← Product Code LQY · K171867

# Drs Finder NEO (K171867)

_Good Doctors Co., Ltd. · LQY · Mar 30, 2018 · DE · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K171867

## Device Facts

- **Applicant:** Good Doctors Co., Ltd.
- **Product Code:** LQY
- **Decision Date:** Mar 30, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** DE

## Intended Use

Dr's Finder NEO is intended for detecting the apex of root canal.

## Device Story

Dr's Finder NEO is a dental apex locator used by clinicians to determine the position of a dental file relative to the root canal apex. The device operates by emitting small electrical currents at frequencies of 400 Hz, 5400 Hz, and 10400 Hz through a standard dental file inserted into the root canal. It measures the electrical impedance between the file and the patient's mouth. The device processes these impedance measurements to calculate the file's proximity to the apex, displaying the result on an OLED screen. This real-time feedback assists the clinician in accurately locating the root canal apex during endodontic treatment, potentially improving procedural accuracy and patient outcomes. The device is a compact, battery-powered unit.

## Clinical Evidence

Bench testing only. Performance was evaluated using extracted teeth and stainless steel files to compare measurement accuracy against the reference device (K090925). Electrical safety and EMC testing were conducted per EN 60601-1, EN 60601-1-2, and related standards. Sterility assurance for accessories was validated using the biological indicator overkill method per ISO 17665.

## Technological Characteristics

Apex locator using multi-frequency impedance measurement (400 Hz, 5400 Hz, 10400 Hz). Powered by a 4.2V rechargeable Lithium Polymer battery. Features an OLED display. Dimensions: 43 x 46 x 16 mm; Weight: 24g. Accessories include file hook, lip hook, and probe cord, which are steam-sterilizable (ISO 17665). Complies with EN 60601-1, EN 60601-1-2, EN 61000-3-2/3, and EN 55011.

## Predicate Devices

- Dr's Finder ([K151274](/device/K151274.md))

## Reference Devices

- MULTIPLE (APEX LOCATOR), MODEL RCM-7 ([K090925](/device/K090925.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 30, 2018

Good Doctors Co., Ltd. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 690 Roosevelt Irvine, California 92620

Re: K171867 Trade/Device Name: Dr's Finder NEO Regulatory Class: Unclassified Product Code: LOY Dated: February 23, 2018 Received: February 28, 2018

#### Dear Priscilla Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Andrew I. Steen -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K171867

Device Name

Dr's Finder NEO

Indications for Use (Describe)

Dr's Finder NEO is intended for detecting the apex of root canal.

Type of Use (Select one or both, as applicable)

| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)             |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

(K171867)

This summary of 510(k) is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: Mar 29, 2018

#### 1. 510K Applicant / Submitter:

Good Doctors Co., Ltd. #208, Woolim Lions Valley B-dong, 283 Bupyeong-daero, Bupyeong-gu, Incheon 403-911, Republic of Korea Tel: +82-32-424-6325 Fax: +82-32-424-6326

#### 2. Submission Contact Person

LK Consulting Group USA, Inc. 690 Roosevelt, Irvine CA 92620 Priscilla Juhee Chung Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com

#### 3. Device

- Proprietary Name: Dr's Finder NEO .
- Common Name: Apex Locator .
- . Classification: Unclassified (Pre-Amendment)
- Product Code: LQY ●

#### 4. Predicate Device

- . Primary Predicate Device: Dr's Finder by Good Doctors Co., Ltd. (K151274)
- Reference Device: . Multiple (Apex Locator), Model RCM-7 by J. MRIT AMFG. CORP (K090925)

#### 5. Description:

The device allows the relative position of a dental file and the apex to be determined electrically. Using a standard dental file inserted into the root canal as an electrode, the 

{4}------------------------------------------------

device emits very small electrical currents having frequencies of 400 Hz, and 10400 Hz. The current between the file and mouth is measured at each of these frequencies, and the readout of the relative proximity to the apex will appear on the meter.

# 8. Indications for Use

Dr's Finder NEO is intended for detecting the apex of root canal.

# 9. Substantial Equivalence Discussion:

The Dr's Finder Neo is substantially equivalent to the Dr's Finder (K151274) and MULTIPLE (APEX LOCATOR), MODEL RCM-7 (K090925). The Dr's Finder Neo and the predicate devices have the same intended use, the same principles of operation and similar technological characteristics.

The major difference between the subject device and the Dr's Finder (K151274) is that the subject device does not have Bluetooth function to transfer the data to the mobile App. The major difference between the subject device and the Apex Locator (K090925) is the frequencies used for measurement.

