← Product Code [JOY](/productcode/JOY) · K171315

# Advanced RBC Application (K171315)

_CellaVision AB · JOY · Aug 1, 2017 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K171315

## Device Facts

- **Applicant:** CellaVision AB
- **Product Code:** [JOY](/productcode/JOY.md)
- **Decision Date:** Aug 1, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5260
- **Device Class:** Class 2
- **Review Panel:** Hematology
- **Attributes:** AI/ML, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K171315 · Aug 1, 2017 | Advanced RBC Application | CellaVision AB | Routine clinical workflow samples from hospital laboratories | Clinical samples collected from routine hospital laboratory workflows were used to conduct a reproducibility study to evaluate the performance of the Advanced RBC Application across three sites. | Reproducibility; Routine clinical workflow; Hospital laboratory samples |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Reproducibility study; Multi-site reproducibility study based on CLSI EP05-A3; Follow-up/Duration: 5 days | Samples collected from routine workflow from hospital laboratories including normal and elevated levels for each of the 21 RBC morphological characteristics; Number of Sites: 3 | Not applicable for this study | Grading reproducibility and total variance components (within-run, between-run, between-day, between-site) |

## Indications for Use

The CellaVision DM1200 with the Advanced RBC Application is an automated cell-locating device, intended for in-vitro diagnostic use. The CellaVision DM1200 with the Advanced RBC Application automatically locates and presents images of blood cells on peripheral blood smears. The operator identifies and verifies the suggested classification of each cell according to type. The CellaVision DM1200 with the Advanced RBC Application is intended for blood samples that have been flagged as abnormal by an automated cell counter. The CellaVision DM1200 with the Advanced RBC Application is intended to be used by skilled operators, trained in the use of the device and in recognition of blood cells. The CellaVision DM96 with the Advanced RBC Application is an automated cell-locating device, intended for in-vitro diagnostic use. The CellaVision DM96 with the Advanced RBC Application automatically locates and presents images of blood cells on peripheral blood smears. The operator identifies and verifies the suggested classification of each cell according to type. The CellaVision DM96 with the Advanced RBC Application is intended for blood samples that have been flagged as abnormal by an automated cell counter. The CellaVision DM96 with the Advanced RBC Application is intended to be used by skilled operators, trained in the use of the device and in recognition of blood cells.

## Device Story

System consists of slide scanning unit (microscope, camera, feeder) and computer with acquisition/classification software. Input: Romanowsky-stained peripheral blood smears on glass slides. Process: System scans slide, locates cells, and uses artificial neural networks (ANNs) to pre-characterize RBCs into 21 morphological categories based on size, color, shape, and inclusions. Output: Overview images and individual cell images presented on screen; suggested morphological grades (0-3) and relative distribution percentages. Operator reviews, verifies, or modifies classifications; signs results. Used in clinical laboratories by trained operators. Benefits: Automates tedious manual RBC morphology review; provides standardized pre-characterization to assist clinical decision-making regarding anemia or other hematological conditions.

## Clinical Evidence

Multi-site study (5 sites, 1598 samples) compared Advanced RBC Application against manual microscopy and automated cell counters. Primary evaluation (color, shape, inclusion) showed efficiencies of 87.7%, 88.5%, and 94.4% respectively. Secondary evaluation of clinically significant morphologies showed 90.6% efficiency. Precision studies (repeatability/reproducibility) across three sites confirmed consistent grading performance across 21 morphological characteristics. No clinical outcomes data; performance based on analytical agreement with expert manual review.

## Technological Characteristics

System includes slide scanning unit (microscope, camera, feeder) and computer. Software utilizes deterministic artificial neural networks (ANNs) for image classification. Connectivity: LIS integration via barcode. Sterilization: Not applicable (slide preparation is external). Dimensions/Form factor: Desktop analyzer. Software: Embedded module within CellaVision DM Software.

## Regulatory Identification

An automated cell-locating device is a device used to locate blood cells on a peripheral blood smear, allowing the operator to identify and classify each cell according to type. (Peripheral blood is blood circulating in one of the body's extremities, such as the arm.)

