K171803 · Hivox Biotek, Inc. · NUH · Nov 29, 2017 · Neurology
Device Facts
Record ID
K171803
Device Name
HIVOX OTC Electrical Stimulator
Applicant
Hivox Biotek, Inc.
Product Code
NUH · Neurology
Decision Date
Nov 29, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
HIVOX OTC Electrical Stimulator, SEM44 - TENS: The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), lower extremities (leg), abdomen and bottom due to strain from exercise or normal household work activities. EMS: The device is designed to be used for stimulate healthy muscles in order to improve and facilitate muscle performance. HIVOX OTC Electrical Stimulator, SEM44-1 - TENS: The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), lower extremities (leg), abdomen and bottom due to strain from exercise or normal household work activities.
Device Story
HIVOX OTC Electrical Stimulator (SEM44, SEM44-1) is a transcutaneous electrical nerve stimulator (TENS) and powered muscle stimulator (EMS). Device uses adhesive electrodes applied to skin to deliver electrical impulses, mimicking body signals to stimulate nerve and muscle fibers. SEM44 provides both TENS and EMS; SEM44-1 provides TENS only. Features two independent channels; users select pre-set or custom programs. Operated by patients in home/OTC setting. Output provides tingling/vibrating sensation to influence nerve/muscle transmission. Benefits include temporary pain relief and improved muscle performance. Healthcare providers do not operate; output is used by patient to manage pain or muscle conditioning.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including shelf life, biocompatibility, software validation, electromagnetic compatibility, electrical safety, and functional performance testing.
Technological Characteristics
ABS housing; biphasic square waveform; 1-150Hz frequency; 4.5V battery power; two independent channels; software-controlled. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, and ISO 10993-5/10. Sterilization not specified.
Indications for Use
Indicated for temporary relief of pain associated with sore/aching muscles in shoulder, waist, back, neck, upper/lower extremities, abdomen, and bottom due to exercise or household work strain. EMS mode indicated for stimulation of healthy muscles to improve/facilitate muscle performance.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size, also in blue. The logo is simple and professional, and it is easily recognizable.
November 29, 2017
Hivox Biotek Inc. Linda Lin Territory Sales 5F., No. 123, Xingde Rd. Sanchong Dist. New Taipei City, 24158 Tw
Re: K171803
Trade/Device Name: HIVOX OTC Electrical Stimulators Model SEM44 and Model SEM44-1 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, NGX Dated: August 21, 2017 Received: August 31, 2017
Dear Linda Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
William J. Heetderks -S 2017.11.29 20:14:45 -05'00'
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### Device Name
HIVOX OTC Electrical Stimulator (SEM44, SEM44-1)
Indications for Use (Describe)
HIVOX OTC Electrical Stimulator, SEM44 -
TENS: The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), lower extremities (leg), abdomen and bottom due to strain from exercise or normal household work activities.
EMS: The device is designed to be used for stimulate healthy muscles in order to improve and facilitate muscle performance.
HIVOX OTC Electrical Stimulator, SEM44-1 -
TENS: The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), lower extremities (leg), abdomen and bottom due to strain from exercise or normal household work activities.
