← Product Code [MQP](/productcode/MQP) · K171686

# Choice Spine Hawkeye™ Vertebral Body Replacement (VBR) System (K171686)

_Choicespine · MQP · Sep 15, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K171686

## Device Facts

- **Applicant:** Choicespine
- **Product Code:** [MQP](/productcode/MQP.md)
- **Decision Date:** Sep 15, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Choice Spine Hawkeye™ Vertebral Body Replacement (VBR) System is intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). The Choice Spine Hawkeye™ Vertebral Body Replacement (VBR) System is intended for use with supplemental fixation and is to be used with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion.

## Device Story

The Hawkeye VBR System consists of oval/trapezoidal implants (spacers) with hollow centers for bone graft placement and ridged surfaces to prevent migration. Implants are available in various heights and lordotic angles to match patient anatomy. Used in the thoracolumbar spine (T1-L5) by surgeons to replace damaged vertebral bodies. The system provides structural support during the fusion process. Implants are provided non-sterile and require steam sterilization by the user. This submission introduces taller implant sizes (51-60mm) and an additive manufacturing process for titanium components.

## Clinical Evidence

Bench testing only. Testing included static axial compression, static torsion, dynamic torsion, dynamic compression (ASTM F2077), subsidence (ASTM F2267), and expulsion testing. Results demonstrated that the subject device is substantially equivalent to the predicate devices.

## Technological Characteristics

Materials: PEEK (ASTM F2026) with Tantalum markers (ASTM F560) or Ti-6Al-4V ELI (ASTM F136 or ASTM F3001). Manufacturing: Additive manufacturing for titanium components. Design: Oval/trapezoidal spacers with hollow centers and ridged surfaces. Sterilization: Steam sterilization required by user. Connectivity: None.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Choice Spine Vertebral Body Replacement (VBR) System ([K120570](/device/K120570.md))
- Choice Spine Vertebral Body Replacement (VBR) System ([K162103](/device/K162103.md))
- Choice Spine Lumbar Spacer System ([K153107](/device/K153107.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 15, 2017

Choice Spine Kim Finch Manager Regulatory Affairs 400 Erin Drive Knoxville, Tennessee 37924

Re: K171686

Trade/Device Name: Choice Spine Hawkeye™ Vertebral Body Replacement (VBR) System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: MOP Dated: August 16, 2017 Received: August 17, 2017

Dear Ms. Finch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K171686

Device Name

Choice Spine Hawkeye™ Vertebral Body Replacement (VBR) System

Indications for Use (Describe)

The Choice Spine HawkeyeTM Vertebral Body Replacement (VBR) System is intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, or unstable vertebral body due to tumor or trauma (i.e., fracture). The Choice Spine Hawkeye™ Vertebral Body Replacement (VBR) System is intended for use with supplemental fixation and is to be used with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion.

| Type of Use (Select one or both, as applicable)                                                          |                                                                                                         |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">✖</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

| Date:                                  | September 13, 2017                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor:                               | Choice Spine, LP<br>400 Erin Drive<br>Knoxville, TN 37919                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Phone:                                 | 865-246-3333                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Fax:                                   | 865-246-3334                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Contact<br>Person:                     | Kim Finch, Manager of Regulatory Affairs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Proposed<br>Proprietary<br>Trade Name: | Choice Spine Hawkeye™ Vertebral Body Replacement (VBR) System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Product Class:                         | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Classification<br>Name:                | 888.3060 - Spinal Intervertebral Body Fixation Orthosis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Common<br>Name:<br>Device              | Spinal Vertebral Body Replacement Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Product Code:                          | MQP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Purpose of<br>Submission:              | The purpose of this submission is to gain clearance for additional sizes of the<br>Choice Spine Hawkeye™ Vertebral Body Replacement (VBR) System implant<br>family in Titanium using a new method of additive manufacturing. The Choice<br>Spine Vertebral Body Replacement (VBR) System has been cleared by the FDA<br>in PEEK (K120570) and in Titanium (K162103).                                                                                                                                                                                                                                             |
| Device<br>Description:                 | The Choice Spine Hawkeye™ Vertebral Body Replacement (VBR) System is<br>composed of implant components which have a basic oval/trapezoidal shape<br>with a hollow center for placement of bone graft. The superior and inferior<br>surfaces have ridges, or "teeth" for resisting migration. The replacement<br>implants, "spacers", are available in an assortment of heights and in multiple<br>angles of lordosis to accommodate different anatomic requirements. This<br>system includes implants made of PEEK (ASTM F2026) with Tantalum markers<br>(ASTM F560) or Ti-6Al-4V ELI (ASTM F136 or ASTM F3001). |
| Intended Use:                          | The Choice Spine Hawkeye™ Vertebral Body Replacement (VBR) System is<br>intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed,<br>damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture).<br>The Choice Spine Hawkeye™ Vertebral Body Replacement (VBR) System is<br>intended for use with supplemental fixation and is to be used with autogenous<br>and/or allogeneic bone graft comprised of cancellous and/or corticocancellous<br>bone graft to facilitate fusion.                                                                                                |

{4}------------------------------------------------

The implant components are available in two material varieties: Materials: polyetheretherketone (PEEK-OPTIMA® LT1, Invibio®) per ASTM F2026 with integral radiopaque markers manufactured from tantalum per ASTM F560, and titanium alloy (Ti-6Al-4V ELI) which conforms to ASTM F136 or ASTM F3001. All chosen materials are commonly used in medical devices. The implants will be provided non-sterile, but will be steam sterilized before use.

Substantial

Equivalence: The current subject devices in this submission are equivalent to the Choice Spine Vertebral Body Replacement (VBR) System K120570 (primary) and K162103, Choice Spine Lumbar Spacer System K153107 as additional predicates. The predicate devices are equivalent in principle of operation, indications for use, material, biocompatibility, manufacturing methods/ post processing steps, sterilization method, stabilization method, anatomic location and approach, product code and classification, and footprints (length & width). The additional taller height sizes in this submission, along with the new additive manufacturing method per ASTM F3001 and slight design change of being more trapezoidal and having window cutouts, introduce a new worst case for design. The performance data is substantially equivalent when compared to the primary predicate.

Conclusion: The additional implants proposed in this submission are identical in: Principle of Operation, Indications for Use, Biocompatibility, Sterilization and Stabilization Method, Anatomic Location and Approach, Intended Use, Product Code and Classification, and Footprints (Length & Width) to the previously cleared implants in this family.

> The addition of the taller heights (51-60mm) represents a new worst case implant design along with a change in manufacturing method. Therefore, the worst-case size of the additional implants proposed in this submission was subjected to mechanical testing, which is discussed in detail in section 22, and results (along with the comparison in section 14.1) showed that the subject device is substantially equivalent to the predicates listed.

Non-clinical The non-clinical testing performed was Static Axial Compression, Static Testing Torsion, Dynamic Torsion, Dynamic Compression per ASTM F2077, Subsidence per ASTM F2267, and Expulsion testing.

---

**Source:** [https://fda.innolitics.com/device/K171686](https://fda.innolitics.com/device/K171686)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
