Sequential Circulator SC-2004FC-OC

K171793 · Bio Compression Systems, Inc. · JOW · Aug 17, 2017 · Cardiovascular

Device Facts

Record IDK171793
Device NameSequential Circulator SC-2004FC-OC
ApplicantBio Compression Systems, Inc.
Product CodeJOW · Cardiovascular
Decision DateAug 17, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.5800
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Bio Compression Systems Sequential Circulator SC-2004FC-OC with associated garments is a sequential, pneumatic compression device intended for the primary or adjunctive treatment of primary or secondary lymphedema. The device is also intended for the additional or alternate treatment of venous insufficiency and chronic venous stasis ulcers associated with venous insufficiency, as well as, general treatment for swelling of the extremities. The device is intended for home or hospital use.

Device Story

Pneumatic compression device (pump) with inflatable garment sleeves; applied to affected extremities. User manually adjusts pressure via front-panel knob; pump inflates segmented garment chambers sequentially (distal to proximal) to provide graduated compression. Pressure held constant across chambers to prevent reverse flow; then simultaneously released. Cycle repeats. Used in home or hospital settings by patients or clinicians. Faster cycle time compared to predicate; maximum pressure remains 80 mmHg. Benefits include treatment of lymphedema, venous insufficiency, and extremity swelling.

Clinical Evidence

Bench testing only. Functional testing included electrical leakage, pressure adjustment, chamber inflation pressure, air pressure display accuracy, and inflation/deflation cycle times. Results demonstrate comparable inflation cycle profiles between subject and predicate devices.

Technological Characteristics

Pneumatic compression pump; graduated/gradient pressurization; maximum pressure 80 mmHg. Sequential inflation distally to proximally. Manually adjusted pressure control. Components include compressor and segmented inflatable garment sleeves. No software-based control described.

Indications for Use

Indicated for primary or secondary lymphedema, venous insufficiency, chronic venous stasis ulcers, and general extremity swelling. Intended for use in home or hospital settings.

