FORTIFY® and FORTIFY® Integrated Corpectomy Spacers are vertebral body replacement devices intended for use in the thoracolumbar spine (T1-L5). FORTIFY® Spacers (titanium) are also intended for use in the cervical spine (C2-T1). All FORTIFY® TPS coated spacers are indicated for the same use as non-coated PEEK versions. When used in the cervical spine (C2-T1), FORTIFY® devices (titanium) are intended for use in skeletally mature patients to replace a diseased or damaged vertebral body caused by tumor fracture or osteomyelitis, or for reconstruction following corpectomy performed to achieve decompression of the spinal tissues in cervical degenerative disorders. These spacers are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion. When used in the thoracolumbar spine (T1-L5), FORTIFY® Integrated devices are intended for use to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). These spacers are designed to provide anterior spinal column support even in the absence of fusion for a prolonged period. The interior of the spacers can be packed with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft as an adjunct to fusion. These devices are intended to be used with FDA-cleared supplemental spinal fixation systems that have been labeled for use in the cervical, thoracic, and/or lumbar spine (i.e., posterior screw and rod systems, anterior plate systems, and anterior screw and rod systems). When used at more than two levels, supplemental fixation should include posterior fixation.
Device Story
Spinal implants (spacers) provide structural stability following discectomy, corpectomy, or vertebrectomy; inserted via anterior, posterior, lateral, or transforaminal approaches. Devices feature superior/inferior ridges to grip vertebral endplates; interior cavities packed with autograft/allogenic bone graft to facilitate fusion. Used in OR by surgeons; provide anterior column support. Subject of 510(k) is update to MRI compatibility labeling; no changes to device design, materials, or function.
Clinical Evidence
Bench testing only. MRI compatibility testing performed per ASTM F2052:2006, ASTM F2119:2007, ASTM F2182:11a, and ASTM F2213:2006. No clinical data required.
Technological Characteristics
Materials: PEEK polymer, titanium alloy, commercially pure titanium, cobalt chromium molybdenum, tantalum markers, hydroxyapatite (HA) coating. Design: Various heights/geometries with superior/inferior ridges for endplate gripping. Connectivity: N/A. Sterilization: Unchanged from predicate.
Indications for Use
Indicated for skeletally mature patients requiring vertebral body replacement (thoracolumbar T1-L5; cervical C2-T1 for titanium) due to tumor, trauma (fracture), osteomyelitis, or degenerative disorders. Used with supplemental fixation systems. Contraindications include patients with insufficient life expectancy for fusion (cervical tumor cases).
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 21, 2017
Globus Medical Inc. Lori Burns Director, Regulatory Affairs 2560 General Armistead Ave. Audubon, Pennsylvania 19403
# Re: K171848
Trade/Device Name: FORTIFY® and FORTIFY® Integrated Corpectomy Spacers, XPand® Corpectomy Spacers, NIKO® Corpectomy Spacers, SUSTAIN® Spacers, COALITION® Spacers, PATRIOT® Lumbar Spacers, PATRIOT® Cervical Spacers, ALTERATM Spacers, RISE® Spacers, CALIBER® Spacers, ELSATM Spacers, LATIS® Spacers, MONUMENT™ Spacers, InterContinental® Plate-Spacers, MAGNIFY™ Spacers, INDEPENDENCE® Spacers Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX, OVD, MOP, PLR, ODP, OVE, KWO Dated: November 17, 2017
Received: November 20, 2017
Dear Ms. Burns:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Katherine D. Kavlock -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171848
Device Name
FORTIF Y ® and FORTIF Y ® Integrated Corpectomy Spacers
### Indications for Use (Describe)
FORTIFY® and FORTIFY® Integrated Corpectomy Spacers are vertebral body replacement devices intended for use in the thoracolumbar spine (T1-L5). FORTIFY® Spacers (titanium) are also intended for use in the cervical spine (C2-T1). All FORTIFY® TPS coated spacers are indicated for the same use as non-coated PEEK versions.
