← Product Code [FMI](/productcode/FMI) · K171982

# DROPLET PEN NEEDLE (K171982)

_Htl-Strefa S.A · FMI · Mar 21, 2018 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K171982

## Device Facts

- **Applicant:** Htl-Strefa S.A
- **Product Code:** [FMI](/productcode/FMI.md)
- **Decision Date:** Mar 21, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5570
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

Droplet® Pen Needles are intended for use with pen injector devices for the subcutaneous injection of drugs.

## Device Story

Droplet® Pen Needles are sterile, single-use, manual hypodermic needles designed for subcutaneous drug delivery via pen injector devices. The device consists of a double-ended stainless steel cannula secured in a plastic hub with adhesive; the hub features internal threads for attachment to pen injectors. A silicone-based lubricant is applied to both the patient and cartridge ends of the cannula to facilitate penetration. The assembly includes an inner needle shield for protection and an outer cover, with a peel-away seal for sterility. Used by patients, caregivers, and healthcare professionals in various settings (e.g., home, clinic). The device functions by screwing onto a pen injector, allowing the cartridge end to pierce the septum and the patient end to deliver the medication. It provides a sterile, compatible interface for drug administration, reducing the risk of needle-related injury and ensuring consistent flow.

## Clinical Evidence

No clinical data was required or provided. Substantial equivalence is supported by bench testing, including compliance with ISO 11608-2 (needle-based injection systems) and comprehensive biocompatibility testing (ISO 10993-5, -10, -11, -4; ASTM F756; USP <151>).

## Technological Characteristics

Materials: Medical grade stainless steel (ISO 9626), plastic resin hubs/shields, medical grade silicone lubricant. Principle: Manual, screw-thread attachment. Dimensions: 29G-32G, 4mm-12mm lengths. Sterilization: Gamma irradiation (SAL 10^-6). Packaging: Polypropylene container with medical grade paper seal.

## Regulatory Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

## Predicate Devices

- BD Pen Needle ([K162516](/device/K162516.md))
- Droplet® Pen Needle ([K143437](/device/K143437.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

March 21, 2018

HTL-STREFA S.A. Aleksandra Prazmowska-Wilanowska Regulatory Affairs Director ul. Adamowek 7 95-035 Ozorkow POLAND

Re: K171982

Trade/Device Name: Droplet® Pen Needles (Models: Droplet® Pen Needles 4mm x 32G, Droplet® Pen Needles 5mm x 32G, Droplet® Pen Needles 6mm x 32G, Droplet® Pen Needles 8mm x 32G, Droplet® Pen Needles 5mm x 31G, Droplet® Pen Needles 6mm x 31G, Droplet® Pen Needles 8mm x 31G, Droplet® Pen Needles 10mm x 29G, Droplet® Pen Needles 12mm x 29G)

Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: February 21, 2018 Received: February 21, 2018

Dear Ms. Aleksandra Prazmowska-Wilanowska:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely, Tina Kiang

-S

Tina Kiang, Ph.D.
Acting Director
Division of Anesthesiology,
General Hospital, Respiratory,

Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K171982

#### Device Name

| Droplet® Pen Needles (Models: | Droplet® Pen Needles 4mm x 32G   |
|-------------------------------|----------------------------------|
| Droplet® Pen Needles          | Droplet® Pen Needles 5mm x 32G   |
| Droplet® Pen Needles          | Droplet® Pen Needles 6mm x 32G   |
| Droplet® Pen Needles          | Droplet® Pen Needles 8mm x 32G   |
| Droplet® Pen Needles          | Droplet® Pen Needles 5mm x 31G   |
| Droplet® Pen Needles          | Droplet® Pen Needles 6mm x 31G   |
| Droplet® Pen Needles          | Droplet® Pen Needles 8mm x 31G   |
| Droplet® Pen Needles          | Droplet® Pen Needles 10mm x 29G  |
| Droplet® Pen Needles          | Droplet® Pen Needles 12mm x 29G) |

