← Product Code [MUE](/productcode/MUE) · K172027

# Adani MammoScan (K172027)

_Adani · MUE · Oct 20, 2017 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K172027

## Device Facts

- **Applicant:** Adani
- **Product Code:** [MUE](/productcode/MUE.md)
- **Decision Date:** Oct 20, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1715
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The MammoScan® Full-Field Digital Mammography System is a device intended to produce planar digital x-ray images of the entire breast. The MammoScan is indicated for generating mammographic images that can be used for screening and diagnosis of breast cancer. The MammoScan is intended to be used in the same clinical applications as traditional film/screen systems.

## Device Story

Adani MammoScan is a full-field digital mammography system; captures planar digital x-ray images of breast. System components: gantry, acquisition station (PC, monitor, keyboard, mouse), protective shield. Image acquisition: x-ray tube (tungsten target, aluminum filtration) generates fan-shaped beam; slit collimator shapes beam; internal swing arm pivots x-ray tube and detector in arc under breast support. Detector uses CCD-TDI technology to convert x-rays to electric signals. Acquisition workstation manages image preparation, pre/post-processing, and transfer to PACS. Used in clinical settings by technologists; gantry motorized for height/rotation adjustments to accommodate patients standing or sitting. Output images used by radiologists for breast cancer screening and diagnosis. Benefits include digital image acquisition replacing traditional film/screen methods.

## Clinical Evidence

No clinical trials performed. Evidence based on bench testing and image attribute evaluation per FDA Class II Special Controls Guidance for Full-Field Digital Mammography Systems. Bench testing included sensitometric response, spatial resolution, noise analysis, DQE, dynamic range, AEC performance, phantom testing, patient radiation dose, and image erasure/fading. Biocompatibility testing (ISO 10993) performed on compression plates. Image attribute evaluation confirmed image quality is acceptable for clinical mammographic usage.

## Technological Characteristics

Full-field digital mammography system. Detector: Cesium iodide (CsI:Tl) with 2 TDI CCDs, 27 µm native pixel size. X-ray source: Tungsten target, aluminum filtration. Generator: High-frequency (Sedecal SHF-1030-M). Connectivity: PACS integration via acquisition workstation. Software: Moderate level of concern. Sterilization: N/A (non-invasive). Standards: IEC 60601-1, IEC 60601-2-45, EN 62220-1-2, EN 62304.

## Regulatory Identification

A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.

## Predicate Devices

- SenoScan® Full-Field Digital Mammography System ([P010017](/device/P010017.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 20, 2017

ADANI % Daniel Kamm, P.E. Principal Engineer Kamm & Associates 8870 Ravello Ct. NAPLES FL 34114

Re: K172027

Trade/Device Name: Adani MammoScan Regulation Number: 21 CFR 892.1715 Regulation Name: Full-field digital mammography system Regulatory Class: II Product Code: MUE Dated: September 18, 2017 Received: September 21, 2017

Dear Mr. Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Robert Ochs

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K17 2027

Device Name Adani MammoScan

Indications for Use (Describe)

The MammoScan® Full-Field Digital Mammography System is a device intended to produce planar digital x-ay images of the entire breast. The MammoScan is indicated for generating mammographic images that can be used for screening and diagnosis of breast cancer. The MammoScan is intended to be used in the same clinical applications as traditional film/screen systems.

| Type of Use (Select one or both, as applicable)                                                 |                                             |
|-------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary K17 ADANI Selitski St., 7 Minsk, Belarus 220075, Tel. Selitski St., 7 Minsk, Belarus 220075 Tel.+375292145846, Contact: Vladimir Klokov, Vice President Date Prepared: June 4, 2017