## Predicate Devices

- Manual Light Microscope with a Romanowsky stain
- CellaVision DM96 ([K033840](/device/K033840.md))
- CellaVision DM1200 Automated Hematology Analyzer ([K092868](/device/K092868.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM

A. 510(k) Number:
K171315

B. Purpose for Submission:
Seeking clearance for Advanced RBC Application, an optional software module delivered with the CellaVision DM96 and DM1200 systems

C. Manufacturer and Instrument Name:
CellaVision AB; CellaVision DM96 and DM1200 with Advanced RBC Application

D. Type of Test or Tests Performed:
White blood cell (WBC) differential, red blood cell (RBC) characterization, and platelet estimation

E. System Descriptions:

1. Device Description:

CellaVision DM96 and DM1200 Systems

The CellaVision DM96 and DM1200 (hereafter DM systems) both consist of a slide scanning unit (SSU) and a computer system. The SSU consists of a microscope, camera and a slide feeder unit; whereas the computer system contains the acquisition and classification software—CellaVision DM Software (CDMS). The DM96 and DM1200 perform the following functionalities:

- Scan a user-defined area on a slide
- Locate and present images of every located cell or object found on the smear
- Store images and results in a database
- Present an overview image of a user-defined area on a slide
- Create digital slides

Advanced RBC Application

The Advanced RBC Application is an optional software module of the CellaVision DM Software (CDMS) that replaces the existing (standard) RBC functionality of the CellaVision DM Software installed on the CellaVision DM96 and DM1200. General functionality of the Advanced RBC Application includes:

{1}

- Present and suggest 21 morphological characteristics in an overview image of red blood cells on a screen
- Allow operators to confirm or modify the pre-characterization of the red blood cell morphology
- Organize, present, and suggest a morphology for every individual red blood cell image on a screen
- Allow operators to confirm or modify the suggested classification of the red blood cells

The results for the red blood cells (RBCs) are presented as absolute cell counts and proportional counts and classified according to the user-predefined 4-grade scale for each morphological characteristic.

The major difference between the standard RBC functionality and the Advanced RBC Application is that the Advanced RBC Application pre-characterizes RBCs based on size, color, shape, and inclusion using trained artificial neural networks (ANNs) into 21 morphological characteristics whereas the standard RBC functionality pre-characterizes into six morphological characteristics and does not have trained artificial neural networks (ANNs). Additionally, with the Advanced RBC Application, pre-characterized RBCs are displayed individually and in an overview screen image whereas the standard RBC functionality displays an overview screen image only.

## 2. Principles of Operation:

Patient blood smears are wedged and stained outside the device on standard rectangular glass slides by standardized staining methods provided in the labeling. Prepared slides are coded with a barcode label and placed in a barcode-labeled magazine, which is then loaded into the DM system. The DM system identifies the barcode on the magazines and the slides. The DM system requests order information from the Laboratory Information System, using the slide barcode information. If the DM System cannot communicate with the LIS, it defines the type of analysis from the information given by the barcode on the magazine and the predefined analysis settings. When a red blood cell image is ordered, an overview image is captured. The Advanced RBC Application pre-characterizes the RBCs into 21 different morphological characteristics and calculates the relative percentage of distribution for each characteristic. Based on the relative percentages, a grade (0–3) is suggested for each characteristic. The operator views the overview or a panel of individual cell images and either verifies the results or performs a manual re-characterization. The operator can also re-characterize each cell into 10 different user defined morphology characteristics. The operator also has the capability of viewing the cells automatically excluded from analysis. A summary of the red blood cell analysis is presented. The operator signs the slide result with user name and password. The results can be sent to the LIS and/or be printed.

2

{2}

3

3. **Modes of Operation:**

Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?

Yes ☐ X ☐ or No ☐

Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?

Yes ☐ or No ☐ X ☐

4. **Specimen Identification:**

The device requires barcode labeled slides for processing.

5. **Specimen Sampling and Handling:**

A peripheral blood sample collected in K₂EDTA or K₃EDTA tubes is mixed manually by inversion. A thin blood film is wedged on a clean dry glass slide (a blood smear) and stained with Romanowsky stain. Body fluid samples are centrifuged onto a glass slide and stained with Romanowsky stain. The slide preparation occurs separately from the device.

6. **Calibration:**

Not applicable

7. **Quality Control:**

A cell location test is used to verify the slide preparation process and integrity of the system hardware. In addition, the device performs self-tests during the software startup, and at certain points (e.g. slide processing) during the operation of the system.

