K171565 · Ge Oec Medical Systems, Inc. · OWB · Aug 10, 2017 · Radiology
Device Facts
Record ID
K171565
Device Name
OEC Elite
Applicant
Ge Oec Medical Systems, Inc.
Product Code
OWB · Radiology
Decision Date
Aug 10, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The OEC Elite Mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical procedures.
Device Story
OEC Elite is a mobile fluoroscopic C-arm imaging system used by surgeons and physicians in clinical settings (OR, clinic). It captures X-ray images of patient anatomy and interventional tools using a C-shaped gantry with an X-ray tube and detector (Flat Panel or Image Intensifier). The workstation provides the user interface, image processing, and display. The system provides real-time visualization for localizing pathology and guiding procedures. The vascular option adds features like subtraction, roadmapping, and digital cine pulse to optimize visualization of contrast-filled vessels, catheters, and stents. Physicians use the workstation to view images, which aids in clinical decision-making and procedural guidance. The device benefits patients by enabling minimally invasive interventional procedures through improved anatomical visualization.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including system performance verification, hazard mitigation, and imaging performance evaluation using anthropomorphic phantoms per FDA guidance for solid-state X-ray imaging devices. Metrics included spatial resolution (MTF), contrast resolution, temporal resolution, and dose rate, showing performance equivalent to the predicate.
Technological Characteristics
Mobile C-arm fluoroscopic X-ray system. Components: high voltage generator, X-ray tube, Flat Panel Detector or Image Intensifier, LCD workstation. Connectivity: digital image processing/recording. Software: vascular-specific profiles and processing. Standards: IEC 60601-1 Ed. 3, IEC 60601-2-54, IEC 60601-2-43, 21 CFR Subchapter J. Software development lifecycle compliant with IEC 60324.
Indications for Use
Indicated for adult and pediatric patient populations during diagnostic, interventional, and surgical procedures, including orthopedic, gastrointestinal, endoscopic, urologic, neurologic, vascular, critical care, and emergency procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 10, 2017
GE OEC Medical Systems, Inc. Rachel Schandel Regulatory Affairs Leader 384 Wright Brothers Drive Salt Lake City, Utah 84116
Re: K171565
Trade/Device Name: OEC Elite Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, JAA, OXO Dated: July 26, 2017 Received: July 27, 2017
Dear Ms. Rachel Schandel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D. O'Hara For
Robert A. Ochs, Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K171565
Device Name OEC Elite
#### Indications for Use (Describe)
The OEC Elite mobile fluoroscopy system is designed to provide fluoroscopic and digital spot images of adult and pediatric patient populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include orthopedic, gastrointestinal, endoscopic, urologic, vascular, critical care, and emergency procedures.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| <div><span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div><span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. There are three small, white, teardrop-shaped elements evenly spaced around the outside of the circle, giving the logo a distinctive and recognizable appearance.
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) Summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h).
| Date: | May 26th, 2017 |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE OEC Medical Systems, Inc (GE Healthcare)<br>384 Wright Brothers Drive<br>Salt Lake City, Utah 84116 |
| Primary Contact: | Rachel Schandel<br>Regulatory Affairs Leader<br>T (801)5364790<br>Rachel.schandel@ge.com |
| Secondary Contact: | Helen Peng<br>Senior Regulatory Affairs Manager<br>GE Medical Systems, LLC<br>Tel: 262-424-8222<br>e-mail: hong.peng@ge.com |
| Secondary Contact: | John Jaeckle<br>Chief Regulatory Affairs Strategist<br>GE Medical Systems, LLC<br>Tel: 262-424-9547<br>e-mail: john.jaeckle@ge.com |
## PRODUCT IDENTIFICATION
| Device Trade Name: | OEC Elite |
|------------------------------------|---------------------------------------------------------------------------------------------------------|
| Regulation Name: | Image-intensified Fluoroscopic x-ray system |
| Regulation: | 21CFR 892.1650 |
| Classification: | Class II |
| Product Code: | OWB, JAA, OXO |
| Manufacturer<br>/ Design Location: | GE OEC Medical Systems, Inc (GE Healthcare)<br>384 Wright Brothers Drive.<br>Salt Lake City, Utah 84116 |
| Manufacturing Location(s): | GE OEC Medical Systems, Inc (GE Healthcare)<br>384 Wright Brothers Drive. |
Salt Lake City, Utah 84116
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. The circle is surrounded by decorative swirls, also in blue, which add a sense of elegance and motion to the design. The logo is simple, yet recognizable, and represents the long-standing brand identity of General Electric.
