K171798 · Media Plus, LLC · GXY · Jun 1, 2018 · Neurology
Device Facts
Record ID
K171798
Device Name
Silverwear Silver Pro Garment Device
Applicant
Media Plus, LLC
Product Code
GXY · Neurology
Decision Date
Jun 1, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Silverwear SilverPro Series Conductive Garments are cutaneous electrodes to be used with legally marketed TENS or NMES devices. The knitted garment electrodes are non-sterile reusable OTC conductive garments that are intended to deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact. These body parts can include hand (glove), wrist (sleeve), elbow or arm (sleeve), knee high stockings, ankle (sleeve), back band, and shoulder band.
Device Story
Silverwear SilverPro garments are conductive, knitted textile electrodes; input consists of electrical stimulation signals from legally marketed TENS or NMES devices. Garments are worn on body parts (hands, feet, arms, legs, back, knees, shoulders) to deliver current uniformly to the skin surface. Operated by patients in home or clinical settings; device connects to stimulator via standard adhesive electrode interface. Provides pain relief through electrostimulation; benefits include uniform current distribution and reusable, washable design. No active processing or software involved.
Clinical Evidence
Bench testing only. Testing confirmed fabric conductivity and resistivity (<7 ohms/inch) maintained through multiple washings. Biocompatibility testing (ISO 10993-5, ISO 10993-10) for cytotoxicity, sensitization, and irritation confirmed safety for limited duration (<24 hours) contact with intact skin.
Technological Characteristics
Knitted garment electrodes composed of silver-plated nylon yarn, polyester sheath yarn with carbon filling. Provides <5 ohms resistance per inch. Passive conductive interface; no internal energy source. Multiple configurations (gloves, sleeves, socks, bands). Washable, reusable, non-sterile. Biocompatibility per ISO 10993-5 and ISO 10993-10.
Indications for Use
Indicated for use with TENS or NMES devices to deliver low-level stimulation current for pain relief to intact skin on hands, feet, arms, legs, low back, knees, and shoulders. Intended for OTC use by patients requiring electrostimulation.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services logo. The logo on the right is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
June 1, 2018
Media Plus, LLC % Sharon Blinkoff Counsel Locke Lord LLP 200 Vesey Street New York, New York 10281
Re: K171798
Trade/Device Name: Silverwear Silver Pro Garment Device Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: March 1, 2018 Received: March 5, 2018
Dear Sharon Blinkoff:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Vivek J. Pinto -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171798
Device Name Silverwear Silver Pro Garment Device
#### Indications for Use (Describe)
The Silverwear SilverPro Series Conductive Garments are cutaneous electrodes to be used with legally marketed TENS or NMES devices. The knitted garment electrodes are non-sterile reusable OTC conductive garments that are intended to deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact. These body parts can include hand (glove), wrist (sleeve), elbow or arm (sleeve), knee high stockings, ankle (sleeve), back band, and shoulder band.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510 (k) Summary
Media Plus LLC Applicant: 560 Sylvan Avenue, 3rd floor Englewood Cliffs, NJ 07632
Contact: Rita Vaccaro
Phone number: 201-541-1222 ext. 300
Distributor: SilverWear USA, LLC 560 Sylvan Avenue, 3rd floor Englewood Cliffs, NJ 07632 (844-884-4499)
- Contact: Rita Vaccaro 201-541-1222 ext. 300
Contact
For 510k
- Submission : Sharon Blinkoff Locke Lord LLP 200 Vesey Street New York NY 10281 Sharon.blinkoff@lockelord.com (232-912-2983)
Date of Submission:
Proprietary Name: Silverwear SilverPro Garment Electrodes Common Name: Cutaneous Electrodes 21 CFR 882.1320 Product Code: GXY Cutaneous Electrodes Panel: Neurological Devices
Trade Name: Silverwear SilverPro model numbers: SP 1001, SP 1002, SP 1002, SP 1003, SP 1004, SP 1005, SP 1006, SP 1007, SP 1008, SP 1009
{4}------------------------------------------------
#### A. Legally Marketed Predicate devices
- 1. Theraknit Garment Electrodes k053214
- 2. Cutaneous Electrodes k141076 secondary predicate
### B. Device Description
The Silverwear SilverPro garment electrodes are conductive garments that are made from material which is knitted from primarily nylon yarn with minor amounts of lycra, spandex to achieve a stretch fabric for a snug garment fit polyester for tactile qualities and polyester sheathed carbon yarn and pure silver coated nylon fibers to provide conductivity. The knitting follows standard knitting procedures with the conductive knitted material being fabricated into wearable electrode garments which are provided in multiple sizes of various garment configurations including; gloves, bands, socks, and sleeves. The stretch characteristics of the material provide sufficient elasticity to ensure firm surface contact with the skin. The design of the devices is such that they can be used skin by reversing the surface contacting the skin. The entire surface of the garment electrode is very conductive having a resistance of less than 5 ohms per inch. This provides low current density with uniform current distribution to enable efficient use of these garment electrodes for use in TENS (transcutaneous electrical nerve stimulators) and NMES (powered muscle stimulator devices). The Silverwear SilverPro garment electrodes are non- sterile external devices which are designed for single patient, for multiple uses and are intended for OTC use with FDA Cleared TENS and NMES class II devices.
