MOR 3.0mm and PUR NP 3.2mm Implant Systems, MOR 2.1x18mm and 2.4x18mm
Applicant
Sterngold Dental, LLC
Product Code
DZE · Dental
Decision Date
Sep 28, 2017
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MOR™ implants are intended to be used for oral rehabilitation of edentulous and partially dentate patients in the maxilla and mandible to support single unit restorations. Implant retained restorations may consist of single crowns or bridges as well as complete or partial dentures. These implants are intended for delayed loading. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. The MOR™ implants are only intended for use with straight abutments. The implant body is intended to be placed such no angle correction is necessary. The PUR Implants are intended to be used for oral rehabilitation of edentulous and partially dentate patients in the maxilla and mandible to support single unit, and multiple unit retained restorations may consist of single crowns or bridges as well as complete or partial dentures. These implants are intended for delayed loading. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. The PUR implants are only intended for use with straight abutments. The implant body is intended to be placed such no angle correction is necessary.
Device Story
Endosseous dental implants (MOR 3.0mm, PUR 3.2mm, MOR 2.1x18mm, 2.4x18mm) designed for oral rehabilitation. Implants are root-form, threaded, and manufactured from Titanium 6AL-4V ELI with a blasted and acid-etched surface to facilitate osseointegration. Used by dentists/oral surgeons in clinical settings to support single or multiple unit restorations. Implants are placed into the maxilla or mandible; prosthetic components (straight abutments, cover screws) attach to the implant body. The device provides a stable foundation for crowns, bridges, or dentures. Benefits include restoration of dental function and aesthetics for edentulous or partially dentate patients. Implants are provided sterile via gamma irradiation; ancillary components are provided non-sterile for user-performed moist heat sterilization.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing: biocompatibility assessment per ISO 10993-1, surface treatment analysis via SEM/EDS, sterilization validation (SAL 10^-6) per ISO 11137 and ISO 17665, shelf-life validation (5 years) per ISO 11607, and mechanical torque testing in high-density urethane foam blocks to verify structural integrity under clinical load conditions.
Technological Characteristics
Materials: Wrought Titanium 6AL-4V ELI. Design: Threaded, root-form endosseous dental implants with internal hex connection. Surface: Blasted with white aluminum oxide and acid-etched. Sterilization: Gamma irradiation (implants); moist heat (abutments/screws). Dimensions: Various diameters (2.1mm-3.2mm) and lengths (8mm-18mm). Connectivity: None (mechanical device).
Indications for Use
Indicated for oral rehabilitation of edentulous and partially dentate patients in maxilla and mandible to support single or multiple unit restorations (crowns, bridges, complete/partial dentures). Intended for delayed loading; immediate loading indicated with good primary stability and appropriate occlusal load. Only for use with straight abutments where no angle correction is required.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{1}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of two main elements: a circular text element and a symbol. The text element reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion. The symbol is a stylized representation of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 28, 2017
Sterngold Dental, LLC Maria Rao Director of OA & Regulatory Affairs 23 Frank Mossberg Drive Attleboro, Massachusetts 02703
Re: K171728
Trade/Device Name: MOR 3.0mm and PUR NP 3.2mm Implant Systems, MOR 2.1 x 18mm and 2.4x18mm Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: August 25, 2017 Received: August 29, 2017
Dear Maria Rao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mary S. Runner -S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171728
#### Device Name
MOR 3.0mm and PUR NP 3.2mm Implant Systems MOR 2.1x18mm and 2.4x18mm
#### Indications for Use (Describe)
The MOR™ implants are intended to be used for oral rehabilitation of edentulous and partially dentate patients in the maxilla and mandible to support single unit restorations. Implant retained restorations may consist of single crowns or bridges as well as complete or partial dentures. These implants are intended for delayed loading. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. The MOR™ implants are only intended for use with straight abutments. The implant body is intended to be placed such no angle correction is necessary.
The PUR Implants are intended to be used for oral rehabilitation of edentulous and partially dentate patients in the maxilla and mandible to support single unit, and multiple unit retained restorations may consist of single crowns or bridges as well as complete or partial dentures. These implants are intended for delayed loading. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.
