K171465 · Stryker · MBI · Aug 24, 2017 · Orthopedic
Device Facts
Record ID
K171465
Device Name
Stryker Self-Punching ICONIX
Applicant
Stryker
Product Code
MBI · Orthopedic
Decision Date
Aug 24, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Stryker Self-Punching ICONIX Anchors are intended to be used for soft-tissue to bone fixation in the elbow, shoulder, knee and hip. Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Repair Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis Hip: Capsular Repair, Acetabular Labral Repair, Gluteal Tendon Repair
Device Story
Self-Punching ICONIX Anchors; all-suture anchors with push-in design; provided sterile on disposable inserter. Polyester sheath interwoven over non-absorbable working sutures; mounted on forked inserter tip. Self-punching mechanism allows direct insertion into bone without drill or guide; sheath bunches upon deployment for fixation. Used in orthopedic surgery (elbow, shoulder, knee, hip); operated by surgeons. Output is mechanical fixation of soft tissue to bone; aids in structural repair of tendons and ligaments; benefits patient through secure tissue reattachment.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical benchtop testing. Testing verified fixation strength at time zero and after cyclic loading (compared to Predicate A) and insertion effort (compared to Predicate B). Results showed statistically higher fixation strength and reduced insertion effort for the subject device.
Technological Characteristics
All-suture anchor; polyester sheath; non-absorbable working sutures; disposable inserter. Self-punching insertion mechanism; no drill/guide required. Sterile barrier system packaging. Materials identical to reference devices.
Indications for Use
Indicated for soft-tissue to bone fixation in the elbow, shoulder, knee, and hip for patients requiring surgical repair of tendons, ligaments, or labral tissues.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 24, 2017
Stryker Katie Farraro Regulatory Affairs Specialist 5900 Optical Court San Jose, California 95138
Re: K171465
Trade/Device Name: Stryker Self-Punching ICONIX Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: July 7, 2017 Received: July 10, 2017
Dear Ms. Farraro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| PSC Publishing Services (301) -111-6740<br>नाना | រ វិច នៃច្បារ | 000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 |
|-------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | to noticellos s, at broadser of benuber in as noved a bus , osnogs 10 loubuos ton year variety and<br>" เครื่อง ที่ได้ BMO bilev ปุ่นตอนมว ย รับธรูปตน ที่ 22 ธนาคม ที่ต้างเทคที่เกิด | |
| | toodse redoo your and elemine sidt guibreger stremmoo bries. noilsemomin to noirely bre<br>:of ,nebrud sint pripuber 107 snoitseggus pribulani ,noitsellos noilsmonni sidi 10<br><br>Office of Chief Information Office<br>Food and Dung Administration<br>Department of Health and Human Services<br>vop.syy.spj@yglelSAAA | |
| | ereldmos based and rishniam bus 19416 , secure etable guiteixe dosea , suoijourismi waiver of smit<br>ədi onların və sinə qalında çəkilməsini ilə sinfinin sinfinə sinfi tot nəticəllər sinfi növündə əni<br>. WOJER 225900A JIAMS 77ATS ASS 3HT OT MAO7 G3T379MOJ SUOY GM32 TON 004<br>See to to to to the noticular and to stremembres of the sellade notices single | |
| | ОЭСЭН ЭГЭЭАЧ ЭТАЯАЧЭГ А ИО ЭНИГТИОЭ | |
| (วันธุ์ เป็นสินค้า เป็นความที่ต้างอิง (วันเทอร์) อาจาก (ว | Type of Use (Sleech one or Doth, as applicable)<br>( Preserpription Use (Part 21 CFR 80 Subgation | |
| | | |
| | | |
| | a problem a production a stadio a labor Caline a consideres and | |
| ອນນຸໄປປົງ ເດຍສອງ , ແຜນການປີ 1ຮປະຈິກ ການສະຫຼຸບັງ ໄຮເອເຮັປີ ລາວເຂົ້າ ການສະຖິງ ໄດ້ຮອງສາກອາຫຼິ້ນ ໂຮງລາວໄປວັນທີ່ 1991 ເຮືອປີ | Lissement Respons, Illiotioisse Band Tenodesis | |
| | risqes Is realolusqs o o fitis as Constitue and Colleges and<br>างมีเราะตุรี เธโนวเซอร์ อัลรออกร์ อยุควร์ สุเธอร์ แต่เรว 1 ค.ศ. 2 เหตุล กระทรซิ เหตุร | |
| | ตอนาวมารถอวรรี เตอนเธรูไป โรเจเรไโดว์ โธเทรนี จารย์ปี สุทธศหวรฝราย : พอปเรื | |
| the elbow | .woled suoitesiin see .qid bas see .qiri bans.<br>ກ່ ແດ່ມີຂະໜີ ອາໄດ້ຮັບໄປສອບນ້ຳ ກິດຂາວນີ ອອນ ອີງ ປະຈຸບາວໄທ ອາຣ ຂາວປິວາ A XINO그I ອູດເຕ່ລົດພາສາຫະລັດສະຫຼຸລາວລາວ | Indications for Use (Describe) |
| | | Stryker Self-Punching ICONIX<br>Device Name |
| | | 377465<br>510(k) Number (if known |
| See PRA Statement below | indications for Use | |
| Expiration Date: January 31, 2017<br>Form Approved: OMB No. 0810-0120 | ЭЭЛУЯЭГ ИАМИН ОНА НТТАЭН ТО ТИЗМТЯАРЭС<br>Food and Drug Administration | |
| לחקרת מקום המו<br>။ ၂၀၂ ခုနှစ်မြို့နယ | | |
{3}------------------------------------------------
# 510(k) Summary
#### I. SUBMITTER
Stryker Endoscopy 5900 Optical Ct San Jose, CA 95138 Katie Farraro, Regulatory Affairs Specialist Contact Person: Phone: 408-754-2285 Fax: 408-754-2507 July 7, 2017 Date Prepared:
#### II. DEVICE
| Name of Device: | Stryker Self-Punching ICONIX |
|-----------------------|-------------------------------------------------------------------------|
| Model Numbers: | 3910-500-921, 3910-500-922 |
| Common or Usual Name: | Suture, Fastener, Fixation, Nondegradable, Soft Tissue |
| Classification Name: | Smooth or threaded metallic bone fixation fastener (21<br>CFR 888.3040) |
| Regulatory Class: | II |
| Product Code: | MBI |
#### III. PREDICATE AND REFERENCE DEVICES
Predicate Devices:
Predicate A (Primary Predicate): Device Name: ICONIX All Suture Anchors Company Name: Stryker 510(k) #: K133671
Predicate B: Device Name: Y-Knot RC All-Suture Anchor, 2.8mm - Double Loaded and Triple Loaded w/ #2 Hi-Fi Sutures Company Name: ConMed Corporation 510(k) #: K133224
Predicate devices have not been subject to a recall.