When used in the cervical spine (C2-T1), FORTIFY® devices (titanium) are intended for use in skeletally mature patients to replace a diseased or damaged vertebral body caused by tumor fracture or osteomyelitis, or for reconstruction following corpectomy performed to achieve decompression of the spinal tissues in cervical degenerative disorders. These spacers are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.
When used in the thoracolumbar spine (T1-L5), FORTIFY® Integrated devices are intended for use to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). These spacers are designed to provide anterior spinal column support even in the absence of fusion for a prolonged period.
The interior of the spacers can be packed with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft as an adjunct to fusion.
These devices are intended to be used with FDA-cleared supplemental spinal fixation systems that have been labeled for use in the cervical, thoracic, and/or lumbar spine (i.e., posterior screw and rod systems, anterior plate systems, and anterior screw and rod systems). When used at more than two levels, supplemental fixation should include posterior fixation.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of infornation is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171848
Device Name XPand® Corpectomy Spacers
#### Indications for Use (Describe)
XPand® Corpectomy Spacers (including XPand®-R) are vertebral body replacement devices intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, or unstable vertebral body due to tumor or trauma (i.e., fracture). XPand® Corpectorny Spacers are intended to be used with supplemental spinal fixation systems that have been labeled for use in the thoracic and/or lumbar spine (i.e., posterior pedicle screw and rod systems, anterior plate system, and anterior screw and rod systems). The interior of the spacer can be packed with bone grafting material. XPand® Corpectomy Spacers are designed to provide anterior spinal column support even in the absence of fusion for a prolonged period. All XPand® TPS coated spacers are indicated for the same use as non-coated PEEK versions.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171848
Device Name NIKO® Corpectomy Spacers
NIKO® Corpectomy Spacers are vertebral body replacement devices intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). NIKO® Corpectomy Spacers are intended to be used with supplemental spinal fixation systems that have been labeled for use in the thoracic and/or lumbar spine (i.e., posterior pedicle screw and rod systems, and anterior screw and rod systems). The interior of the spacers can be packed with bone grafting material. NIKO® Corpectomy Spacers are designed to provide anterior spinal column support even in the absence of fusion for a prolonged period. All NIKO® TPS coated spacers are indicated for the same use as non-coated PEEK versions.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171848
Device Name SUSTAIN® Spacers
### Indications for Use (Describe)
When used as lumbar intervertebral body fusion devices, SUSTAIN® Spacers (including SUSTAIN® R, and SUSTAIN®-IR) are intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of nonoperative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis at the involved level(s). SUSTAIN®, SUSTAIN® R and SUSTAIN®-IR Spacers are to be used with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. SUSTAIN® TPS Spacers are to be used with autogenous bone graft material. These devices are intended to be used with supplemental fixation, such as the CREO®, REVERE®, REVOLVE®, or BEACON® Stabilization Systems.
When used as cervical intervertebral body fusion devices, SUSTAIN® Spacers (including SUSTAIN® R) are intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and at least six (6) weeks of non-operative treatment. SUSTAIN® Spacers are to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation, such as the ASSURE®, PROVIDENCE®, or XTEND® Anterior Cervical Plate Systems. All SUSTAIN® TPS coated spacers are indicated for the same use as non-coated PEEK versions.
When used as vertebral body replacement devices, SUSTAIN® Spacers (including SUSTAIN® R TPS) are intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, or unstable vertebral body due to tumor or trauma (i.e., fracture). The spacers are intended to be used with supplemental spinal frave been labeled for use in the thoracic and/or lumbar spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). The interior of the spacers can be packed with bone grafting material. SUSTAIN® Spacers are designed to provide anterior spinal column support even in the absence of fusion for a prolonged period. All SUSTAIN® TPS coated spacers are indicated for the same use as non-coated PEEK versions.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{6}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{7}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171848
Device Name COALITION® Spacers
#### Indications for Use (Describe)
COALITION® Spacers are interbody fusion devices intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) for one or two contiguous levels, depending on the system. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. These devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone, depending on the system.
