Hunan Handlike Minimally Invasive Surgery Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Feb 13, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Endoscopic Instruments are designed to cut, dissect, manipulate and/or cauterize various tissues during endoscopic/laparoscopic, general surgical procedures.
Device Story
Endoscopic surgical instruments (disposable and reusable) used in endoscopic/laparoscopic general surgery. Devices include monopolar electrosurgical instruments for tissue manipulation, cutting, and coagulation, and non-powered instruments for manipulation and dissection. Instruments feature a handle, shaft, and jaw (dissector, grasper, or scissor). Operated by surgeons in clinical settings. Monopolar variants connect to electrosurgical generators to deliver high-frequency current for cauterization/coagulation. Disposable versions provided sterile; reusable versions require steam sterilization by end-users after initial use. Output is mechanical tissue interaction or thermal coagulation. Benefits include minimally invasive access to internal tissues for surgical intervention.
Clinical Evidence
No clinical data. Substantial equivalence based on bench testing, including electrical safety (IEC 60601-1, 60601-1-2, 60601-2-2, 60601-2-18), biocompatibility (ISO 10993), and performance testing (appearance, dimensions, corrosion resistance, holding forces, cutting efficacy, and thermal damage zones on porcine tissue).
Technological Characteristics
5mm diameter, 340mm length. Monopolar electrosurgical or non-powered mechanical. Materials comply with ISO 10993. Sterilization: EO (disposable/reusable) and Steam (reusable). Standards: IEC 60601-1 (General safety), IEC 60601-1-2 (EMC), IEC 60601-2-2 (HF surgical equipment), IEC 60601-2-18 (Endoscopic equipment).
Indications for Use
Indicated for cutting, dissecting, manipulating, and/or cauterizing soft tissue during endoscopic or laparoscopic general surgical procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Endoscopic Monopolar Instruments and Accessories (K102921)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 13, 2018
Hunan Handlike Minimally Invasive Surgery Co., Ltd % Ms. Elly Xu Shenzhen Joyantech Consulting Co., Ltd Room 1122, No.55 Shizhou Middle Road, Nanshan District Shenzhen, Guangdong, P.R.China
Re: K171825
Trade/Device Name: Endoscopic Surgical Instrument Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: December 10, 2017 Received: December 18, 2017
Dear Ms. Xu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{2}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Jennifer R. Stevenson -
# 53
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171825
Device Name Endoscopic Surgical Instrument
Indications for Use (Describe)
Endoscopic Instruments are designed to cut, dissect, manipulate and/or cauterize various tissues during endoscopic/ laparoscopic, general surgical procedures.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Product: Endoscopic Surgical Instruments
Version: A/0
## VOL 005_510(k) Summary
This 510(K) Summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
#### 1. Submission Sponsor
| Applicant Name | Hunan Handlike Minimally Invasive Surgery Co.,Ltd |
|----------------|-----------------------------------------------------------------------------|
| Address | A4, Lugu Yuyuan, #27, Wenxuan Rd, High-tech Zone,<br>Changsha, Hunan, CHINA |
| Post Code | 410205 |
| Phone No. | 86-731-89727782 |
| Fax No. | 86-731-88830694 |
| Contact Person | Hanhan Hu |
| Email | huhanhan@handlike.cn |
| Date Prepared | 2018-2-10 |
#### 2. Submission correspondent
Image /page/3/Picture/10 description: The image shows a company logo with the text "Name" and "Address" above it. The logo is green and features a stylized animal head with wings. Below the logo, there are four Chinese characters, which translate to "Zhuo Yuan Tian Cheng."