We have performed side by side performance test on the subject device and the reference device (K090925), and the test results show that the subject device would perform as well as the reference device. We evaluated the reference device (K090925) instead of the primary predicate device (K151274) to compare the performance with the leading model in the market. The reference device has been marketed and used longer than the primary predicate device.

|                        | Subject Device                                                          | Primary Predicate<br>Device                                         | Reference Device                                                                                  |
|------------------------|-------------------------------------------------------------------------|---------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| 510(K) #               | K171867                                                                 | K151274                                                             | K090925                                                                                           |
| Device<br>Name         | Dr's Finder neo<br>(AL-DFA20)                                           | Dr's Finder<br>(AL-DF10)                                            | MULTIPLE (APEX<br>LOCATOR),<br>MODEL RCM-7                                                        |
| Manufacturer           | Good Doctors Co.,<br>Ltd.                                               | Good Doctors Co., Ltd.                                              | J. MRITAMFG. CORP                                                                                 |
| Product<br>Code        | LQY                                                                     | LQY                                                                 | LQY                                                                                               |
| Design                 | Image: Dr's Finder neo                                                  | Image: Dr's Finder                                                  | Image: Apex Locator RCM-7                                                                         |
| Indications<br>for Use | Dr's Finder Neo is<br>intended for detecting<br>the apex of root canal. | Dr's Finder is intended for<br>detecting the apex of root<br>canal. | RCM-7 is a dental device,<br>Apex Locator.<br>It can be used to detect the<br>apex of root canal. |

{5}------------------------------------------------

| Method of<br>calculating<br>location of<br>root | Comparison of<br>impedance<br>multi frequencies | Comparison of<br>impedance<br>multi frequencies | Comparison of<br>impedance at multi<br>frequencies |
|-------------------------------------------------|-------------------------------------------------|-------------------------------------------------|----------------------------------------------------|
| Measurement<br>Current                          | 30 uA, maximum                                  | 30 uA, maximum                                  | 30 uA, maximum                                     |
| Measurement<br>Power                            | Lithium Polymer DC<br>4.2V (rechargeable)       | Lithium Polymer DC 4.2V<br>(rechargeable)       | DC 4.5V LR03<br>(AAA size X3ea) batteries          |
| Power<br>Consumption                            | 0.46 W                                          | 0.29 W                                          | 0.2 W                                              |
| Frequencies<br>used<br>for<br>comparison        | 400 Hz, 5400Hz, and<br>10400 Hz                 | 400 Hz, 5400Hz, and<br>10400 Hz                 |                                                    |
| Display                                         | OLED                                            | OLED                                            | LCD                                                |
| Accessory                                       | File hook, Lip hook,<br>Probe cord              | File hook, Lip hook, Probe<br>cord              | File hook, Lip hook, Probe<br>cord                 |
| Accuracy                                        | ±0.2mm                                          | ±0.2mm                                          |                                                    |
| Dimensions<br>(W x L x H)                       | 43 x 46 x 16 mm                                 | 55 x 85 x 21mm                                  | 60 x 103 x57mm                                     |
| Weight                                          | 24g                                             | 56g                                             | 110g                                               |
| Bluetooth<br>transfer to<br>mobile app          | No                                              | Yes                                             | No                                                 |

## 10. Performance Tests (Non-clinical)

- The comparison testing on Dr's Finder NEO and the predicate device (RCM-7, K090925) ● was performed to compare the measurement performance of locating apex root using extracted tooth and a stainless steel file. Based on the test results, we conclude that the Dr's Finder NEO is substantially equivalent to the reference device.
- . The EMC and electrical safety testing was conducted on the Dr's Finder NEO in accordance with the following standards.

- EN60601-1:2006/A1:2013 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance

- EN60601-1-2:2015 Medical electrical equipment. General requirements for basic safety and essential performance. Collateral Standard. Electromagnetic disturbances. Requirements and tests

- EN 61000-3-2:2014 Electromagnetic compatibility (EMC). Limits. Limits for harmonic current emissions (equipment input current ≤ 16 A per phase) - EN 61000-3-3:2013 Electromagnetic compatibility (EMC). Limits. Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems, for equipment with rated current ≤ 16 A per phase and not subject to conditional connection

{6}------------------------------------------------

- EN 55011:2009/A1:2010 Industrial, scientific and medical equipment. Radiofrequency disturbance characteristics. Limits and methods of measurement

- To verify the sterility assurance level (10-6) for steam sterilization procedure of the file . hook and the lip hook, the biological indicator (BI) overkill method was used in accordance to ISO 17665-1 and ISO 17665-2. In addition, dry time under full cycle was validated.
- Software Validation Test: It is implemented by referring to IEC 60601-1:2005+A1:2012 ● Clause 14 and FDA software validation guidance "General Principles of Software Validation; Final Guidance for Industry and FDA Staff , Document issued on: January 11, 2002".

The test results of non-clinical tests performed on the subject device supported that it is substantially equivalent to the predicate devices despite the differences.

## 11. Conclusions:

Based on the information provided in this premarket notification, Good Doctors Co., Ltd. concludes that the Dr's Finder NEO is substantially equivalent to the predicate device as described herein in.

---

**Source:** [https://fda.innolitics.com/device/K171867](https://fda.innolitics.com/device/K171867)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