- C. Intended Use
The Silverwear SilverPro garment cutaneous electrodes are provided in various sizes of several forms of garments such as gloves, socks, sleeves knee pads, pads and bands are intended for use with legally marketed electrostimulation TENS and NMES devices, to deliver uniformly to the skin low level stimulation current for pain relief. The garment electrodes can be used on external body parts of the body with external application to hands, feet, arms legs, low back, knee and shoulder for pain relief using electrostimulation. The devices are intended for OTC use, for external application to intact skin of low dose current density with uniform distribution which follows the conductive materials knitted into the fabric.
| Elements of<br>comparison | Subject Device | Primary Predicate | Secondary Predicate | |
|---------------------------|------------------------------|-------------------------|----------------------------------------|----|
| Company | Silverwear SilverPro Garment | Neurotron Medical, Inc. | Shenzhen Konmed<br>Technology Co., Ltd | -- |
### D. Comparison to Predicate Device
{5}------------------------------------------------
| Device Name | Silverwear SilverPro wearable therapy garments | TheralKnit garment | Electrodes with silver conductive Model:<br>KM-406: Glove Style<br>KM-407: Socks Style<br>KM-408: Wristbands Style<br>KM-409: Elbow pads Style<br>KM-410 : Knee Pads Style | |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Regulation Number | 882.1320 | 882.1320 | 882.1320 | same |
| Product k number | Applying | K053214 | K171721 | -- |
| code | GXY | GXY | GXY | same |
| OTC / Rx | OTC | Rx | OTC | similar |
| Intended Use /<br>Indications for Use | The Silverwear SilverPro Series<br>Conductive Garments are<br>cutaneous electrodes to be used<br>with legally marketed TENS or<br>NMES devices. The knitted<br>garment electrodes are non-<br>sterile reusable OTC conductive<br>garments that are intended to<br>deliver the stimulation signals<br>generated by the stimulator to<br>the body surface with which<br>they are in contact. These body<br>parts can include hand (glove),<br>wrist (sleeve), elbow or arm<br>(sleeve), knee or leg (sleeve),<br>knee high stockings, ankle<br>(sleeve), back band, and<br>shoulder band. | The TheralKnit Garment<br>electrodes are cutaneous<br>to be used with legally<br>marketed TENS<br>stimulating device. The<br>knitted garment electrodes<br>will deliver the<br>stimulation signals<br>generated by the<br>stimulator to the body<br>surface with which they<br>are in contact. These body<br>parts can include hand<br>(glove), feet (socks),<br>elbow or knee (sleeve),<br>arm, leg, shoulder, back<br>(pads). | Electrodes with silver<br>conductive as Glove style,<br>Socks style, Wristbands<br>Style, Wrist sleeve, Elbow<br>Pads Style and Knee Pads<br>Style, Elbow Sleeve, are<br>intended for use with<br>legally marketed TENS<br>stimulating device. The<br>electrodes with silver<br>conductive will deliver<br>stimulation signals<br>generated by the stimulator<br>to the body surface with<br>which they are in contact.<br>These body parts can<br>include such as hands<br>(gloves), feet (socks),<br>wrist, elbow and knee. | same |
| Design (Shape) | Electrode A: Glove Style<br>Electrode B: Wrist Sleeve<br>Electrode C: Elbow/Arm Sleeve<br>Electrode D: Knee/Leg Sleeve<br>Electrode E: Knee High Socks<br>Electrode F: Ankle Sleeve<br>Electrode G: Back Band<br>Electrode H: Shoulder Band | Electrode A: Glove Style<br>Electrode B: Socks Style<br>Electrode C: Sleeve Style<br>Electrode D: Pads Style | KM-406: Glove Style<br>KM-407: Socks Style<br>KM-408: Wristbands Style<br>KM-409: Elbow pads Style<br>KM-410: knee Pads Style | similar |
| Size | All Sizes | All Sizes | One size<br>Below is size for when<br>without stretched:<br>KM-406: 200 (cm2)<br>KM-407: 285 (cm2)<br>KM-408: 95 (cm2)<br>KM-409: 160 (cm2)<br>KM-410: 236 (cm2) | similar |
| Impedance | 7 ohms resistance per inch | 7 ohms resistance per inch | 2 ohms resistance per inch | similar |
{6}------------------------------------------------
| Washable or not | Washable | Washable | Washable | same |
|--------------------------------|-----------------------------------------|--------------------------------------------|---------------------------------------------|------|
| Re-usable | For Single Patient | For Single Patient | For Single Patient | same |
| Biocompatibility | Complied with ISO 10993-5, ISO 10993-10 | Complied with ISO<br>10993-5, ISO 10993-10 | Complied with ISO 10993-<br>5, ISO 10993-10 | same |
| Patient Contacting<br>Material | Silver plated nylon | Silver plated nylon | Silver plated nylon | same |