## Predicate Device:
| Device Name: | OEC Elite |
|------------------|-------------------------------------------------------|
| 510(k) number: | K170752 |
| Manufacturer: | GE OEC Medical Systems, Inc (same as proposed device) |
| Regulation Name: | Image-intensified Fluoroscopic x-ray system |
| Regulation: | 21CFR 892.1650 |
| Classification: | Class II |
| Product Code: | OWB, JAA, OXO |
## Marketed Device:
OEC Elite with Vascular option is a Mobile Fluoroscopic C-arm Imaging System with modification to the predicate device OEC Elite (K170752) to offer the users the vascular features. It is of comparable type and substantially equivalent to the predicate device OEC Elite. This 510(k) submission for the OEC Elite system with vascular option includes the same intended use and substantially equivalent indications for use as its predicate device. The existing indications for use of OEC Elite are expanded to include the vascular capability which is one of the current clinical uses of mobile c-arm systems that are well established in the medical community.
The system is labeled as the OEC Elite.
## Device Description:
The OEC Elite is a Mobile Fluoroscopic C-arm Imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic X-Ray images during diagnostic, interventional, and surgical procedures. These images help the physician visualize the patient's anatomy and interventional tools. This visualization helps to localize clinical regions of interest and pathology. The images provide real-time visualization and records of pre-procedure anatomy, in vivo-clinical activity and post-procedure outcomes. The system is composed of two primary physical components. The first is referred to as the "C - Arm" because of its "C" shaped image gantry; the second is referred to as the "Workstation", which is the primary interface for the user to interact with the system.
The C-arm is a stable mobile platform capable of performing linear motions (vertical, horizontal) and rotational motions (orbital, lateral, wig-wag) that allow the user to position the X-Ray image chain at various and distances with respect to the patient anatomy to be imaged. The C - arm is mechanically balanced allowing for ease of movement and capable of being "locked" in place using a manually activated lock. The C-Arm is comprised of the high voltage generator, software, X-ray control, and a "C" shaped image gantry, which supports an X-ray tube and a Flat Panel Detector or Image Intensifier, depending on the choice of detector configuration desired.
The workstation is a stable mobile platform with an articulating arm supporting a color image, high resolution, LCD display monitor. It also includes image processing equipment/software, recording devices, data input/output devices and power control systems.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The circle is surrounded by a series of curved lines that resemble water droplets or waves. The logo is simple and recognizable, and it is associated with a well-known and established company.
The primary purpose of the mobile fluoroscopy system is to provide fluoroscopic images of the patient during diagnostic, interventional, and surgical procedures such as orthopedic, gastrointestinal, endoscopic, urologic, neurologic, vascular, critical care and emergency procedures.
The primary technology change for the subject device compared to the unmodified predicate, OEC Elite, is the introduction of the vascular capabilities.
# Intended Use
The OEC Elite Mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical procedures.
# Indications for Use:
The OEC Elite mobile fluoroscopy system is designed to provide fluoroscopic and digital spot images of adult and pediatric patient populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include orthopedic, gastrointestinal, endoscopic, urologic, neurologic, vascular, critical care, and emergency procedures.
# Technology:
The OEC Elite with Vascular option employs the same fundamental scientific technology as that of the predicate device OEC Elite (K170752).
The image chain including the X-Ray source and generator remain unchanged from that of the predicate OEC Elite. The primary change was in the software to implement the vascular features such as Subtraction, Roadmapping, and Digital cine pulse to help perform vascular procedures with the easiest workflow and least intervention by the user. A suite of supporting features such as Peak opacification, Cine Recording/playback, Cine Automatic image playback, Cine Frame-by-frame review, Reregistration, Variable landmarking, and Mask save/recall, were added to further enhance the vascular workflows. Additionally a new vascular profile was developed to optimize the visualization of positive contrast agent-filled vessels, catheters, stents, and other tools in vascular sperformed in thick anatomy, while a Bolus Chase profile was created to optimize the visualization of the run-off procedures and interventional procedures using small guidewires.
The modifications however, were built upon the existing robust and extensible software architecture, following the same design control process and software development lifecycle process that is compliant to IEC60324. The changes described above do not change the control mechanism, operating principle, energy type, or intended use from the predicate device.
