K171429 · Levita Magnetics International Corp · PNL · Jun 6, 2017 · General, Plastic Surgery
Device Facts
Record ID
K171429
Device Name
Magnetic Surgical System
Applicant
Levita Magnetics International Corp
Product Code
PNL · General, Plastic Surgery
Decision Date
Jun 6, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4815
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Magnetic Surgical System is designed to grasp and retract the body and the fundus of the gallbladder in laparoscopic cholecystectomy procedures to facilitate access and visualization of the surgical site. The device is indicated for use in patients within a BMI range of 20 to 34 kg/m2.
Device Story
System comprises hand-held Magnetic Grasper Device and external Magnetic Controller; used in laparoscopic cholecystectomy. Grasper consists of detachable grasper and delivery/retrieval shaft; inserted via ≥10 mm laparoscopic port. Once attached to gallbladder tissue, grasper detaches from shaft; external Magnetic Controller placed on patient skin; magnetic field attraction between controller and grasper enables tissue retraction/manipulation. Distance between controller and grasper modulates magnetic attraction strength. Operated by surgeons in OR. Benefits include improved visualization and access to surgical site during minimally invasive procedures.
Clinical Evidence
No new clinical data provided. Substantial equivalence is supported by previously conducted in vivo performance testing and non-clinical performance data (magnetic field characterization, coupling/recoupling ability, biocompatibility, sterility, and human factors validation) from the predicate device.
Technological Characteristics
System consists of a detachable grasper and external magnetic controller. Operates via magnetic field attraction for tissue manipulation. Compatible with ≥10 mm laparoscopic ports. Reusable components require validated reprocessing. Labeling includes magnetic field safe zones and ferromagnetic screening checklists.
Indications for Use
Indicated for patients (BMI 20-34 kg/m2) undergoing laparoscopic cholecystectomy to grasp and retract the gallbladder body and fundus for improved surgical site access and visualization.
Regulatory Classification
Identification
A magnetic surgical instrument system is a prescription device used in laparoscopic surgical procedures consisting of several components, such as surgical instruments, and a magnetic controller. The magnetic controller is provided separately from the surgical instrument and is used outside the patient. The external magnetic controller is magnetically coupled with the internal surgical instrument(s) at the surgical site to grasp, hold, retract, mobilize, or manipulate soft tissue and organs.
Special Controls
In combination with the general controls of the FD&C Act, the Magnetic Surgical Instrument System is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) In vivo performance data must demonstrate that the device performs as intended under anticipated conditions of use. Testing must demonstrate the ability of the device to grasp, hold, retract, mobilize, or manipulate soft tissue and organs.
(2) Non-clinical performance data must demonstrate that the system performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Magnetic field strength testing characterization to identify the distances from the magnet that are safe for patients and users with ferromagnetic implants, devices, or objects.
(ii) Ability of the internal surgical instrument(s) to be coupled, de-coupled, and re-coupled with the external magnet over the external magnet use life.
(3) The patient-contacting components of the device must be demonstrated to be biocompatible.
(4) Performance data must demonstrate the sterility of the device components that are patient-contacting.
(5) Methods and instructions for reprocessing reusable components must be validated.
(6) Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components and device functionality over the labeled shelf life.
(7) Training must be developed and validated by human factors testing and analysis to ensure users can follow the instructions for use to allow safe use of the device.
(8) Labeling must include:
(i) Magnetic field safe zones.
(ii) Instructions for proper device use.
(iii) A screening checklist to ensure that all patients and operating staff are screened from bringing ferromagnetic implants, devices, or objects near the external magnet.
(iv) Reprocessing instructions for any reusable components.
(v) Shelf life.
(vi) Use life.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W O66-G609 Silver Spring, MD 20993-0002
June 6, 2017
Levita Magnetics International Corp. % Ms. Cindy Domecus Domecus Consulting Services LLc 1171 Barroilhet Drive Hillsborough, California 94010
Re: K171429
Trade/Device Name: Magnetic Surgical System Regulation Number: 21 CFR 21 CFR §878.4815 Regulation Name: Magnetic surgical instrument system Regulatory Class: Class II Product Code: PNL Dated: May 11, 2017 Received: May 15, 2017
Dear Ms. Domecus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-
Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized human figure with three faces in profile, representing the department's focus on health and well-being.
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely, Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K171429
Device Name Magnetic Surgical System
Indications for Use (Describe)
The Magnetic Surgical System is designed to grasp and the fundus of the gallbladder in laparoscopic cholecystectory procedures to facilitate access and visualization of the surgical site. The device is indicated for use in patients within a BMI range of 20 to 34 kg/m2.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-----------------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Special 510(k) Summary (21 CFR § 807.92(c))
#### I: SUBMITTER INFORMATION
| • Submitter: | Levita Magnetics International Corporation<br>1730 S Amphlett Blvd, Suite 240<br>San Mateo, CA 94402 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| • Contact: | Cindy Domecus, R.A.C. (US & EU)<br>Principal, Domecus Consulting Services LLC<br>DomecusConsulting@comcast.net<br>(650) 343-4813 |
| Date Summary Prepared: | 04 June 17 |
#### II: SUBJECT DEVICE INFORMATION
| Device Trade Name: | Magnetic Surgical System (Model No.: FGA1051 and FGA1055) |
|----------------------------|-----------------------------------------------------------|
| Common Name: | Magnetic Surgical System |
| Classification Name: | Magnetic Surgical Instrument System |
| | (21 CFR §878.4815) |
| Product Code: | PNL |
| Regulatory Classification: | Class II |
#### III: PREDICATE DEVICE INFORMATION
Predicate Device: Magnetic Surgical System (DEN150007)
No reference devices were used in this submission.
