The Bladder Scanner (model: PadScan Z3, PadScan Z3, PadScan Z5) projects ultrasound energy through the lower abdomen of the patient to obtain images of the bladder which is used to calculate bladder Volume noninvasively. The Bladder Scanner is intended to be used only by qualified medical professionals.
Device Story
Handheld, battery-operated B-mode ultrasound device; projects ultrasound energy through lower abdomen to image bladder; calculates bladder volume noninvasively. Operated by qualified medical professionals in clinical settings. Features two operation modes (Expert/Easy), touch screen, and thermal printer. Provides real-time pre-scan imaging; outputs bladder volume and images. Data can be uploaded via USB or Bluetooth. Benefits include fast, non-invasive, and reliable bladder volume assessment, aiding clinical decision-making regarding urinary retention or catheterization.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), acoustic output (NEMA UD 2), and biocompatibility (ISO 10993-1, 5, 10).
Technological Characteristics
Mechanical sector probe; 2.5 MHz resonant frequency; 120° sector angle; 12 scan planes. PE skin-contact material. Battery or AC/DC adapter powered. 7-8 inch TFT-LCD display. Connectivity via USB/Bluetooth. Class II equipment, Type B applied part. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD 2.
Indications for Use
Indicated for non-invasive bladder volume calculation in patients via B-mode pulsed-echo ultrasound. Contraindicated for fetal use, pregnant patients, patients with ascites, patients with open/damaged skin, or wounds in the suprapubic region. Intended for use by qualified medical professionals.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
January 24, 2018
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
AvantSonic Technology Co., Ltd. % Mr. Kevin Wang Consultant Chonconn Medical Device Consulting Co., Ltd. 22A. Haijing Square. No. 18. Taizi Road Nanshan District, Shenzhen, 518067 CHINA
Re: K171528
Trade/Device Name: Bladder Scanner (model: PadScan ZS3. PadScan Z3. PadScan Z5) Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO, ITX Dated: December 18, 2017 Received: December 18, 2017
Dear Mr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171528
### Device Name
Bladder Scanner (model: PadScan DS3, PadScan Z3, PadScan Z5)
Indications for Use (Describe)
The Bladder Scanner (model: PadScan Z3, PadScan Z3, PadScan Z5) is B-mode pulsed-echo ultrasound device. It intended as a handheld battery-operated device. The Bladder Scanner projects ultrasound energy through the lower abdomen of the patient to obtain images of the bladder which is used to calculate bladder Volume noninvasively. The Bladder Scanner is intended to be used only by qualified medical professionals.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 14px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
*PRAStaff@fda.hhs.gov*
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
System: PadScan DS3
Transducer: MP2A
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Clinical Application | Mode of Operation | | | | | | |
|-----------------------------|---------------------------------|-------------------|---|----|----|-------|--------------------|-----------------|
| General | Specific | B | M | PW | CW | Color | Combined (Specify) | Other (Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Fetal / Obstetrics | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro logical) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) * | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Musculo-skeletal(Conventional) | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Bladder) | N | | | | | | |
| | Adult Cardiac | | | | | | | |
| Cardiac | Pediatric Cardiac | | | | | | | |
| | Intravascular(Cardiac) | | | | | | | |
| | Trans-esoph.(Cardiac) | | | | | | | |
| | Intra- cardiac | | | | | | | |
| Peripheral<br>vascular | Peripheral vascular | | | | | | | |
| | Other (Specify) | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
{4}------------------------------------------------
## System: PadScan Z3
## Transducer: MP2
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Clinical Application | Mode of Operation | | | | | | |
|--------------------|---------------------------------|-------------------|---|----|----|-------|-----------|-----------|
| General | Specific | B | M | PW | CW | Color | Combined | Other |
| | | | | | | | (Specify) | (Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal / Obstetrics | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro logical) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| Fetal | Neonatal Cephalic | | | | | | | |
| Imaging<br>& Other | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Musculo-skeletal(Conventional) | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Bladder) | N | | | | | | |
| | Adult Cardiac | | | | | | | |
| Cardiac | Pediatric Cardiac | | | | | | | |
| | Intravascular(Cardiac) | | | | | | | |
| | Trans-esoph.(Cardiac) | | | | | | | |
| | Intra- cardiac | | | | | | | |
| Peripheral | Peripheral vascular | | | | | | | |
| vascular | Other (Specify) | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
{5}------------------------------------------------
## System: PadScan Z5
## Transducer: MP2
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|---------------------------------|----------|-------------------|---|----|----|-------|--------------------|-----------------|
| | General | Specific | B | M | PW | CW | Color | Combined (Specify) | Other (Specify) |
| | | | | | | | | | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal / Obstetrics | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intra-operative (Specify) | | | | | | | | |
| | Intra-operative (Neuro logical) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (Specify) | | | | | | | | |
| Fetal<br>Imaging<br>& Other | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Musculo-skeletal(Conventional) | | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other (Bladder) | N | | | | | | | |
| Cardiac | Adult Cardiac | | | | | | | | |
| | Pediatric Cardiac | | | | | | | | |
| | Intravascular(Cardiac) | | | | | | | | |
| | Trans-esoph.(Cardiac) | | | | | | | | |
| | Intra- cardiac | | | | | | | | |
| Peripheral<br>vascular | Peripheral vascular | | | | | | | | |
| | Other (Specify) | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
{6}------------------------------------------------
## 510(K) Summary
Prepared in accordance with the requirements of 21 CFR Part 807.92
## Prepared Date: 2017/05/08
1. Submission sponsor
Name: AvantSonic Technology Co., Ltd. Address: No.394, Jingdongfang Avenue, Beibei District, Chongqing, P.R. China 400714 Contact person: Cuiling Wu Title: R&D manager E-mail: wucuiling@sina.com Tel: +86-755 88271976