8. **Software:**

FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types:

Yes ☐ X ☐ or No ☐

F. **Regulatory Information:**

1. **Regulation section:**

21 CFR 864.5260

{3}

2. Classification:
Class II

3. Product code:
JOY

4. Panel:
Hematology (81)

G. Intended Use:

1. Indication(s) for Use:

CellaVision DM1200

The CellaVision DM1200 with the Advanced RBC Application is an automated cell-locating device, intended for in-vitro diagnostic use.

The CellaVision DM1200 with the Advanced RBC Application automatically locates and presents images of blood cells on peripheral blood smears. The operator identifies and verifies the suggested classification of each cell according to type.

The CellaVision DM1200 with the Advanced RBC Application is intended for blood samples that have been flagged as abnormal by an automated cell counter.

The CellaVision DM1200 with the Advanced RBC Application is intended to be used by skilled operators, trained in the use of the device and in recognition of blood cells.

CellaVision DM96

The CellaVision DM96 with the Advanced RBC Application is an automated cell-locating device, intended for in-vitro diagnostic use.

The CellaVision DM96 with the Advanced RBC Application automatically locates and presents images of blood cells on peripheral blood smears. The operator identifies and verifies the suggested classification of each cell according to type.

The CellaVision DM96 with the Advanced RBC Application is intended for blood samples that have been flagged as abnormal by an automated cell counter.

The CellaVision DM96 with the Advanced RBC Application is intended to be used by skilled operators, trained in the use of the device and in recognition of blood cells.

4

{4}

2. Special Conditions for Use Statement(s):

For prescription use only

H. Substantial Equivalence Information:

1. Predicate Device Name(s) and 510(k) numbers:

Manual Light Microscope with a Romanowsky stain, Class I Exempt

CellaVision DM96; K033840

CellaVision DM1200 Automated Hematology Analyzer; K092868

2. Comparison with Predicate Device:

CellaVision DM96 and DM1200 with Advanced RBC Application have the same similarities and differences with respect to the items outlined in the comparison tables below. The comparison tables are completed with representative data from CellaVision DM1200 with Advanced RBC Application.

|  Similarities  |   |   |   |
| --- | --- | --- | --- |
|  Item | CellaVision DM1200 with Advanced RBC Application | CellaVision® DM1200 Automated Hematology Analyzer | Manual light microscopy with Romanowsky stain  |
|  Intended use | DM1200 with the Advanced RBC Application is an automated cell-locating device, intended for in-vitro diagnostic use.
DM1200 with the Advanced RBC Application automatically locates and presents images of blood cells on peripheral blood smears. The operator identifies and verifies the suggested classification of each cell according to type.
DM1200 with the Advanced RBC Application is intended for blood samples that have been flagged as abnormal by an automated cell counter.
DM1200 with the Advanced RBC Application is intended to be used by skilled | DM1200 is an automated cell-locating device.
DM1200 automatically locates and presents images of blood cells on peripheral blood smears. The operator identifies and verifies the suggested classification of each cell according to type.
DM1200 is intended to be used by skilled operators, trained in the use of the device and in recognition of blood cells. | Manual method for cell-locating and identification of red blood cells from peripheral blood smears.
Verification of results by skilled human operator.  |

