Creo Medical Electrosurgical System including Speedboat RS2 Surgical Accessory

K171983 · Creo Medical, Ltd. · KNS · Aug 17, 2017 · Gastroenterology, Urology

Device Facts

Record IDK171983
Device NameCreo Medical Electrosurgical System including Speedboat RS2 Surgical Accessory
ApplicantCreo Medical, Ltd.
Product CodeKNS · Gastroenterology, Urology
Decision DateAug 17, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.4300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Intended for use in the cutting of soft tissue using radiofrequency current, the coagulation (hemostasis, cauterization) of soft tissue using microwave energy, and the delivery and injection of solutions for endoscopic surgical procedures within the gastrointestinal tract.

Device Story

System comprises electrosurgical generator, Speedboat RS2 surgical instrument, footswitch, and interface cable. Used in endoscopic GI procedures by physicians. Generator delivers bipolar RF (400 kHz) for cutting and microwave energy (5.8 GHz) for coagulation; only one energy source delivered at a time. Speedboat RS2 instrument features gold-metallized ceramic bipolar electrode (blade) and stainless-steel hull; includes retractable needle for fluid injection via proximal slider control. Generator software controls voltage, power, current, duty cycle, and duration. System provides integrated cutting, coagulation, and injection capabilities in single device, potentially improving procedural efficiency and patient outcomes by reducing instrument exchanges.

Clinical Evidence

Bench testing only. Evaluated mechanical, cutting, coagulation, and injection performance in ex-vivo and in-vivo models. Compared thermal penetration and spread against predicate devices. Confirmed compliance with electrical safety (IEC 60601), biocompatibility (ISO 10993), and sterility (ISO 11135) standards.

Technological Characteristics

Generator delivers bipolar RF (400 kHz) and microwave (5.8 GHz) energy via coaxial connector. Speedboat RS2 (2.3m length) features gold-metallized ceramic bipolar electrode and stainless-steel hull. Connectivity via interface cable with push-fit latching. Software-controlled energy parameters. Sterilization via ISO 11135.

Indications for Use

Indicated for cutting soft tissue via radiofrequency current, coagulation (hemostasis, cauterization) via microwave energy, and delivery/injection of solutions during endoscopic gastrointestinal procedures.

