← Product Code [GEX](/productcode/GEX) · K171648

# BM.IRIS (K171648)

_Bluecore Co., Ltd. · GEX · Sep 8, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K171648

## Device Facts

- **Applicant:** Bluecore Co., Ltd.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Sep 8, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

IRIS Q-switched Nd : YAG Laser System is indicated for the incision, ablation, vaporization of soft tissues for general dermatologic and general surgical procedures for coagulation and hemostasis. 532nm Wavelength Tattoo removal light ink (red, tan, purple, orange, sky blue,) Removal of Benign Epidermal Pigmented Lesions, Minor Benign Vascular Lesions, Talangiectasias Treatment of Lentigines, Cafe-Au-Lait, Sebrorrheic Keratoses, Becker's Nevi. Freckles Treatment of Post Inflammatory Hyperpigmentation (PIH) 1064nm Wavelength: Tattoo removal: dark ink (black, blue, green ) Removal of Nevus of Ota Removal or lightening of unwanted hair with or without adjuvant preparation Treatment of Common Nevi, Melasma, Skin resurfacing procedures for the treatment of acne scars, wrinkle

## Device Story

BM.IRIS is a Q-switched Nd:YAG laser system for dermatologic and surgical soft tissue procedures. System utilizes Xe-gas lamp to excite Nd:YAG crystal, producing 1064nm and 532nm wavelengths; optional dye handpieces enable 585nm or 650nm output. Energy delivered via articulated arm and multi-spot handpiece. Physician operates system via front-panel LCD and foot switch to control fluence and spot size. Device provides coagulation, hemostasis, ablation, and vaporization. Used in clinical settings by physicians to treat pigmented lesions, tattoos, vascular lesions, and for skin resurfacing. Benefits include precise tissue interaction for aesthetic and surgical applications.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing, including IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-22 (laser safety), and IEC 60825-1 (equipment classification).

## Technological Characteristics

Q-switched Nd:YAG laser; wavelengths 1064nm/532nm (optional 585nm/650nm via dye handpiece). Components: Xe-gas lamp, Nd:YAG crystal, articulated arm, multi-spot handpiece. Handpiece tips: S.S.304. Cooling system included. Controlled via LCD panel and foot switch. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Spectra Nd:YAG ([K113588](/device/K113588.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the address of the Food and Drug Administration. The address is 10903 New Hampshire Avenue, Document Control Center - WO66-G609, Silver Spring, MD 20993-0002. The text is left-aligned and in a simple font.

September 8, 2017

Bluecore Co.,Ltd. % Young Chi C.E.O Bio-Med USA Inc 111 Ellison Street Paterson, New Jersey 07446

Re: k171648

Trade/Device Name: Bm.iris Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: May 28, 2017 Received: June 5, 2017

Dear Young Chi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely. Jennifer R. Stevenson -S3

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K171648

## Device Name

#### IRIS Indications for Use (Describe)

IRIS Q-switched Nd : YAG Laser System is indicated for the incision, ablation, vaporization of soft tissues for general dermatologic and general surgical procedures for coagulation and hemostasis.

### 532nm Wavelength

Tattoo removal light ink (red, tan, purple, orange, sky blue,) Removal of Benign Epidermal Pigmented Lesions, Minor Benign Vascular Lesions, Talangiectasias Treatment of Lentigines, Cafe-Au-Lait, Sebrorrheic Keratoses, Becker's Nevi. Freckles Treatment of Post Inflammatory Hyperpigmentation (PIH)

### 1064nm Wavelength:

Tattoo removal: dark ink (black, blue, green ) Removal of Nevus of Ota Removal or lightening of unwanted hair with or without adjuvant preparation Treatment of Common Nevi, Melasma, Skin resurfacing procedures for the treatment of acne scars, wrinkle

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# BlueCore Co., Ltd

#### Q-Switched Nd:YAG laser

# 510 (K) Summary

As required by CFR 807.92(c)

### 1. Manufacturer. Submitter

Prepared Aug 20, 2017

BlueCore Co., Ltd. Acehigh 21, 48 Centum-Jungang ro 12 Haeundae Gu, Busan, Rep of Korea T: 82 51 747 4318, F: 82 51 747 4319