#### E. Substantial Equivalence Summary
The Silverwear SilverPro device garment cutaneous electrodes are medical devices that have the same design features, construction, intended use, conductivity, electrical connection and target population as the legally marketed predicate devices. The Silverwear SilverPro devices have the same technological characteristics as the predicate devices. Both the Silverwear SilverPro and the predicate devices receive electrostimulation signals from legally marketed TENS or NMES devices through a standard electrical connection of an adhesive electrode which adhesively connects to the Silverwear SilverPro device and is wired to the TENS or NMES device. Both the Silverwear SilverPro and the predicate devices are washable and intended for multiple uses by a single patient with intact skin.
- F. Technological Characteristics
The Silverwear SilverPro garment electrodes are made from silver plated nylon yarn and a polyester sheath yarn with carbon filling, using a conventional knitting process. The fabric is highly conductive and provides less than 5 ohms resistance per inch. The actual devices which are fabricated into multiple different garment forms with each form being designed to accept an adhesively applied electrode component from the NMES or TENS device which is the source of the current that is delivered by the garment electrodes to the target skin tissue. The Silverwear SilverPro devices are non- sterile multiple use devices which are washable using conventional detergents. Bench tests of the fabric show that the garment electrodes do not change their inherent conductivity with multiple washings so that there is no significant adverse effect on the conductivity of the device or its inherent ability to deliver low dose current density uniformly to the skin of the wearer even after multiple washings. The Silverwear SilverPro garment electrode provide a snug elastic fit against the skin based upon the low mild elastic properties of the yarn used to fabricate the devices, these elastic proper a snug fit for delivery of the current generated by the TENS or NMES devices.
- G. Software
There is no software used in the Silverwear SilverPro devices
### H. Testing
Bench testing has been conducted on the fabric used to fabricate the various forms of the Silverwear SilverPro devices to demonstrate that the fabric and the devices meet design controls in terms of conductivity and resistivity, and uniform delivery of low doses of current all
{7}------------------------------------------------
consistent with that of the predicate devices. Resistivity testing was conducted using standard industry testing to confirm that the resistivity met the standards of less than 7 ohms per inch through multiple washings to confirm the multiple use aspects of the device.
Biocompatibility tests were conducted on the SilverWear SilverPro devices, which are intended for external use only, on intact skin of limited body areas for less than 24 hour. Biocompatibility was conducted using tests methods to determine Cytotoxicity, Sensitization and Irritation of Cutaneous Reactivity.
#### l. The subject device:
- a. is OTC only
- b. Intended for limited external exposure only not for internal use
- ls for application to intact skin and not for application on mucosal areas ﻥ
- d. Is for application to limited areas of skin and for limited duration of use under 24 hours and as prescribed in the TENS and NMES devices for which use of these wearable electrodes are intended. These devices are not intended for use in mucosal areas
#### J. Conclusions
The subject Silverwear SilverPro devices are comparable in design specifications to the predicate garment devices and like the predicate garment devices come in multiple versions all of which have the same intended use, are intended to be used on the same patient population, with the same technical electrical performance to deliver low doses of current to provide electrical stimulation to the skin, which like the predicate devices are intended to be used with the same FDA cleared OTC TENS or NMES predicate devices as the predicate devices. Testing has demonstrated that this SilverPro device is in terms of conductivity and resistivity relies on the same principles of Technology as the comparable predicate devices and in terms of design, material of fabrication, method of application, and intended use and performance is comparable to its predicate devices in delivering the uniform delivery of low dose current generated by FDA cleared OTC TENS or NMES devices. The proposed device presents no new technological, safety or effectiveness differences from those of the predicate class II garment electrode devices. Thus, the Silverwear SilverPro garment electrodes are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.