K171680 · Xiros, Ltd. · GAT · Jun 30, 2017 · General, Plastic Surgery
Device Facts
Record ID
K171680
Device Name
Infinity-Lock 3, Infinity-Lock 5
Applicant
Xiros, Ltd.
Product Code
GAT · General, Plastic Surgery
Decision Date
Jun 30, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Infinity-Locks are single use devices intended to be used for soft tissue approximation, including Achilles tendon repair in patients with acute rupture of Achilles tendon.
Device Story
Infinity-Lock 3 and 5 are non-absorbable, sterile, woven poly(ethylene terephthalate) surgical tapes; designed for soft tissue approximation, specifically Achilles tendon repair. Devices feature an integral loop for cow-hitch knotting, tapered ends for easier tissue passage, and green-striped tails for identification. Used by surgeons in clinical settings; tapes are sutured through tendon ends (e.g., Bunnell technique) to draw separated ends together, facilitating healing. Devices are single-use implants. Clinical benefit derived from improved usability features compared to predicate, aiding surgical workflow without altering fundamental mechanical performance or safety profile.
Clinical Evidence
Bench testing and cadaveric validation only. Mechanical testing (ultimate tensile strength) performed on knotted and unknotted samples per internal method QT010 and USP 39 NF33:2016. Clinician validation using cadaveric materials confirmed suitability for intended use. No clinical trial data presented.
Technological Characteristics
Material: Poly(ethylene terephthalate) (polyester) fibers. Construction: Woven textile tape. Form factor: Two sizes (3 and 5) with integral loop, tapered ends, and green identification stripe. Sterilization: Sterile. Performance: Complies with USP 39 NF33:2016 requirements for non-absorbable surgical sutures.
Indications for Use
Indicated for soft tissue approximation, including Achilles tendon repair in patients with acute rupture of the Achilles tendon.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Teleflex Medical Inc., polyethylene terephthalate suture (Cottony Il green suture polyester) (K021019)
Submission Summary (Full Text)
{1}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 30, 2017
Xiros, Ltd Mr. Stephen Seed Springfield House Whitehouse Lane, Leeds LS19 7UE England
Re: K171680
Trade/Device Name: Infinity-Lock 3, Infinity-Lock 5 Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable Poly(ethylene terephthalate) Surgical Suture Regulatory Class: Class II Product Code: GAT Dated: June 2, 2017 Received: June 6, 2017
Dear Mr. Seed:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{2}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# David Krause -S
- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K171680
Device Name Infinity-Lock 3 Infinity-Lock 5
#### Indications for Use (Describe)
The Infinity-Looks are single use devices intended to be used for soft tissue approximation, including Achilles tendon repair in patients with acute rupture of the Achilles tendon.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "XIROS" in a rounded rectangle. The text is in a sans-serif font and is colored in gray. The background is white. There is a small line extending from the right side of the rounded rectangle.
#### SECTION 7. Special 510(k) SUMMARY
In accordance with 21 CFR 807.92, the following information constitutes the Special 510(k) summary for the Infinity-Locks 3 and 5.
## 1.SUBMITTER INFORMATION
| Submitter`s Name: | Xiros Limited |
|-------------------|-------------------------------------------------------------------|
| Address: | Springfield House<br>Whitehouse Lane<br>Leeds LS19 7UE<br>England |
| Telephone Number: | +44 (0)1132387200 |
| Fax Number: | +44 (0)113 2387201 |
| Contact Person 1: | Stephen Seed (Compliance Director) |
| Contact Person 2: | Manuela Rodrigues (Principal Regulatory Affairs Associate) |
| Date: | 02 May 2017 |
### II. DEVICE IDENTIFICATION
| Name of Device(s): | Infinity-Lock 3<br>Infinity-Lock 5 |
|-----------------------|-------------------------------------------------|
| Trade Name(s): | Infinity-Lock 3<br>Infinity-Lock 5 |
| Common or usual name: | Non-absorbable Polyester Surgical Tape |
| Classification Name: | Suture, Non-absorbable, Synthetic, Polyethylene |
| Regulatory Class: | Class II |
| Product Code: | GAT |
| Panel: | General & Plastic Surgery |
### III. PREDICATE/UNMODIFIED DEVICE
Predicate: Xiros, Ltd, Poly-Tapes K002172.