Regulatory Classification

Identification

A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Public Health Service Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping to create a sense of depth. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 17, 2017 Bio Compression Systems, Inc. Barbara Whitman Director, Regulatory Affairs & Quality Assurance 120 W Commercial Ave Moonachie, New Jersey 07074 Re: K171793 Trade/Device Name: Sequential Circulator SC-2004FC-OC Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible limb sleeve Regulatory Class: Class II Product Code: JOW Dated: July 18, 2017 Received: July 19, 2017 Dear Barbara Whitman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device {1}------------------------------------------------ related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. # Nicole G. Ibrahim -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ | Form Approved: OMB No. 0910-0120 | |-----------------------------------| | Expiration Date: January 31, 2017 | | See PRA Statement below. | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration **Indications for Use** | 510(k) Number (if known) | K171793 | |--------------------------|------------------------------------| | Device Name | Sequential Circulator SC-2004FC-OC | Indications for Use (Describe) The Bio Compression Systems Sequential Circulator SC-2004FC-OC with associated garments is a sequential, pneumatic compression device intended for the primary or adjunctive treatment of primary or secondary lymphedema. The device is also intended for the additional or alternate treatment of venous insufficiency and chronic venous stasis ulcers associated with venous insufficiency, as well as, general treatment for swelling of the extremities. The device is intended for home or hospital use. | Type of Use (Select one or both, as applicable) | <div> <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> </div> <div> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> </div> | |-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff [PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov) "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | FORM FDA 3881 (8/14) | Page 1 of 1 | |----------------------|-------------| |----------------------|-------------| {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Bio Compression Systems, Inc. Below the logo is the text "Special 510(k) Device Modification". The last line of text says "Sequential Circulators SC-2004FC-OC". # 510(k) Summary of Substantial Equivalence | Contact: | Barbara J. Whitman<br>Director, Regulatory Affairs and Quality Assurance | | | |--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|----------------------------------| | | Bio Compression Systems, Inc.<br>120 W. Commercial Ave.<br>Moonachie, NJ 07074 USA<br>Phone: 201-340-6422<br>Fax: 201-939-4503<br>Email: barbaraw@biocompression.com<br>Web: www.biocompression.com | | | | Date Prepared: | August 15, 2017 | | | | Device Name/Classification: | Sequential Circulators, SC-2004FC-OC<br>Class II<br>Cardiovascular and Respiratory Devices<br>21 CFR Part 870.5800<br>JOW – Compressible Limb Sleeve | | | | Trade Name of<br>Proposed Device(s): | SC-2004FC-OC Sequential Circulator | | | | Predicate Devices: | 510(k) # | Device Name | Manufacturer | | | K150953 | SC-2004-OC and SC-2008-OC<br>Sequential Circulators | Bio Compression<br>Systems, Inc. | | Manufacturer and<br>Distributor: | Bio Compression Systems, Inc.<br>120 W. Commercial Ave., Moonachie, NJ 07074 USA | |----------------------------------|----------------------------------------------------------------------------------| |----------------------------------|----------------------------------------------------------------------------------| #### Proposed Device Description The Bio Compression Systems, Inc. SC-2004FC-OC Sequential Circulator is a pneumatic compression device ('pump') with associated compressible garment sleeves. The pump is manually adjusted by the user to produce air pressure to inflate and deflate segmented inflatable chambers of garment sleeves. The garment sleeves are externally applied over affected extremities. The pumps consist of compressors capable of producing a maximum pressure of 80 mmHg and provide graduated or gradient pressurization to the chambers of the garments. The sequential inflation is applied distally to proximally with distal chambers inflated to a greater pressure than proximal ones. {4}------------------------------------------------ As each chamber is inflated, the pressure is held constant until all chambers are inflated, in order to prevent reverse gradient flow. Once all of the chambers are inflated, they are then simultaneously released and the cycle repeats. The user can adjust the Pressure Knob on the front of the device for desired pressure. Associated garments are available in a variety of standard sizes. ### Indications For Use The Indications For Use of the modified subject device are identical to those of the predicate devices. Below are the Indications For Use for the subject device, as well as, their intended use. The Bio Compression Systems Sequential Circulator SC-2004FC-OC with associated garments is a sequential, pneumatic compression device intended for the primary or adjunctive treatment of primary or secondary lymphedema. The device is also intended for the additional or alternate treatment of venous insufficiency and chronic venous stasis ulcers associated with venous insufficiency, as well as, general treatment for swelling of the extremities. The device is intended for home or hospital use. ### Description of Device Modifications/Technological Characteristics: The Sequential Circulator SC-2004FC-OC is technologically identical to the predicate devices, SC-2004-OC and SC-2008-OC, Sequential Circulators (K150953). The modification to the predicate device involved replacing the motor in the SC-2004-OC with the exact motor used in the SC-2008-OC, hence deriving the subject device, SC-2004FC-OC. This change enables the subject device to operate in a faster cycle time, which some patients prefer. There are no maximum pressure changes, as they identically match those of the predicate devices. There is no additional risk introduced by this change. The model number on the front of the pump will change to identify the subject device as SC-2004FC-OC. The modification to the subject devices has not altered the fundamental technology of the predicate devices. There have been no prior submissions for the subject device. ## Performance Data Before being released to market, every device is tested and must meet all performance specifications. In addition to aesthetic acceptance criteria, functional testing includes electrical leakage, pressure adjustment, inflation pressure in each chamber, air pressure display accuracy, and inflation/deflation cycle times. The results demonstrate comparable inflation cycle profiles (rise times, inflation pressures, deflation times, and cycle times) between the applicant and predicate devices. #### Statement of Substantial Equivalence: Based upon risk management, safety & performance testing, compliance with voluntary standards, and comparison to predicate devices, the Sequential Circulator SC-2004FC-OC is substantially equivalent to the predicate devices Sequential Circulators SC-2004-OC and SC-2008-OC. The modified device does not introduce any new potential safety risks and is substantially equivalent to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...