← Product Code [OWN](/productcode/OWN) · K171345

# Cellvizio 100 Series Systems with Confocal Miniprobes (K171345)

_Mauna Kea Technologies · OWN · Aug 4, 2017 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K171345

## Device Facts

- **Applicant:** Mauna Kea Technologies
- **Product Code:** [OWN](/productcode/OWN.md)
- **Decision Date:** Aug 4, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Cellvizio® 100 Series Systems with Confocal Miniprobes™ are confocal laser systems with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues.

## Device Story

CelioFlex™ UHD Confocal Miniprobes™ are fiber-optic probes used with Cellvizio® 100 Series systems for real-time imaging of internal tissue microstructure. Probes feature a distal 'grabule' designed for manipulation by laparoscopic or robotic grasping accessories. During robotic-assisted procedures, the surgeon uses robotic graspers to position the probe tip against target tissue. The system provides visual feedback to the physician on a monitor, aiding in the assessment of tissue during surgery. The device is intended for use in clinical settings by qualified physicians. The primary benefit is the ability to perform in-vivo microscopic visualization of tissues during minimally invasive procedures.

## Clinical Evidence

No clinical data. Evidence consists of bench testing, including biocompatibility, reprocessing efficacy, laser/electrical safety, and imaging quality. Additional validation testing for robotic-assisted use included 100 grasp/release cycles and 40 bend/flexibility cycles using representative robotic grasping tools, confirming no loss of imaging performance or structural integrity.

## Technological Characteristics

Confocal laser imaging system using fiber optic probes. Features a distal 'grabule' for robotic/laparoscopic grasper compatibility. Compatible with trocars ≥ 5 mm. System performs optical sectioning via laser scanning. Biocompatible materials. Reprocessing-resistant design.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- CelioFlex™ UHD 5 ([K151593](/device/K151593.md))
- CelioFlex™ UHD 5-C ([K160416](/device/K160416.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 7. Premarket Notification 510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

# 510(k) Number:

# Applicant Information:

| Date Prepared:    | May 5, 2017                                                  |
|-------------------|--------------------------------------------------------------|
| Name:             | Mauna Kea Technologies                                       |
| Address:          | 9 Rue d'Enghien<br>F-75010 Paris, France                     |
| Phone:            | +33 1 48 24 03 45                                            |
| Fax:              | +33 1 48 24 12 18                                            |
| Contact Person:   | Michael A. Daniel, Consultant<br>madaniel@clinregconsult.com |
| Phone Number:     | (415) 407-0223                                               |
| Office:           | (775) 392-2970                                               |
| Facsimile Number: | (610) 545-0799                                               |

# Device Information:

| Device Trade Name:       | CelioFlex™ UHD Confocal Miniprobes to be used with |
|--------------------------|----------------------------------------------------|
|                          | Cellvizio® 100 Series Systems                      |
| Common Name:             | Endoscope and Accessories                          |
| Classification Name(s):  | Confocal Optical Imaging                           |
| Product Code/Regulation: | OWN/GCJ 21 CFR876.1500                             |
| Classification:          | Class II                                           |

# Predicate devices:

The CelioFlex™ UHD 5 has been cleared with K151593

The CelioFlex™ UHD 5-C has been cleared with K160416

# Device Description:

CelioFlex™ UHD Confocal Miniprobes™ are used with Cellvizio® 100 Series systems to provide imaging of the internal microstructure of body cavities, organs, and canals through contact of their distal tip with the tissues. They are designed for use during laparoscopic surgical procedures, including robot-assisted procedures.

CelioFlex™ UHD Confocal Miniprobes™ have a "grabule" (previously cleared in K151593 and K160416) on their distal tip. The "grabule" has been designed as a grip for laparoscopic grasping accessories, facilitating user control of the miniprobe distal tip during conventional laparoscopic procedures while preventing potential wear to the Miniprobe tip that repeated handling by forceps might cause. The diameter of the "grabule" is fitted for insertion of 

{1}------------------------------------------------

CelioFlex™ UHD Confocal Miniprobes™ through trocars with diameter larger or equal to 5 mm.

CelioFlex™ UHD Confocal Miniprobes are also compatible with laparoscopy robot-assisted procedures. Robotic arm graspers can be used to manipulate CelioFlex™ UHD Confocal Miniprobes.

During robotic-asisted procedures, CelioFlex™ UHD Confocal Miniprobes™ are inserted through a pre-installed trocar. When the CelioFlex™ UHD distal end appears on the monitor. the surgeon uses the robotic accessory to grasp the distal part of the Confocal Miniprobe™ or "grabule", with the laparoscopic grasper to position the Confocal Miniprobe-TM facing the tissue to be imaged.