Email
Shenzhen Joyantech Consulting Co., Ltd Room 1122, No.55 Shizhou Middle Road, Nanshan District, Shenzhen, Guangdong, P.R.China Post Code 518000 Phone No. 86-755-86069197 Contact Person Elly Xu; Field Fu
elly@cefda.com; cefda13485@163.com
#### 3. Devices Identification
| Trade name | Endoscopic Surgical Instrument |
|-----------------------|------------------------------------------------------|
| Common name | Endoscopic Surgical Instrument |
| Model | Disposable Endoscopic Surgical Instrument |
| | Reusable Endoscopic Surgical Instrument |
| Classification | II |
| Classification name | Electrosurgical, Cutting & Coagulation & Accessories |
| Regulation number | 878.4400 |
| Product code | GEI |
| 510(k) review panel | General & Plastic Surgery |
| Performance standards | The performance and safety was evaluated in |
{5}------------------------------------------------
#### K171825
| Hunan Handlike Minimally Invasive Surgery Co.,Ltd | | VOL_005:001_510(K) Summary |
|---------------------------------------------------|--|--------------------------------------------------------------------------------------------|
| Product: Endoscopic Surgical Instruments | | Version: A/0 |
| | | accordance with Electrosurgical Device Guidance<br>of FDA. |
| | | Biocompatibility tests were done in conformance<br>with relevant requirements of ISO10993. |
#### 4. Leqally Marketed Predicate Devices
| Trade Name | Endoscopic Surgical Instruments |
|-------------------|------------------------------------------------------|
| Regulation number | 878.4400 |
| Regulation class | II |
| Regulation name | Electrosurgical, Cutting & Coagulation & Accessories |
| 510(k) Number | K132771 |
| Product Code | GEI |
| Manufacturer | Qingdao Kingston Medical Devices Ltd. |
| Trade Name | Endoscopic Surgical Instruments |
|-------------------|------------------------------------------------------|
| Regulation number | 878.4400 |
| Regulation class | II |
| Regulation name | Electrosurgical, Cutting & Coagulation & Accessories |
| 510(k) Number | K102921 |
| Product Code | GEI |
| Manufacturer | BEMA GmbH & Co. KG |
#### 5. Device Description
Endoscopic Surgical Instrument is a kind of device used to cut, dissect, manipulate and/or cauterize various tissues during endoscopic/laparoscopic, general surgical procedures. The device includes Disposable Endoscopic Surgical Instrument and Reusable Endoscopic Surgical Instrument. Both of the disposable and reusable endoscopic instruments are provided sterile when they're released. After used for the first time, Reusable Endoscopic Instrument needs to be sterilized by the end users. The instruments are 5mm in diameter as well as 340mm in length.
#### 6. Indications for Use Statement
Endoscopic Instruments are designed to cut, dissect, manipulate and/or cauterize various tissues during endoscopic/laparoscopic, general surgical procedures.
#### 7. Substantial Equivalence Discussion
#### 7.1 Comparison between Disposable Endoscopic Surgical Instrument and K132771
| Characteristic | Disposable Endoscopic<br>Surgical Instrument | Endoscopic Surgical<br>Instrument | Comments |
|----------------|----------------------------------------------|-----------------------------------|----------|
| Indication for | Disposable Endoscopic | The Endoscopic Surgical | Same |
{6}------------------------------------------------
Hunan Handlike Minimally Invasive Surgery Co.,Ltd
Product: Endoscopic Surgical Instruments
VOL_005:001_510(K) Summary Version: A/0
| use/Intended use | Surgical is designed to cut, dissect, manipulate and/or cauterize various tissues during endoscopic/laparoscopic, general surgical procedures. | Instruments are intended for cutting, grasping, dissecting and coagulation of soft tissue in endoscopic surgical procedures. They are intended for single patient use. | Same | | |
|------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|------------------------------------------------|------|
| Operation mode | Devices are either monopolar endoscopic instruments for tissue manipulation and cutting/coagulation, or non-powered endoscopic instruments for tissue manipulation dissection, and cutting. | Devices are either monopolar endoscopic instruments for tissue manipulation and cutting/coagulation, or non-powered endoscopic instruments for tissue manipulation dissection, and cutting. | | | |
| Structure | Handle, shaft and dissector/grasper/scissor jaw | Handle, shaft and dissector/grasper/scissor jaw | | | |
| Diameter | 5mm/34cm | 5mm/33cm (or 42cm) | Similar | | |
| Sterilization | EO | EO | Same | | |
| Single Use | Yes | Yes | Same | | |
| Biocompatibility | Comply with<br>ISO 10993-5<br>ISO 10993-7<br>ISO 10993-10<br>ISO 10993-11<br>ISO 10993-12 | Comply with<br>ISO 10993-5<br>ISO 10993-7<br>ISO 10993-10<br>ISO 10993-11<br>ISO 10993-12 | Same | | |
| | Electric Safety, EMC and<br>Performance Test | IEC 60601-1,<br>IEC 60601-1-2<br>IEC 60601-2-2<br>IEC 60601-2-18 | | IEC 60601-1,<br>IEC 60601-1-2<br>IEC 60601-2-2 | Same |
## 7.2 Comparison between Reusable Endoscopic Surgical Instrument and K102921
| Characteristic | Reusable Endoscopic<br>Surgical Instrument | Endoscopic Monopolar<br>Instruments and<br>Accessories | Comments |