Indications for Use (Describe)

Droplet® Pen Needles are intended for use with pen injector devices for the subcutaneous injection of drugs.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K171982 510(k) Summary

As required by the Medical Devices Act of 1990 and in accordance with 21 CFR ട്ല807.92(a).

| Summary                |                                                                                                                                                                                                                                                                                                                           |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| [807.92 (a)(1,2)]      |                                                                                                                                                                                                                                                                                                                           |
| Date Prepared:         | March 15, 2018                                                                                                                                                                                                                                                                                                            |
| Submitted By:          | HTL-STREFA S.A.<br>ul. Adamówek 7<br>95-035 Ozorków<br>POLAND                                                                                                                                                                                                                                                             |
| Primary Contact:       | Aleksandra Prażmowska-Wilanowska<br>RA Director<br>Phone: +48 42 270 00 10<br>ola.prazmowska-wilanowska@htl-strefa.pl                                                                                                                                                                                                     |
| Secondary Contact:     | Izabela Banaś<br>Senior RA Specialist<br>Phone: +48 42 270 00 10<br>izabela.banas@htl-strefa.pl                                                                                                                                                                                                                           |
| Trade Name:            | Droplet® Pen Needles                                                                                                                                                                                                                                                                                                      |
| Device models:         | Droplet® Pen Needles 4 mm x 32G<br>Droplet® Pen Needles 5 mm x 32G<br>Droplet® Pen Needles 6 mm x 32G<br>Droplet® Pen Needles 8 mm x 32G<br>Droplet® Pen Needles 5 mm x 31G<br>Droplet® Pen Needles 6 mm x 31G<br>Droplet® Pen Needles 8 mm x 31G<br>Droplet® Pen Needles 10 mm x 29G<br>Droplet® Pen Needles 12 mm x 29G |
| Common Name:           | Pen Needle                                                                                                                                                                                                                                                                                                                |
| Regulation Number:     | 21 CFR § 880.5570                                                                                                                                                                                                                                                                                                         |
| Product Code:          | FMI                                                                                                                                                                                                                                                                                                                       |
| Device Classification: | II                                                                                                                                                                                                                                                                                                                        |
| Review Panel:          | 80 General Hospital                                                                                                                                                                                                                                                                                                       |
| Predicate Device:      |                                                                                                                                                                                                                                                                                                                           |

{4}------------------------------------------------

[807.92(a)(3)]

#### The legally marketed device to which substantial equivalence is claimed is:

| Manufacturer         | Trade Name    | 510(k) Number |  |
|----------------------|---------------|---------------|--|
| Becton Dickson & Co. | BD Pen Needle | K162516       |  |

#### Secondary Predicate Device

| Manufacturer Name | Trade Name          | 510(k) Number |
|-------------------|---------------------|---------------|
| HTL-Strefa S.A.   | Droplet® Pen Needle | K143437       |

The secondary predicate device cleared under K143437 has been included in this submission to cover all gages and lengths of the new device. All design features remain unchanged with regard to the device submitted and cleared under K143437.

#### Description of Device:

#### [807.92(a)(4)]

The Droplet® Pen Needles are sterile, single use needles intended for use with pen injector devices for the subcutaneous injection of drugs. Pen needles are used by consumers, caregivers and healthcare professionals.

The pen needle assembly consists of a double-ended cannula that is assembled into an injection molded hub using adhesive. The hub has internal threads, which allow it to be screwed onto the pen injector device. This allows the cartridge end of the cannula to penetrate through the rubber septum of the cartridge. The patient end and the cartridge end of the cannula are lubricated using a silicone based lubricant for ease of injection and rubber septum penetration.

There is an inner needle shield assembled over the patient end of the cannula to protect the needle point from damage and accidental needle sticks. There is also an outer cover. Each pen needle assembly is protected with a peel away seal to provide a sterility barrier.

