FuckWater is a water-based personal lubricant for penile and/or rectal and/or rectal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.
Device Story
FuckWater is a non-sterile, water-based personal lubricant for over-the-counter home use. Applied topically to penis, vagina, or rectum to reduce friction during sexual activity. Supplements natural body lubrication. Compatible with natural rubber latex and polyisoprene condoms; incompatible with polyurethane condoms. Packaged in 2oz to 16oz bottles. Benefits include enhanced comfort and ease of intimate activity. No clinical data; safety established via biocompatibility and bench performance testing.
Clinical Evidence
No clinical data were submitted. Evidence consists of non-clinical bench testing, including biocompatibility (cytotoxicity, sensitization, vaginal/rectal irritation, acute systemic toxicity per ISO 10993), condom compatibility (ASTM D7661-10), and shelf-life stability (ASTM F1980-16).
Indicated for individuals requiring a personal lubricant for penile, vaginal, or rectal application to moisturize, lubricate, and enhance comfort during sexual activity. Compatible with natural rubber latex and polyisoprene condoms; contraindicated for use with polyurethane condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of three human profiles facing to the right, with flowing lines above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 7, 2017
Non-Friction Products Inc. % James W. Monroe CEO Monroe Medical Device Consulting, LLC 319 Shilling Drive Somerset, NJ 08873
Re: K171730 Trade/Device Name: FuckWater Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: June 9, 2017 Received: June 12, 2017
Dear James W. Monroe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
For
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## 510k Summary: FuckWater Personal Lubricant
## Submitter Information
#### Submitter:
Marc Barrette President, Non-Friction Products Inc. 551 Church Street Toronto, ON M4Y 2E2 Canada Phone: (416) 817-1745. Facsimile: (416) 967-0224 marc@nonfrictionproducts.com
#### Contact Person:
James W. Monroe CEO, Monroe Medical Device Consulting, LLC 319 Shilling Drive Somerset, NJ 08873 USA Phone: (908) 809-0081, Facsimile: (888) 834-1696 wmonroe@monroemdc.com
Date Prepared: August 23, 2017
#### Device Information
Device Name: FuckWater Common or Usual Name: Personal lubricant Classification: 21 CFR 884.5300; condom Product Code: NUC; lubricant, personal Device Class: II
Predicate: Glycerin and Paraben Free Astroglide, K072647
The predicate has not been the subject of a design-related recall.
# Intended Use / Indications for Use
FuckWater is a water-based personal lubricant for penile and/or rectal and/or rectal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.
{3}------------------------------------------------
# Device Description
FuckWater is a non-sterile, home-use, water-based, personal lubricant. It is packaged in 20z, 40z, 8oz and 16oz bottles. The device is intended to be applied to the penis, vagina or rectum to reduce friction during sexual activity. The device is compatible with natural rubber latex and polyisoprene condoms. The device is not compatible with polyurethane condoms. The device has the following specifications:
| Parameter | Specification |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Appearance | Viscous, Semi-Opaque Fluid |
| Color | Off-White |
| Odor | Light Medicinal |
| pH | 7.2 - 7.6 |
| Osmolality | 2,300-3,400 mOsm/kg |
| Viscosity | 20,000 - 26,000 cP |
| Antimicrobial effectiveness<br>(USP <51>) | Bacteria: Not less than 2.0 log reduction from the initial count<br>at 14 days, and no increase from the 14 days' count at 28 days |
| | Yeast and Mold: No increase from the initial calculated count<br>at 14 and 28 days |
| Total Aerobic Microbial<br>Count (USP <61>, <62>) | < 10 cfu/g |
| | Total Yeast and Mold Count<br>(USP <61>, <62>) |
| Absence of Pathogenic<br>Organisms (USP <61>, <62>) | |
# Comparison to Predicate Device
FuckWater is substantially equivalent to the predicate device listed below with respect to intended use/indications for use, principles of operation and technological characteristics.