The COALITION® Spacer is a stand-alone interbody fusion device intended for use at one or two levels of the cervical spine (C2-T1) and is to be used with two titanium alloy screws which accompany the implant. The spacer is to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone.
The COALITION MIS™ Spacer is an interbody fusion device and is to be used with two titanium alloy screws or anchors which accompany the implants. When used with screws, COALITION MIS™ Spacers are stand-alone interbody fusion devices intended for use at one or two levels of the cervical spine (C2-T1). When used with anchors, COALITION MIS™ Spacers are intended for use at one level of the cervical spine (C2-T1) with additional supplemental fixation such as posterior cervical screw fixation. The spacer is to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone.
The COALITION AGX™ Spacer is intended to be used with supplemental fixation, such as anterior cervical plates or posterior cervical screw fixation. When used with the COALITION AGX™ Plate, the plate-spacer assembly takes on the indications for use of the COALITION AGX™ Spacer, with the COALITION AGX™ Plate acting as the supplemental fixation. The COALITION AGX™ Plate and Spacer assembly is a stand-alone device intended for use at one level of the cervical spine (C2-T1) and is to be used with two titanium alloy screws which accompany the implant. The spacer is to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone.
The COALITION AGX™ Plate is intended for anterior screw fixation to the cervical spine (C2-C7) for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), turnors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudoarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis.
COALITION® TPS Spacers are stand-alone interbody fusion devices intended for use at one level of the cervical spine (C2-T1) and are to be used with two titanium alloy screws which accompany the implants. The spacers are to be filled with autogenous bone graft.
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> □ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED
Form Approved: OMB No. 0910-0120
{8}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments reqarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{9}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171848
Device Name PATRIOT® Lumbar Spacers
### Indications for Use (Describe)
PATRIOT® Spacers (including Constitution®, Signature®, TransContinental®, and TransContinental® M) are interbody fusion devices intents with degenerative disc disease (DDD) at one or two continuous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis at the involved level(s). PATRIOT® Spacers are to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine (e.g. posterior pedicle screw and rod systems, anterior plate systems, and rod systems). Hyperlordotic interbody devices (>20° lordosis) must be used with at least anterior supplemental fixation. All PATRIOT® TPS coated spacers are indicated for the same use as non-coated PEEK versions.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
| _ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{10}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171848
Device Name PATRIOT® Cervical Spacers
### Indications for Use (Describe)
PATRIOT® Spacers (including COLONIAL®) are interbody fusion devices intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) at one level. DDD is defined as discogenic pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. All PATRIOT® TPS coated spacers are indicated for the same use as non-coated PEEK versions.
PATRIOT® Spacers are to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation, such as the ASSURE® or PROVIDENCE® Anterior Cervical Plate Systems.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{11}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171848
Device Name ALTERATM Spacers
### Indications for Use (Describe)
The ALTERA™ Spacer is an interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).
The ALTERA™ Spacer is to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{12}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171848
Device Name RISE® Spacers
Indications for Use (Describe)
The RISE® Spacer is a lumbar interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).
The RISE® Spacer is to be filled with autogenous bone graft material. This device is intended to be used with supplemental fixation, such as the REVERE® or REVOLVE® Stabilization Systems.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{13}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171484
Device Name CALIBER® Spacers
### Indications for Use (Describe)
CALIBER® Spacers are interbody fusion devices intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). All CALIBER® TPS coated spacers are indicated for the same use as non-coated PEEK versions.
CALIBER® Spacers are to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation, such as the CREO®, REVERE® or REVOLVE® Stabilization Systems.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{14}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171848
Device Name ELSA™ Spacers
### Indications for Use (Describe)
The ELSA™ Spacer is an interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The ELSA™ Spacer is to be filled with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone, may be used with two bone screws, and is to be used with supplemental fixation. Hyperlordotic (≥20°) implants must be used with the two bone screws and supplemental fixation in addition to the bone screws.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{15}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171848
Device Name LATIS® Spacers
### Indications for Use (Describe)
LATIS® Spacers are interbody for intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).