| Type of Use (Select <i>one or both</i> , as applicable) |
|---------------------------------------------------------|
|---------------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Traditional 510(k), K171803/AIR Section 5 - 510(k) Summary
# 510(k) SUMMARY
| 5.1 | Type of Submission: | Traditional |
|-----|---------------------|------------------------------------------------------------------------------------|
| 5.2 | Date of Summary: | November 29, 2017 |
| 5.3 | Submitter: | HIVOX BIOTEK INC. |
| | Address: | 5F., No. 123, Xingde Rd., Sanchong Dist., New<br>Taipei City 24158, Taiwan, R.O.C. |
| | Phone: | +886-2-8511-2668 |
| | Fax: | +886-2-8511-2669 |
| | Contact: | Mr. Linda Lin<br>(linda.lin@hivox-biotek.com) |
#### 5.4 Identification of the Device:
| Proprietary/Trade name: | HIVOX OTC Electrical Stimulator |
|------------------------------|--------------------------------------------------------------------------------------------------|
| Model Number: | SEM44, SEM44-1 |
| Classification Product Code: | NUH |
| Subsequent Product Code: | NGX |
| Regulation Number: | 1) 882.5890<br>2) 890.5850 |
| Regulation Description: | 1) Transcutaneous electrical nerve<br>stimulator for pain relief<br>2) Powered muscle stimulator |
| Review Panel: | 1) Neurology<br>2) Physical Medicine |
| Device Class: | II |
#### 5.5 Identification of the Predicate Device:
| Predicate Device Name: | OTC Electrical stimulator |
|------------------------|-----------------------------------------|
| Model Number: | MT9001, LT3060 |
| Manufacturer: | Shenzhen Dongdixin Technology Co., Ltd. |
{4}------------------------------------------------
| Classification Product Code: | NUH |
|------------------------------|-----------------------|
| Subsequent Product Code: | NGX |
| Regulation number: | 882.5890 |
| Device Class: | II |
| 510(k) Number: | K130802 |
| Predicate Device Name: | Tyece OTC TENS Device |
| Model Number: | SEM44 |
| Manufacturer: | SAVIA Ltd. |
| Classification Product Code: | NUH |
| Regulation number: | 882.5890 |
| Device Class: | II |
| 510(k) Number: | K150386 |
# 5.6 Intended Use/ Indications for Use of the Device
HIVOX OTC Electrical Stimulator, SEM44 -
TENS: The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), lower extremities (leg), abdomen and bottom due to strain from exercise or normal household work activities.
EMS: The device is designed to be used for stimulate healthy muscles in order to improve and facilitate muscle performance.
# HIVOX OTC Electrical Stimulator, SEM44-1 -
TENS: The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), lower extremities (leg), abdomen and bottom due to strain from exercise or normal household work activities.
{5}------------------------------------------------
# 5.7 Device Description
HIVOX OTC Electrical Stimulator - SEM44 and SEM44-1, falls into the electro stimulation device category.
SEM44 provides two basic functions, TENS/EMS; SEM44-1 provides one basic function, TENS:
1. Electrical stimulation of nerve tracts (TENS)
2. Electrical stimulation of muscle tissue (EMS)
The two models also feature two independent stimulation channels and four adhesive electrodes which FDA cleared k number is K132588. For TENS, SEM44 and SEM44-1 are designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), lower extremities (leg), abdomen and bottom due to strain from exercise or normal household work activities .; For EMS, SEM44 is designed to be used for stimulating healthy muscles in order to improve and facilitate muscle performance. For these purposes, the user can either choose from pre-set programs or specify their own to suit the user's individual needs.
The principle of electro stimulation units is based on the imitation of impulses in our bodies that are transferred to nerve and muscle fibers with electrodes via our skin. The electrodes can be applied to many parts of the body. In certain applications the user will merely notice a slight tingling or vibrating sensation. For SEM44, the electrical impulses that are sent into the tissue influence the transmission of stimulation into nerves, nerve centers and muscle groups in the application area; for SEM44-1, the electrical impulses that are sent into the tissue influence the transmission of stimulation into nerves and nerve centers in the application area.
{6}------------------------------------------------
# 5.8 Non-clinical Testing
A series of safety and performance tests were conducted on the subject device, HIVOX OTC Electrical Stimulator (SEM44, SEM44-1).
- Shelf Life
- . Biocompatibility
- Software Validation
- Electromagnetic compatibility and electrical safety
- Function test
All the test results demonstrate HIVOX OTC Electrical Stimulator (SEM44, SEM44-1) meets the requirements of its pre-defined acceptance criteria and intended uses, and is substantially equivalent to the predicate devices.
# 5.9 Clinical Testing
No clinical test data was used to support the decision of substantial equivalence.
# 5.10 Substantial Equivalence Determination
The HIVOX OTC Electrical Stimulator (SEM44, SEM44-1) submitted in this 510(k) file is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the cleared OTC Electrical stimulator Models MT9001, LT3060 (K130802) and Tyece OTC TENS Device, Model SEM44 (K150386). Differences between the devices cited in this section do not raise any new issue of substantial equivalence.