5

{5}

|  Similarities  |   |   |   |
| --- | --- | --- | --- |
|  Item | CellaVision DM1200 with Advanced RBC Application | CellaVision® DM1200 Automated Hematology Analyzer | Manual light microscopy with Romanowsky stain  |
|   | operators, trained in the use of the device and in recognition of blood cells. |  |   |
|  Specimen type | Peripheral blood | Peripheral blood | Peripheral blood  |
|  Sample preparation | Romanowsky stained blood film on glass slides of peripheral whole blood. | Romanowsky stained blood film on glass slides of peripheral whole blood. | Romanowsky stained peripheral blood film on glass slides.  |
|  Differences  |   |   |   |
| --- | --- | --- | --- |
|  Item | DM1200 with Advanced RBC Application | CellaVision® DM1200 Automated Hematology Analyzer | Manual light microscopic process  |
|  Analysis technique | The device presents an overview image. The cell images are pre-characterized into 21 different morphological characteristics based on size, color, shape and inclusion using segmentation, feature calculation and the deterministic artificial neural networks (ANNs) trained to distinguish between morphology characteristics of red blood cells: - Polychromatic cells; - Hypochromatic cells; - Anisocytosis; - Microcytes; - Macrocytes; - Poikilocytosis; - Target cells; - Schistocytes; - Helmet cells; - Sickle cells; - Spherocytes; - Elliptocytes; - Ovalocytes; - Tear drop cells; - Stomatocytes; - Acanthocytes; - Echinocytes; - Howell-Jolly bodies; | The device presents an overview image. The cell images are pre-characterized into 6 different morphological characteristics based on size, color and shape using segmentation, feature calculation and classification (decision tree): - Polychromatic cells; - Hypochromatic cells; - Anisocytosis; - Microcytes; - Macrocytes; and - Poikilocytosis. The operator can further characterize into 10 morphology characteristics: The operator verifies the suggested morphological characteristics by accepting or re-characterizing | The examiners manually characterize red blood cell morphology based on size, color, shape and inclusion.  |

{6}

|  Differences  |   |   |   |
| --- | --- | --- | --- |
|  Item | DM1200 with Advanced RBC Application | CellaVision® DM1200 Automated Hematology Analyzer | Manual light microscopic process  |
|   | - Pappenheimer bodies;
- Basophilic stippling; and
- Parasites.

The operator can further characterize into 10 user defined characteristics.

The operator verifies the suggested morphological characteristics by accepting or re-characterizing. |  |   |

I. Special Control/Guidance Document Referenced (if applicable):

CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition

CLSI EP12-A2 User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline – Second Edition

CLSI EP17-A2 Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition

CLSI H20-A2 Reference Leukocyte Differential Count (Proportional) and Evaluation of Instrumental Method; Approved Standard – Second Edition

General Principles of Software Validation; Final Guidance for industry and FDA Staff, January 11, 2002

Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005

In Vitro Diagnostic devices: Guidance for the Preparation of 510(k) Submissions, Food and Drug Administration, January 1997

Guidance for Industry, FDA Reviewer and Compliance on Off-The-Shelf Use in Medical Device, September 9, 1999

Guidance for Industry – Cybersecurity for Networked Medical Devices Containing Off-the-Shelf (OTS) Software, January 14, 2005

Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, October 2, 2014

{7}

Statistical Guidance on Reporting Results from Studies Evaluating Diagnostic Tests, March 13, 2007

## J. Performance Characteristics:

### 1. Analytical Performance:

#### a. Accuracy:

Method comparison studies were performed at five clinical sites including two U.S. sites and three sites outside the U.S. The Advanced RBC Application on CellaVision DM96 and CellaVision DM1200 was compared with manual microscopy (reference method) and results from automated differential cell counters. An automated differential cell counter was used at four of the five clinical sites to characterize the RBC size (i.e., macrocytes, microcytes, or anisocytosis). At three of the five sites both the automated differential cell counter and manual microscopy were used for the comparison study. At the remaining two sites, one site used manual microscopy as the comparison method and the other site used the cell counter method as the comparison method to evaluate group size. For one of the five sites, an expert operator reviewed all images and then judged the morphology for each class by overriding (or confirming) the suggested grading. The results for the remaining sites were achieved by re-characterization of individual cell images into appropriate cell characteristics, thereby altering the suggested grading. At each study site, venous whole blood was collected in K₃EDTA Vacutainer tubes. From each sample one slide was prepared. A total of 1598 samples were collected and tested for RBC characterization on DM96 and DM1200 at five different laboratories. The study population consisted of 47.3% males and 52.4% females including pediatric (i.e. neonates, infants, children, and adolescents) and adult subjects up to 101 years of age. The gender for 0.3% of the study population (5 samples) was reported as unknown. In addition to healthy individuals, subjects with the following clinical conditions were included in the studies: iron deficiency anemia, thalassemia, megaloblastic anemia, myelodysplastic syndrome (including sideroblastic anemia and myelofibrosis), hemolytic anemia, sickle cell anemia, disseminated intravascular coagulation, thrombotic microangiopathy (thrombotic thrombocytopenic purpura, hemolytic uremic syndrome), post-splenectomy and malaria.

