← Product Code [FDF](/productcode/FDF) · K171673

# Focus Cap (K171673)

_Gce Medical Corporation · FDF · Jan 19, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K171673

## Device Facts

- **Applicant:** Gce Medical Corporation
- **Product Code:** [FDF](/productcode/FDF.md)
- **Decision Date:** Jan 19, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Focus Cap has been designed to be attached to the distal end of a colonoscope. The Focus Cap is intended for the following: For use in keeping the suitable depth of the colonoscope's field of view.

## Device Story

Focus Cap is a short, transparent tube attachment for the distal end of a colonoscope. It functions as a mechanical spacer, maintaining a fixed distance between the scope camera and the patient's anatomy to prevent image obstruction. The device is used by clinicians during colonoscopy procedures. It is available in 11mm and 13.5mm sizes and attaches to the scope via compression. The design features a slanted edge to facilitate fluid exit, replacing the side-hole design found in predicate devices. By ensuring a consistent field of view, the device assists the clinician in visualizing the colon, potentially improving diagnostic or therapeutic outcomes during the procedure.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including mechanical verification and biocompatibility validation, which met all pre-determined acceptance criteria.

## Technological Characteristics

Transparent tube attachment; 11mm and 13.5mm sizes; compression-fit mechanism; slanted design for fluid management. Non-powered, passive mechanical accessory. No software or electronic components.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Olympus Distal Attachment ([K984358](/device/K984358.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 19, 2018

GCE Medical Corporation % Stephen M. Page President MedReg Associates Inc. 228 Hull Cove Farm Road Jamestown, Rhode Island 02835

Re: K171673 Trade/Device Name: Focus Cap Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDF Dated: January 10, 2018 Received: January 16, 2018

Dear Stephen M. Page:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows the name "Charles Viviano -S" in a large, sans-serif font. The text is black and appears to be the main focus of the image. The background is plain and does not distract from the text.

For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known)

K171673

Device Name Focus Cap

Indications for Use (Describe)

The Focus Cap has been designed to be attached to the distal end of a colonoscope. The Focus Cap is intended for the following:

* For use in keeping the suitable depth of the colonoscope's field of view.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

PSC Publishing Services (301) 443-6740 EF

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

K171653

# 510(k) Summary

## Submitter Information

- A. Company Name: GCE Medical Corporation
- B. Company Address: 2723 West Tenaya, Fresno CA 93711
- Company Phone: 559-439-5757 C.
- Contact Person: Dr. Stanley Chang D.
- E. Summary Prepared on: January 10, 2018

### Device Identification

- Device Trade Name: Focus Cap A.
- B. Device Common Name: Transparent Cap
- Classification Name: Endoscope and accessories C.
- Classification Number: 21 CFR Part 876.1500 D.
- Product Code: FDF D.
- E. Device Class: II

### Identification of Predicate Device

| Trade Name               | Manufacturer | 510(k)  |
|--------------------------|--------------|---------|
| Distal Distal Attachment | Olympus      | K984358 |

### Indications for Use

The GCE Medical Focus Cap has been designed to be attached to the distal end of a colonoscope. The Focus Cap is intended for the following:

- For use in keeping the suitable depth of the colonoscope's field of view.
### Device Description

The GCE Focus Cap has the shape of a short transparent tube which is attached to the distal end of the colonoscope. The Focus Cap is designed to facilitate the scope's view during colonoscopy procedures.

{4}------------------------------------------------

The Focus Cap will be offered in two sizes:

- 11mm (Catalog #11 FC 1701)
- 13.5mm (Catalog #13 FC 1702)

The 11mm Focus Cap will fit 11.6mm and 12.8mm scopes. The 13.5mm Focus Cap will fit 13.2mm and 13.9mm scopes.

## Comparison to Predicate Device:

| Characteristic                          | Comparison to Predicate<br>Olympus Distal<br>Attachment (K984358) | Differences                                                                                                                                                                                                                                                                                                                                                                             |
|-----------------------------------------|-------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                            | Identical                                                         | N/A                                                                                                                                                                                                                                                                                                                                                                                     |
| Indications for Use                     | Identical                                                         | N/A                                                                                                                                                                                                                                                                                                                                                                                     |
| Fundamental<br>scientific<br>technology | Identical                                                         | N/A                                                                                                                                                                                                                                                                                                                                                                                     |
| Operating<br>principles                 | Similar                                                           | The operating principle for both<br>devices is that they attach to a scope<br>by compression. Although there are<br>slight differences in their relative<br>adhesion strengths, those differences<br>are minimal and have no impact on<br>safety or effectivenss.                                                                                                                       |
| Mechanism of<br>action                  | Similar                                                           | Both the predicate and proposed<br>devices act as simple spacers. They<br>establish a space between the end of<br>the scope and patient's anatomy,<br>thereby preventing the camera image<br>from being blocked. Although the two<br>devices have slightly different<br>dimensional characteristics, those<br>differences are minimal and have no<br>impact on safety or effectiveness. |
| Technological<br>aspects                | Similar                                                           | The Olympus Distal Attachment<br>incorporates a side hole which is not<br>present on the Focus Cap. In addition,<br>the Focus Cap incorporates a slant<br>design while the Olympus design is<br>flat The slant design of the Focus                                                                                                                                                      |

{5}------------------------------------------------

| Cap, which essentially takes the place   |
|------------------------------------------|
| of the side hole, allows for a wide      |
| open surface for the exit of fluid which |
| is many times larger than the side hole  |
| on the predicate device. Therefore.      |
| these differences have no impact on      |
| safety or effectiveness.                 |

## Performance Testing

Performance testing has been completed for the GCE Focus Cap to demonstrate substantial equivalence to the predicate Olympus Distal Attachment (K984358). The Focus Cap has been subjected to the following verification and validation testing: Mechanical and biocompatibility. All test requirements were met as specified by test protocols and pre-determined acceptance criteria.

## Conclusions

The subject Focus Cap and predicate Olympus Distal Attachment (K984358) share the same intended use and indications for use. Furthermore, the subject and predicate devices share the same fundamental technology and key device characteristics. Differences in design and technological characteristics between the subject and predicate devices do not raise any new questions of safety and effectiveness. The results of verification and validation support the safety and effectiveness of the Focus Cap for its intended use and provide reasonable assurance of the substantial equivalence of the subject device to the predicate Olympus Distal Attachment (K984358).

---

**Source:** [https://fda.innolitics.com/device/K171673](https://fda.innolitics.com/device/K171673)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
