K171931 · Medcomp (Medical Components, Inc.) · LJS · Oct 26, 2017 · General Hospital
Device Facts
Record ID
K171931
Device Name
Celerity ECG Cable Accessory Pack
Applicant
Medcomp (Medical Components, Inc.)
Product Code
LJS · General Hospital
Decision Date
Oct 26, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Indications for Use
The Celerity System is indicated for the positioning of Peripherally Inserted Central Catheters (PICC) in adult patients. It provides real-time catheter tip location information by using the patient's cardiac electrical activity. The Celerity System is indicated for use as an alternative method to chest x-ray or fluoroscopy confirmation of PICC tip placement in adult patients. Note: In general, devices that utilize ECG technique to observe P-wave are limited, but not contraindicated for patients where cardiac rhythms may change presentation of the P-wave; including: - Atrial fibrillation - Atrial flutter - Severe tachycardia - Pacemaker-driven rhythm - Chronic obstructive pulmonary disease (COPD) Such patients are easily identified prior to PICC insertion. Use of an additional confirmation method is necessary to confirm catheter tip location.
Device Story
Accessory pack for Celerity System; includes sterile remote cover and specialized alligator PICC clip. Device facilitates continuous ECG waveform display to guide PICC placement in right atrium. Used in clinical settings by healthcare providers. Clip conducts ECG signal from PICC stylet to Celerity remote control; remote cover provides sterile barrier for remote control cable. Output allows real-time visualization of P-wave changes, aiding clinicians in optimal catheter tip positioning. Benefits include reduced reliance on X-ray/fluoroscopy for confirmation.
Clinical Evidence
Bench testing only. Testing included accelerated aging (1 and 3 years) for alligator clips and accessory packs, equipment interaction testing, and ISTA 2A shipping tests. No clinical data provided.
Technological Characteristics
Components: nickel-plated steel alligator clip with PVC insulator; latex-free polyethylene remote cover. Principle: ECG signal conduction from PICC stylet to remote control. Sterilization: Ethylene Oxide. Connectivity: Wired accessory to Celerity System. Non-patient contacting.
Indications for Use
Indicated for PICC positioning in adult patients (18+). Provides real-time tip location via cardiac electrical activity as an alternative to chest x-ray or fluoroscopy. Not contraindicated but limited in patients with atrial fibrillation, atrial flutter, severe tachycardia, pacemaker-driven rhythms, or COPD; additional confirmation methods required for these populations.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceutical, and other products.
October 26, 2017
Medical Components, Inc. (dba Medcomp®) Colton Muraira Regulatory Affairs 1499 Delp Dr. Harleysville, Pennsylvania 19438
Re: K171931
Trade/Device Name: Celerity™ ECG Cable Accessory Pack Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: Class II Product Code: LJS Dated: September 15, 2017 Received: September 26, 2017
Dear Colton Muraira:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Tina Kiang
S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K171931
#### Device Name
Celerity ECG Cable Accessory Pack
#### Indications for Use (Describe)
The Celerity System is indicated for the positioning of Peripherally Inserted Central Catheters (PICC) in adult patients. It provides real-time catheter tip location information by using the patient's cardiac electrical activity. The Celerity System is indicated for use as an alternative method to chest x-ray or fluoroscopy confirmation of PICC tip placement in adult patients.
Note: In general, devices that utilize ECG technique to observe P-wave are limited, but not contraindicated for patients where cardiac rhythms may change presentation of the P-wave; including:
- · Atrial fibrillation
- · Atrial flutter
- · Severe tachycardia
- · Pacemaker-driven rhythm
- · Chronic obstructive pulmonary disease (COPD)
Such patients are easily identified prior to PICC insertion. Use of an additional confirmation method is necessary to confirm catheter tip location.