The PUR implants are only intended for use with straight abutments. The implant body is intended to be placed such no angle correction is necessary.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------|--|
| <div style="display:flex; align-items:center;"><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | |
| <div style="display:flex; align-items:center;"><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summarv
#### Submitter's Information: 1.
| Name: | Sterngold Dental, LLC |
|----------------------|----------------------------------------------------------------------|
| Address: | 23 Frank Mossberg Drive<br>Attleboro, MA 02703 |
| Phone: | 508-226-5660 ext 1206 |
| Contact: | Maria Rao, QA/RA Director |
| Date Prepared: | September 28, 2017 |
| Device Name: | |
| Trade Name: | MOR 3.0mm and PUR 3.2mm Implant Systems<br>MOR 2.1x18mm and 2.4x18mm |
| Regulation Number: | 21 CFR 872.3640 |
| Product Code: | DZE, NHA |
| Classification Name: | Endosseous dental implant |
| Classification: | Class II |
#### 3. Predicate Devices:
2.
#### Primary Predicates
Prismatik Inclusive Tapered implant - K153099 IMTEC MDI MII One-Piece Implant 2.9mm - K081653
#### Reference Predicates
Sterngold PUR Implant System - K151928 Implant Direct Spectra System Dental Implants 2008 - K090234 Sterngold MOR 2.1mm and 2.4mm - K153173
#### 4. Product Description:
#### PUR 3.2mm Implant System
The PUR 3.2mm Implant System consists of 3.2mm threaded endosseous dental implants, as well as prosthetic components
The prosthetic components include a Straight Narrow Platform Slim Abutment with a 1.5mm and 3.0mm cuffs and a 0.050' Hex Prosthetic Screw. The purpose of the Slim Abutments is to provide surgical and prosthetic options for smaller spaces. They attach directly to the implant with the aid of the prosthetic screw and provide the transitional link between the head of the implant and the restorative components. These abutments are straight and not intended for angulation.
The PUR Straight Slim Abutments are similar to FDA cleared PUR Straight Abutments Narrow and Regular Platforms 1.5mm and 3.0mm cuffs (K151928). The difference is the body of the proposed Straight Narrow Platform Slim Abutment is more tapered. The abutments and prosthetic screw are made of Titanium 6AL-4V ELI. The ancillary component is a 3.2mm cover screw, which is flush with the top surface of the implant to prevent tissue and bone from growing inside
{5}------------------------------------------------
the implant. It is screwed onto the immediately after implant placement. It is made of Titanium 6AL-4V ELI.
The PUR 3.2mm implant is an endosseous dental implant made of Titanium 6AL-4V ELI with a treated roughened surface (blasted and etched). The implant body is tapered with double-lead threads and10 start micro threads at the collar. The implant lengths are 8mm, 10mm, 12mm and 14 mm. This implant The PUR 3.2mm implant features the same prosthetic platform as the 3.5mm and 4.3mm implants already cleared by FDA (K151928). The purpose of the 3.2mm PUR implant is to provide surgical and prosthetic options for smaller spaces. The PUR 3.2mm Implants will be provided "Sterile" using Gamma sterilization.
### MOR 3.0mm Implant System
The MOR 3.0mm Implant System consists of 3.0mm endosseous dental implants. The MOR 3.0mm implant is a one-piece endosseous dental implant, manufactured of Titanium 6AL-4V ELI. The portion of the implant that is submerged in the bone is a treated roughened surface (blasted and etched) to facilitate osseointegration. The MOR 3.0mm implant is similar in design to the MOR 2.4mm implant previously cleared by FDA. The implant lengths are10mm, 13mm, 15mm, and 18mm. The implant has two prosthetic head design options - 0-ball identical to Sterngold MOR 2.4mm Implant (K153173), and a Tapered Abutment similar to the IMTEC MDI MII One Piece Implant 2.9mm (K081653). This implant has the same design threads as the MOR 2.4mm Implant. The MOR Implants will be provided "Sterile" using Gamma sterilization.
#### ട്. Indications for Use:
The MOR" implants are intended to be used for oral rehabilitation of edentulous and partially dentate patients in the maxilla and mandible to support single unit, and multiple unit restorations. Implant retained restorations may consist of single crowns or bridges as well as complete or partial dentures. These implants are intended for delayed loading. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.
The MOR" implants are only intended for use with straight abutments. The implant body is intended to be placed such no angle correction is necessary.
The PUR Implants are intended to be used for oral rehabilitation of edentulous and partially dentate patients in the maxilla and mandible to support single unit, and multiple unit restorations. Implant retained restorations may consist of single crowns or bridges as well as complete or partial dentures. These implants are intended for delayed loading. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.
The PUR implants are only intended for use with straight abutments. The implant body is intended to be placed such no angle correction is necessary.
#### 6. Substantial Equivalence:
The device comparison outlined on Table 2 provides the similarities and substantial equivalency of the proposed devices to its predicates. The comparative data demonstrates that the MOR 3.0mm, PUR 3.2mm Dental Implant Systems, and 2.1x18mm and 2.4 x18mm Implants are substantially equivalent to the predicate devices. They have the same or similar indications for use, design and technological characteristics.