# Intended Use:

Cellvizio® 100 Series Systems with Confocal Miniprobes™ are confocal laser systems with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues.

# Indications for Use:

CelioFlex™ UHD Confocal Miniprobes™ are intended to be used by qualified physicians to provide visualization of body cavities, organs, and canals during endoscopic and laparoscopic and laparoscopic surgical procedures, including robotic-assisted procedures.

# Comparison to previous Devices:

CelioFlex™M UHD Confocal Miniprobes™ are already cleared for conventional laparoscopic procedures and have been thoroughly tested for this indication (K151593).

The subject devices are strictly identical to the predicate devices in terms of design, technological characteristics, and operating principle. The performances of the subject devices are thus unchanged in terms of imaging performances, biocompatibility, reprocessing, and laser safety, electrical safety and EMC when used with the Cellvizio® 100 Series systems. The Intended Use of the subject devices also remains the same as that of predicate devices.

Robotic accessories are more and more commonly used during laparoscopic procedures and the only difference in robot-assisted procedures is that the Confocal Miniprobe is handled by robotic grasping accessories instead of hand-held grasping accessories.

# Testing Completed:

The table below summarizes the verification and validation testing performed on predicate devices (CelioFlex™ UHD 5 and CelioFlex™ UHD 5-C Confocal Miniprobes™) that have been provided in previous submissions K151593 and K160416 respectively.

{2}------------------------------------------------

| Test performed                                                                                                                       | Acceptance Criteria  | Test Results |
|--------------------------------------------------------------------------------------------------------------------------------------|----------------------|--------------|
| Biocompatibility (Cytotoxicity, Sensitization, Irritation or<br>Intracutaneous reactivity & Systemic toxicity)                       | All requirements met | Pass         |
| Resistance to reprocessing methods: mechanical resistance,<br>tensile strength assessment, functional testing, image quality<br>etc. | All requirements met | Pass         |
| Efficacy of reprocessing methods                                                                                                     | All requirements met | Pass         |
| Laser safety                                                                                                                         | All requirements met | Pass         |
| Imaging quality                                                                                                                      | All requirements met | Pass         |
| Compatibility of the CelioFlex™ UHD Confocal Miniprobes with<br>Trocars commonly used during laparoscopic procedures.                | All requirements met | Pass         |

The potential risks associated with the use of CelioFlex™ UHD Confocal Miniprobes™ during robot-assisted laparoscopy procedures have been assessed and additional testing with representative worst case robot grasping tools was performed to validate that the CelioFlex™ UHD Confocal Miniprobes™ can be used during robot-assisted laparoscopic procedures without change of safety, performances and increase of residual risks. The result from these tests, summarized in the table below, have demonstrated that the use of CelioFlexI™ UHD Confocal Miniprobes™ during robot-assisted laparoscopic procedures does not change performance or safety and effectiveness, and does not affect image quality and does not change the global residual risk.

| Test performed                                                                                      | Acceptance Criteria                                                                                                                           | Result |
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Grasp and Release Test with grasping<br>robotic accessory (100 handling cycles with<br>8 samples)   | The CelioFlexTM UHD Confocal Miniprobes must complete the<br>testing without loss of imaging performance or broken and/or<br>loose components | Pass   |
| Bend and Flexibility Test with grasping<br>robotic accessory (40 handling cycles with 8<br>samples) | The CelioFlexTM UHD Confocal Miniprobes must complete the<br>testing without loss of imaging performance or broken and/or<br>loose components | Pass   |

# Summary:

Based upon the Intended Use, Indications for Use, product technical information, performances, safety and effectiveness, and biocompatibility provided in this submission and in the submissions of the previously cleared CelioFlex UHD 5 (K151593) and CelioFlex UHD 5-C (K160416), CelioFlex™ UHD Confocal Miniprobes™ used in robotic-assisted surgery procedures are substantially equivalent to predicate devices. The subject devices can safely and effectively be used to visualize and image the internal microstructure of tissues during endoscopic and laparoscopic surgical procedures, including robotic-assisted procedures.

{3}------------------------------------------------

# Please wait...

If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.

You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.

For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.

Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 4, 2017

Mauna Kea Technologies % Michael Daniel President Daniel &Daniel Consulting 340 Jones Lane Gardnerville. Nevada 89460

Re: K171345

Trade/Device Name: Cellvizio 100 Series Systems with Confocal Miniprobes Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: OWN, GCJ Dated: May 5, 2017 Received: May 8, 2017

Dear Michael Daniel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{5}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

---

**Source:** [https://fda.innolitics.com/device/K171345](https://fda.innolitics.com/device/K171345)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
