used for myofascial pain/trigger points. Provides temporary pain relief via thermal energy reduction. No software or electronic components.

Clinical Evidence

Bench testing only. Comparative study performed to evaluate temperature and output performance between direct spray application and application via cotton ball, cotton swab, or gauze. Results confirmed safety and effectiveness profile equivalent to predicate.

Technological Characteristics

Pressurized aerosol dispensing system; non-flammable HFC 245fa and HFC 134a propellant mixture. Mechanical positive shut-off valve. Biocompatibility per ISO 10993; microbial limits per USP <61> and <62>. Boiling point 44.6°F (7.0°C).

Indications for Use

Indicated for topical pain control during minor surgical procedures, injections, and minor sports injuries. Medium Spray also indicated for myofascial pain, trigger points, and muscle tension. Use on skin, mucous membranes, and minor open wounds. Not recommended for patients under four years old without pediatrician consultation.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 27, 2017 Gebauer Company Brittney Schmies Regulatory Affairs Specialist 4444 East 153rd Street Cleveland, Ohio 44128 Re: K172028 Trade/Device Name: Gebauer's Pain Ease Topical Anesthetic Skin Refrigerant (Mist Spray and Medium Spray) Regulatory Class: Unclassified Product Code: MLY Dated: September 25, 2017 Received: September 26, 2017 Dear Ms. Schmies: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. Vivek J. Pinto -S for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K172028 ### Device Name Gebauer's Pain Ease Topical Anesthetic Skin Refrigerant (Mist Spray and Medium Spray) ### Indications for Use (Describe) Gebauer's Pain Ease Topical Anesthetic Skin Refrigerant (Mist Spray and Medium Spray): a vapocoolant (skin refrigerant) intended for topical application to skin, mucous membranes and minor open wounds. Gebauer's Pain Ease controls pain associated with minor surgical procedures (such as lancing boils, incisions, drainage of small abscesses, and sutures), injections (venipuncture, IV starts, cosmetic procedures) and the temporary relief of minor sports injuries (sprains, bruising, cuts and abrasions). The Medium Spray is also intended for the treatment of myofascial pain caused by trigger points, restricted motion and muscle tension. Type of Use (Select one or both, as applicable): | <span style="font-family: sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |---------------------------------------------------------------------------------------------------------------------------------------| | <span style="font-family: sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows a black and white drawing of a spiral shape. The spiral has a hole in the center and a small extension on the right side. The spiral is filled with small dots, giving it a textured appearance. There is also a small circle with a dot in the center to the right of the spiral. # GEBAUER COMPANY # 510(k) Summary - K172028 This 510(k) Summary is being submitted in accordance of 21 CFR Part 807.92. #### I. SUBMITTER | Owner: | Gebauer Company<br>4444 East 153rd Street<br>Cleveland, OH 44128<br>(216) 581-3030 | |-----------------|--------------------------------------------------------------------------------------| | Contact Person: | Brittney Schmies<br>Gebauer Company<br>4444 East 153rd Street<br>Cleveland, OH 44128 | Date Prepared: October 27, 2017 (216) 581-3030 #### II. DEVICE Trade Name: Skin Refrigerant (Mist Spray and Medium Spray) Gebauer's Pain Ease Topical Anesthetic Refrigerant Topical, Vapocoolant Common Name: Classification Name: Product Code: MLY Cold Spray #### III. PREDICATE DEVICE: | Primary: | Gebauer's Pain Ease (Mist Spray and<br>Medium Spray) | |----------|------------------------------------------------------| | | K032671 | | | Legally marketed medical device | {4}------------------------------------------------ #### IV. DESCRIPTION Gebauer's Pain Ease (Mist and Medium Spray) Topical Anesthetic Skin Refrigerant is a prescription device designed to deliver HFC 245fa high purity (1,1,1,3,3-Pentafluoropropane) and HFC 134a pharmaceutical grade (1,1,1,2-Tetrafluoroethane) in a mist and medium spray. This mixture self-propels itself from the delivery system. which is designed to account for its low vapor pressure. The device delivery system is specifically designed to deliver a medium and mist spray of the