{5}------------------------------------------------

#### Indications for Use:

### [807.92(a)(5)]

Droplet® Pen Needles are intended for use with pen injector devices for the subcutaneous injection of drugs.

#### Technological Characteristics:

[807.92(a)(6)]

A comparison of characteristics of Droplet® Pen Needles, the predicate device and the secondary predicate device is shown in the table below:

#### Device Comparison between New Device, Predicate Device and the Secondary Predicate Device Including Indications for Use

|        |                                      | New Device                                                                                                      | Predicate Device                                                                                      | Secondary Predicate Device                                                                                        |
|--------|--------------------------------------|-----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
|        | Manufacturer                         | HTL-STREFA S.A.                                                                                                 | Becton Dickinson                                                                                      | HTL-STREFA S.A.                                                                                                   |
|        | 510(k) Number                        | K171982                                                                                                         | K162516                                                                                               | K143437                                                                                                           |
|        | Product Code                         | FMI                                                                                                             | FMI                                                                                                   | FMI                                                                                                               |
| Design | Primary Container                    | Image: Primary Container New Device                                                                             | Image: Primary Container Predicate Device                                                             | Image: Primary Container Secondary Predicate Device                                                               |
|        | Needle Shield                        | Image: Needle Shield New Device                                                                                 | Image: Needle Shield Predicate Device                                                                 | Image: Needle Shield Secondary Predicate Device                                                                   |
|        | Needle Tube and Hub                  | Image: Needle Tube and Hub New Device                                                                           | Image: Needle Tube and Hub Predicate Device                                                           | Image: Needle Tube and Hub Secondary Predicate Device                                                             |
|        | Indications for Use                  | The Droplet® Pen Needle is intended for use with a pen injector device for the subcutaneous injection of drugs. | BD Pen Needle is intended for use with a pen injector device for the subcutaneous injection of drugs. | The Droplet® Pen Needle is intended for use with a pen injector device for the subcutaneous injection of insulin. |
|        | Principle of operation               | Manual                                                                                                          | Manual                                                                                                | Manual                                                                                                            |
|        | Method of attachment to pen injector | Screw threads                                                                                                   | Screw threads                                                                                         | Screw threads                                                                                                     |

{6}------------------------------------------------

| Length                                        | 4mm, 5mm, 6mm,<br>8mm, 10mm, 12mm                                      | 4mm, 5mm, 8mm,<br>12.7mm                                               | 4mm, 5mm, 6mm,<br>8mm, 10mm, 12mm                                      |
|-----------------------------------------------|------------------------------------------------------------------------|------------------------------------------------------------------------|------------------------------------------------------------------------|
| Gage                                          | 29G, 31G, 32G                                                          | 29G, 30G, 31G, 32G                                                     | 29G, 31G, 32G                                                          |
| Biocompatibility                              | Conforms to ISO<br>10993-1                                             | Conforms to ISO<br>10993-1                                             | Conforms to ISO<br>10993-1                                             |
| Sterility                                     | SAL = 10-6                                                             | SAL = 10-6                                                             | SAL = 10-6                                                             |
| Sterilization<br>method                       | Gamma irradiation                                                      | Gamma irradiation                                                      | Gamma irradiation                                                      |
| Shelf Life                                    | 5 years                                                                | 5 years                                                                | 3 years                                                                |
| Unit Packaging                                | Polypropylene<br>container with seal<br>made of medical<br>grade paper | Polypropylene<br>container with seal<br>made of medical<br>grade paper | Polypropylene<br>container with seal<br>made of medical<br>grade paper |
| User Packaging                                | Cardboard sales<br>box                                                 | Cardboard sales box                                                    | Cardboard sales box                                                    |
| Materials                                     |                                                                        |                                                                        |                                                                        |
| • Needle Tube                                 | Medical Grade<br>Stainless Steel                                       | Medical Grade<br>Stainless Steel                                       | Medical Grade<br>Stainless Steel                                       |
| • Hub, Primary<br>Container,<br>Needle Shield | Plastic resins                                                         | Plastic resins                                                         | Plastic resins                                                         |
| • Lubricant                                   | Medical grade<br>silicone                                              | Medical grade<br>silicone                                              | Medical grade<br>silicone                                              |

The new device, the predicate device and the secondary predicate device are classified under 21 CFR 880.5570, which states: "A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin."