LATIS® Spacers are to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{16}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171848
Device Name MONUMENT™ Spacers
### Indications for Use (Describe)
The MONUMENT™ Spacer is an interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The MONUMENT™ Spacer is to be filled with autogenous bone graft material, and is to be used with four titanium alloy screws that accompany the implant. The device is intended to be used with supplemental fixation (i.e. pedicle screws, facet fixation).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{17}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171848
Device Name InterContinental® Plate-Spacers
#### Indications for Use (Describe)
InterContinental® Plate-Spacers are lateral lumbar interbody fusion devices intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be sketally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). InterContinental® Plate-Spacers are to be filled with autogenous bone graft material, and are to be used with two titanium alloy screws which accompany the implant. These devices are intended to be used with supplemental fixation (e.g. pedicle or facet screw systems) in addition to the integrated screws. All InterContinental® TPS coated spacers are indicated for the same use as non-coated PEEK versions.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
|X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{18}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171848
Device Name MAGNIFY™ Spacers
### Indications for Use (Describe)
The MAGNIFY™ Spacer is an interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The MAGNIFY™ Spacer is to be filled with autogenous bone graft material, and is to be used with supplemental fixation, such as the CREO®, REVERE® or REVOLVE® Stabilization Systems.
The MAGNIFY™-S Spacer is a stand-alone interbody fusion device in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The MAGNIFY™-S Spacer is to be filled with autogenous bone graft material, and is to be used with three titanium alloy screws that accompany each implant.
Type of Use (*Select one or both, as applicable*)
| <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------|------------------------------------------------------------|
|-------------------------------------------------------------|------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{19}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171848
Device Name INDEPENDENCE® Spacers
### Indications for Use (Describe)
INDEPENDENCE® (including INDEPENDENCE MIS™, and INDEPENDENCE MIS AGX™ Spacers) are interbody fusion devices intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1), DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis at the involved level(s). INDEPENDENCE® Spacers are to be filled with autograft bone graft composed of cancellous and/ or corticocancellous bone. All INDEPENDENCE® TPS coated spacers are indicated for the same use as non-coated PEEK versions.
INDEPENDENCE® are stand-alone interbody fusion devices intended to be used with three titanium alloy screws which accompany the implants.
INDEPENDENCE MIS™ are interbody fusion devices intended to be used with three titanium alloy screws or anchors which accompany the implants. When used with screws, these devices are stand-alone interbody fusion devices. When used with anchors, these devices are intended for use with supplemental fixation (e.g. facet screws or posterior fixation), Hyperlordotic implants (≥25° lordosis) are intended for use with supplemental fixation (e.g. facet screws or posterior fixation).
INDEPENDENCE MIS AGX™ Integrated Spacer and INDEPENDENCE MIS AGX™ Integrated Ti Spacer are interbody fusion devices that may be used with three titanium alloy screws or anchors which accompany the implants. When used with screws, these devices are stand-alone interbody fusion devices. When used with anchors, these devices are intended for use with supplemental fixation (e.g. facet screws or posterior fixation). Hyperlordotic implants (≥25° lordosis) are intended for use with supplemental fixation (e.g. facet screws or posterior fixation).
INDEPENDENCE MIS AGX™ Spacers are C-shaped, non-integrated PEEK spacers that are intended to be used with supplemental fixation (e.g. facet screws or posterior fixation). When used in conjunction with the INDEPENDENCE MIS AGX™ Integrated Ti Spacer, these devices become the INDEPENDENCE MIS AGX™ Integrated Spacer.
