← Product Code [DQX](/productcode/DQX) · K171897

# Approach CTO Microwire Guide (K171897)

_Cook Incorporated · DQX · Aug 21, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K171897

## Device Facts

- **Applicant:** Cook Incorporated
- **Product Code:** [DQX](/productcode/DQX.md)
- **Decision Date:** Aug 21, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K171897 · Aug 21, 2017 | Approach CTO Microwire Guide | Cook Incorporated | Published clinical literature (case reports and clinical experience reviews) | The literature review provides evidence of the device's performance in routine clinical settings, specifically for facilitating catheter delivery and crossing chronic total occlusions in the peripheral vasculature. | Literature review; Case report; Clinical experience; Peripheral vasculature; Chronic total occlusion |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Mustapha, J.A. (2014) - Lower Extremity CTO Crossing with the Saber Catheter; Case report | Patients with peripheral vasculature CTOs | Not applicable for this study | Successful crossing of a CTO cap |
| Rundback, John H. et al. (2013) - Transpedal Interventions for Critical Limb Ischemia; Clinical experience review | Patients with critical limb ischemia | Not applicable for this study | Clinical experience with tools for CTO |

## Indications for Use

The Approach CTO Microwire Guide is intended for use in facilitating delivery of percutaneous catheters into the peripheral vasculature and is also indical for the intraluminal placement of percutaneous catheters or other therapeutic devices beyond stenotic lesions (including chronic total occlusions) in the peripheral vasculature prior to further percutaneous intervention.

## Device Story

The Approach CTO Microwire Guide is a sterile, single-use catheter guide wire consisting of a core mandril and a distal spring coil. It is available in 0.014-inch diameter and lengths of 135-300 cm, with flexible straight tips featuring 6-25 gram tip loads. The device is used by physicians in clinical settings to facilitate the delivery of percutaneous catheters and to navigate beyond stenotic lesions or chronic total occlusions in the peripheral vasculature. It is supplied with an insertion cannula to assist in passing through valve assemblies. By providing a trackable, pushable guide, the device enables the placement of therapeutic tools, facilitating subsequent percutaneous interventions and potentially improving outcomes for patients with peripheral vascular disease.

## Clinical Evidence

Bench testing included acute performance evaluations in animal and cadaver models. The animal study (n=not specified) evaluated preparation, introduction, pushability, trackability, flexibility, radiopacity, and device interaction; 100% of articles met acceptance criteria with no perforations or dissections. The cadaver model evaluated similar parameters, achieving 82% crossing success and 100% clinical success. Literature review provided additional support via case reports and proof-of-concept studies demonstrating successful use in peripheral CTO crossing.

## Technological Characteristics

Class II catheter guide wire; 0.014" diameter; 135-300 cm length; core mandril and distal spring coil construction; straight flexible tip (6-25g load); supplied with insertion cannula; sterile, single-use.

## Regulatory Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- CiTop 0.014" Guidewire ([K070212](/device/K070212.md))

## Reference Devices

- Approach CTO Microwire Guide ([K081337](/device/K081337.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 21, 2017

Cook Incorporated Ms. Jennifer Allman Regulatory Affairs Specialist 750 Daniels Wav Bloomington, IN 47404

Re: K171897

Trade/Device Name: Approach CTO Microwire Guide Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX, PDU Dated: June 23, 2017 Received: June 26, 2017

Dear Ms. Allman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K171897

Device Name Approach CTO Microwire Guide

#### Indications for Use (Describe)

The Approach CTO Microwire Guide is intended for use in facilitating delivery of percutaneous catheters into the peripheral vasculature and is also indical for the intraluminal placement of percutaneous catheters or other therapeutic devices beyond stenotic lesions (including chronic total occlusions) in the peripheral vasculature prior to further percutaneous intervention.

| Type of Use (Select one or both, as applicable)    |                                             |
|----------------------------------------------------|---------------------------------------------|
| X   Prescription   Ise (Part 21 CER 801 Subnart D) | Over-The-Counter Use (21 CER 801 Subnart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in a smaller, white, sans-serif font on a red background.

COOK INCORPORA 750 DANIELS WAY, P.O. BOX BLOOMINGTON, IN 47402-0 339.2235 TOLL FREE: 8 WWW.COOKMEDIC

K171897 Page 1 of 4

### 5.0 510(k) SUMMARY

# Approach® CTO Microwire Guide 21 CFR §807.92 Date Prepared: August 17, 2017

# Submitted By:

| Applicant:            | Cook Incorporated      |
|-----------------------|------------------------|
| Contact:              | Jennifer L. Allman     |
| Applicant Address:    | Cook Incorporated      |
|                       | 750 Daniels Way        |
|                       | Bloomington, IN 47404  |
| Contact Phone Number: | (812) 335-3575 x104280 |
| Contact Fax Number:   | (812) 332-0281         |

### Device Information:

| Trade Name:              | Approach® CTO Microwire Guide |
|--------------------------|-------------------------------|
| Common Name:             | Wire, Guide, Catheter         |
| Classification Name:     | Catheter guide wire           |
| Regulation/Product Code: | 21 CFR §870.1330/DQX, PDU     |
| Device Class/Panel:      | Class II/Cardiovascular       |

# Predicate Device:

The CiTop" 0.014" Guidewire (Ovalum Ltd., K070212) is intended to facilitate the intraluminal placement of the wire guide beyond stenotic lesions (including chronic total occlusions) in the peripheral vasculature prior to further percutaneous intervention.