{7}------------------------------------------------
| Item | Subject device | Predicate device I | Substantial equivalence determination |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name | HIVOX OTC Electrical Stimulator | OTC Electrical stimulator | N/A |
| Model | SEM44<br>SEM44-1 | MT9001<br>LT3060 | |
| 510(k) No. | (to be assigned) | K130802 | |
| Intended Use | SEM44 –<br>TENS: The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), lower extremities (leg), abdomen and bottom due to strain from exercise or normal household work activities.<br>EMS: The device is designed to be used for stimulate healthy muscles in order to improve and facilitate muscle performance. | MT9001 -<br>TENS: The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.<br>EMS: The device is designed to be used for stimulate healthy muscles in order to improve and facilitate muscle performance.<br>LT3060 -<br>TENS: The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder waist back neck | Similar.<br>Both devices utilize TENS to release pain associated with sore and aching muscles, and EMS to improve and facilitate muscle performance. |
{8}------------------------------------------------
| | | shoulder, waist, back, neck, upper extremities (arm), upper extremities (arm), lower extremities (leg) abdomen and bottom due to strain from exercise or normal household work activities. | and lower extremities (leg) due to strain from exercise or activities. normal household work activities. | | |
|----------------------------------------|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|--|
| | Type of use | OTC | OTC | Same | |
| | | Basic Unit Characteristics | | | |
| Power Source | | 4.5V (batteries, 3x1.5V AAA) | 9V batteries | Different but does not adversely impact safety and effectiveness of subject device | |
| Method of Line Current Isolation | | N/A | N/A | Same | |
| Patient Leakage Current | Normal condition | N/A | 0.61μA | This parameter is not applicable to subject device. | |
| | Single fault condition | N/A | 0.68μA | | |
| Number of Output Modes | | TENS: 15<br>EMS: 35 | TENS: 1<br>EMS: 1 | Different but does not adversely impact safety and effectiveness of subject device | |
| Number of Output Channels | Synchronous or Alternating | 2 Synchronous | Alternating | Different but does not adversely impact safety and effectiveness of | |
| | | Method of<br>Channel<br>Isolation | | subject device | |
| | | By electrical circuit and<br>software | By electrical circuit and<br>software | Same | |
| | Regulated Current or<br>Regulated Voltage? | Regulated voltage | Current control | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device | |
| | Software/Firmware/<br>Microprocessor Control? | Yes | Yes | Same | |
| | Automatic Overload Trip? | Yes | Yes | Same | |
| | Automatic No-Load Trip? | Yes | Yes | Same | |
| Automatic Shut Off? | | Yes | Yes | Same | |
| User Override Control? | | Yes | Yes | Same | |
| Indicator<br>Display | On/Off<br>Status | Yes | Yes | Same | |
| | Low Battery | Yes | Yes | Same | |
| | Voltage/<br>Current<br>Level | Yes | Yes | Same | |
| Timer Range (minutes) | | 5-100 minutes | 1-60 minutes | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device | |
| Compliance with Voluntary<br>Standards | | IEC 60601-1,<br>IEC 60601-1-2. | IEC 60601-1,<br>IEC 60601-1-2. | Same | |
{9}------------------------------------------------
{10}------------------------------------------------
| | IEC 60601-2-10,<br>ISO10993-5/10 | IEC 60601-2-10,<br>ISO10993-5/10 | | | |
|-------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|----------------------------------|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Compliance with 21 CFR<br>898-7 | Yes | Yes | Same | | |
| Weight | 89 g (including belt clip,<br>without batteries),<br>123 g (including belt clip<br>and batteries) | 128g (including batteries) | Slightly different<br>but does not<br>impact safety and<br>effectiveness of<br>subject device | | |
| Dimensions [L x W x T] | 132 x 63 x 29.5 mm<br>(including belt clip) | 117 x 60 x 34 mm | Slightly different<br>but does not<br>impact safety and<br>effectiveness of<br>subject device | | |
| Housing Materials and<br>Construction | ABS | ABS | Same | | |
| Output Specifications | | | | | |
| Waveform (e.g., pulsed<br>monophasic, biphasic) | Biphasic | Biphasic | Same | | |
| Shape (e.g., rectangular,<br>spike, rectified sinusoidal) | Square | Square | Same | | |