Type of Use (Select one or both, as applicable):
| | <span> <b> <span style="text-decoration: overline;">X</span> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--|------------------------------------------------------------------------------------------------------------------------------------|
| | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
{20}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{21}------------------------------------------------
# 510(k) Summary: MRI Compatibility for Intervertebral Fusion and Vertebral Body Replacement Implants
| Company: | Globus Medical Inc.<br>2560 General Armistead Ave.<br>Audubon, PA 19403<br>610-930-1800 |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact: | Lori Burns<br>Director of Regulatory Affairs |
| Secondary Contact: | Kelly J. Baker, Ph.D.<br>Senior Vice President, Regulatory and Clinical Affairs |
| Date Prepared: | December 20, 2017 |
| Device Name: | FORTIFY® and FORTIFY® Integrated<br>Corpectomy Spacers<br>XPand® Corpectomy Spacers<br>NIKO® Corpectomy Spacers<br>SUSTAIN® Spacers<br>COALITION® Spacers<br>PATRIOT® Lumbar Spacers<br>PATRIOT® Cervical Spacers<br>ALTERA™ Spacers<br>RISE® Spacers<br>CALIBER® Spacers<br>ELSA™ Spacers<br>LATIS® Spacers<br>MONUMENT™ Spacers<br>InterContinental® Plate-Spacers<br>MAGNIFY™ Spacers<br>INDEPENDENCE® Spacers |
| Classification: | Per 21 CFR as follows:<br>§888.3060 Spinal Intervertebral Body Fixation Orthosis<br>and/or<br>§888.3080 Intervertebral Body Fusion Device |
| | PLR FORTIFY® Corpectomy Spacers |
| | MQP FORTIFY® and FORTIFY® Integrated Corpectomy Spacers |
| | XPand® Corpectomy Spacers |
| | NIKO® Corpectomy Spacers |
| | SUSTAIN® Spacers (Cervical & Lumbar) |
| | MAX SUSTAIN® Spacers (Lumbar) |
| | PATRIOT® Spacers (Lumbar) |
{22}------------------------------------------------
TransContinental® M ALTERA™ Spacers RISE® Spacers CALIBER® Spacers LATIS® Spacers MAGNIFY™ Spacers INDEPENDENCE® Spacers ODP SUSTAIN® Spacers (Cervical) COALITION AGX™ Spacers
- PATRIOT® Spacers (Cervical) OVE COALITION® Spacers
- COALITION AGX™ Spacers KWQ COALITION AGX™ Spacer
- OVD ELSA™ Spacers MONUMENT™ Spacers InterContinental® Plate-Spacer MAGNIFY™ Spacers INDEPENDENCE® Spacers
Requlatory Class: II, Panel Code: 87
Primary Predicate: FORTIFY® I Corpectomy Spacers (K121107)
Additional
FORTIFY® Corpectomy Spacers (K112756) Predicates: XPand® Corpectomy Spacers (K060665) NIKO® Corpectomy Spacers (K072465) SUSTAIN® Spacers (K130478 & K151665) COALITION® Spacers (K151939) PATRIOT® Spacers (Lumbar) (K122097, K161223) PATRIOT® Spacers (Cervical) (K072991) ALTERA® Spacers (K140411) RISE® Spacers (K113447) CALIBER® Spacers (K123231) ELSA™ Spacers (K161379) LATIS® Spacers (K123913) MONUMENT® Spacers (K132559) InterContinental® Plate-Spacer (K103382, K161223) MAGNIFY™ Spacers (K142498) INDEPENDENCE® Spacers (K160597, K170157) TPS Spacers (K143578)
{23}------------------------------------------------
## Purpose:
The purpose of this 510(k) is to update the labeling for MRI compatibility.
# Device Description:
Globus Medical's interbody fusion and vertebral body replacement devices are cervical, thoracic or lumbar devices used to provide structural stability in skeletally mature individuals following discectomy, corpectomy or vertebrectomy and may be inserted using an anterior, posterior, lateral or transforaminal approach. The devices are available in various heights and geometric options to accommodate surgical approaches and patient anatomy. Ridges on the superior and inferior surfaces of each device grip the endplates of the adjacent vertebrae to resist expulsion. These devices are to be used in conjunction with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft to be packed inside the device.
These devices are made from radiolucent PEEK polymer with titanium alloy or tantalum markers or titanium alloy, commercially pure titanium, titanium alloy, and cobalt chromium molybdenum al. Implants are also available with hydroxyapatite (HA) coating.