# Reference Device:

The Approach® CTO Microwire Guide, manufactured by Cook Inc., was cleared for commercial distribution under K081337 on August 8, 2008. The Approach® CTO Microwire Guide is intended for use in facilitating delivery of percutaneous catheters into the peripheral vasculature.

{4}------------------------------------------------

# Device Description:

The Approach® CTO Microwire Guide, subject of this submission, is a Class II device according to 21 CFR §870.1330; product code DQX (Wire, Guide, Catheter). The subject device consists of a core mandril and a distal spring coil. The Approach® CTO Microwire Guide is available with an outside diameter of 0.014 inches and lengths ranging from 135 - 300 centimeters. The flexible tip portion of the wire guide is straight with tip loads ranging from 6 - 25 grams.

The Approach® CTO Microwire Guide is supplied with an insertion cannula which facilitates the insertion of the wire guide through valve assemblies.

The Approach® CTO Microwire Guide is a packaged, sterile device intended for single patient use.

### Intended Use:

The Approach® CTO Microwire Guide is intended for use in facilitating delivery of percutaneous catheters into the peripheral vasculature and is also indicated for the intraluminal placement of percutaneous catheters or other therapeutic devices beyond stenotic lesions (including chronic total occlusions) in the peripheral vasculature prior to further percutaneous intervention.

# Comparison to Predicate:

The subject device is identical in indicated therapeutic effect to the predicate device, CiTop™ 0.014" Guidewire (K070212), which is intended to facilitate the intra-luminal placement of the wire guide beyond stenotic lesions (including chronic total occlusions) in the peripheral vasculature prior to further percutaneous intervention.

# Comparison to Reference Device:

The subject device Approach® CTO Microwire Guide is identical in design, method of operation, fundamental technological characteristics, manufacturing processes, and intended use to the reference device, the Approach® CTO Microwire Guide (K081337).

{5}------------------------------------------------

# Performance Data:

The following tests were performed to demonstrate that the Approach® CTO Microwire Guide met applicable design and performance requirements and support a determination of substantial equivalence.

- Acute Performance Evaluation of the Cook Approach® CTO Microwire Guide in . an Animal Model - This testing evaluated the preparation, introduction, push ability, track ability, flexibility, radiopacity, withdrawal, reintroduction, and interaction with supporting devices of the test articles under conditions intended to simulate clinical use. All (100%) of the test articles evaluated met the predetermined acceptance criteria for all of the performance parameters. No perforations or dissections were observed in this study.
- Acute Performance Evaluation of the Cook Approach® CTO Microwire Guide in . a Cadaver Model - This testing evaluated the preparation, introduction, push ability, track ability, radiopacity, and interaction with supporting devices of the test articles under conditions intended to simulate clinical use. All test articles evaluated met the predetermined acceptance criteria for all of the performance parameters. Individual test article crossing success for this evaluation was 82% with a clinical success rate of 100%.

# Summary of Literature Review

The following references demonstrate the use of the subject device. Approach® CTO Microwire Guide, in facilitating delivery of percutaneous catheters into the peripheral vasculature and for the intra-luminal placement of percutaneous catheters or other therapeutic devices beyond stenotic lesions (including chronic total occlusions) in the peripheral vasculature prior to further percutaneous intervention as well as supporting a determination of substantial equivalence to the predicate device:

- Mustapha, J.A., MD (2014). Lower Extremity CTO Crossing with the Saber™ . Catheter. Supplement to Endovascular Today, October issue, pp. 19- pp. 22.
This supplement includes a case report in which the subject device was used to cross a CTO cap in antegrade fashion.

{6}------------------------------------------------

- Rundback, John H., MD, FAHA, FSVM, Herman, Keven C., MD (2013). ● Transpedal Interventions for Critical Limb Ischemia [Vascular Disease Management 2013:10(8):E152-E158]. Retrieved from www.vasculardiseasemanagement.com
This reference discusses favorable experiences with tools recommended in the setting of chronic total occlusion, one of which is the Approach CTO Microwire Guide.

- Scheinowitz, M. et al. (2009). Crossing Chronic Total Occlusions with a New . 0.014" CiTop Guidewire: Proof of Concept. Catheterization and Cardiovascular Interventions 74 (278-285). DOI 10.1002/ccd.
This reference demonstrates equivalent crossing success of the predicate device as compared to the success demonstrated by the subject device in the performance evaluation determined in the cadaver model, discussed above.

## Conclusion:

The results of the performance evaluation in animal and human models along with the literature review support a conclusion that the Approach® CTO Microwire Guide, subject of this submission, met the acceptance criteria based on the intended use and support the conclusion that the intended use for the subject device does not raise new questions of safety or effectiveness as compared to the predicate device, the CiTop™ 0.014" Guidewire (K070212).

---

**Source:** [https://fda.innolitics.com/device/K171897](https://fda.innolitics.com/device/K171897)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