Gebauer's Pain Ease (Mist Spray and Stream Spray) mixture. The medium and mist spray is an appropriate mode of application when users follow directions for use, cooling the skin through rapid evaporation of the non-medicated volatile propellants. The new device, Gebauer's Pain Ease, is identical in all aspects to the predicate device, Gebauer's Pain Ease (Mist Spray and Medium Spray) 510(k) K032671, except that the product can be applied to the skin for pre-injection anesthesia by cotton swab or gauze on intact skin. (If the skin is breached use this application method only with STERILE, disposable cotton balls, cotton swabs or gauze. The cotton balls, cotton swabs or gauze are not supplied with the device.) Both the new and predicate devices are indicated for use to control pain for pre-injection anesthesia, minor surgery, and minor sports injuries. The Medium Spray for the new device, like the predicate medium spray device, is also indicated for the management of myofascial pain caused by trigger points. restricted motion and muscle tension by using the Spray and Stretch® technique. #### V. INDICATIONS FOR USE The indications for Use are the same as the predicate device. Gebauer's Pain Ease Topical Anesthetic Skin Refrigerant (Mist Spray and Medium Spray): a vapocoolant (skin refrigerant) intended for topical application to skin, intact mucous membranes (oral cavity, nasal passage ways and the lips) and minor open wounds. Gebauer's Pain Ease controls pain associated with minor surgical procedures (such as lancing boils, incisions, drainage of small abscesses, and sutures), injections (venipuncture. IV starts, cosmetic procedures) and the temporary relief of minor sports injuries (sprains, bruising, cuts and abrasions). The Medium Spray is also intended for the treatment of myofascial pain caused by trigger points, restricted motion and muscle tension. # VI. - TECHNICAL SUMMARY As with the predicate device, the cooling action experienced by the patient is caused by the evaporation of the chemical mixture from the patient's skin. The user applies pressure to the nozzle to dispense the aerosol product onto the skin, or saturates a {5}------------------------------------------------ cotton ball, cotton swab or gauze with the product and applies to the skin. The material is contained in a can, filled under pressure, and dispensed using standard aerosol nozzle technology. See table below for the comparison between the subject device and the predicate: | Comparison Chart – Technological Characteristics | | | |--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Trade Name | Gebauer's Pain Ease<br>Predicate | Gebauer's Pain Ease<br>Subject Device | | Product Design | Pressurized dispensing<br>container which includes<br>the vapocoolant, aerosol<br>can, valve and actuator. | Pressurized dispensing<br>container which includes<br>the vapocoolant, aerosol<br>can, valve and actuator. | | Intended Use | A vapocoolant (skin<br>refrigerant) intended for<br>topical application to<br>skin, mucous<br>membranes and minor<br>open wounds.<br>Gebauer's Pain Ease<br>controls pain associated<br>with minor surgical<br>procedures (such as<br>lancing boils, incisions,<br>drainage of small<br>abscesses, and sutures),<br>injections<br>(venipuncture, IV starts,<br>cosmetic procedures)<br>and the temporary relief<br>of minor sports injuries<br>(sprains, bruising, cuts<br>and abrasions). The<br>Medium Spray is also<br>intended for the<br>treatment of myofascial<br>pain caused by trigger<br>points, restricted motion<br>and muscle tension. | A vapocoolant (skin<br>refrigerant) intended for<br>topical application to<br>skin, mucous<br>membranes and minor<br>open wounds.<br>Gebauer's Pain Ease<br>controls pain associated<br>with minor surgical<br>procedures (such as<br>lancing boils, incisions,<br>drainage of small<br>abscesses, and sutures),<br>injections<br>(venipuncture, IV starts,<br>cosmetic procedures)<br>and the temporary relief<br>of minor sports injuries<br>(sprains, bruising, cuts<br>and abrasions). The<br>Medium Spray is also<br>intended for the<br>treatment of myofascial<br>pain caused by trigger<br>points, restricted motion<br>and muscle tension. | | Product Fill Volume | 3.5 oz. (103.5 mL) | 3.5 oz. (103.5 mL) | | Vapocoolant Composition | 1,1,1,3,3<br>Pentafluoropropane (HFC<br>245fa high purity) and<br>1,1,1,2 Tetrafluoroethane | 1,1,1,3,3<br>Pentafluoropropane (HFC<br>245fa high purity) and<br>1,1,1,2 