- 1. Identification of the Device: Proprietary-Trade Name: Adani MammoScan® Classification Name: Full-Field Digital Mammography System Product Code: MUE Regulation 892.1715 Common/Usual Name: Digital Mammography System.
- Equivalent legally marketed device: P010017, SenoScan® Full-Field Digital Mammography System, Fischer 2. Imaging Corporation Denver, CO 80026. Classification Name: Full-Field Digital Mammography System Product Code: MUE Regulation 892.1715 Common/Usual Name: Digital Mammography System.
- 3. Indications for Use (intended use): The MammoScan® Full-Field Digital Mammography System is a device intended to produce planar digital x-ray images of the entire breast. The MammoScan is indicated for generating mammographic images that can be used for screening and diagnosis of breast cancer. The MammoScan is intended to be used in the same clinical applications as traditional film/screen systems.
- Description of the Device: The Adani MammoScan is a full-field type digital mammography system comprised ধ . of an image acquisition system, a gantry and an acquisition station, equipped with PC computer, keyboard, mouse, monitor and a protective shield. The image acquisition system includes a built-in digital detector of CCD-TDI technology, x-ray tube (with tungsten target and aluminum filtration), high voltage generator, compression mechanism, and slit collimator. The slit-collimator shapes the beam and limits the x-ray beam to the active width of the detector, which prevents image degradation caused by scattered radiation. During image acquisition, the internal swing arm is motor driven to pivot at the x-ray tube with the slit collimator, causing the detector (at the other end of the swing arm and on the far side of the breast support) to swing in an arc under the breast support from one edge of the breast support to the other. The x-ray, which is shaped into a flat fan-shaped beam by the slit collimator, passes through the compressed breast tissue as it sweeps across the breast support and falls on the image detector underneath. The acquisition workstation is the user interface for preparing and initiating image acquisition, image pre- and post-processing, and image transfer to the desired destination (e.g. PACS) for interpretation and archiving. The MammoScan® Full-Field Digital Mammography System consists of a high-voltage generator, gantry, an acquisition work station with a technologist shield and accessories. The patient is imaged at the gantry contains the diagnostic source (x-ray tube, filter, and collimator), built-in solid state image detector, breast support, and compression assembly. The gantry can be raised, lowered, and rotated, under motorized control, to accommodate patients of all statures, standing and/or sitting, to produce images in all standard views.
- 5. Safety and Effectiveness, comparison to predicate device. The Adani MammoScan system employs the same fundamental technological characteristics as its predicate device and The X-ray technology is substantially equivalent to the Fischer SenoScan Full Field Digital Mammography System, approved by FDA via P010017. Clinical uses for which Adani MammoScan was designed are equivalent to those cleared for Fischer SenoScan.

{4}------------------------------------------------

The Adani MammoScan and Fischer SenoScan use same technology for converting x-rays to the electric signal. The results of bench, clinical, and standards testing indicates that the new device is as safe and effective as the predicate device. Risk analysis and software validation has been performed. An examination of the substantial equivalence chart below reveals the use of nearly identical or equivalent technology throughout.