### Indications for Use:
# FORTIFY® and FORTIFY® Integrated Corpectomy Spacers
FORTIFY® and FORTIFY® Integrated Corpectomy Spacers are vertebral body replacement devices intended for use in the thoracolumbar spine (T1-L5). FORTIFY® Spacers (titanium) are also intended for use in the cervical spine (C2-T1). All FORTIFY® TPS coated spacers are indicated for the same use as noncoated PEEK versions.
When used in the cervical spine (C2-T1), FORTIFY® devices (titanium) are intended for use in skeletally mature patients to replace a diseased or damaged vertebral body caused by tumor fracture or osteomyelitis, or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in cervical degenerative disorders. These spacers are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. with bone graft used at the surgeon's discretion.
When used in the thoracolumbar spine (T1-L5), FORTIFY® and FORTIFY® Integrated devices are intended for use to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). These spacers are designed to provide anterior spinal column support even in the absence of fusion for a prolonged period.
The interior of the spacers can be packed with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft as an adjunct to fusion.
{24}------------------------------------------------
These devices are intended to be used with FDA-cleared supplemental spinal fixation systems that have been labeled for use in the cervical, thoracic, and/or lumbar spine (i.e., posterior screw and rod systems, anterior plate systems, and anterior screw and rod systems). When used at more than two levels, supplemental fixation should include posterior fixation.
# XPand® Corpectomy Spacers
XPand® Corpectomy Spacers (including XPand®-R) are vertebral body replacement devices intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). XPand® Corpectomy Spacers are intended to be used with supplemental spinal fixation systems that have been labeled for use in the thoracic and/or lumbar spine (i.e., posterior pedicle screw and rod systems, anterior plate system, and anterior screw and rod systems). The interior of the spacer can be packed with bone grafting material. XPand® Corpectomy Spacers are designed to provide anterior spinal column support even in the absence of fusion for a prolonged period. All XPand®TPS coated spacers are indicated for the same use as non-coated PEEK versions.
# NIKO® Corpectomy Spacers
NIKO® Corpectomy Spacers are vertebral body replacement devices intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). NIKO® Corpectomy Spacers are intended to be used with supplemental spinal fixation systems that have been labeled for use in the thoracic and/or lumbar spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). The interior of the spacers can be packed with bone grafting material. NIKO® Corpectomy Spacers are designed to provide anterior spinal column support even in the absence of fusion for a prolonged period. All NIKO® TPS coated spacers are indicated for the same use as non-coated PEEK versions.
### SUSTAIN® Spacers
When used as lumbar intervertebral body fusion devices, SUSTAIN® Spacers (including SUSTAIN® R, and SUSTAIN®-IR) are intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of nonoperative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). SUSTAIN®, SUSTAIN® R and SUSTAIN®-IR Spacers are to be used with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. SUSTAIN® TPS Spacers are to be used with autogenous bone graft material. These devices are intended to be used with supplemental fixation, such as the CREO®, REVERE®, REVOLVE®, or BEACON® Stabilization Systems.
{25}------------------------------------------------
When used as cervical intervertebral body fusion devices, SUSTAIN® Spacers (including SUSTAIN® R) are intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. SUSTAIN® Spacers are to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation, such as the ASSURE®, PROVIDENCE®, or XTEND® Anterior Cervical Plate Systems. All SUSTAIN® TPS coated spacers are indicated for the same use as non-coated PEEK versions.
When used as vertebral body replacement devices, SUSTAIN® Spacers (including SUSTAIN® and SUSTAIN® R TPS) are intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaqed, or unstable vertebral body due to tumor or trauma (i.e., fracture). The spacers are intended to be used with supplemental spinal fixation systems that have been labeled for use in the thoracic and/or lumbar spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). The interior of the spacers can be packed with bone grafting material. SUSTAIN® Spacers are designed to provide anterior spinal column support even in the absence of fusion for a prolonged period. All SUSTAIN® TPS coated spacers are indicated for the same use as non-coated PEEK versions.
### COALITION® Spacers
COALITION® Spacers are interbody fusion devices intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) for one or two contiquous levels, depending on the system. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. These devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone, depending on the system.