The PUR 3.2mm Dental Implants with lengths of 8, 10, 12 and 14mm are substantially equivalent to predicate devices previously cleared under K090234. The 3.2 x 8mm is similar to the 3.2 x 8mm implant approved under K090234 with same or similar indications for use. Any differences between the proposed devices and the predicate devices do not introduce new concerns in safety and efficacy and does not render the device NSE. Technological characteristics demonstrate the device is as safe and as effective as the predicate devices.
{6}------------------------------------------------
| Attributes | Proposed Devices<br>Ø3.0mm MOR Implant | Predicate Device<br>IMTEC MDI Ø2.9mm<br>Implant<br>(K081653) | Reference Predicate<br>Sterngold<br>MOR Dental Implant<br>System (K153173) | Attributes | Proposed Device<br>Ø3.2mm PUR Implant | Predicate Device<br>(K153099)<br>Prismatik Inclusive<br>Tapered Implant | Reference Predicate<br>Implant Direct<br>Spectra System Dental<br>Implants 2008<br>(K090234) | Reference Predicate<br>Sterngold<br>PUR Dental Implant<br>System<br>(K151928) |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Picture of<br>Implant | Image: Proposed Devices Ø3.0mm MOR Implant | Image: Predicate Device IMTEC MDI Ø2.9mm Implant (K081653) | Image: Reference Predicate Sterngold MOR Dental Implant System (K153173) | Picture of<br>Implant | Image: Proposed Device Ø3.2mm PUR Implant | Image: Predicate Device (K153099) Prismatik Inclusive Tapered Implant | Image: Reference Predicate Implant Direct Spectra System Dental Implants 2008 (K090234) | Image: Reference Predicate Sterngold PUR Dental Implant System (K151928) |
| Implant<br>Design | Threaded, Root-form implant | Threaded, Root-form implant | Threaded, Root-form implant | Implant<br>Design | Threaded, Root-form<br>implant | Threaded, Root-form<br>implant | Threaded, Root-form<br>implant | Threaded, Root-form<br>implant |
| Implant<br>Connection | Internal Hex | Internal Hex | Internal Hex | Implant<br>Connection | Internal Hex | Internal Hex | Internal Hex | Internal Hex |
| Implant Body<br>Geometry | Screw Type | Screw Type | Screw Type | Implant Body<br>Geometry | Screw Type | Screw Type | Screw Type | Screw Type |
| Intended Use | The MOR™ implants are<br>intended to be used for oral<br>rehabilitation of edentulous and<br>partially dentate patients in the<br>maxilla and mandible to<br>support single unit, and<br>multiple unit restorations.<br>Implant retained restorations<br>may consist of single crowns or<br>bridges as well as complete or<br>partial dentures. These implants<br>are intended for delayed<br>loading. Immediate loading is<br>indicated when there is good<br>primary stability and an<br>appropriate occlusal load.<br>The MOR™ implants are only<br>intended for use with straight<br>abutments. The implant body is<br>intended to be placed such no<br>angle correction is necessary | Intended to be used for oral<br>rehabilitation of edentulous and<br>partially dentate patients in the<br>maxilla and mandible to support<br>single unit, and multiple unit<br>restorations. Implant retained<br>restorations may consist of<br>single crowns or bridges as well<br>as complete or partial dentures.<br>These implants are intended for<br>delayed loading. Immediate<br>loading is indicated when there<br>is good primary stability and an<br>appropriate occlusal load.<br>The PUR implants are only<br>intended for use with straight<br>abutments. The implant body is<br>intended to be placed such no<br>angle correction is necessary | Intended for long-term<br>applications in the bone of<br>the patient's upper or lower<br>arch. The MOR implants may<br>also be used for inter-<br>radicular transitional<br>applications.<br>These devices will permit<br>immediate splinting stability<br>and long-term fixation of<br>new or existing crown and<br>bridge installations, for full<br>or partial edentulous cases,<br>and employing minimally<br>invasive surgical<br>intervention. | Intended Use | The PUR Implants are<br>intended to be used for<br>oral rehabilitation of<br>edentulous and partially<br>dentate patients in the<br>maxilla and mandible to<br>support single unit, and<br>multiple unit restorations.<br>Implant retained<br>restorations may consist of<br>single crowns or bridges as<br>well as complete or partial<br>dentures. These implants<br>are intended for delayed<br>loading. Immediate<br>loading is indicated when<br>there is good primary<br>stability and an<br>appropriate occlusal load.<br>The PUR implants are<br>only intended for use with<br>straight abutments.