{4}------------------------------------------------
Reference Devices:
Reference Device 1 Device Name: Stryker ICONIX TT All Suture Anchor Company Name: Stryker 510(k) #: K170098
Reference Device 2 Device Name: Force Fiber Polyethylene Non-Absorbable Suture Company Name: Teleflex Medical 510(k) #: K063778
Reference Device 3 Device Name: Force Fiber Blue Polyethylene Non-Absorbable Surgical Suture Company Name: Beere Precision Medical Instruments, Kmedic, Telef 510(k) #: K092533
Reference Device 4 Device Name: XBraid TT Suture Tape Company Name: Stryker 510(k) #: K162310
Reference devices have not been subject to a recall.
#### IV. DEVICE DESCRIPTION
The Self-Punching ICONIX Anchors (herein referred to as the proposed device(s)) are all suture anchors with a push-in design, provided sterile and preloaded on a disposable inserter. The anchors consist of a polyester sheath interwoven over non-absorbable working sutures, which are folded over and mounted on the forked tip of the inserter. The anchor is inserted into the bone using a self-punching mechanism, and the polyester sheath bunches as the anchor is deployed to allow for fixation in bone. The anchor preloaded on the inserter is packaged in a single-use sterile barrier system (SBS).
#### V. INTENDED USE
The Stryker Self-Punching ICONIX Anchors are intended to be used for soft-tissue to bone fixation in the elbow, shoulder, knee and hip. See indications below.
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
{5}------------------------------------------------
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Repair
Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
Hip: Capsular Repair, Acetabular Labral Repair, Gluteal Tendon Repair
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE AND REFERENCE DEVICES
The Self-Punching ICONIX Anchors are substantially equivalent to Predicate A (the Primary Predicate) with regard to intended use, materials of construct, anchor design and deployment method, and performance attributes.
The proposed devices are substantially equivalent to Predicate B in regard to operational principle (specifically, insertion method). The Self-Punching ICONIX products utilize a self-punching insertion mechanism, such that the device can be directly inserted into bone without the use of a drill or guide.
Reference Devices 1-4 were selected to support scientific methodology throughout this submission, as these legally marketed devices utilize identical components to the proposed devices. Reference Device 1 is another suture anchor in the ICONIX product family that utilizes identical suture tape (XBraid TT) material and anchor sheath material, and Reference Devices 2-4 represent each of the suture components used in the proposed devices as standalone medical devices.
The differences between the proposed products and the predicate devices are summarized below, and do not raise new questions of safety and effectiveness.
Proposed and Predicate Device Differences:
#### Suture Configurations
The proposed devices include 2.3mm diameter polyester sheaths that are identical to the 2.3mm anchors of Predicate A. However, another suture configuration is available in the proposed devices that includes flat sutures woven into the sheath. This configuration is unique compared to Predicate A.
#### Inserter
In order to achieve the self-punching insertion mechanism, the inserter of Predicate A has been modified to include a larger diameter, sharper tips, and a longer handle. The
{6}------------------------------------------------
working length of the inserter shaft also features an increased shaft diameter compared to Predicate B.
### VII. PERFORMANCE DATA
Non-clinical benchtop testing was performed to verify the fixation strength at time zero as well as following cyclic loading of the Self-Punching ICONIX products. This fixation strength was compared to that of Predicate A through statistical analysis. The comparison indicates that the proposed devices provide statistically higher fixation strength than the devices of Predicate A. and will be functional within the intended use.
Benchtop testing was also performed to verify the insertion effort required for use of the Self-Punching ICONIX products. Due to the equivalent self-punching insertion mechanism, this was compared to that of Predicate B. The comparison indicates that the proposed devices require reduced insertion effort compared to the devices of Predicate B, and will be functional within the intended use.
Clinical testing was not required to demonstrate substantial equivalence for this submission.
# VIII. CONCLUSIONS
The information presented within this traditional premarket submission demonstrates that the Stryker Self-Punching ICONIX products are substantially equivalent to the predicate devices and will perform safely and effectively within the intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.