The COALITION® Spacer is a stand-alone interbody fusion device intended for use at one or two levels of the cervical spine (C2-T1) and is to be used with two titanium alloy screws which accompany the implant. The spacer is to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone.
The COALITION MIS™ Spacer is an interbody fusion device and is to be used with two titanium alloy screws or anchors which accompany the implants. When used with screws, COALITION MIS™ Spacers are stand-alone interbody fusion devices intended for use at one or two levels of the cervical spine (C2-T1). When used with anchors, COALITION MIS™ Spacers are intended for use at one level of the cervical spine (C2-T1) with additional supplemental fixation such as posterior
{26}------------------------------------------------
cervical screw fixation. The spacer is to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone.
The COALITION AGX™ Spacer is intended to be used with supplemental fixation, such as anterior cervical plates or posterior cervical screw fixation. When used with the COALITION AGX™ Plate, the plate-spacer assembly takes on the indications for use of the COALITION AGX™ Spacer, with the COALITION AGX™ Plate acting as the supplemental fixation. The COALITION AGX™ Plate and Spacer assembly is a stand-alone device intended for use at one level of the cervical spine (C2-T1) and is to be used with two titanium alloy screws which accompany the implant. The spacer is to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone.
The COALITION AGX™ Plate is intended for anterior screw fixation to the cervical spine (C2-C7) for the following indications: deqenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudoarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis.
COALITION® TPS Spacers are stand-alone interbody fusion devices intended for use at one level of the cervical spine (C2-T1) and are to be used with two titanium alloy screws which accompany the implants. The spacers are to be filled with autogenous bone graft.
# PATRIOT® Lumbar Spacers
PATRIOT® Spacers (including Constitution®, Signature®, Continental®, TransContinental®, and TransContinental® M) are interbody fusion devices intended for use in patients with degenerative disc disease (DDD) at one or two contiquous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). PATRIOT® Spacers are to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine (e.g. posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation. All PATRIOT® TPS coated spacers are indicated for the same use as noncoated PEEK versions.
### PATRIOT® Cervical Spacers
PATRIOT® Spacers (including COLONIAL®) are interbody fusion devices intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) at one level. DDD is defined as discogenic pain with
{27}------------------------------------------------
degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of nonoperative treatment. All PATRIOT® TPS coated spacers are indicated for the same use as non-coated PEEK versions.
PATRIOT® Spacers are to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation, such as the ASSURE® or PROVIDENCE® Anterior Cervical Plate Systems.
# ALTERA™ Spacers
The ALTERA™ Spacer is an interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of nonoperative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).
The ALTERA™ Spacer is to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation.
# RISE® Spacers
The RISE® Spacer is a lumbar interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of nonoperative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).
The RISE® Spacer is to be filled with autogenous bone graft material. This device is intended to be used with supplemental fixation, such as the REVERE® or REVOLVE® Stabilization Systems.
### CALIBER® Spacers
CALIBER® Spacers are interbody fusion devices intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of nonoperative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). All CALIBER® TPS coated spacers are indicated for the same use as non-coated PEEK versions.
{28}------------------------------------------------
CALIBER® Spacers are to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation, such as the CREO®, REVERE® or REVOLVE® Stabilization Systems.
# ELSA™ Spacers
The ELSA™ Spacer is an interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of nonoperative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The ELSA™ Spacer is to be filled with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone, may be used with two bone screws, and is to be used with supplemental fixation. Hyperlordotic (≥20°) implants must be used with the two bone screws and supplemental fixation in addition to the bone screws.
# LATIS® Spacers
LATIS® Spacers are interbody fusion devices intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of nonoperative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).
LATIS® Spacers are to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation.
### MONUMENT® Spacers
The MONUMENT® Spacer is an interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of nonoperative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The MONUMENT® Spacer is to be filled with autogenous bone graft material, and is to be used with four titanium alloy screws that accompany the implant. The device is intended to be used with supplemental fixation (i.e. pedicle screws, facet fixation).