Type of Use (Select one or both, as applicable)
× | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## K171931 510(K) SUMMARY
| Manufacturer's Name: | Medical Components, Inc. (dba Medcomp®)<br>1499 Delp Drive<br>Harleysville, PA 19438 |
|---------------------------|--------------------------------------------------------------------------------------|
| Corresponding Official: | Colton Muraira<br>Regulatory: North America and EU |
| Telephone Number: | (215) 256-4201, x 2285 |
| Email: | CMuraira@Medcompnet.com |
| Preparation Date: | October 23, 2017 |
| Trade Name: | Celerity™ ECG Cable Accessory Pack |
| Common or Usual Name: | PICC Placement Accessory |
| Regulation Name: | Percutaneous, implanted, long-term<br>intravascular catheter |
| Regulation Number: | 21 CFR 880.5970 |
| Product Code: | LJS |
| Device Class: | Class II |
| Primary Predicate Device: | K142889 - AngioDynamics Inc. Celerity™ PICC Tip Confirmation<br>System |
### Device Description:
The Celerity™ ECG Cable Accessory Pack consists of a sterile remote cover and a specialized alligator PICC clip. The Celerity™ ECG Cable Accessory Pack is designed to be used in conjunction with the Celerity™ System to provide a continuous display of electrocardiograph waveforms to guide placement of peripherally-inserted central catheters in the patient's right atrium of the heart.
### Indications For Use
The Celerity™ System is indicated for the positioning of Peripherally Inserted Central Catheters (PICC) in adult patients. It provides real-time catheter tip location information by using the patient's cardiac electrical activity. The Celerity System is indicated for use as an alternative method to chest x-ray or fluoroscopy confirmation of PICC tip placement in adult patients.
{6}------------------------------------------------
Note: In general, devices that utilize ECG technique to observe P-wave are limited, but not contraindicated for patients where cardiac rhythms may change presentation of the P-wave; including:
- । Atrial fibrillation
- Atrial flutter -
- -Severe tachycardia
- Pacemaker-Driven Rhythm -
- -Chronic obstructive pulmonary disease (COPD)
Such patients are easily identified prior to PICC insertion. Use of an additional method is necessary to confirm catheter tip location.
### Substantial Equivalence Discussion
The subject device is an accessory pack to the Celerity System. The subject device accessory pack consists of a sterile remote cover and a specialized alligator PICC clip. Both devices use changes in cardiac electrical activity to provide real-time catheter tip location.
Celerity™ ECG Cable Accessory Pack has been compared to the cleared "Celerity™ PICC Tip Confirmation System" (510(k)# K142889) as a reference for substantial equivalence. A table comparing the two devices is provided as follow:
| Attribute | Subject Device – K171931:<br>Celerity ECG Cable Accessory<br>Pack | Predicate Device – K142889 :<br>Celerity PICC Tip Confirmation<br>System | Comparison | |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| 510(k) Holder | Medcomp, Inc | AngioDynamics | N/A | |
| Design | The Celerity ECG Cable<br>Accessory Pack is designed to be<br>used in conjunction with the<br>Celerity™ System to provide a<br>continuous display of<br>electrocardiograph [ECG]<br>waveforms to guide placement of<br>peripherally-inserted central<br>catheters [PICC] in the patient's<br>right atrium of the heart.<br><br>Sterile remote cover: covers a<br>sufficient length of the remote<br>control cable to provide a sterile<br>barrier capable of being pierced by<br>the remote control PICC<br>connector.<br><br>Sterile PICC Clip Connector: | The Celerity PICC Tip<br>Confirmation System is an<br>accessory to peripherally-inserted<br>central catheters (PICC). It aids a<br>clinician in placing PICC<br>optimally, in the superior vena<br>cava near the entrance to the right<br>atrium of the heart. | Components<br>Same | |
| | Capable of conducting an ECG<br>signal from the PICC stylet to the<br>Celerity remote control. | To provide real time tip location<br>information of a central venous<br>catheter by utilization of ECG to<br>observe P-wave changes as the tip<br>approaches the right atrium of the<br>heart via the superior vena cava. | To provide real time tip location<br>information of a central venous<br>catheter by utilization of ECG to<br>observe P-wave changes as the tip<br>approaches the right atrium of the<br>heart via the superior vena cava. | Same |
| Intended use | | | | |
