Compressyn™ Staple

K171473 · Dallen Medical, Inc. · JDR · Nov 7, 2017 · Orthopedic

Device Facts

Record IDK171473
Device NameCompressyn™ Staple
ApplicantDallen Medical, Inc.
Product CodeJDR · Orthopedic
Decision DateNov 7, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Compressyn™ Staple is intended for: 1) hand and foot bone fragment and osteotomy fixation and joint arthrodesis, 2) fixation of proximal tibial metaphysis osteotomy, 3) adjunctive fixation of small bone fragments. These fragments may be located in long bones such as the femur, fibula and tibia in the lower extremity; the humerus, ulna or radius in the upper extremities; the clavicle and ribs; and in the flat bones such as the pelvis, scapula and sternum.

Device Story

Compressyn™ Staple is a metallic bone fixation system; consists of stainless steel staple preloaded on handheld mechanical delivery device (ABS/polycarbonate). Used by surgeons for bone fragment/osteotomy fixation and joint arthrodesis. Device placed into two tissue segments via pre-drilled holes; barbs prevent back-out; provides stabilized compression. Provided sterile in disposable tray with drill bit, drill guide, placement pins, and sizing guide. Benefits patient by providing stabilized fixation for bone healing.

Clinical Evidence

Bench testing only. Testing included staple delivery and compressive force evaluation. Testing performed to ASTM F564-10 standards.

Technological Characteristics

Stainless steel staple; handheld delivery device made of ABS and polycarbonate. Mechanical fixation via barbs. Provided sterile. Tested per ASTM F564-10.

Indications for Use

Indicated for patients requiring bone fragment fixation, osteotomy fixation, or joint arthrodesis in the hand, foot, long bones (femur, fibula, tibia, humerus, ulna, radius), clavicle, ribs, and flat bones (pelvis, scapula, sternum).

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ November 7, 2017 Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized design. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Dallen Medical, Inc. Al Memmolo Chief Operating Officer 1046 Calle Recodo. Suite G San Clemente, California 92673 ## Re: K171473 Trade/Device Name: Compressyn™ Staple Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: JDR Dated: September 29, 2017 Received: October 3, 2017 Dear Al Memmolo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Katherine D. Kavlock -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K171473 Device Name Compressyn™ Staple Indications for Use (Describe) The Compressyn™ Staple is intended for: 1) hand and foot bone fragment and osteotomy fixation and joint arthrodesis, 2) fixation of proximal tibial metaphysis osteotomy, 3) adjunctive fixation of small bone fragments. These fragments may be located in long bones such as the femur, fibula and tibia in the lower extremity; the humerus, ulna or radius in the upper extremities; the clavicle and ribs; and in the flat bones such as the pelvis, scapula and sternum. Type of Use (Select one or both, as applicable) [X] Prescription Use (Part 21 CFR 801 Subpart D) [ ] Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/14) PSC Publishing Services (301) 443-6740 вн Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {3}------------------------------------------------ K171473 PAGE 1 OF 2 This 510(k) summary information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. ## GENERAL INFORMATION | APPLICANT: | Dallen Medical, Inc.<br>1046 Calle Recodo, Suite G<br>San Clemente, CA 92673<br>(949) 218-0030<br>(949) 218-0040 Fax | |------------------------|----------------------------------------------------------------------------------------------------------------------| | CONTACT PERSON: | Al Memmolo<br>Chief Operating Officer | | DATE PREPARED: | September 29, 2017 | | DEVICE DESCRIPTION: | | | TRADE NAME: | Compressyn™ Staple | | GENERIC/COMMON NAME | Fixation Staple | | CLASSIFICATION NAME: | Single/multiple component metallic bone fixation appliances<br>and accessories, CFR 888.3030 | | PRODUCT CODE: | JDR | | DEVICE CLASSIFICATION: | Class II | | PREDICATE DEVICES: | Dallen Medical Compressyn Staple, Preloaded Sterile System<br>(K140358)<br>New Deal SA Uni-Clip Staple (K991482) | ## Product Description: The Compressyn Staple is a fully integrated system that consists of a stainless steel staple preloaded on a handheld ABS and polycarbonate mechanical delivery device. The staple is offered in 14 sizes with barbs to prevent back out. It is a compression staple fixation device that is placed into two tissue segments using pre-drilled holes to provide stabilized fixation. The Compressyn Staple is provided sterile in a disposable tray with a drill bit, drill guide, placement pins and staple sizing guide. {4}------------------------------------------------ ## Indications for Use: The Compressyn Staple is intended for: 1) hand and foot bone fragment and osteotomy fixation and joint arthrodesis, 2) fixation of proximal tibial metaphysis osteotomy, 3) adjunctive fixation of small bone fragments. These fragments may be located in long bones such as the femur, fibula and tibia in the lower extremity; the humerus, ulna or radius in the upper extremities; the clavicle and ribs; and in the flat bones such as the pelvis, scapula and sternum. ## Technical Characteristics and Performance Data: The Compressyn Staple is identical to the predicate and therefore has the same physical and technical characteristics. The Compressyn Staple was tested for staple delivery and compressive force and was shown to be equivalent to the predicate. Testing to ASTM F564-10 was also performed for the Compressyn Staple, Preloaded Sterile System. ## Basis for Determination of Substantial Equivalence: Upon reviewing the documentation provided in this submission and comparing intended use, principle of operation, performance data, and overall technological characteristics, the Compressyn Staple is determined by Dallen Medical to be substantially equivalent to existing legally marketed devices.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%