The evaluation was divided into two parts, a primary evaluation at group level and a secondary evaluation for individual morphologies. The secondary evaluation was in turn divided into two sub-levels as described below:

- Primary evaluation of the groups color, size, shape and inclusions
- Secondary evaluation of individual morphologies:
- Part 1: Evaluation of the group of clinically significant morphologies and parasites: schistocytes (+helmet cells), target cells, sickle cells, tear drop cells and parasites.
- Part 2: Evaluation of the remaining individual morphologies: polychromatic cells, hypochromatic cells, anisocytosis, microcytes, macrocytes,

{8}

poikilocytosis, spherocytes, elliptocytes, ovalocytes, stomatocytes, acanthocytes, echinocytes, Howell-Jolly bodies, Pappenheimer bodies and basophilic stippling.

To determine agreement between the Advanced RBC Application and manual microscopy reference method, the efficiency, analytical sensitivity and analytical specificity were calculated along with the 95% confidence intervals for color, size and shape. Comparison between the DM system with Advanced RBC Application and the automated differential cell counter method for size was provided as overall agreement, positive percent agreement (PPA) and negative percent agreement (NPA) along with the 95% confidence interval. The results met the predefined acceptance criteria.

## Primary evaluation on group level – all sites

|  Morphological group | N | Efficiency (95% CI) | Sensitivity (95% CI) | Specificity (95% CI)  |
| --- | --- | --- | --- | --- |
|  Color | 1181 | 87.7% (85.7% - 89.5%) | 82.1% (78.0% - 85.6%) | 90.5% (88.3% - 92.4%)  |
|  Shape | 1181 | 88.5% (86.5% - 90.2%) | 86.6% (83.3% - 89.3%) | 89.9% (87.4% - 91.9%)  |
|  Inclusion | 1181 | 94.4% (93.0% - 95.6%) | 87.1% (80.3% - 91.8%) | 95.3% (93.9% - 96.4%)  |
|  Morphological group | N | Overall agreement (95% CI) | PPA (95% CI) | NPA (95% CI)  |
| --- | --- | --- | --- | --- |
|  Size¹ | 948 | 81.3% (78.7% - 83.7%) | 80% (76.3% - 83.2%) | 82.9% (79.1% - 86.2%)  |

¹Data from site #1 not included because manual microscopy was used at this site.

## Summary of Secondary Evaluations – clinically significant morphologies – all sites

|  Morphological group | N | Efficiency (95% CI) | Sensitivity (95% CI) | Specificity (95% CI)  |
| --- | --- | --- | --- | --- |
|  Clinically significant morphologies + Parasites | 1181 | 90.6% (88.8% - 92.1%) | 85% (81.3% - 88.1%) | 93.6% (91.7% - 95.1%)  |

## b. Precision/Reproducibility:

Precision performance using the Advanced RBC Application on CellaVision DM96 and CellaVision DM1200 was evaluated by performing repeatability and reproducibility studies across three clinical laboratories.

## Repeatability

The repeatability study was performed at three clinical laboratories. The samples included both normal and elevated levels for each of the 21 RBC morphological characteristics included in the Advanced RBC Application. At two of the three sites, six samples were run on DM96 and seven samples were run on DM1200, respectively. At the third site, seven samples were run on DM96 and eight samples

{9}

were run on DM1200. From each sample, two slides were prepared as replicates. Slides were run two times a day with two replicates per run for 20 days on both DM96 and DM1200. The study was designed to provide repeatability data for both qualitative results (e.g. grade 0, 1, 2 and 3) and quantitative results (e.g. proportional count) for each morphological characteristic. The grading agreement was calculated for each slide and for each sample by determining the relation between the occurrence of true grade and the total number of runs for each morphological characteristic. The results met the predefined acceptance criteria. Representative data from one site is presented below.