| Indications for<br>use | The Celerity ECG Cable<br>Accessory Pack is indicated for the<br>positioning of Peripherally<br>Inserted Central Catheters (PICC)<br>in adult patients. It provides real-<br>time catheter tip location<br>information by using the patient's<br>cardiac electrical activity. The<br>Celerity System is indicated for<br>use as an alternative method to<br>chest x-ray or fluoroscopy<br>confirmation of PICC tip<br>placement in adult patients.<br><br>Note: In general, devices that<br>utilize ECG technique to observe<br>P-wave are limited, but not<br>contraindicated for patients where<br>cardiac rhythms may change<br>presentation of the P-wave;<br>including:<br>- Atrial fibrillation<br>- Atrial flutter<br>- Severe tachycardia<br>- Pacemaker-Driven Rhythm<br>- Chronic obstructive<br>pulmonary disease (COPD)<br><br>Such patients are easily identified<br>prior to PICC insertion. Use of an<br>additional method is necessary to<br>confirm catheter tip location. | The Celerity System is indicated<br>for the positioning of Peripherally<br>Inserted Central Catheters (PICC)<br>in adult patients. It provides real-<br>time catheter tip location<br>information by using the patient's<br>cardiac electrical activity. The<br>Celerity System is indicated for<br>use as an alternative method to<br>chest x-ray or fluoroscopy<br>confirmation of PICC tip<br>placement in adult patients.<br><br>Note: In general, devices that<br>utilize ECG technique to observe<br>P-wave are limited, but not<br>contraindicated for patients where<br>cardiac rhythms may change<br>presentation of the P-wave;<br>including:<br>- Atrial fibrillation<br>- Atrial flutter<br>- Severe tachycardia<br>- Pacemaker-Driven Rhythm<br>- Chronic obstructive<br>pulmonary disease (COPD)<br><br>Such patients are easily identified<br>prior to PICC insertion. Use of an<br>additional method is necessary to<br>confirm catheter tip location. | Same | |
| Target<br>Population | Adults (18 years or older) | Adults (18 years or older) | Same | |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Same | |
| Materials | ECG Clip Cable: nickel plated<br>steel with a white PVC insulator.<br>The audio connector consists of a<br>nylon and polyethylene insulator<br>and gold or nickel plated contacts<br>Remote Cover: latex-free<br>polyethylene | ECG Clip Cable: nickel plated<br>steel with a white PVC insulator.<br>The audio connector consists of a<br>nylon and polyethylene insulator<br>and gold or nickel plated contacts<br>Remote Cover: latex-free<br>polyethylene | Same | |
{7}------------------------------------------------
{8}------------------------------------------------
The indications for use statement and the intended use of the subject device are identical to the predicate device. The differences between the subject device and the predicate device are:
- 1. Proposed device is an accessory pack containing separate components to be used with the predicate Celerity System, which already contains stated components.
- 2. Different manufacturer of the ECG Clip Cable and Remote Cover
Based on the aforementioned modifications to the subject device, the subject device does not raise different types of safety and effectiveness questions when compared to the predicate device.
# Bench/Performance/Non-Clinical Testing
The following testing was conducted to demonstrate substantial equivalence of the Celerity™ ECG Cable Accessory Pack to the predicate devices:
| Item | Test Name |
|------|---------------------------------------------------------------------------------|
| 1 | Alligator Clips – 1 year accelerated aging |
| 2 | Alligator Clips – 3 year accelerated aging |
| 3 | Equipment Interaction of Alligator Clips,<br>ECG Electrodes, and ECG Snap Leads |
| 4 | Probe Cover – 1 year accelerated aging |
| 5 | Celerity Accessory Pack – 1 year accelerated aging |
| 6 | Celerity Accessory Pack – 3 year accelerated aging |
| 7 | Celerity Accessory Packs – ISTA 2A Shipping Test |
{9}------------------------------------------------
#### Biocompatibility
The subject device is non-patient contacting and therefore, biocompatibility testing is not applicable.
## Conclusion
The modifications to the subject device include the change of the manufacture of the ECG Clip Cable and the Remote Cover and the packaging of the ECG Clip Cable and Remote Cover as an accessory pack. The modifications do not raise new or different questions of safety and effectiveness and are supported by non-clinical testing.
The Celerity ECG Cable Accessory Pack is substantially equivalent to the Celerity PICC Tip Confirmation System cleared under K142889.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.