K171348 · C.R. Bard, Inc. · ITX · May 22, 2017 · Radiology
Device Facts
Record ID
K171348
Device Name
Pinpoint GT Needle Guide Kits
Applicant
C.R. Bard, Inc.
Product Code
ITX · Radiology
Decision Date
May 22, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1570
Device Class
Class 2
Indications for Use
Pinpoint™ GT Needle Guide Kits are intended to be used with Site~Rite® Ultrasound Systems. Pinpoint™ GT Needle Guides provide guidance for a needle to intersect an ultrasound beam at a fixed distance below the skin to assist the medical practitioner in placing the tip of a needle in a specific structure. Site~Rite® Probe Covers sheathe the transducer and isolate a site of surgical penetration from microbial and other contamination.
Device Story
Pinpoint™ GT Needle Guide Kits are sterile, single-use accessory kits for Site~Rite® Ultrasound Systems; used by medical practitioners to facilitate ultrasound-guided needle placement. Kits include needle guides (0.5cm–3.5cm depths), conductive gel, probe covers (48" or 96"), and elastic bands. Needle guides feature insertion funnels and stabilizing notches to guide needles to a fixed depth below the skin; probe covers isolate the transducer from the surgical site. Device provides mechanical guidance; no electronic processing or software involved. Clinical benefit includes improved accuracy and sterility during percutaneous needle procedures.
Clinical Evidence
Bench testing only. No clinical data. Verification and validation included needle guide depth accuracy, insertion/removal force testing, needle retention, and packaging integrity. Biocompatibility testing performed per ISO 10993 (cytotoxicity, irritation, sensitization, systemic toxicity). Sterilization validated per ISO 11135.
Technological Characteristics
Materials: Low-Density Polyethylene (needle guides), Linear Low-Density Polyethylene (colorant base). Components: Needle guides with insertion funnels/stabilizing notches, conductive gel, polyethylene probe covers, elastic bands. Sterilization: Ethylene Oxide (ISO 11135). Packaging: ISO 11607 compliant. No electronic or software components.
Indications for Use
Indicated for use with Site~Rite® Ultrasound Systems to provide needle guidance for intersecting an ultrasound beam at a fixed depth below the skin, assisting medical practitioners in needle tip placement within specific structures, and providing a sterile barrier for the transducer.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
Predicate Devices
Site~Rite® Needle Guide Kits and Probe Cover Kit (K042445)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 22, 2017
C.R. Bard, Inc. Bard Access Systems, Inc % Mr. Christopher Phillips Regulatory Affairs Associate Manager 650 North 5600 West SALT LAKE CITY UT 84116
Re: K171348
Trade/Device Name: Pinpoint™ GT Needle Guide Kits Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: ITX Dated: May 5, 2017 Received: May 8, 2017
Dear Mr. Phillips:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. D'Hara
For
Robert Ochs, Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K171348
Device Name Pinpoint™ GT Needle Guide Kits
Indications for Use (Describe)
Pinpoint™ GT Needle Guide Kits are intended to be used with Site~Rite® Ultrasound Systems.
Pinpoint™ GT Needle Guides provide guidance for a needle to intersect an ultrasound beam at a fixed distance below the skin to assist the medical practitioner in placing the tip of a needle in a specific structure.
Site~Rite® Probe Covers sheathe the transducer and isolate a site of surgical penetration from microbial and other contamination.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for BARD Access Systems. The word "BARD" is in large, bold, sans-serif font. Below the word "BARD" is the phrase "ACCESS SYSTEMS" in a smaller, sans-serif font. The logo is black and white.
## 510(k) Summary 21 CFR 807.92(a)
| General<br>Provisions | Submitter Name:<br>Address:<br>Contact Person:<br>Telephone Number:<br>Fax Number:<br>Date of Preparation: | Bard Access Systems, Inc.<br>605 North 5600 West<br>Salt Lake City, UT 84116<br>Christopher M. Phillips<br>801.522.5000 ext 5965<br>801.522.5091<br>05 May 2017 |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject<br>Device Kits | Trade Name:<br>Common Name:<br>Regulation Name:<br>Product Code:<br>Regulation Number:<br>Regulatory Class:<br>Classification Panel: | Pinpoint™ GT Needle Guide Kits<br>Transducer, Ultrasonic, Diagnostic<br>Diagnostic Ultrasonic Transducer<br>ITX<br>21 CFR 892.1570<br>Class II<br>Radiology |
| Predicate<br>Device Kits | Trade Name:<br>Common Name:<br>Regulation Name:<br>Premarket Notification:<br>Product Code:<br>Regulation Number:<br>Regulatory Class:<br>Manufacturer:<br>Classification Panel: | Site~Rite® Needle Guide Kits and Probe Cover Kit<br>Transducer, Ultrasonic, Diagnostic<br>Diagnostic Ultrasound Transducer<br>K042445 (cleared Oct. 19, 2004)<br>ITX<br>21 CFR 892.1570<br>Class II<br>Bard Access Systems, Inc.<br>Radiology |
{4}------------------------------------------------
Pinpoint™ GT Needle Guide Kits consist of the following single use sterile disposables: a set of Pinpoint™ GT Needle Guides, conductive gel, a Site~Rite® Probe Cover, and elastic bands to secure the probe cover to the ultrasound probe.