## 2. Contact person

Bio-Med USA Inc. Young Chi, President. 27 New England Drive, Ramsey, NJ 07446. U.S.A. t: 1-973 278 5222 f: 1 201 934 6030 e mail: biomedusa@msn.com

### 3. Name of Device

| Trade name :           | BM.IRIS                            |
|------------------------|------------------------------------|
| Classification name :  | Powered, Laser surgical instrument |
| Common name :          | Nd:YAG Q-switched Surgical Laser   |
| Regulation :           | 878.4810<br>Class II               |
| Classification Panel : | General and Plastic Surgery.       |
| Product Code :         | GEX                                |
| Type of submission :   | Traditional                        |

### 4. Legally marketed Predicate Device

K113588 Spectra Nd:YAG Lutronic Corp

BM.IRIS Q-Switched Nd: Y AG laser system produce same two wave length (1064nm, 532nm), and same characteristics such as Design, Construction, Energy rate, Pulse Duration, Cooling system and intended use as already cleared predicate device K113588 by Lutronic,

## 5. Device Description

The IRIS Q-Switched Nd: YAG laser system produces a two pulsed beam, 1064 nm Infrared and 532nm long pulse laser, and using different Handpiece able to control various treatment fluence.

laser tube : placed in the mixed crystals of copper pipe to the heater and produces a laser beam, Resonator : amplifies the beam, through the Xe-gas contained lamp lamp : Xe-gas contains high pressure lamp to increase specific laser beam 

{4}------------------------------------------------

This converted light energy creates the Nd: Y AG crystal and exhaust from the crystal is amplified into a specific wave length. Laser energy produced is delivered to the Tissue by means of an articulated arm and a specially designed multi spot Hand Piece.

The Physician can optimize the effect for different applications by controlling the energy of the laser pulse and the spot size of the treatment beam, and is able to activate laser emission using Foot Switch. Laser parameters and other system features are controlled from a display panel located on the front of the power supply unit.

#### This system also consist of

Optic main Bench assembly, Articulated Arm Hand pieces, LCD control panel, Cooling system, Foot Pedal Switch

This device has optional 180mm Dye Hand Pieces to able to switch to 585nm or 650nm from 532nm.

## 6. Performance test

Clinical data is not required in this submission, but manufactured in accordance with below both mandatory and voluntary standards, performance test data attached.

IEC60601-1 part 1 : General requirement for basic safety and essential performance. IEC60601-1-2: 2007 Electro Magnetic Compatibility test IEC60601-2-22 Part 2, Particular requirements for safety of diagnostic and Therapeutic laser IEC60825-1 :2nd ED, Equipment classification and requirement.

## 7. Indication for use

IRIS Q-switched Nd : YAG Laser System is indicated for the incision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis.

#### 532nm Wavelength

Tattoo removal light ink (red, tan, purple, orange, sky blue,) Removal of Benign Epidermal Pigmented Lesions, Minor Benign Vascular Lesions, Talangiectasias Treatment of Lentigines, Cafe-Au-Lait, Sebrorrheic Keratoses, Becker's Nevi, Freckles Treatment of Post Inflammatory Hyperpigmentation (PIH)

#### 1064nm Wavelength:

Tattoo removal: dark ink (black, blue, green ) Removal of Nevus of Ota Removal or lightening of unwanted hair with or without adjuvant preparation Treatment of Common Nevi, Melasma, Skin resurfacing procedures for the treatment of acne scars, wrinkle

{5}------------------------------------------------

## 8. Biocompatibility

This device are meant non-contacted mode. Hand piece tips is made by S.S.304 same as predicate device.

# 9. Conclusion.

BM.IRIS Q-Switched Nd: YAG laser system, in this submission, is substantially equivalent to several already cleared predicate device in respect to the Intended use, Main function, Technology, Principal operation and performance. And every Safety test report show it as safe and effective as predicate device and it does not raise any additional issues for safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/device/K171648](https://fda.innolitics.com/device/K171648)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
