K171757 · Straumann USA, LLC (On Behalf of Institut Straumann Ag) · NHA · Nov 27, 2017 · Dental
Device Facts
Record ID
K171757
Device Name
Straumann Screw Retained Abutments
Applicant
Straumann USA, LLC (On Behalf of Institut Straumann Ag)
Product Code
NHA · Dental
Decision Date
Nov 27, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
Straumann® Screw Retained Abutments are indicated to be placed into the implants of the Straumann® Dental Implant System to provide support for prosthetions such as crowns, bridges and bass. The final processed devices have the purpose of restoring chewing function. Straumann® Screw Retained Abutments are indicated for screw-retained restorations.
Device Story
Straumann® Screw Retained Abutments consist of one-piece straight or angled (17°/30°) abutments, basal screws, and carrier pins. Devices are placed into Straumann® Dental Implant System implants by dental professionals to support prosthetic reconstructions (crowns, bridges, bars) for restoring chewing function. Subject device is identical to predicate K133421 in design, material, and operating principle, but is supplied sterile via gamma irradiation, whereas predicates were non-sterile. Packaging ensures sterility maintenance. Clinical benefit is restoration of chewing function via screw-retained prosthetic support.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Material: Ti-6Al-7Nb titanium alloy. Design: One-piece straight and angled (17°, 30°) abutments with basal screws and carrier pins. Connection: Engaging Narrow CrossFit (NC) and Regular CrossFit (RC). Sterilization: Gamma irradiation (25 kGy minimum) validated per ISO 11137-1 and ISO 11137-2 to SAL 10^-6.
Indications for Use
Indicated for patients requiring prosthetic reconstruction (crowns, bridges, bars) supported by Straumann Dental Implant System implants to restore chewing function; indicated for screw-retained restorations.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Straumann® Magellan™ Screw Retained Abutment System (K133421)
Reference Devices
Straumann Bone Level NC Angled Screw Retained Abutments (K141871)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right is written in a clear, sans-serif font with the FDA acronym in a blue box.
Straumann USA, LLC (on behalf of Institut Straumann AG) % Jennifer M. Jackson, MS Director, Regulatory Affairs & Quality Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810
November 27, 2017
Re: K171757
Trade/Device Name: Straumann® Screw Retained Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: October 25, 2017 Received: October 26, 2017
Dear Jennifer M. Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Mary S. Runner -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K171757
Device Name Straumann® Screw Retained Abutments
Indications for Use (Describe)
Straumann® Screw Retained Abutments are indicated to be placed into the implants of the Straumann® Dental Implant System to provide support for prosthetions such as crowns, bridges and bass. The final processed devices have the purpose of restoring chewing function. Straumann® Screw Retained Abutments are indicated for screw-retained restorations.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div>×</div> |
|----------------------------------------------|--------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Straumann® Screw Retained Abutments
510(k) Summary
#### 5 510(k) Summary
#### 5.1 Submitter
Straumann USA, LLC (on behalf of Institut Straumann AG)
60 Minuteman Road
Andover, MA 01810
| Phone Number: | 1-978-747-2614 |
|---------------|----------------|
|---------------|----------------|
Fax Number: 1-978-747-0041
Contact Person: Chanrasmey White
Date of Submission: 10/25/2017
## 5.2 Device
| Trade Name: | Straumann® Screw Retained Abutments |
|----------------------|-------------------------------------|
| Common Name: | Endosseous Dental Implant Abutments |
| Classification Name: | Endosseous Dental Implant Abutments |
| Regulatory Class: | II (21 CFR §872.3630) |
| Product Code: | NHA (21 CFR §872.3630) |
## 5.3 Predicate Device
Primary Predicate:
- (K133421) Straumann® Magellan™ Screw Retained Abutment System ●
Reference Predicate:
- (K141871) Straumann Bone Level NC Angled Screw Retained Abutments ●
- (K151247) Straumann Screw Retained Abutments ●
- (K150814) Straumann Screw-Retained Abutment .
- (K130808) Straumann Healing Abutments, Healing Caps, Closure Screws ●
{4}------------------------------------------------
## Straumann® Screw Retained Abutments
510(k) Summary
#### Device Description 5.4
Straumann® Screw Retained Abutments include one-piece straight and angled (17° and 30°) abutments, basal screws and abutment carrier pin. The subject device is identical to the primary predicate, K133421, and subsequent premarket notification submissions K141871, K150814 and K151247 with the exception that the subject device will be delivered to the user sterile as opposed to non-sterile and a change in packaging to ensure sterility of the devices.
### 5.5 Indications for Use
Straumann® Screw Retained Abutments are indicated to be placed into the implants of the Straumann® Dental Implant System to provide support for prosthetic reconstructions such as crowns, bridges and bars. The final processed devices have the purpose of restoring chewing function. Straumann® Screw Retained Abutments are indicated for screw-retained restorations.
#### Technoloqical Characteristics 5.6
The proposed Straumann® Screw Retained Abutments intended use, material, fundamental operating principles and overall design are identical to the primary predicate device. The subject device is identical to the primary predicate. K133421, and subsequent premarket notification submissions K141871, K150814 and K151247 with the exception that the subject device will be delivered to the user sterile as opposed to non-sterile and a change in packaging to ensure sterility of the devices. Packaging materials and configuration is equivalent to reference predicate device, Straumann® Straumann Healing Abutments, Healing Caps, Closure Screws, K130808. The technological characteristics of the subject devices are compared to the non-sterile predicate devices (K133421, K141871, K150814 and K151247) in the following table.