<br>The<br>implant body is intended<br>to be placed such no angle<br>correction is necessary. | Indicated for use in<br>maxillary and mandibular<br>partially or fully<br>edentulous cases, to<br>support single, multiple-<br>unit, and overdenture<br>restorations. The implants<br>are to be used for<br>immediate loading only in<br>the presence of primary<br>stability and appropriate<br>occlusal loading. | The Spectra-System Dental<br>Implant 2008 system is<br>comprised of dental<br>implant fixtures and<br>prosthetic devices that<br>compose a two-piece<br>implant system. The Dental<br>Implants are intended for<br>use in the mandible and<br>maxilla, in support of<br>single unit or multiple unit<br>cement or screw-receiving<br>restorations and for the<br>retention and support of<br>overdentures. The implants<br>are intended for immediate<br>placement and function for<br>the support of single-tooth<br>or multiple-tooth<br>restorations, recognizing<br>bone stability and<br>appropriate occusal load<br>requirements. | It can be used in dental<br>implant applications, for<br>oral rehabilitation of<br>edentulous and partially<br>dentate patients in the<br>maxilla and mandible.<br>Implant retained<br>restorations may consist of<br>single crowns or bridges as<br>well as complete or partial<br>dentures. The PUR Implant<br>is intended for delayed<br>loading. It is also indicated<br>for immediate loading with<br>good primary stability and<br>appropriate occlusal<br>loading. |
| Implant<br>Material | Wrought Titanium 6AL-4V<br>ELI | Wrought Titanium 6AL-4V ELI | Wrought Titanium 6AL-4V<br>ELI | Implant<br>Material | Wrought Titanium 6AL-<br>4V ELI | Wrought Titanium 6AL-<br>4V ELI | Wrought Titanium 6AL-<br>4V ELI | Wrought Titanium 6AL-<br>4V ELI |
| Implant<br>Surface | Roughened - blasted and acid<br>etched | Roughened - blasted and acid<br>etched | Roughened - blasted and acid<br>etched | Implant<br>Surface | Roughened - blasted and<br>acid etched | Roughened - blasted and<br>acid etched | Roughened - blasted and<br>acid etched | Roughened - blasted and<br>acid etched |
| Ø Diameters<br>(mm) | 3.0mm<br>2.1mm and 2.4mm | 2.9mm | 2.1mm, 2.4mm | Ø Diameters<br>(mm) | 3.2mm | 3.2mm | 3.2, 3.7, 4.2, 4.7,5.2, 5.7,<br>7.0mm | 3.5mm, 4.3mm, 5.0mm,<br>6.0mm |
| Lengths (mm) | 10mm, 13mm, 15mm, 18mm | 10mm, 13mm, 15mm, 18mm | 10mm, 13mm, 15mm | Lengths (mm) | 8mm, 10mm,12mm,<br>14mm | 11.5mm, 13mm, 16mm | 8.0mm | 8.0mm, 10mm, 12mm,<br>14mm |
| Sterilization | Gamma irradiation | Gamma irradiation | Gamma irradiation | | | | | |
# Device Comparison – MOR 3.0mm Implant System
{7}------------------------------------------------
{8}------------------------------------------------
| Sterilization | Gamma irradiation | Gamma irradiation | Gamma irradiation | Gamma irradiation |
|---------------------------|---------------------------------|---------------------------------|---------------------------------|---------------------------------|
| Abutment<br>Angulation | No Angulation | 0° - 15° | No Angulation | No Angulation |
| Abutment<br>Material | Wrought Titanium 6AL-<br>4V ELI | Wrought Titanium 6AL-<br>4V ELI | Wrought Titanium 6AL-<br>4V ELI | Wrought Titanium 6AL-<br>4V ELI |
| Construction | Machined | Machined | Machined | Machined |
| Abutment<br>Sterilization | Moist Heat | Moist Heat | Moist Heat | Moist Heat |
# PUR NP Slim Abutments and Prosthetic Screw
| Attributes | Proposed Devices<br>PUR NP Straight Abutment Slim 1.5mm<br>& 3.0mm and<br>.050" Hex Prosthetic Screw | Reference Predicate<br>Sterngold<br>PUR Straight Abutment 1.5mm and 3.0mm<br>and<br>0.48" Prosthetic Screw<br>(K151928) |
|---------------------|------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Design | Image: Proposed Devices | Image: Reference Predicate |
| Material | Wrought Titanium 6AL-4V ELI | Wrought Titanium 6AL-4V ELI |
| Surface | Smooth | Smooth |
| Ø Diameters<br>(mm) | 1.5mm, 3.0mm | 1.5mm, 3.0mm |
| Hex | .050" | .048" |
| Sterilization | Moist Heat | Moist Heat |
| Single Use | Yes | Yes |
# PUR NP Cover Screw
| Attributes | Proposed Devices<br>PUR NP Cover Screw 3.2mm | Reference Predicate<br>Sterngold<br>PUR Cover Screw (K151928) |
|---------------------|-------------------------------------------------|----------------------------------------------------------------|
| Design | Image: Proposed Device PUR NP Cover Screw 3.2mm | Image: Reference Predicate Sterngold PUR Cover Screw (K151928) |
| Material | Wrought Titanium 6AL-4V ELI | Wrought Titanium 6AL-4V ELI |
| Surface | Smooth | Smooth |
| Ø Diameters<br>(mm) | 3.2mm | 3.5, 4.3, 5.0, 6.0mm |
| Sterilization | Gamma | Gamma |
| Single Use | Yes | Yes |
{9}------------------------------------------------
### 7.0 Summary of Non-Clinical Performance Testing
Non-clinical test data was used to support substantial equivalency.