Regulatory Classification

Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles three stylized human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 17, 2017 Creo Medical Ltd. % Keith Penny Director Quality Assurance and Regulatory Affairs Creo Medical Ltd. Riverside Court, Beaufort Park Chepstow, NP16 5UH United Kingdom Re: K171983 > Trade/Device Name: Creo Medical Electrosurgical System including Speedboat RS2 Surgical Accessory Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: II Product Code: KNS Dated: June 29, 2017 Received: June 30, 2017 Dear Keith Penny: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K171983 ### Device Name Creo Medical Electrosurgical System including Speedboat RS2 Surgical Accessory Indications for Use (Describe) Intended for use in the cutting of soft tissue using radiofrequency current, the coagulation (hemostasis, cauterization) of soft tissue using microwave energy, and the delivery and injection of solutions for endoscopic surgical procedures within the gastrointestinal tract. | Type of Use (Select one or both, as applicable) | |--------------------------------------------------------------------------| | <span></span> <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NECESSARY # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## SECTION 5 510(k) SUMMARY #### 1. SUBMITTER Creo Medical Ltd Riverside Court Beaufort Park Chepstow NP16 5UH United Kingdom Phone: +44 (0) 1291 606005 Fax: +44 (0) 1291 606015 Contact: Keith Penny Director Regulatory Affairs Date Prepared: June 28, 2017 #### DEVICE 2. | Device Name: | Creo Medical Electrosurgical System including Speedboat RS2 surgical | |----------------------|----------------------------------------------------------------------| | | accessory | | Common Name: | Electrosurgical System | | Classification Name: | Unit, Electrosurgical, Endoscopic (With Or Without Accessories) | | Regulation Number: | 21 CFR 876.4300 | | Product Code: | KNS | | Classification: | Class II | #### 3. PREDICATE DEVICES Olympus Hook Knife, single-use electrosurgical knife K092309 K133933 Boston Scientific Injection Gold Probe, bipolar electrohemostasis catheter ERBE VIO 300 D electrosurgical generator K083452 #### 4. SYSTEM DESCRIPTION The Creo Medical Electrosurgical System comprises: - Speedboat RS2 surgical instrument (7-RS2-001) ● - Electrosurgical Generator (7-EMR-050) . - Footswitch (2-EMR-032) . - . Interface Cable (7-RS2-210) The Electrosurgical Generator is designed to deliver bipolar radiofrequency (RF) energy and microwave energy for the purpose of cutting and coagulating tissue. The Electrosurgical Generator output is actuated via a two-pedal Footswitch. One pedal activates the bipolar RF energy output for cut; the other pedal activates the microwave energy output for coagulation. The Electrosurgical Generator incorporates proprietary software developed by Creo Medical for generating and {5}------------------------------------------------ controlling the two energies delivered. The Electrosurgical Generator and Footswitch are nonsterile and reusable. The Interface Cable connects electrosurgical instruments to the Electrosurgical Generator and is for single-use and is supplied with a sterile sheath that is fitted over its distal end during connection to the surgical accessory. Speedboat RS2 is an electrosurgical instrument for use with the Creo Medical Electrosurgical Generator only. Speedboat RS2 is for endoscopic use and provides cutting, coagulation and injection of fluids incorporated in a single device. Speedboat RS2 is for single-use only and provided sterile. #### 5. INDICATION FOR USE STATEMENT Intended for use in the cutting of soft tissue using radiofrequency current, the coagulation (hemostasis, cauterization) of soft tissue using microwave energy, and the delivery and injection of solutions for endoscopic surgical procedures within the gastrointestinal tract. #### 6. TECHNOLOGICAL CHARACTERISTCS The Electrosurgical Generator comprises two distinct energy sources for the independent generation of bipolar radiofrequency (RF) current at 400 kHz and microwave current at 5.8 GHz. Both energy sources are delivered through the same output coaxial connector. These generator outputs are controlled in terms of voltage, power, current, duty cycle, and duration by means of widely used and long-established design and engineering techniques. Via these controls and system design, only one energy source can be delivered to the output coaxial connector at any one time. The Speedboat RS2 surgical instrument is 2.3 m long and used through the working channel of a compatible endoscope. The distal tip comprises: - a qold-metallized ceramic bipolar electrode assembly (Blade) for delivery of cutting and . coagulation energies - . a stainless-steel rounded contour component (Hull) mounted to the Blade - . an extendable and retractable needle for delivery and injection of fluids that is useroperated by a slider control located on the proximal handle of the instrument The proximal handle of the Speedboat RS2 incorporates a Luer-lock port for connection of a usersupplied syringe containing injection solution and a coaxial connector for connection to the Interface Cable. The Interface Cable has push-fit latching connectors on each end designed to deliver energy at both RF and microwave frequencies. | Difference | Performance Testing | |-------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Cut: Energy<br>The predicate device (Hook Knife)<br>uses monopolar RF energy at<br>350 kHz while the subject device<br>uses bipolar RF energy at 400<br>kHz | Bench testing:<br>Comparison with the predicate device for the penetration<br>and spread of thermal effects in gastrointestinal tissue.<br>The bench testing showed substantial equivalence with the<br>predicate device.<br>The difference in cut energy raises no new questions<br>regarding safety and efficacy. | #### DIFFERENCE BETWEEN SUBJECT AND PREDICATE DEVICES 7. {6}------------------------------------------------ | Difference | Performance Testing | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Cut: Design distal tip<br>The predicate device (Hook Knife)<br>has a L- shaped stainless steel<br>distal tip while the subject device<br>has a blade shaped gold-<br>metallized ceramic tip with a<br>metal hull | Bench testing:<br>Comparison with the predicate device for the penetration<br>and spread of thermal effects in gastrointestinal tissue.<br>The bench testing showed substantial equivalence with the<br>predicate device.<br>The difference in the design of the distal tip raises no new<br>questions regarding safety and efficacy. | | Coagulation: Energy<br>The predicate device (Gold<br>Probe) uses bipolar RF energy at<br>350 kHz while the subject device<br>uses microwave energy at 5.8<br>GHz | Bench testing:<br>Comparison with the predicate device for the penetration<br>and spread of thermal effects in gastrointestinal tissue.<br>The bench testing showed substantial equivalence with the<br>predicate device.<br>The difference in coagulation energy raises no new<br>questions regarding safety and efficacy. | | Coagulation: Design distal tip<br>The predicate device (Gold<br>Probe) has a dome shaped gold-<br>metallized ceramic distal tip while<br>the subject device has a blade<br>shaped gold-metallized ceramic<br>tip with a metal hull | Bench testing:<br>Comparison with the predicate device for the penetration<br>and spread of thermal effects in gastrointestinal tissue.<br>The bench testing showed substantial equivalence with the<br>predicate device.<br>The difference in the design of the distal tip raises no new<br>questions regarding safety and efficacy. | | Shaft and distal tip materials<br>The materials used in the distal<br>tip and shaft of the subject device<br>are similar but not identical to the<br>predicate devices | Bench testing:<br>Comparison with the predicate devices for the penetration<br>and spread of thermal effects in gastrointestinal tissue.<br>The bench testing showed substantial equivalence with the<br>predicate device.<br>Biocompatibility:<br>Meets all relevant requirements of ISO10993.<br>The difference in the shaft and distal tip materials raise no<br>new questions regarding safety and efficacy. | #### PERFORMANCE DATA 8. The system meets all design specifications, design-risk analysis, and medical device standards for electrical safety and EMC (IEC 60601), biocompatibility (ISO 10993) and sterility (ISO 11135). The mechanical, cut, coagulation and injection performance in ex-vivo and in-vivo meets the design specification and shows substantial equivalence to the predicate devices. #### CONCLUSION 9. The Creo Medical Electrosurgical System is substantially equivalent to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...