This predicate has not been subject to a design related recall.
Reference device: Teleflex Medical Inc., polyethylene terephthalate suture (Cottony Il green suture polyester) K021019.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "XIROS" in a rounded rectangle. The text is in a sans-serif font and is a light gray color. The rounded rectangle has a thick, light gray border. The background is white.
## IV. DEVICE DESCRIPTION
The Infinity-Locks are non-absorbable, sterile, poly(ethylene terephthalate) sutures. They are prepared from fibres of high molecular weight, long chain, linear polyesters having recurrent aromatic rings as an integral component. Infinity-Locks differ from USP Sutures in being in the form of woven tapes and in exceeding all USP sizes.
The product comprises two sizes of textile implant based on the Neoligaments poly-tape design. The tapes have an integral loop at one end which are used to tie a cowhitch through one end of a ruptured tendon. The tapes have two tails which are then sutured through the separated end (using methods such as the Bunnell technique), thus drawing the two ends together in order to facilitate healing of the tear. One of the tails has a green stripe to aid identification and both taper into a cord to assist with passing through soft tissues.
#### V. INTENDED USE
The Infinity-Locks are single use devices intended to be used for soft tissue approximation, including Achilles tendon repair in patients with acute rupture of Achilles tendon.
The device subject to modification (Poly-Tapes) has exactly the same intended use as the modified device. The Poly-Tapes are single use devices intended to be used for soft tissue approximation, including Achilles tendon repair in patients with acute rupture of Achilles tendon.
### VI. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The substantial equivalence of the Infinity-Locks is supported by the fact that they have the same intended use and they have common technological characteristics to the Poly-Tapes. The same material is used in the manufacture of the proposed modified device (both are made of polyester). Both Poly-Tapes and Infinity-Locks have the same construction; they are woven from polyester, and have the same performance characteristics in accordance with USP requirements.
The modified device retains all the features of the original poly-tapes but includes some new ones to aid usability. These are the: -
- . Tapered ends that are intended to aid threading which was previously achieved by adding sutures in theatre.
- . Tails that are marked with green weft to aid identification, previously both were plain.
- Addition of an integral loop to make it easier for surgeons to tie a cow hitch knot . when required. The previous device was un-looped and the modified device can be simply un-looped when required.
These minor differences in the design of the Infinity-Lock variants do not bring any new questions of safety and effectiveness in accordance with performance data hence do not constitute a change in the fundamental technological characteristics.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the word "XIROS" in a rounded rectangle. The text is in a sans-serif font and is gray. The background is white. The rounded rectangle has a thick border.
### VII. PERFORMANCE DATA
Mechanical testing was performed on the modified device and original poly-tape in both knotted and unknotted conditions to determine the ultimate tensile strengths using an internal method, QT010, and following USP 39 NF33:2016 for non-absorbable surgical sutures, where applicable. The modified devices were found to be equivalent or superior in all cases.
Validation was also undertaken by clinicians using cadaveric materials to demonstrate the suitability of the modified device for the intended use which was found to be satisfactory.
The performance data benefit/risk analysis concluded that the differences encountered do not affect the safety and efficacy of the new device variants in relation to the predicate.
Summary: Based on the pre-clinical testing performance, the Infinity-Lock 3 and Infinity-Lock 5 are found to have the same safety and effectiveness profile to the predicate device (Poly-Tapes).
#### VIII.CONCLUSION
Xiros conclude the Infinity-Lock 3 and Infinity-Lock 5 are substantially equivalent to the Poly-Tapes.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.