Tetrafluoroethane | | | (HFC 134a pharmaceutical grade). | (HFC 134a pharmaceutical grade). | | Energy Delivered | Thermal energy via<br>Refrigerant Spray. | Thermal energy via<br>Refrigerant Spray. | | Vapocoolant Discharge<br>Method | Depress the actuator to<br>release the vapocoolant. | Depress the actuator to<br>release the vapocoolant | | Method of Application | Direct spray application to<br>intact skin, intact oral<br>mucous membranes and<br>minor open wounds. | Direct spray application to<br>intact skin, intact oral<br>mucous membranes, and<br>minor open wounds, or<br>application via cotton ball,<br>cotton swab or gauze on<br>intact skin only. If the skin<br>is breached this application<br>method is to be used only<br>with STERILE, disposable<br>cotton balls, cotton swabs<br>or gauze. | | Environmental<br>Compatibility | Non-Flammable. | Non-Flammable. | | Mechanical Safety | Mechanism has positive<br>shut-off release. | Mechanism has positive<br>shut-off release. | | Manufacturing<br>Environment | Controlled Environment | Controlled Environment | | Microbial Limits Testing | Tested in accordance with<br>USP <61> and <62>. | Tested in accordance with<br>USP <61> and <62>. | | Biocompatibility Testing | In accordance with ISO<br>10993 for dermal irritation,<br>sensitization, cytotoxicity,<br>oral mucosal irritation,<br>acute dermal toxicity. | In accordance with ISO<br>10993 for dermal irritation,<br>sensitization, cytotoxicity,<br>oral mucosal irritation,<br>acute dermal toxicity. | | Target Population | General | As the safety for use of the<br>device in pediatric patients<br>has not been established, it<br>is recommended the device<br>should not be used on<br>patients under four without<br>consultation of a<br>pediatrician. | | Boiling Point | 44.6°F (7.0°C) | 44.6°F (7.0°C) | | Storage Temperature | Do not store at temperatures<br>above 50°C (120°F) | Do not store at temperatures<br>above 50°C (120°F) | | Shelf Life | 3 years | 3 years | {6}------------------------------------------------ {7}------------------------------------------------ #### DETERMINATION OF SUBSTANTIAL EQUIVELANCE VII. This premarket notification 510(k) is being submitted to expand the application methods in the instructions for use for Gebauer's Pain Ease, Mist and Medium Stream. There is demonstrated equivalency in basic product design and technology, in indications for use, target population, and risk factors. As stated above, the new device is identical in formulation, delivery system and packaging to the predicate device Gebauer's Pain Ease, Mist and Medium Stream previously cleared for market under 510(k) K032671. The new device has expanded instructions for use to apply the product by either spraying directly on the skin or saturating a cotton ball, cotton swab or gauze on intact skin. (If the skin is breached use this application method only with STERILE, disposable cotton balls, cotton swabs or gauze. The cotton balls, cotton swabs or gauze are not supplied with the device.) Both the new and predicate devices are indicated for use to temporarily control pain associated with pre-injection anesthesia, minor surgery and minor sports injuries. The medium stream spray for both the predicate and new device has the identical indication for the treatment of myofascial pain caused by trigger points, restricted motion and muscle tension. #### VIII. PERFORMANCE DATA The predicate and subject devices use the same materials, design and energy source and no technological differences exist. Tests were selected and performed to ensure the subject device's performance, with the new method of application, is equivalent as the predicate. To demonstrate equivalence the following test was performed: # Side-by-Side Temperature & Output Bench Testing Comparative testing was conducted to demonstrate equivalence between direct topical application via spraving and topical application using cotton ball, cotton swab or gauze. The testing was conducted to determine the temperature and output of the predicate and subject devices ## Summarv Based on the temperature data generated, the subject device was found to have a safety and effectiveness profile that is similar to the predicate device. #### CONCLUSION IX. Based on the information above and within this submission, it is concluded that the subiect Pain Ease device is safe and effective for its intended use and is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...