## Determination of Substantial Equivalence:
The main change in the proposed device is the introduction of the vascular feature. The table below summarizes the substantive feature/technological differences between the predicate device and the proposed device:
{6}------------------------------------------------
| | Predicate Device<br>OEC Elite (K170752) | Proposed Device<br>OEC Elite with Vascular Option |
|-----------------------------------------|---------------------------------------------------|---------------------------------------------------|
| Imaging Modes | > Continuous – Fluoroscopy | > Continuous – Fluoroscopy |
| | ○ Normal dose | ○ Normal dose |
| | ○ High Level dose | ○ High Level dose |
| | ○ Low dose | ○ Low dose |
| | > Pulsed Fluoroscopy | > Pulsed Fluoroscopy |
| | ○ Normal dose | ○ Normal dose |
| | ○ High Level dose | ○ High Level dose |
| | ○ Low dose | ○ Low dose |
| | > Digital Spot | > Digital Spot |
| | | > Digital Cine Pulse |
| | | ○ Normal dose |
| | | ○ Low dose |
| | | > Roadmap |
| | | ○ Normal dose |
| | | ○ Low dose |
| | | > Subtraction |
| | | ○ Normal dose |
| | | ○ Low dose |
| Imaging Features | > Auto X-Ray technique control | > Auto X-Ray technique control |
| | > Noise and motion reduction | > Noise and motion reduction |
| | > Auto/Manual Brightness and Contrast Control | > Auto/Manual Brightness and Contrast Control |
| | > Negate | > Negate |
| | > Swap and auto-swap | > Swap and auto-swap |
| | > Save and auto-save | > Save and auto-save |
| | > Last image hold | > Last image hold |
| Predicate Device<br>OEC Elite (K170752) | Proposed Device<br>OEC Elite with Vascular Option | |
| Edge enhancement | Edge enhancement | |
| Zoom & Roam | Zoom & Roam | |
| Image rotation | Image rotation | |
| Image flip/ invert | Image flip/ invert | |
| Manual/Auto Smart Metal | Manual/Auto Smart Metal | |
| AutoTrak | AutoTrak | |
| Window/Level | Window/Level | |
| Patient Annotation | Patient Annotation | |
| Markers | Markers | |
| Measurement Functions | Measurement Functions | |
| | Peak opacification | |
| | Cine Recording/playback | |
| | Cine Automatic image playback | |
| | Cine Frame-by-frame review | |
| | Re-registration | |
| | Variable landmarking | |
| | Mask save/recall | |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a blue circle. There are decorative swirls around the outer edge of the circle, giving it a classic and recognizable appearance.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a blue circle. There are decorative white swirls around the perimeter of the circle, adding a touch of elegance to the design.
GE Healthcare believes the OEC Elite with the Vascular option is of comparable type and substantially equivalent to our currently marketed system OEC Elite.
The substantial equivalence was also based on software documentation for a "Moderate" level of concern device.
#### Summary of Additional Testing
In addition to the verification and validation testing successfully completed as required by GE OEC Medical System's Quality Management System, additional engineering (non-Clinical) testing was performed to provide the requisite data to substantiate performance claims, the revised indications, and ultimately substantial equivalence.
#### Non-Clinical Testing
Verification and validation, including hazard mitigation has been executed with results demonstrating the OEC Elite mobile fluoroscopy system met design input and user needs.
The system has been tested and is compliant with the IEC 60601-1 Ed. 3 series, including IEC60601-2-54 and IEC 60601-2-43. All applicable 21CFR Subchapter J performance standards are met. The OEC Elite system was developed under the GE OEC Medical Systems Quality Management System, including design controls, risk management and software development life cycle processes. The following quality assurance measures were applied to the development of the system:
- Risk Analysis
- · Required Reviews
- · Design Reviews
- · Testing on unit level (Sub System verification)
- · Integration testing (System verification)
- · Performance testing (Verification)
- · Safety testing (Verification)
- · Simulated use testing (Validation)
Additional engineering bench testing was performed including demonstration of system performance; and an imaging performance evaluation using anthropomorphic phantoms. All the image quality/performance testing identified for fluoroscopy found in FDA's "Information for Industry: X-ray Imaging Devices- Laboratory Image Quality and Dose Assessment, Tests and Standards" was performed with acceptable results.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized script, enclosed within a blue circle. The circle has a white border with decorative swirls, giving it a classic and recognizable appearance.
The OEC Elite with Vascular option was also evaluated for its performance per the FDA "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices'' (SSXI). It was determined the modified system continues to comply to this guidance. The evaluation includes but not limited to the following performance metrics identified in the SSXI guidance, showing identical performance of the subject device to the predicate device OEC Elite:
- DOE ●
- Dynamic Range
- Spatial Resolution (MTF, Limiting Resolution) ●
- Temporal Resolution
- Contrast Resolution
- Beam Alignment
- Dose Rate ●
- Stability of the device characteristics over time
- brightness uniformity
- . Fluoroscopy Frame Rate
- Reuse Rate .
# Clinical Testing
Because OEC Elite with the vascular option does not change the system's intended use and represents equivalent technological characteristics, this type of change supports using scientific, established/standardized, engineering/physics-based performance testing, without inclusion of clinical images. Therefore, for OEC Elite with Vascular option configurations, clinical images are not required to demonstrate the substantial equivalence to the predicate device.
# Substantial Equivalence Conclusion:
The differences discussed in this submission do not introduce any adverse effects nor raise new questions of safety and effectiveness. Based on the successful verification and validation testing, additional engineering bench testing, conformance to standards, and development under GE OEC Medical System's Quality Management System, we believe that the OEC Elite is of comparable type and substantially equivalent to the predicate device OEC Elite (K170752), and therefore is safe and effective for its intended use.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.