### InterContinental® Plate-Spacers
InterContinental® Plate-Spacers are lateral lumbar interbody fusion devices intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic
{29}------------------------------------------------
studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). InterContinental® Plate-Spacers are to be filled with autogenous bone graff material, and are to be used with two titanium alloy screws which accompany the implant. These devices are intended to be used with supplemental fixation (e.g. pedicle or facet screw systems) in addition to the integrated screws. All InterContinental® TPS coated spacers are indicated for the same use as noncoated PEEK versions.
### MAGNIFY™ Spacers
The MAGNIFY™ Spacer is an interbody fusion device intended for use in patients with deqenerative disc disease (DDD) at one or two contiquous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of nonoperative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The MAGNIFY™ Spacer is to be filled with autogenous bone graft material, and is to be used with supplemental fixation, such as the CREO®, REVERE® or REVOLVE® Stabilization Systems.
The MAGNIFY™-S Spacer is a stand-alone interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of nonoperative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The MAGNIFY™-S Spacer is to be filled with autogenous bone graft material, and is to be used with three titanium alloy screws that accompany each implant.
# INDEPENDENCE® Spacers
INDEPENDENCE® (including-INDEPENDENCE MIS™, and INDEPENDENCE MIS AGX™ Spacers) are interbody fusion devices intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of nonoperative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). INDEPENDENCE® Spacers are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. All INDEPENDENCE® TPS coated spacers are indicated for the same use as non-coated PEEK versions.
{30}------------------------------------------------
INDEPENDENCE® are stand-alone interbody fusion devices intended to be used with three titanium alloy screws which accompany the implants.
INDEPENDENCE MIS™ are interbody fusion devices intended to be used with three titanium alloy screws or anchors which accompany the implants. When used with screws, these devices are stand-alone interbody fusion devices. When used with anchors, these devices are intended for use with supplemental fixation (e.g. facet screws or posterior fixation). Hyperlordotic implants (≥25° lordosis) are intended for use with supplemental fixation (e.g. facet screws or posterior fixation).
INDEPENDENCE MIS AGX™ Integrated Spacer and INDEPENDENCE MIS AGX™ Integrated Ti Spacer are interbody fusion devices that may be used with three titanium alloy screws or anchors which accompany the implants. When used with screws, these devices are stand-alone interbody fusion devices. When used with anchors, these devices are intended for use with supplemental fixation (e.g. facet screws or posterior fixation). Hyperlordotic implants (≥25° lordosis) are intended for use with supplemental fixation (e.g. facet screws or posterior fixation).
INDEPENDENCE MIS AGX™ Spacers are C-shaped, non-integrated PEEK spacers that are intended to be used with supplemental fixation (e.g. facet screws or posterior fixation). When used in conjunction with the INDEPENDENCE MIS AGX™ Integrated Ti Spacer, these devices become the INDEPENDENCE MIS AGX™ Integrated Spacer.
# Summary of Technological Characteristics as Compared to the Predicate
The technological characteristics remain the same between the subject devices and the predicate devices. No changes have been made to any of the devices. This submission is only to update the labeling to include Magnetic Resonance Imaging (MRI) information.
### Performance Data:
MRI testing was performed on the subject devices per the following ASTM Standards:
- ASTM F2052:2006 Standard Test Method for Measurement of Magnetically ● Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment
- ASTM F2119:2007 Standard Test Method for Evaluation of MR Image ● Artifacts from Passive Implants
- ASTM F2182:11a Standard Test Method of Measurement of Radio ● Frequency Induced Heating Near Passive Implants During Magnetic Resonance Imaging
- ASTM F2213:2006 Standard Test Method for Measurement of Magnetically . Induced Torque on Medical Devices in the Magnetic Resonance Environment
{31}------------------------------------------------
No further device performance testing was required for this submission. The performance testing remains the same for the subject and predicate devices.
No further sterilization, biocompatibility and endotoxin evaluation and/or testing were required for this submission. The sterilization, biocompatibility and endotoxin testing remains the same for the subject and predicate devices.
## Basis of Substantial Equivalence:
The subject devices are substantially equivalent to the predicate devices with respect to technical characteristics, performance, and intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.