# Summary of the repeatability data for each of the 21 RBC morphological characteristics obtained using CellaVision DM96 with the Advanced RBC Application

Polychromatic cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT42 | 1 | 79 | 0 | 0 | 1 | 99%  |
|  ReT43 | 68 | 12 | 0 | 0 | 0 | 85%  |
|  ReT44 | 35 | 45 | 0 | 0 | 1 | 56%  |
|  ReT45 | 0 | 0 | 80 | 0 | 2 | 100%  |
|  ReT46 | 0 | 80 | 0 | 0 | 1 | 100%  |

Hypochromatic Cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT42 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT43 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT44 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT45 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT46 | 30 | 50 | 0 | 0 | 1 | 63%  |

Anisocytosis

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 100% | 0 | 100%  |
|  ReT42 | 0 | 80 | 0 | 100% | 1 | 100%  |
|  ReT43 | 0 | 30 | 50 | 63% | 2 | 63%  |
|  ReT44 | 0 | 80 | 0 | 100% | 1 | 100%  |
|  ReT45 | 0 | 0 | 36 | 55% | 3 | 55%  |
|  ReT46 | 31 | 49 | 0 | 61% | 1 | 61%  |

{10}

Microcytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT42 | 0 | 80 | 0 | 0 | 1 | 100%  |
|  ReT43 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT44 | 69 | 11 | 0 | 0 | 0 | 86%  |
|  ReT45 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT46 | 0 | 80 | 0 | 0 | 1 | 100%  |

Macrocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT42 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT43 | 0 | 0 | 0 | 80 | 3 | 100%  |
|  ReT44 | 16 | 64 | 0 | 0 | 1 | 80%  |
|  ReT45 | 0 | 0 | 80 | 0 | 2 | 100%  |
|  ReT46 | 80 | 0 | 0 | 0 | 0 | 100%  |

Poikilocytosis

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 0 | 80 | 0 | 0 | 1 | 100%  |
|  ReT42 | 0 | 0 | 80 | 0 | 2 | 100%  |
|  ReT43 | 0 | 80 | 0 | 0 | 1 | 100%  |
|  ReT44 | 0 | 19 | 61 | 0 | 2 | 76%  |
|  ReT45 | 0 | 79 | 1 | 0 | 1 | 99%  |
|  ReT46 | 0 | 0 | 80 | 0 | 2 | 100%  |

Target cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT42 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT43 | 0 | 80 | 0 | 0 | 1 | 100%  |
|  ReT44 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT45 | 6 | 74 | 0 | 0 | 1 | 93%  |
|  ReT46 | 80 | 0 | 0 | 0 | 0 | 100%  |

Schistocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT42 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT43 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT44 | 71 | 9 | 0 | 0 | 0 | 89%  |
|  ReT45 | 0 | 80 | 0 | 0 | 1 | 100%  |
|  ReT46 | 80 | 0 | 0 | 0 | 0 | 100%  |

{11}

Helmet cells*

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT42 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT43 | 77 | 3 | 0 | 0 | 0 | 96%  |
|  ReT44 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT45 | 79 | 1 | 0 | 0 | 0 | 99%  |
|  ReT46 | 80 | 0 | 0 | 0 | 0 | 100%  |

Schistocytes+Helmet cells*

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT42 | 60 | 20 | 0 | 0 | 0 | 75%  |
|  ReT43 | 58 | 22 | 0 | 0 | 0 | 73%  |
|  ReT44 | 34 | 46 | 0 | 0 | 1 | 58%  |
|  ReT45 | 0 | 77 | 3 | 0 | 1 | 96%  |
|  ReT46 | 80 | 0 | 0 | 0 | 0 | 100%  |

*Helmet cells can be relevant if Schistocytes are present in the sample; therefore, the results of all three are presented

Sickle cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT42 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT43 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT44 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT45 | 0 | 0 | 22 | 58 | 3 | 73%  |
|  ReT46 | 80 | 0 | 0 | 0 | 0 | 100%  |

Spherocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT42 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT43 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT44 | 30 | 50 | 0 | 0 | 1 | 63%  |
|  ReT45 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT46 | 79 | 1 | 0 | 0 | 0 | 99%  |

Elliptocytes*

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT42 | 0 | 10 | 65 | 5 | 2 | 81%  |
|  ReT43 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT44 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT45 | 79 | 1 | 0 | 0 | 0 | 99%  |
|  ReT46 | 0 | 80 | 0 | 0 | 1 | 100%  |