There are no significant differences in the design, principle of operation, method of attachment to pen injector, biocompatibility, sterilization method, materials, or intended use between the Droplet® pen needle device and the predicate device.

The Droplet® pen needle differs from the predicate device as follows:

1. The Droplet® pen needle has slightly different molded features of its primary container, needle shield, and hub.

2. The Droplet® pen needle product line does not include a 30G model.

3. The Droplet® pen needle product line includes an additional model with a 6 mm needle length.

4. The Droplet® pen needle product line's longest needle length is 12 mm, while the predicate's longest needle length is 12.7 mm.

{7}------------------------------------------------

There are no significant differences in the design, principle of operation, method of attachment to pen injector, biocompatibility, sterilization method, materials, or intended use between the Droplet® pen needle device and the secondary predicate device.

The Droplet® pen needle differs from the secondary predicate device as follows:

- 1. In the indications for use the clarification to the intended use of the Droplet® pen needle does not introduce critical differences or new risks to the intended use of the device.
The differences do not raise new questions of safety or effectiveness for the subject device. Droplet® pen needle product line and the predicate products line cover a substantially equivalent range of options in gage and needle lengths and the few specific differences between the product lines do not have any impact on performance.

Based upon the above comparisons to the predicate device. Droplet® Pen Needles are substantially equivalent to the predicate device.

#### Non-Clinical Performance Data:

#### [(807.92(b)(1)]

Droplet® pen needles successfully passed all the required non-clinical testing which included the following:

- Testing for compliance with the requirements of 11608-2:2012 Needle-based injection systems for medical use -- Requirements and test methods -- Part 2: Needles

| Test Parameter                                                                        | Requirement - ISO 11608-2:2012                                                                                                                                                                                                                  | Result          |
|---------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Materials                                                                             | The needle shall be made of tubing materials<br>specified in ISO 9626.                                                                                                                                                                          | Requirement met |
| Dimensions                                                                            | The needles shall fit the test apparatus specified in<br>item 7.3 of ISO 11608-2.                                                                                                                                                               | Requirement met |
| Determination of flow<br>rate through the needle<br>(not required by the<br>standard) | The needle was tested in accordance with Annex A<br>to ISO 11608-2 to determine flow rate through the<br>needle.                                                                                                                                | Requirement met |
| Bond between hub and<br>needle tube                                                   | The union of the hub and needle tube shall not<br>break when tested in accordance with Clause 9 of<br>ISO 11608-2.                                                                                                                              | Requirement met |
| Freedom from defects                                                                  | The needle tube shall fulfill the requirements of<br>ISO 7864, clause11.3.                                                                                                                                                                      | Requirement met |
| Lubrication                                                                           | The needle tube should be lubricated at both the<br>patient end and the cartridge end. The lubricant<br>shall not, under normal or corrected-to-normal<br>vision, be visible as droplets of fluid on the outside<br>surface of the needle tube. | Requirement met |

{8}------------------------------------------------

| Dislocation of<br>measuring point at<br>patient end                                       | Dislocation of the cannula point at the patient end<br>shall be in accordance with Table 2 when tested in<br>accordance with Clause 8 (of ISO 11608-2).                           | Requirement met |
|-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Determination of<br>functional<br>compatibility with<br>needle-based injection<br>systems | Compatibility with any NIS (Needle-based<br>Injection System) shall be claimed only after<br>testing in accordance with Clause 11.                                                | Requirement met |
| Ease of assembly and<br>disassembly                                                       | Attachment of the needle shall be possible without<br>removing the needle from its opened unit<br>packaging. Compliance is checked according to<br>the requirements of Clause 11. | Requirement met |
| Pre-conditioning of<br>needles                                                            | All requirements of the standard related to<br>preconditioning of needles were met.                                                                                               | Requirement met |