# 6. Substantial Equivalence Chart, MammoScan® vs. Predicate

| Parameter                                            | Predicate device, SenoScan®, P010017                                                                                                                                                                                                                                                                                                                                                         | New device, MammoScan®, this<br>submission                                                                                                                                                                                                                                                                                                                                                             |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications                                          | The SenoScan® Full-Field Digital Mammography<br>System is a dedicated mammography system<br>intended to produce radiographic images of the<br>human breast for the purpose of diagnostic and<br>screening mammography. The SenoScan® Full-<br>Field Digital Mammography system is intended<br>to be used in the same clinical applications as<br>traditional film-based mammographic systems | The MammoScan® Full-Field Digital<br>Mammography System is a device intended to<br>produce planar digital x-ray images of the entire<br>breast. The MammoScan is indicated for<br>generating mammographic images that can be<br>used for screening and diagnosis of breast cancer.<br>The MammoScan is intended to be used in the<br>same clinical applications as traditional<br>film/screen systems. |
| Detector Information                                 |                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                        |
| Type                                                 | Cesium iodide with 4 TDI CCD's                                                                                                                                                                                                                                                                                                                                                               | Cesium iodide with 2 TDI CCD's                                                                                                                                                                                                                                                                                                                                                                         |
| Detector area                                        | 21x1 cm                                                                                                                                                                                                                                                                                                                                                                                      | 22x0.7 cm                                                                                                                                                                                                                                                                                                                                                                                              |
| Maximum field size                                   | 22x30 cm                                                                                                                                                                                                                                                                                                                                                                                     | 22x30 cm                                                                                                                                                                                                                                                                                                                                                                                               |
| Pixel dimensions                                     | Native pixel size 27 µm<br>54 µm in normal operation mode                                                                                                                                                                                                                                                                                                                                    | Native pixel size 27 µm<br>54 µm in normal operation mode                                                                                                                                                                                                                                                                                                                                              |
| Fill factor                                          | 100 %                                                                                                                                                                                                                                                                                                                                                                                        | 100 %                                                                                                                                                                                                                                                                                                                                                                                                  |
| Matrix size                                          | 4096x5625 pixels                                                                                                                                                                                                                                                                                                                                                                             | 4096x5560 pixels                                                                                                                                                                                                                                                                                                                                                                                       |
| X-Ray interaction<br>material                        | Csl:TI                                                                                                                                                                                                                                                                                                                                                                                       | Csl:TI                                                                                                                                                                                                                                                                                                                                                                                                 |
| A/D conversion bit<br>depth                          | 12 bits                                                                                                                                                                                                                                                                                                                                                                                      | 16 bits                                                                                                                                                                                                                                                                                                                                                                                                |
| Scanning rate                                        | -                                                                                                                                                                                                                                                                                                                                                                                            | 4.6 Cm/sec                                                                                                                                                                                                                                                                                                                                                                                             |
| Decay rate of the<br>phosphor afterglow              | less than 3 µs                                                                                                                                                                                                                                                                                                                                                                               | less than 3 µs                                                                                                                                                                                                                                                                                                                                                                                         |
| Image read-out<br>mechanism.                         | CCD TDI                                                                                                                                                                                                                                                                                                                                                                                      | CCD TDI                                                                                                                                                                                                                                                                                                                                                                                                |
| Method of detector<br>cooling                        | Air/Fan                                                                                                                                                                                                                                                                                                                                                                                      | Air/Fan                                                                                                                                                                                                                                                                                                                                                                                                |
| Power source                                         | 220 VAC                                                                                                                                                                                                                                                                                                                                                                                      | 220 VAC                                                                                                                                                                                                                                                                                                                                                                                                |
| X-Ray Parameters:                                    |                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                        |
| X-Ray tube model<br>name                             | Varian RAD73                                                                                                                                                                                                                                                                                                                                                                                 | Varian RAD70 or IAE XM 1016T                                                                                                                                                                                                                                                                                                                                                                           |
| Focal Spot Size                                      | 0.3 mm                                                                                                                                                                                                                                                                                                                                                                                       | 0.3 mm                                                                                                                                                                                                                                                                                                                                                                                                 |
| Inherent filtration                                  | 0.76 mm Be                                                                                                                                                                                                                                                                                                                                                                                   | 0.76 mm Be or 0.5 mm Be                                                                                                                                                                                                                                                                                                                                                                                |
| Target material                                      | Rhenium-tungsten facing on molybdenum                                                                                                                                                                                                                                                                                                                                                        | Rhenium-tungsten                                                                                                                                                                                                                                                                                                                                                                                       |
| Heat capacity                                        | 600 KHU                                                                                                                                                                                                                                                                                                                                                                                      | 600 kHU or300 kHU                                                                                                                                                                                                                                                                                                                                                                                      |
| Parameter                                            | Predicate device, SenoScan®, P010017                                                                                                                                                                                                                                                                                                                                                         | New device, MammoScan®, this submission                                                                                                                                                                                                                                                                                                                                                                |
| Additional<br>Equivalent Filtration<br>in X-ray Beam | 0,5 mm Al in normal operation<br>2,0 mm Al for calibration – automatically selected                                                                                                                                                                                                                                                                                                          | 0,5 mm Al in normal operation<br>2,0 mm Al for calibration - automatically selected                                                                                                                                                                                                                                                                                                                    |
| Collimator operation                                 | Fixed at front and sides, motorized rear blade                                                                                                                                                                                                                                                                                                                                               | Fixed at front and sides, motorized rear blade                                                                                                                                                                                                                                                                                                                                                         |
| Projected X-Ray beam                                 | 1x22cm                                                                                                                                                                                                                                                                                                                                                                                       | 0,7x22 cm                                                                                                                                                                                                                                                                                                                                                                                              |
| Generator Information                                |                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                        |
| Type                                                 | High frequency                                                                                                                                                                                                                                                                                                                                                                               | High frequency, Model name :<br>Sedecal SHF-1030-M                                                                                                                                                                                                                                                                                                                                                     |
| Maximum power                                        | 12 kW                                                                                                                                                                                                                                                                                                                                                                                        | 10 kW                                                                                                                                                                                                                                                                                                                                                                                                  |
| Output voltage<br>range                              | 20-45 kV with 1 kV step<br>maximum ripple 3%                                                                                                                                                                                                                                                                                                                                                 | 20-50 kV with 0.1 kV step<br>maximum ripple 1%                                                                                                                                                                                                                                                                                                                                                         |
| Output current<br>range                              | 80-200 mA                                                                                                                                                                                                                                                                                                                                                                                    | 5-250 mA, From 5 mA to 80 mA in 0.1 mA steps<br>and from 80 mA to 250 mA in 1 mA steps                                                                                                                                                                                                                                                                                                                 |
| Output mAs range                                     | -                                                                                                                                                                                                                                                                                                                                                                                            | 0.1 to 1200 mAs                                                                                                                                                                                                                                                                                                                                                                                        |
| Operation of AEC                                     | -                                                                                                                                                                                                                                                                                                                                                                                            | AEC is implemented using pre-scan (not on<br>generator level)                                                                                                                                                                                                                                                                                                                                          |
| Backup timer:                                        | Yes                                                                                                                                                                                                                                                                                                                                                                                          | Yes                                                                                                                                                                                                                                                                                                                                                                                                    |
| Photo                                                | Image: SenoScan                                                                                                                                                                                                                                                                                                                                                                              | Image: MammoScan                                                                                                                                                                                                                                                                                                                                                                                       |