#### Substantial Equivalence Table
| Attribute | Fuck-Water<br>(Subject Device) | Glycerin and Paraben<br>Free Astroglide<br>(Predicate) | Equivalence |
|-------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | K171730 | K072647 | N/A |
| Product Code | NUC | NUC | Same |
| Indications for Use | FuckWater is a water-<br>based personal<br>lubricant, for penile,<br>vaginal and/or rectal<br>application, intended to<br>moisturize and<br>lubricate, enhance the<br>ease and comfort of<br>intimate sexual activity<br>and supplement the<br>body's natural<br>lubrication. This<br>product is compatible<br>with natural rubber<br>latex and polyisoprene<br>condoms. This product<br>is not compatible with<br>polyurethane condoms. | Glycerin and Paraben<br>Free Astroglide® is a<br>personal lubricant, for<br>penile, anal, or vaginal<br>application, intended to<br>moisturize and<br>lubricate, to enhance<br>the ease and comfort of<br>intimate sexual activity<br>and supplement the<br>body's natural<br>lubrication. This<br>product is compatible<br>with latex condoms. | Similar; minor<br>differences do<br>not change the<br>intended use<br>and do not<br>raise different<br>questions of<br>safety and<br>effectiveness. |
| Condom<br>Compatibility | Natural Rubber Latex<br>Polyisoprene | Natural Rubber Latex | Different;<br>differences in<br>condom<br>compatibility<br>do not raise<br>different<br>questions of<br>safety and<br>effectiveness. |
| Over-The-Counter<br>Use | Yes | Yes | Same |
| Ingredients | Water, Propylene<br>Glycol, Isopropyl<br>Palmitate,<br>Dimethicone,<br>Hydroxyethyl<br>Cellulose, Polysorbate<br>20, Sorbitan Stearate,<br>Cetearyl Alcohol,<br>Glyceryl Stearate,<br>Sodium Benzoate,<br>Gluconolactone | Water, Butylene<br>Glycol, Xylitol,<br>Propylene Glycol,<br>Polyquaternium 15 | Different;<br>differences in<br>formulation do<br>not raise<br>different<br>questions of<br>safety and<br>effectiveness. |
| Biocompatibility<br>Tested | Yes | Yes | Same |
| Sterile | No | No | Same |
| Microbial Quality<br>and Antimicrobial<br>Effectiveness<br>Tested | Yes | Yes | Same |
{4}------------------------------------------------
{5}------------------------------------------------
As noted in the table above, the differences between the subject and predicate devices include the indications for use, condom compatibility and ingredients. These differences do not alter the intended use or the fundamental technology of the subject device as compared to the predicate, and do not raise different questions of safety or effectiveness. The technological differences between the subject device and predicate were evaluated through performance testing as described in the section below.
# Summary of Non-Clinical Performance Data
## Biocompatibility
The subject device is intended to come into direct contact with the mucosa of patients for a limited (< 24hr) duration. Therefore, biocompatibility testing was conducted to evaluate vaginal and rectal irritation, systemic toxicity, hypersensitivity and cytotoxicity in accordance with ISO 10993 as described below.
### Cytotoxicity
The purpose of this test was to examine the test article under evaluation for cytotoxic potential. This test was performed per ISO 10993-5:2009. Based on qualitative evaluation of the mouse fibroblast cells exposed to the test article, the test article was demonstrated to be noncytotoxic.
### Hypersensitivity
The purpose of this test was to determine to what extent a test article has the potential to act as a contact sensitizer in guinea pigs. The assay was conducted according to the methods detailed in the ISO 10993-10:2010, Guinea Pig Maximization. Results from treated animals support that the test article is non-sensitizing. Under similar treatment conditions, all positive control animals used in the historical positive control study exhibited a strong sensitization response to the challenge dose compared to that of the control animals. The results from the historical positive control study demonstrate that the guinea pigs in the study reacted as expected when exposed to a known sensitizer (i.e., DNCB) and the results validate the sensitivity of this test.
#### Vaginal Mucosal Irritation
The purpose of this test was to assess the potential of the test material to produce irritation of the vaginal mucosal tissue in New Zealand White rabbits. This test was performed according to ISO 10993-10:2010. Vaginal Irritation Test. Based on the Irritation Index, the test article was considered a non-irritant to the vaginal mucosa.
#### Rectal Irritation
The purpose of this test was to assess the potential of the test material to produce irritation of the rectal mucosal tissue in New Zealand White rabbits. This test was performed according to ISO 10993-10:2010, Rectal Irritation Test. Based on the Irritation Index, the test article was considered a non-irritant to the rectal mucosa.
#### Acute Systemic Toxicity
The purpose of this test was to evaluate systemic responses to the test article following injection into mice. This test was conducted according to ISO 10993-11:2006, Acute Systemic Toxicity. Testing demonstrated that the subject device is not acutely toxic.
{6}------------------------------------------------
#### Bench Performance Testing
The following non-clinical testing was conducted: Condom compatibility in accordance with ASTM D7661:10 and age testing (ASTM F1980-16) to justify claims of 522 days shelf-life.
#### Shelf Life
Shelf life for the subject device is claimed to be 522 days. Accelerated aging testing was conducted per ASTM F1980-16, which demonstrated that the device met specifications for the proposed shelf-life.
#### Condom Compatibility
Condom compatibility was conducted in accordance with ASTM D7661-10 - Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex. The subject device was found to be compatible with male natural rubber latex and polyisoprene condoms and incompatible with male polyurethane condoms.
## Summary of Clinical Performance Data
No clinical data were submitted to support substantial equivalence.
## Conclusion
The subject device has very similar indications for use to the predicate. Minor differences in the indications statements do not alter the intended use of the subject device. The subject device and the predicate have the same basic technological characteristics. Minor differences in technology do not alter the intended use or fundamental technology, and do not raise different questions of safety and effectiveness. The technological differences between the subject and predicate device were evaluated through performance testing, including biocompatibility, condom compatibility, and shelf-life. The results from performance testing support that the subject device is as safe and effective as the predicate. Therefore, the subject device is substantially equivalent to the predicate device.
{7}------------------------------------------------
# Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.