A summary of the sterile, single-use ultrasound accessories included in subject device kits is included in the Table below.
| | Summary of Sterile, Single-Use Ultrasound Accessories included in<br>Subject Device Kits (Pinpoint™ GT Needle Guide Kits) | |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Description | Shallow Pinpoint™ GT<br>Needle Guide Kit with a<br>96" Site~Rite® Probe<br>Cover | A set of shallow Pinpoint™ GT Needle Guides<br>(0.5cm, 1.0cm, 1.5cm, and 2.0cm depths)*, Conductive gel, A 96" 1 mil Site~Rite® Probe Cover*, and Elastic bands |
| | Deep Pinpoint™ GT<br>Needle Guide Kit with a<br>96" Site~Rite® Probe<br>Cover | A set of deep Pinpoint™ GT Needle Guides<br>(2.5cm, 3.0cm, and 3.5cm depths)*, Conductive gel, A 96" 1 mil Site~Rite® Probe Cover*, and Elastic bands |
| | Shallow Pinpoint™ GT<br>Needle Guide Kit with a<br>48" Site~Rite® Probe<br>Cover | A set of shallow Pinpoint™ GT Needle Guides<br>(0.5cm, 1.0cm, 1.5cm, and 2.0cm depths)*, Conductive gel, A 48" 1 mil Site~Rite® Probe Cover, and Elastic bands |
| | Deep Pinpoint™ GT<br>Needle Guide Kit with a<br>48" Site~Rite® Probe<br>Cover | A set of deep Pinpoint™ GT Needle Guides<br>(2.5cm, 3.0cm, and 3.5cm depths)*, Conductive gel, A 48" 1 mil Site~Rite® Probe Cover, and Elastic bands |
*Indicates a change from the predicate device kits.
| Intended Use | Pinpoint™ GT Needle Guide Kits are intended to be used with Site~Rite®<br>Ultrasound Systems. |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications For<br>Use | Pinpoint™ GT Needle Guide Kits are intended to be used with Site~Rite®<br>Ultrasound Systems. |
| | Pinpoint™ GT Needle Guides provide guidance for a needle to intersect an<br>ultrasound beam at a fixed distance below the skin to assist the medical<br>practitioner in placing the tip of a needle in a specific structure. |
| | Site~Rite® Probe Covers sheathe the transducer and isolate a site of surgical<br>penetration from microbial and other contamination. |
{5}------------------------------------------------
Technological Characteristics of the subject device kits. Pinpoint™ GT Needle Guide Kits, are substantially equivalent to the basic design and function to those of the cited predicate device kits. The differences do not alter the intended use of the subject device kits and do not raise new questions of equivalence.
Modifications made to the subject device kits when compared to the predicate device kits are as follows:
- (1) New packaging configuration:
- New packaging configuration for the subject device kits (Pinpoint™ ● GT Needle Guide Kits).
- (2) Modified needle quide offering:
- New subject Pinpoint™ GT Needle Guides will be offered in the ● subject device kits (Pinpoint™ GT Needle Guide Kits). The subject Pinpoint™ GT Needle Guides are modified from the predicate Site~Rite® Needle Guides (which were included in predicate device kits), and include a needle insertion funnel and stabilizing notches. The subject Pinpoint™ GT Needle Guides also utilize different colorant formulations than the predicate Site~Rite® Needle Guides, though they are comprised of the same base materials (Low-Density Polyethylene) and the same colorant additive base materials (Linear Low-Density Polyethylene) in the same blend.
- (3) Additional probe cover offering:
- In addition to the predicate 48" 1 mil Site~Rite® probe cover (which ● were included in predicate device kits), new subject 96" 1 mil Site~Rite® Probe Cover will be offered in the subject device kits (Pinpoint™ GT Needle Guide Kits). The subject 96" 1 mil Site~Rite® Probe Cover is modified from the predicate 48" long, 1 mill Site~Rite® Probe Cover, and includes dimensional changes along with a different material formulation.