{5}------------------------------------------------

- 7. Description of non-clinical (bench) testing: The device has been evaluated for as electrical, electromagnetic, radiation, and mechanical safety, and has been found to conform to the following medical device safety standards:

{6}------------------------------------------------

EN 60601-1:2006/ A1:2013 General Requirements for Safety

EN 60601-1-2:2007/ A:2010 EMC Electromagnetic compatibility-Requirements and tests EN 60601-1-3:2008/ A:2010 Collateral Standard: Radiation protection in diagnostic X-ray equipment EN 60601-1-6:2010/ A1:2015 Collateral Standard: Usability

IEC 60601-2-45:2011 Particular requirements for the safety of mammographic X-ray equipment and mammographic stereotactic devices

EN ISO 15223-1:2012 Symbols to be used with medical device labeling and information to be supplied EN ISO 14971:2012 - Application of risk management to medical devices

EN 62220-1-2:2010 Determination of the detective quantum efficiency - Detectors used in mammography EN 62304:2006 + AMD1:2015 CSV Medical device software life-cycle processes

EN 62366:2008/ A1:2015 Medical devices – Application of usability engineering to medical devices

EN 1041:2008/ A1:2013 Information supplied by the manufacturer of medical devices

EN ISO 14155:2011 Clinical investigation of medical devices for human subjects – Good clinical practice

Performance data from non-clinical testing of the Adani MammoScan covering Sensitometric response, Spatial resolution, Noise analysis, Signal-to-noise - DQE, Dynamic range, Automatic exposure control performance, Phantom testing, Patient radiation dose, and Image erasure, fading and repeated exposure was compared with data from the PMA Summary of Safety and effectiveness of the predicate device. This comparison showed that the Adani MammoScan device performed as well as or better than the predicate devices in all relevant areas.

Biocompatibility testing according to ISO 10993 was performed for the patient contact material, the compression plate: cytotoxicity, irritation, sensitization.

### 8. SUMMARY OF THE CLINICAL IMAGE ATTRIBUTE EVALUATION

An image attribute evaluation was conducted in accordance with the Guidance for Industry and FDA Staff, Class II Special Controls Guidance Document: Full-Field Digital Mammography System [issued: March 27, 2012] which concluded that the images were of sufficiently acceptable quality for clinical mammographic usage.

#### 9. SAFETY INFORMATION

The Adani MammoScan system introduces no new safety or efficacy issues other than those already identified with the predicate device. The results of the Risk analysis combined with the appropriate preventive measures taken indicate that the device is of moderate concern as per the May 11, 2005 issue of the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and is consistent with the level of concern indicated in the "Class II Special Controls Guidance Document: Full-Field Digital Mammography System" document issued on: March 27, 2012.

- 10. Conclusion: The Adani MammoScan described in this substantially equivalent to the predicate device in respects of indication for use and image quality. The proposed and predicate devices utilize similar technology and materials, comparable safety and effectiveness features, and are similar in design and construction. The information submitted in this application shows that none of the technical differences between the systems raises new questions of safety and effectiveness. All collected performance data demonstrate that the devices are substantially equivalent. Our conclusion is that Adani MammoScan is as safe and effective as the legally marketed predicate device. After analyzing bench, clinical, risk analysis, software validation and standards testing data, it is the conclusion of Adani that the MammoScan® Full Field Digital Mammography System is as safe and effective as the predicate device, has no significant technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K172027](https://fda.innolitics.com/device/K172027)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