#### Submission correspondent 2.
Name: Chonconn Medical Device Consulting Co., Ltd. Address: 22A, Haijing Square, No. 18, Taizi Road, Nanshan District, Shenzhen, P.R. China 518067 Contact person: Kevin Wang E-mail: kevin@chonconn.com Tel: +86-755 33941160
| Trade/Device Name | Bladder Scanner |
|-------------------|-----------------------------------------------------------------------------------------------------------------|
| Model | PadScan DS3, PadScan Z3, PadScan Z5 |
| Common Name | Diagnostic Ultrasound System with Accessories |
| Regulatory Class | Class II |
| Classification | 21CFR 892.1560/Ultrasonic Pulsed Echo Imaging System/IYO<br>21CFR 892.1570/Diagnostic Ultrasound Transducer/ITX |
| Submission type | Traditional 510(K) |
#### 3. Subject Device Information
#### Predicate Device 4.
| 510(K) number | K131227 |
|---------------|-----------------------------------|
| Device Name | PadScan HD series Bladder Scanner |
| Manufacturer | Caresono Technology Co., Ltd |
| Product Code | IYO, ITX |
### Device Description 5.
The PadScan Series manufactured by AvantSonic Technology Co., Ltd. provides non-invasive volume bladder measurement utilizing real-time ultrasound imaging and measurement. The equipment consists of the main unit, 3D probe, battery and adapter.
It features:
Bladder Scanner– 510(k) submission
{7}------------------------------------------------
- . Two Operation Modes: Expert Mode and Easy Mode
- . Non-invasive, comfortable, correct, reliable, fast and simple operation
- . Printouts with ultrasound images and various parameters through PC software
- . Touch screen keyboard operation
- . Urine volume setting and alarm setting
- . Multi-language selection
- . Combined power supply with AC adapter and a built-in battery
The difference between these models is the size of the LCD screen and enclosure structure. PadScan DS3 is provided 7-inch LCD screen. PadScan Z3 is provided 7-inch LCD screen and LCD screen stand. PadScan Z5 is provided 8-inch LCD screen and LCD screen which has a handle.
#### Intended use 6.
The Bladder Scanner (model: PadScan Z3, PadScan Z3, PadScan Z5) projects ultrasound energy through the lower abdomen of the patient to obtain images of the bladder which is used to calculate bladder volume noninvasively. The Bladder Scanner is intended to be used only by qualified medical professionals.
#### 7. Indication for use
The Bladder Scanner (model: PadScan Z3, PadScan Z3, PadScan Z5) is B-mode pulsed-echo ultrasound device. It intended as a handheld batterv-operated device. The Bladder Scanner projects ultrasound energy through the lower abdomen of the patient to obtain images of the bladder which is used to calculate bladder Volume noninvasively. The Bladder Scanner is intended to be used only by qualified medical professionals.
#### 8. Contraindications
Do not use the Bladder Scanner on following cases:
- a) Fetal use or pregnant patients
- b) Patients with ascites
- c) Patients with open or damaged skin.