Performance testing of the MOR 3.0mm, and PUR 3.2mm Implant Systems, MOR 2.1x18mm and MOR 2.4x18mm was implemented the FDA guidance Class II Special Controls Guidance Document: Rootform Endosseous Dental Implants and Endosseous Dental Abutments.
### 7.1 Biocompatibility
A risk assessment was performed to identify biological hazards and evaluate estimated risks of known hazards. The evaluation of the biological safety of the proposed devices and their materials consisted of intended use, device characteristics, biological hazard identification, and exposure assessment. A biological evaluation was then performed in accordance with EN ISO 10993-1. The final device was considered during evaluation as well as materials, processes, physical characteristics and properties of finished device, intended use, contact type and duration.
The subject device was compared to the new device and the old submission biocompatibility was sufficient to support SE. There were no changes to material composition or manufacturing process.
#### 7.2 Surface Treatment Analysis
MOR 3.0mm, and PUR 3.2mm Implant Systems, MOR 2.1x18mm and MOR 2.4x18mm are blasted with white aluminum oxide particles followed by acid etching. SEM and EDS analysis was performed on the blasted implant surface of a final device.
SEM photographs and surface analysis confirm that residue and contaminants from the blasting material were not present or detectable in any form and implant surface was homogeneous.
### 7.3 Sterilization and Shelf-Life Validation
MOR 3.0mm, and PUR 3.2mm Implant Systems, MOR 2.1x18mm and MOR 2.4x18mm are provided sterile and are sterilized using Cobalt 60 gamma radiation VD max25 method that has been validated to ensure a SAL of 10° per BS EN ISO 11137-1: 2015 and BS EN ISO 11137-2: 2015. The Abutments, temporary Cap, prosthetic screw and drills are provided non-sterile and sterilized prior to use by the end user using moist heat sterilization. Validation was conducted according to ANSI/AAMI/ISO 17665-1: 2006 and 17665-2: 2009. The overkill method was used to develop a cycle (time and temperature) that would provide a Sterility Assurance Level (SAL) of 10°. The devices were tested to a SAL of at least106 using the biological indicator overkill method.
The MOR 3.0mm, and PUR 3.2mm Implant Systems, MOR 2.1x18mm and MOR 2.4x18mm are Jabeled sterile with a 5-year shelf life supported by validation performed on Sterngold's own devices. Validation testing using the Accelerated Aging method was conducted and results indicated a successful sterility; dye, burst and aerosol challenge tests showed that the packaging is able to maintain a sterile barrier for up to five (5) years. Shelf life validation was performed according to AAMI/ ANSI /ISO 11607-1: 2006/(R) 2010.
LAL testing was conducted as per the FDA Guidance "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile".
{10}------------------------------------------------
### 7.4 Torque Testing
Testing consisted of a torque test, which simulates the intended use of the proposed devices. Each implant size was tested to determine how they would perform being inserted in different density test blocks by measuring their final insertion torque values. Very dense urethane foam blocks were used to place the implants in so as to attain high torque values to test for breakage.
All implants were placed without any breakage. The testing goal was to greatly exceed the torque that would be encountered clinically. It was observed that even at 60Ncm, the proposed implants performed successfully with no evidence of breakage. Torque testing and implant to abutment compatibility demonstrated conformance to design input and indications for use.
#### 8.0 Conclusion:
The proposed devices have the same technological characteristics, performance specifications and intended use as of its predicates. Performance testing data and intended use supports a determination of substantial equivalence of the proposed devices to the predicate and the reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.