- (4) Labeling updates:
- The Indications for Use statement for the subject device kit has been ● clarified from that of the predicate device kits to reflect Pinpoint™ GT and Site~Rite® branding, and
- An IFU and labels were generated for the subject device kits with ● content modified from the predicate device kits' IFU and labels, to provide clarity and to ensure safe and effective use of the subject device kit.
The differences identified above between the subject and predicate device kits do not alter the intended use of the subject device kits and do not raise any new questions of equivalence.
Technological Characteristics
{6}------------------------------------------------
Verification and validation tests were designed and performed in accordance with Design Controls as per 21 CFR §820.30. The following tests were conducted per guidance documents and standards in conjunction with internal test protocols to determine appropriate methods for evaluating the performance of the device:
### Validation of Pinpoint™ GT Needle Guide Use
- Loading the Needle Guide ●
- Force to Insert a Needle
- Force to Remove a Needle
### Verification Testing of the Pinpoint™ GT Needle Guide Kits and of the Ultrasound Accessories contained therein, including Pinpoint™ GT Needle Guides and 96" 1 mil Site~Rite® Probe Covers
- Needle Guide Depth Accuracy ●
- Needle Guide Depth Distinction and Material Compatibility
- Needle Guide ID Dimension ●
- Needle Guide Damage during Needle Insertion ●
- Needle Guide Removal Force
- Force to Remove a Needle from a Needle Guide ●
- Needle Retention ●
- Force to Insert a Needle ●
- Packaging Integrity ●
- Kit Component Integrity ●
- Packaging Minimum Seal Width ●
- Packaging Sterile Barrier Integrity ●
#### Safety & Performance Tests
The following guidance documents and standards were used to determine appropriate methods for evaluating the performance of the device:
- AAMI / ANSI / ISO 10993-5:2009/(R)2014, Biological Evaluation Of ● Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
- AAMI / ANSI ST72:2011, Bacterial Endotoxins Test Methods, Routine ● Monitoring, And Alternatives To Batch Testing
- ASTM D4332-14, Standard Practice For Conditioning Containers, ● Packages, Or Packaging Components For Testing
- . ASTM F1980-16, Standard Guide For Accelerated Aging Of Sterile Barrier Systems For Medical Devices
- ISO 10993-1: 2009, Biological Evaluation of Medical Devices; Part 1 -● Evaluation and Testing
- ISO 10993-7:2008, Biological Evaluation of Medical Devices Part 7: ● Ethylene Oxide Sterilization Residuals
- ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation Of ● Medical Devices - Part 10: Tests For Irritation And Skin Sensitization
- ISO 10993-11 Second Edition 2006-08-15, Biological Evaluation Of ● Medical Devices - Part 11: Tests For Systemic Toxicity
- ISO 10993-12 Fourth Edition 2012-07-01, Biological Evaluation Of Medical Devices - Part 12: Sample Preparation And Reference Materials
- ISO 11135: 2014. Sterilization of healthcare products Ethylene Oxide ● - Requirements For Development, Validation And Routine Control Of A Sterilization
{7}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| Safety &<br>Performance<br>Tests | ISO 11607-1 First Edition 2006-04-15, Packaging For Terminally<br>Sterilized Medical Devices - Part 1: Requirements For Materials,<br>Sterile Barrier Systems And Packaging Systems [Including:<br>Amendment 1 (2014)] ISO 11607-2 First Edition 2006-04-15, Packaging For Terminally<br>Sterilized Medical Devices - Part 2: Validation Requirements For<br>Forming, Sealing And Assembly Processes [Including: Amendment 1<br>(2014)] USP 39-NF34:2016, <85>, Bacterial Endotoxins Test USP 39-NF34:2016, <161>, Transfusion And Infusion Assemblies<br>And Similar Medical Devices The subject device kits, Pinpoint™ GT Needle Guide Kits, met all<br>predetermined acceptance criteria derived from the above listed tests and<br>demonstrated substantially equivalent performance as compared to the cited<br>predicate device kits.<br>Risk management, including failure mode and effects analysis (FMEA) of the<br>subject device kits (and the ultrasound accessories included therein) was<br>conducted in accordance with BS EN ISO 14971:2012, Medical Devices –<br>Application of risk management to medical devices. |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of<br>Substantial<br>Equivalence | Based on the intended use, technological characteristics, and performance<br>testing, the subject device kits, Pinpoint™ GT Needle Guide Kits, meet the<br>requirements that are considered sufficient for their intended use and<br>demonstrate substantial equivalence to the cited predicate device kits. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.