- d) Wounds in the suprapubic region
#### 9. Comparison to Predicate Device
Comparison to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate device as summarized in the following table
| SE Comparisons | PadScan DS3, PadScan Z3,<br>PadScan Z5 | PadScan HD series Bladder<br>Scanner | Note |
|-----------------|----------------------------------------|--------------------------------------|------|
| Manufacturer/K# | AvantSonic/Present<br>application | Caresono/K131227 | -- |
Tahle 1
Bladder Scanner- 510(k) submission
{8}------------------------------------------------
# AvantSonic
| Intended Use | The PadScan series Bladder<br>Scanner projects ultrasound<br>energy through the lower<br>abdomen of the patient to<br>obtain images of the<br>bladder which is used to<br>calculate bladder volume<br>noninvasively. The<br>PadScan HD series Bladder<br>Scanner is intended to be | The PadScan HD series<br>Bladder Scanner projects<br>ultrasound energy through<br>the lower abdomen of the<br>patient to obtain images of<br>the bladder which is used to<br>calculate bladder volume<br>noninvasively. The PadScan<br>HD series Bladder Scanner<br>is intended to be used only | Same |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| | used only by qualified<br>medical professionals. | by qualified medical<br>professionals. | |
| Contraindications | Do not use the Bladder<br>Scanner on following cases:<br>a) Fetal use or pregnant<br>patients<br>b) Patients with ascites<br>c) Patients with open or<br>damaged<br>skin.<br>d) Wounds in the<br>suprapubic region | Do not use the PadScan<br>HD2 Bladder<br>Scanner on following cases:<br>a) Fetal use or pregnant<br>patients<br>b) Patients with ascites<br>c) Patients with open or<br>damaged<br>skin.<br>d) Wounds in the suprapubic<br>region | Same |
| Modes of operation | B mode | B mode | Same |
| System<br>Characteristics | a) Portable<br>b) LCD Display<br>c) Thermal Printer<br>d) Power source: Battery or<br>AD-DC adapter | a) Portable<br>b) LCD Display<br>c) Thermal Printer<br>d) Power source: Battery or<br>AD-DC adapter | Same |
| Display | PadScan DS3: 7" TFT-LCD<br>PadScan Z3: 7" TFT-LCD<br>PadScan Z5: 8" TFT-LCD | PadScan HD3: 7" TFT-LCD<br>PadScan HD5: 8" TFT-LCD | Same |
| Controls for Change<br>of acoustic output<br>during scan | No | No | Same |
| Transducer Type | Mechanical Sector Probe | Mechanical Sector Probe | Same |
| Measurement place | Abdomen | Abdomen | Same |
| Transducer Resonant<br>Frequency | 2.5Mhz | 2.5Mhz | Same |
| Number of elements | 1 | 1 | Same |
| Sector Angle | 120° | 120° | Same |
| Number of Scan<br>Planes | 12 | 12 | Same |
| Patient Contacting<br>Material | PE (skin contact) | PE (skin contact) | Same |
| Volume measurement<br>range | 0ml-999ml | 0ml-999ml | Same |
| Volume measurement<br>accuracy | ±10%, ±10ml | ±15%, ±15ml | Different |
| Classification of<br>protection against<br>electric shock | Class II equipment | Class II equipment | Same |
| Applied part type | B type | B type | Same |
| Real-time scanning | Yes (Pre-scan) | Yes (Pre-scan) | Same |
| PC Data Upload | USB, Bluetooth | SD card | Different |
| Power | AC/DC Adapter:<br>Input: AC 100V-240V,<br>50/60Hz,<br>Output: DC13.5V±0.5V<br>Battery: Li-ion<br>rechargeable | AC/DC Adapter:<br>Input: AC100-240V,<br>50/60Hz,<br>Output: DC14V±0.5V<br>Battery: Li-ion rechargeable | Different |
| Standards compliance | IEC 60601-<br>1:2005+A1:2012 | IEC 60601-1:2005<br>+CORR.1(2006)<br>+CORR.2(2007) | Different |
| | IEC 60601-1-2:2014 | IEC 60601-1-2:2007 | |
| | IEC 60601-2-37:2015 | IEC 60601-2-37:2007 | Different |
| FDA limit | Track 1 | Track 1 | Same |
Bladder Scanner– 510(k) submission
{9}------------------------------------------------
## AvantSonic
All the differences don't affect the safety and effectiveness which is concluded after all the required testing, so no safety and effectiveness issues relating to the system come into conclusion.
## 10. Performance Data
## Clinical test:
Clinical testing is not required.
## Non-clinical data
The PadScan DS3, PadScan Z3 and PadScan Z5 Bladder Scanner comply with:
(1) IEC 60601-1 Electrical Safety
(2) IEC 60601-1-2 Electromagnetic Compatibility
(3) IEC 60601-2-37 Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
Bladder Scanner– 510(k) submission
{10}------------------------------------------------
(4) NEMA UD 2 Standard for real-time display of thermal and mechanical acoustic output indies on diagnostic ultrasound equipment.
(5) Acoustic output testing as per the guideline "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers" dated September 9, 2008.
The following biocompatibility standards are conducted on the subject device: (1) ISO 10993-1, ISO 10993-5 and ISO 10993-10
The tests were selected to show substantial equivalence between the subject device and the predicate.
## 11. Conclusion
Based on the safety and performance testing and compliance with acceptable voluntary standards, we believe that the PadScan DS3, PadScan Z3 and PadScan Z5 Bladder Scanner are substantially equivalent to their predicate device in K131227 and does not raise any new safety and/or effectiveness issues.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.