#### IV: DEVICE DESCRIPTION
The Magnetic Surgical System is composed of two hand-held instruments: a Magnetic Grasper Device and an external Magnetic Controller. The Magnetic Grasper Device is comprised of two main components: a Detachable Grasper and a Delivery/Retrieval Shaft. Once the Magnetic Grasper Device is inserted and the Grasper is attached to the desired tissue, the Detachable Grasper can be detached from the Delivery/Retrieval Shaft and controlled externally using the Magnetic Controller. Traction of the tissue is maintained through the magnetic field attraction between the Grasper and the Magnetic Controller. The Magnetic Grasper Device is compatible with a ≥ 10 mm laparoscopic port. The Magnetic Controller is a single unit with handles that is held external to the body and emits a magnetic field that attracts the Detachable Grasper. Once the Detachable Grasper is attached to the desired tissue and detached from the Delivery/Retrieval Shaft, the Magnetic Controller is placed external to the body to magnetically attract the Grasper to manipulate the target tissue. Adjusting the distance between the Magnetic Controller and the Detachable Grasper will modulate the magnetic attraction used for tissue retraction/manipulation.
#### V: INDICATIONS FOR USE
The Magnetic Surgical System is designed to grasp and retract the body and the fundus of the gallbladder in laparoscopic cholecystectomy procedures to facilitate access and visualization of the surgical site. The device is indicated for use in patients within a BMI range of 20 to 34 kg/m².
{4}------------------------------------------------
## VI: MODIFICATIONS OF SUBJECT DEVICE
No modifications are being made to the subject device or the cleared indications for use. This 510(k) is being submitted for a labeling change to the Instructions for Use. Specifically, the subject change is for an update regarding the use life statement.
## VII:. DESIGN CONTROL ACTIVITIES
A review of the Risk Analysis previously conducted in support of DEN 150007 was performed to assess the impact of the proposed modification to the use life statement. This assessment included a review of the hazard analysis, design FMEA and use FMEA. The review confirmed that the proposed change does not adversely change the risk profile of the device. The anticipated hazards and harms remain adequately addressed by the controls already in place and no additional testing was deemed necessary.
## VIII: SPECIAL CONTROLS
The Magnetic Surgical System is subject to the special controls described in §21 CFR 878.4815. These special controls are stated below. Compliance with these requirements has been met as noted in the italicized text below each requirement.
- (1) In vivo performance data must demonstrate that the device performs as intended under anticipated conditions of use. Testing must demonstrate the ability of the device to grasp, hold, retract, mobilize or manipulate soft tissue and organs.
The modifications that are the subject of this submission do not change the performance characteristics of the device. Therefore, previously conducted in vivo performance testing that demonstrated that the device performs as intended under the anticipated conditions of use applies here.
- (2) Non-clinical performance data must demonstrate that the system performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
- (a) Magnetic field strength testing characterization to identify the distances from the magnet that are safe for patients and users with ferromagnetic implants, devices or objects.
- (b) Ability of the internal surgical instrument(s) to be coupled, and recoupled with the external magnet over the external magnet use life.
The modifications that are the subject of this submission do not change the performance characteristics of the device. Therefore, previously conducted nonclinical performance testing that demonstrated that the device performs as intended under the anticipated conditions of use applies here.
- (3) The patient-contacting components of the device must be demonstrated to be biocompatible.
The modifications that are the subject of this submission do not change any aspect of the device. Therefore, previously conducted testing that demonstrated that the device is biocompatible applies here.
{5}------------------------------------------------
- (4) Performance data must demonstrate the sterility of the device components that are patient contacting.
The modifications that are the subject of this submission do not change any aspect of the device. Therefore, previously conducted testing that demonstrated the sterility of the device applies here.
- (5) Methods and instructions for reprocessing reusable components must be validated.
The modifications that are the subject of this submission do not change any aspect of the device. Therefore, previously conducted testing that validated the methods and instructions for reprocessing reusable components applies here.
- (6) Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components and device functionality over the labeled shelf life.
The modifications that are the subject of this submission do not change any aspect of the device. Therefore, previously conducted testing that validated the shelf life of the device applies here.
- (7) Training must be developed and validated by human factors testing and analysis to ensure users can follow the instructions for use to allow safe use of the device.
The modifications that are the subject of this submission do not change any aspect of the device. Therefore, previously conducted human factors testing which validated that users can follow the instructions to allow safe use of the device applies here.
- (8) Labeling must include:
- (a) Magnetic field safe zones.
- (b) Instructions for proper device use.
- (c) A screening checklist to ensure that all patients and operating staff are screened from bringing ferromagnetic implants, devices or objects near the external magnet.
- (d) Reprocessing instructions for any reusable components.
- (e) Shelf life.
- (f) Use life.
The labeling complies with the special controls stated above.
### IX: CONCLUSION
Modification of the use life statement has no impact on the safety and effectiveness of the Magnetic Surgical System. As such, the modified device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.