#### Biocompatibility testing: •

# Biocompatibility Test Summary – no adverse biocompatibility effects were observed:

| Test method                                                                                               | Compliance with                                                                                                                                                                                                         | Result                                                                                                                                                                                                               |
|-----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity Study<br>Using the ISO<br>Elution Method                                                     | ISO 10993-5 - Biological<br>evaluation of medical devices -<br>Part 5: Tests for <i>in vitro</i><br>cytotoxicity                                                                                                        | The test article showed no evidence<br>of causing cell lysis or toxicity.                                                                                                                                            |
| ISO Guinea Pig<br>Maximization<br>Sensitization Test                                                      | ISO 10993-10 - Biological<br>evaluation of medical devices --<br>Part 10: Tests for irritation and<br>skin<br>sensitization                                                                                             | The test article extracts showed no<br>evidence of causing delayed dermal<br>contact sensitization in the guinea<br>pig. The test article was not<br>considered a sensitizer in the guinea<br>pig maximization test. |
| ISO Intracutaneous<br>Study in Rabbits                                                                    | ISO 10993-10 - Biological<br>evaluation of medical devices --<br>Part 10: Tests for irritation and<br>skin sensitization                                                                                                | The test article met the requirements<br>of the test.                                                                                                                                                                |
| ASTM Hemolysis<br>Study                                                                                   | ASTM F756, Standard Practice<br>for Assessment of Hemolytic<br>Properties of Materials and ISO<br>10993-4 - Biological evaluation<br>of medical devices -- Part 4:<br>Selection of tests for interactions<br>with blood | Both the test article in direct contact<br>with blood and the test article extract<br>were non-hemolytic.                                                                                                            |
| ISO Two Week<br>Toxicity Study in the<br>Rat, Repeated<br>Parenteral<br>Administration of<br>Two Extracts | ISO 10993-11 - Biological<br>evaluation of medical devices --<br>Part 11: Tests for systemic<br>toxicity                                                                                                                | There were no microscopic changes<br>considered to be a test article<br>related response. Parenteral<br>administration of the test article<br>extract did not produce systemic<br>toxicity in rats.                  |
| ISO Systemic<br>Toxicity Study in<br>Mice                                                                 | ISO 10993-11 - Biological<br>evaluation of medical devices --<br>Part 11: Tests for systemic<br>toxicity                                                                                                                | There was no mortality or evidence<br>of systemic toxicity from the extracts<br>injected into mice. Each test article<br>met the requirements of the study.                                                          |

{9}------------------------------------------------

| Test method                                        | Compliance with                                                                                                                                                         | Result                                                                                                                                                                  |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| USP Rabbit Pyrogen<br>Study, Material-<br>mediated | ISO 10993-11 Biological<br>evaluation of medical devices --<br>Part 11: Tests for systemic<br>toxicity                                                                  | The total rise of rabbit temperatures<br>during the 3 hour observation period<br>was within acceptable USP limits.<br>The test article was judged as non-<br>pyrogenic. |
| USP Pyrogen Study -<br>Material Mediated           | USP, General Chapter <151>,<br>Pyrogen Test as recommended by<br>ISO 10993-11 Biological<br>evaluation of medical devices --<br>Part 11: Tests for systemic<br>toxicity | The test article was judged as<br>nonpyrogenic.                                                                                                                         |

#### Clinical Performance Data:

# [(807.92(b)(2)]

Clinical data is not required.

#### Conclusion:

## [(807.92(b)(3)]

Droplet® Pen Needles are concluded to be substantially equivalent in the intended use, technology/principle of operation, materials and performance to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K171982](https://fda.innolitics.com/device/K171982)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