| Maximum<br>Output Voltage<br>(Volts, Vpp) | @500Ω | 100±10% | 96±20%<br>(48±20% (Vp)) | Different but | |
| | @2KΩ | 180±10% | 200±20%<br>(100±20%<br>(Vp))<br>228±20%<br>(114±20%<br>(Vp)) | does not<br>adversely impact<br>safety and | |
| | @10KΩ | 250±10% | 210±20%<br>(105±20%<br>(Vp))<br>230±20%<br>(115±20%<br>(Vp)) | effectiveness of<br>subject device | |
| Maximum | @500Ω | 200±10% | 96±20% | Different but | |
| | | | | | |
| Output<br>Current (mA) | @2KΩ | 90±10% | 50±20% | 57±20% | does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device |
| | @10KΩ | 25±10% | 10.5±20% | 11.5±20% | |
| Duration of primary phase | | 50-450μS | | 50-300μS | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device |
| Pulse Duration | | 50-450μS | | 50-300μS | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device |
| Frequency | | 1-150Hz | 1-150Hz | | Same |
| For interferential modes<br>only – Beat Frequency | | N/A | N/A | | Same |
| For<br>multi-program<br>waveforms<br>only – | Symmetrical<br>Phases? | N/A | Yes | | This parameter is<br>not applicable to<br>subject device. |
| | Phase<br>Duration | N/A | 50-300μS | | |
| Net charge (micro coulombs<br>µC) (per pulse) | | 0.001μC@500Ω | 0μC@500Ω | | Same |
| method of achieving zero<br>net charge | | N/A | biphasic and leading<br>polarity alternates for each<br>successive, Pulse + and<br>pulse – pulse channel | | This parameter is<br>not applicable to<br>subject device. |
| Max. phase charge (mC) | | 0.045@500Ω | 0.0288@500Ω | | Different but<br>does not |
| | | | | | adversely impact<br>safety and<br>effectiveness of<br>subject device |
| | | | | | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device |
| Max. current Density<br>(mA/cm², r.m.s) | | 0.667@500Ω | | 1.15@500Ω | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device |
| Max. Average current<br>(average absolute value),<br>mA | | 13.5@500Ω | | 4.32@500Ω | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device |
| Max. Average Power<br>Density, W/cm² (using<br>smallest electrode<br>conductive surface area) | | 0.0046@500Ω | | 0.373@500Ω | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device |
| Burst<br>Mode<br>(i.e.<br>pulse<br>trains) | (a) Pulses per burst | 3 | 7 | N/A | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device |
| | (b) Bursts per<br>second | 2/60Hz | 0.5/1/2/3/4/5<br>Hz | N/A | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device |
| | (c) Burst duration | 36ms | 70ms | N/A | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device |
| | (d) Duty cycle | 36ms/390ms | 35ms/350ms | N/A | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device |
| ON Time (seconds) | | 2 | N/A | 1-30 | Similar and does<br>not impact safety<br>and effectiveness<br>of subject device |
| OFF Time (seconds) | 2 | N/A | 1-60 | Similar and does<br>not impact safety<br>and effectiveness<br>of subject device | |
| Electrode area | 20.25sqcm x4 (81sqcm) | 25sqcm x2 (100sqcm) | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device | | |
| Average DC current through<br>electrodes when device is on<br>but no pulses are being<br>applied (µA) | 0 | 0 | Same | | |
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
| Item | Subject device | Predicate device II | Substantial<br>equivalence<br>determination |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Proprietary Name | HIVOX OTC Electrical<br>Stimulator | Tyece OTC TENS Device | N/A |
| Model | SEM44-1 | SEM44 | N/A |
| 510(k) No. | (to be assigned) | K150386 | N/A |
| Intended Use | TENS: The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), lower extremities (leg), abdomen and bottom due to strain from exercise or | The Tyece OTC TENS Device, Model SEM44 is to be used for temporary relief of pain associated with sore and aching muscles in the lower back, arms, or legs due to strain from exercise or normal household activities. | Similar.<br>Both devices utilize TENS to release pain associated with sore and aching muscles. |
{14}------------------------------------------------
| | | household<br>normal<br>work | | | | | |
|-----------------------------------------------|-------------------------------------|---------------------------------------|-----------------------------------|---------------------------------------------------------------------------------------------------|--|--|--|
| | | activities. | | | | | |
| Type of use | | OTC | OTC | Same | | | |
| | Basic Unit Characteristics | | | | | | |