|-------------------------------------|-----------------------------------------------------------------------------------|--------------------------------------------------------------------|--------------|
| Indication for use/<br>Intended use | Reusable Endoscopic<br>Surgical is designed to cut,<br>dissect, manipulate and/or | Endoscopic Monopolar<br>Instruments and<br>Accessories are used in | Same |
| | | | |
| | cauterize various tissues | laparoscopic and other | |
| | during | minimally invasive | |
| | endoscopic/laparoscopic, | procedures for cutting, | |
| | general surgical | dissection, fixation and | |
| | procedures. | taking of biopsy samples, | |
| | | depending on the design | |
| | | of the tip. They are also | |
| | | intended to control | |
| | | bleeding by use of | |
| | | monopolar high-frequency | |
| | | electrical current. | |
| | Devices are either | Devices are either | |
| | monopolar endoscopic | monopolar endoscopic | |
| | instruments for tissue | instruments for tissue | |
| | manipulation and | manipulation and | |
| Operation mode | cutting/coagulation, or | cutting/coagulation, or | Same |
| | non-powered endoscopic | non-powered endoscopic | |
| | instruments for tissue | instruments for tissue | |
| | manipulation dissection, | manipulation dissection, | |
| | and cutting. | and cutting. | |
| | Handle, shaft and | | |
| Structure | dissector/grasper/scissor | Handle, shaft and tip | Same |
| | jaw | | |
| | Assembly type, | Two and Three-piece | Similar, see |
| Design | Tip-detachable type | modular system, | Note 1 |
| | | Single-piece instrument | |
| Diameter | 5mm/34cm | Unknown | See Note 2 |
| Sterilization | EO and Steam | Steam | See Note 3 |
| | Comply with | Comply with | |
| Biocompatibility | ISO 10993-5 | ISO 10993-5 | |
| | ISO 10993-7 | ISO 10993-7 | Same |
| | ISO 10993-10 | ISO 10993-10 | |
| | ISO 10993-11 | ISO 10993-11 | |
| | ISO 10993-12 | ISO 10993-12 | |
| | IEC 60601-1 | IEC 60601-1 | |
| Electric Safety,<br>EMC and | IEC 60601-1-2 | | Same |
| | IEC 60601-2-2 | IEC 60601-1-2 | |
| Performance Test | IEC 60601-2-18 | IEC 60601-2-2 | |
{7}------------------------------------------------
Hunan Handlike Minimally Invasive Surgery Co.,Ltd
Product: Endoscopic Surgical Instruments
VOL_005:001_510(K) Summary Version: A/0
Note 1: The Assembly type and Tip-detachable type is similar with Two-piece modular system. Both the subject device and the predicate device can be detached into two pieces.
Note 2: The diameter of Reusable Endoscopic Surgical Instrument is similar to Page 4 of 5
{8}------------------------------------------------
| Hunan Handlike Minimally Invasive Surgery Co.,Ltd | VOL_005:001_510(K) Summary |
|---------------------------------------------------|----------------------------|
| Product: Endoscopic Surgical Instruments | Version: A/0 |
Endoscopic Surgical Instrument (K132711).
Note 3: The reusable endoscopic instruments are provided sterile when they're released from the manufacture. After used for the first time, Reusable Endoscopic Instrument needs to be steam sterilized by the end users. Both the steam sterilization have been validated, and the test reports show that the sterilization effect of the proposed device can achieve a Sterility Assurance Level (SAL) of 10°.
#### 8. Non-clinical Testing
Evidence of safety and effectiveness was obtained from electrical safety testing and performance testing as following:
Testing in accordance with IEC 60601-1
Testing in accordance with IEC 60601-1-2
Testing in accordance with IEC 60601-2-2
Testing in accordance with IEC 60601-2-18
Moreover, other performance testing of Endoscopic Surgical Instruments were conducted included appearance, basic dimension, corrosion resistance, transverse holding force, longitudinal holding force, ratchet holding force, cutting efficacy and thermal effects on tissue. The thermal effect have been performed on the fresh pig's liver, kidney and muscle tissue. And the thermal damage zone is measured under magnification using histology after performing cauterizing. All the testing results show that the Endoscopic Surgical Instruments meet the specification and performance characteristics as identified in internal design control procedures.
Biocompatibility testing of the Endoscopic Surgical Instrument confirmed that the device meets the applicable requirements of the FDA Blue Book Memorandum G95-11 entitled Use of International Standards ISO 10993-1:2009 Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing within a risk management process. All testings above were conducted per Electrosurgical Device Guidance of FDA.
#### 9. Clinical Testing
Substantial equivalence does not depend on clinical test data.
#### 10. Conclusions
Based on device comparison information and non-clinical testing, the differences between Endoscopic Surgical Instrument and predicate devices will be not raise any new issues of safety and effectiveness, Endoscopic Surgical Instrument is substantially equivalent to legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.