{5}------------------------------------------------
# Straumann® Screw Retained Abutments
## 510(k) Summary
| Feature | Subject Device | Primary PREDICATE<br>Device | Reference DEVICE | Reference DEVICE | Reference DEVICE |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | | K133421 | K141871 | K150814 | K151247 |
| Indication | Straumann® Screw<br>Retained Abutments are<br>indicated to be placed into<br>the implants of the<br>Straumann® Dental Implant<br>System to provide support<br>for prosthetic<br>reconstructions such as<br>crowns, bridges and bars.<br>The final processed devices<br>have the purpose of<br>restoring chewing function.<br>Straumann® Screw<br>Retained Abutments are<br>indicated for screw-retained<br>restorations. | Straumann® Magellan ™<br>Screw Retained Abutments<br>are indicated to be placed<br>into the implants of the<br>Straumann® Dental Implant<br>System to provide support<br>for prosthetic<br>reconstructions such as<br>crowns, bridges and bars.<br>The final processed devices<br>have the purpose of<br>restoring chewing function.<br>Straumann® Screw<br>Retained Abutments are<br>indicated for screw-retained<br>restorations. | Straumann® Screw<br>Retained Abutments are<br>indicated to be placed into<br>the implants of the<br>Straumann® Dental Implant<br>System to provide support<br>for prosthetic<br>reconstructions such as<br>crowns, bridges and bars.<br>The final processed devices<br>have the purpose of<br>restoring chewing function.<br>Straumann® Screw<br>Retained Abutments are<br>indicated for screw-retained<br>restorations. | Straumann® Screw<br>Retained Abutments are<br>indicated to be placed into<br>the implants of the<br>Straumann® Dental Implant<br>System to provide support<br>for prosthetic<br>reconstructions such as<br>crowns, bridges and bars.<br>The final processed devices<br>have the purpose of<br>restoring chewing function.<br>Straumann® Screw<br>Retained Abutments are<br>indicated for screw-retained<br>restorations. | Straumann® Screw<br>Retained Abutments are<br>indicated to be placed into<br>the implants of the<br>Straumann® Dental Implant<br>System to provide support<br>for prosthetic<br>reconstructions such as<br>crowns, bridges and bars.<br>The final processed devices<br>have the purpose of<br>restoring chewing function.<br>Straumann® Screw<br>Retained Abutments are<br>indicated for screw-retained<br>restorations. |
| Implant-to-Abutment<br>Connection | Narrow CrossFit (NC)<br>Regular CrossFit (RC) | Narrow CrossFit (NC)<br>Regular CrossFit (RC) | Narrow CrossFit (NC) | Narrow CrossFit (NC) | Narrow CrossFit (NC)<br>Regular CrossFit (RC) |
| Interface Type | Engaging | Engaging | Engaging | Engaging | Engaging |
| Platform Diameter(s) | NC Straight and Angled: | NC Straight: | NC Angled:<br>Ø3.5 mm, Ø4.6 mm | NC Straight:<br>Ø3.5 mm, Ø4.6 mm | NC Angled:<br>Ø3.5 mm, Ø4.6 mm<br>RC Angled: |
| Feature | Subject Device | Primary PREDICATE Device | Reference DEVICE | Reference DEVICE | Reference DEVICE |
| K Number | | K133421 | K141871 | K150814 | K151247 |
| | Ø3.5 mm, Ø4.6 mm<br>RC Straight and Angled:<br>Ø4.6 mm | Ø3.5 mm, Ø4.6 mm<br>RC Straight and Angled:<br>Ø4.6 mm | | | Ø4.6 mm |
| Abutment Angulation(s) | 0°, 17°, 30° | 0°, 17°, 30° | 17°, 30° | 0° | 17°, 30° |
| Gingival Height(s) | NC and RC Straight:<br>1.0, 2.5 and 4.0 mm<br>NC and RC Angled:<br>2.5, 4.0 and 5.5 mm | NC and RC Straight:<br>1.0, 2.5 and 4.0 mm<br>RC Angled:<br>2.5 and 4.0 mm | NC Angled:<br>2.5 and 4.0 mm | NC Straight:<br>1.0 mm | NC and RC Angled:<br>5.5 mm |
| Orientation of Angulation<br>to Engagement Features | Type A (45°), Type B (0°) | Type A (45°), Type B (0°) | Type A (45°), Type B (0°) | N/A | Type A (45°), Type B (0°) |
| Device Material | Ti-6Al-7Nb titanium alloy | Ti-6Al-7Nb titanium alloy | Ti-6Al-7Nb titanium alloy | Ti-6Al-7Nb titanium alloy | Ti-6Al-7Nb titanium alloy |
| Sterilization | End user receives product<br>sterilized per Gamma<br>Irradiation, 25 kGy minimum<br>Validated per ISO 11137-1<br>and ISO<br>11137-2 to an SAL of 10 x<br>10-6 | End user to sterilize product<br>per IFU Method: Autoclave<br>moist heat fractionated<br>vacuum or gravity<br>displacement<br>Conditions: 134°C (273°F)<br>for 5 minutes | End user to sterilize product<br>per IFU Method: Autoclave<br>moist heat fractionated<br>vacuum or gravity<br>displacement<br>Conditions: 134°C (273°F)<br>for 5 minutes | End user to sterilize product<br>per IFU Method: Autoclave<br>moist heat fractionated<br>vacuum or gravity<br>displacement<br>Conditions: 134°C (273°F)<br>for 5 minutes | End user to sterilize product<br>per IFU Method: Autoclave<br>moist heat fractionated<br>vacuum or gravity<br>displacement<br>Conditions: 134°C (273°F)<br>for 5 minutes |
{6}------------------------------------------------
# Straumann® Screw Retained Abutments
510(k) Summary
Table 1: Comparison of Subject Devices to non-sterile Predicates
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.