| Power Source | | 4.5V (batteries, 3x1.5V | 4.5V (batteries, 3x1.5V | Same | | | |
| | | AAA) | AAA) | | | | |
| | Method of Line Current<br>Isolation | N/A | N/A | Same | | | |
| Patient | Normal condition | N/A | N/A | Same | | | |
| Leakage<br>Current | Single fault<br>condition | N/A | N/A | | | | |
| Number of Output Modes | | TENS: 15 | 15 Modes (01-15) | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device | | | |
| | Synchronous<br>Or<br>Alternating | 2 Synchronous | 2 Synchronous | Same | | | |
| Number of<br>Output<br>Channels | Method of<br>Channel<br>Isolation | By electrical circuit and<br>software | PCB Insulation<br>Boost Isolation | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device | | | |
| Regulated Current or<br>Regulated Voltage? | | Regulated voltage | Regulated voltage | Same | | | |
| Software/Firmware/<br>Microprocessor Control? | | Yes | Yes | Same | | | |
| Automatic Overload Trip? | | Yes | No | Different but | | | |
{15}------------------------------------------------
| | | | | does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device |
|-----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Automatic No-Load Trip? | | Yes | Yes | Same |
| Automatic Shut Off? | | Yes | Yes | Same |
| User Override Control? | | Yes | Yes | Same |
| | On/Off<br>Status | Yes | Yes | Same |
| | Low Battery | Yes | Yes | Same |
| Indicator<br>Display | Voltage/<br>Current<br>Level | Yes | Yes for voltage | Similar and does<br>not impact safety<br>and effectiveness<br>of subject device |
| Timer Range (minutes) | | 5-100 minutes | 5-100 minutes | Same |
| Compliance with Voluntary<br>Standards | | IEC 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-2-10,<br>ISO10993-5/10 | IEC 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-2-10,<br>ISO10993-5/10 | Same |
| Compliance with 21 CFR<br>898-7 | | Yes | Yes | Same |
| Weight | | 89 g (including belt clip,<br>without batteries),<br>123 g (including belt clip<br>and batteries) | 101 g (including belt chip) | Slightly different<br>but does not<br>impact safety and<br>effectiveness of<br>subject device |
| Dimensions [W x H x D] | | 132 x 63 x 29.5 mm<br>(including belt clip) | 132 x 63 x 29.5 mm<br>(including belt clip) | Same |
| Housing Materials and<br>Construction | | ABS | ABS | Same |
| Output Specifications | | | | |
| Waveform (e.g., pulsed<br>monophasic, biphasic) | | Biphasic | Biphasic | Same |
| Shape (e.g., rectangular,<br>spike, rectified sinusoidal) | | Square | Square | Same |
| Maximum<br>Output Voltage<br>(Volts, Vpp) | @500Ω | 100±10% | 70±15% | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device |
| | @2KΩ | 180±10% | 110±15% | |
| | @10KΩ | 250±10% | 190±15% | |
| Maximum<br>Output Current<br>(mA) | @500Ω | 200±10% | 86±15% | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device |
| | @2KΩ | 90±10% | 23.3±15% | |
| | @10KΩ | 25±10% | 3.75±15% | |
| Duration of primary phase | | 50-450μS | 0 | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device |
| Pulse Duration | | 50-450μS | 50-360μS | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device |
| Frequency | | 1-150Hz | 1-150Hz | Same |
| For<br>multi-program | Symmetrical<br>Phases? | N/A | N/A | Same |
| waveforms<br>only | Phase<br>Duration | | | |
| Net charge (micro coulombs<br>μC) (per pulse) | $0.001μC@500Ω$ | $0.001μC @500Ω$ | Same | |
| Max. phase charge (mC) | $0.045@500Ω$ | $0.0454@500Ω$ | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device | |
| Max. current Density<br>(mA/cm², r.m.s) | $0.667@500Ω$ | $0.790@500Ω$ | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device | |
| Max. Average current<br>(average absolute value),<br>mA | $13.5@500Ω$ | $16.0@500Ω$ | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device | |
| Max. Average Power<br>Density, W/cm² (using<br>smallest electrode<br>conductive surface area) | $0.0046@500Ω$ | $0.00632@500Ω$ | Different but<br>does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device | |
| Burst<br>Mode<br>(i.e.<br>pulse | (a) Pulses per burst | 3 | 4 | Different but<br>does not<br>adversely impact<br>safety and |
| | (b) Bursts per<br>second | 2/60Hz | 4/83Hz | adversely impact<br>safety and |
| | (c) Burst duration | 36ms | 0.18ms | safety and |
| trains) | (d) Duty cycle | 36ms/390ms | 35ms/60ms | effectiveness of<br>subject device |
| | ON Time (seconds) | 2 | 2 | Same |
| | OFF Time (seconds) | 2 | 2 | Same |
|…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.