{12}

Ovalocytes*

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT42 | 0 | 70 | 10 | 0 | 1 | 88%  |
|  ReT43 | 24 | 56 | 0 | 0 | 1 | 70%  |
|  ReT44 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT45 | 1 | 79 | 0 | 0 | 1 | 99%  |
|  ReT46 | 0 | 38 | 42 | 0 | 2 | 53%  |

Elliptocytes+Ovalocytes*

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT42 | 0 | 0 | 59 | 21 | 2 | 74%  |
|  ReT43 | 9 | 71 | 0 | 0 | 1 | 89%  |
|  ReT44 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT45 | 0 | 80 | 0 | 0 | 1 | 100%  |
|  ReT46 | 0 | 12 | 68 | 0 | 2 | 85%  |

*The results for Elliptocytes and Ovalocytes are presented separately but were also evaluated together due to the fact that the names of these morphologies are used interchangeably in the daily practice.

Tear drop cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT42 | 0 | 0 | 0 | 80 | 3 | 100%  |
|  ReT43 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT44 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT45 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT46 | 0 | 0 | 0 | 80 | 3 | 100%  |

Stomatocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT42 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT43 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT44 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT45 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT46 | 0 | 80 | 0 | 0 | 1 | 100%  |

Acanthocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT42 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT43 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT44 | 0 | 58 | 22 | 0 | 1 | 73%  |
|  ReT45 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT46 | 80 | 0 | 0 | 0 | 0 | 100%  |

{13}

14

# Echinocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 0 | 63 | 17 | 0 | 1 | 79%  |
|  ReT42 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT43 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT44 | 0 | 37 | 43 | 0 | 2 | 54%  |
|  ReT45 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT46 | 80 | 0 | 0 | 0 | 0 | 100%  |

# Howell-Jolly bodies

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 0 | 37 | 43 | 0 | 2 | 54%  |
|  ReT42 | 43 | 37 | 0 | 0 | 0 | 54%  |
|  ReT43 | 59 | 21 | 0 | 0 | 0 | 74%  |
|  ReT44 | 4 | 13 | 63 | 0 | 2 | 79%  |
|  ReT45 | 2 | 5 | 73 | 0 | 2 | 91%  |
|  ReT46 | 80 | 0 | 0 | 0 | 0 | 100%  |

# Pappenheimer bodies

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT42 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT43 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT44 | 2 | 75 | 3 | 0 | 1 | 94%  |
|  ReT45 | 0 | 0 | 0 | 80 | 3 | 100%  |
|  ReT46 | 80 | 0 | 0 | 0 | 0 | 100%  |

# Basophilic stippling

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT42 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT43 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT44 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT45 | 79 | 0 | 1 | 0 | 0 | 99%  |
|  ReT46 | 80 | 0 | 0 | 0 | 0 | 100%  |

# Parasites

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT41 | 0 | 79 | 1 | 0 | 1 | 99%  |
|  ReT42 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT43 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT44 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT45 | 79 | 1 | 0 | 0 | 0 | 99%  |
|  ReT46 | 80 | 0 | 0 | 0 | 0 | 100%  |

{14}

# Summary of the repeatability data for each of the 21 RBC morphological characteristics obtained using CellaVision DM1200 with the Advanced RBC Application

Polychromatic cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 0 | 80 | 0 | 0 | 1 | 100%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 6 | 74 | 0 | 0 | 1 | 93%  |
|  ReT54 | 79 | 1 | 0 | 0 | 0 | 99%  |
|  ReT55 | 0 | 0 | 62 | 18 | 2 | 78%  |
|  ReT56 | 80 | 0 | 0 | 0 | 0 | 100%  |

Hypochromatic Cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT54 | 68 | 12 | 0 | 0 | 0 | 85%  |
|  ReT55 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT56 | 0 | 0 | 0 | 80 | 3 | 100%  |

Anisocytosis

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 0 | 36 | 44 | 0 | 2 | 55%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 0 | 80 | 0 | 0 | 1 | 100%  |
|  ReT54 | 0 | 0 | 80 | 0 | 2 | 100%  |
|  ReT55 | 0 | 0 | 13 | 67 | 3 | 84%  |
|  ReT56 | 0 | 3 | 77 | 0 | 2 | 96%  |

Microcytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 0 | 80 | 0 | 0 | 1 | 100%  |
|  ReT54 | 0 | 80 | 0 | 0 | 1 | 100%  |
|  ReT55 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT56 | 0 | 1 | 78 | 1 | 2 | 98%  |

{15}

Macrocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 0 | 2 | 78 | 0 | 2 | 98%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT54 | 0 | 80 | 0 | 0 | 1 | 100%  |
|  ReT55 | 0 | 0 | 80 | 0 | 2 | 100%  |
|  ReT56 | 80 | 0 | 0 | 0 | 0 | 100%  |

Poikilocytosis

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 0 | 16 | 51 | 13 | 2 | 64%  |
|  ReT52 | 3 | 77 | 0 | 0 | 1 | 96%  |
|  ReT53 | 0 | 1 | 79 | 0 | 2 | 99%  |
|  ReT54 | 0 | 30 | 50 | 0 | 2 | 63%  |
|  ReT55 | 0 | 27 | 53 | 0 | 2 | 66%  |
|  ReT56 | 1 | 79 | 0 | 0 | 1 | 99%  |

Target cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 58 | 22 | 0 | 0 | 0 | 73%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT54 | 10 | 70 | 0 | 0 | 1 | 88%  |
|  ReT55 | 0 | 80 | 0 | 0 | 1 | 100%  |
|  ReT56 | 80 | 0 | 0 | 0 | 0 | 100%  |

Schistocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 79 | 1 | 0 | 0 | 0 | 99%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT54 | 0 | 0 | 51 | 29 | 2 | 64%  |
|  ReT55 | 79 | 1 | 0 | 0 | 0 | 99%  |
|  ReT56 | 0 | 0 | 80 | 0 | 2 | 100%  |

Helmet cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 78 | 2 | 0 | 0 | 0 | 98%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT54 | 79 | 1 | 0 | 0 | 0 | 99%  |
|  ReT55 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT56 | 79 | 1 | 0 | 0 | 0 | 99%  |

{16}

Schistocytes+Helmet cells*

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 37 | 43 | 0 | 0 | 1 | 54%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT54 | 0 | 0 | 21 | 59 | 3 | 74%  |
|  ReT55 | 78 | 2 | 0 | 0 | 0 | 98%  |
|  ReT56 | 0 | 0 | 73 | 7 | 2 | 91%  |

*Helmet cells can be relevant if Schistocytes are present in the sample; therefore, the results of all three are presented

Sickle cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT54 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT55 | 0 | 0 | 3 | 77 | 3 | 96%  |
|  ReT56 | 80 | 0 | 0 | 0 | 0 | 100%  |

Spherocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 78 | 2 | 0 | 0 | 0 | 98%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 79 | 1 | 0 | 0 | 0 | 99%  |
|  ReT54 | 5 | 75 | 0 | 0 | 1 | 94%  |
|  ReT55 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT56 | 79 | 1 | 0 | 0 | 0 | 99%  |

Elliptocytes*

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 0 | 80 | 0 | 0 | 1 | 100%  |
|  ReT54 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT55 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT56 | 0 | 80 | 0 | 0 | 1 | 100%  |

Ovalocytes*

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 16 | 64 | 0 | 0 | 1 | 80%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 0 | 0 | 80 | 0 | 2 | 100%  |
|  ReT54 | 3 | 77 | 0 | 0 | 1 | 96%  |
|  ReT55 | 0 | 76 | 4 | 0 | 1 | 95%  |
|  ReT56 | 79 | 1 | 0 | 0 | 0 | 99%  |

{17}

Elliptocytes+Ovalocytes*

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 15 | 65 | 0 | 0 | 1 | 81%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 0 | 0 | 76 | 4 | 2 | 95%  |
|  ReT54 | 0 | 80 | 0 | 0 | 1 | 100%  |
|  ReT55 | 0 | 65 | 15 | 0 | 1 | 81%  |
|  ReT56 | 0 | 80 | 0 | 0 | 1 | 100%  |

*The results for Elliptocytes and Ovalocytes are presented separately but were also evaluated together due to the fact that the names of these morphologies are used interchangeably in the daily practice.

Tear drop cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 0 | 0 | 1 | 79 | 3 | 99%  |
|  ReT54 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT55 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT56 | 0 | 70 | 10 | 0 | 1 | 88%  |

Stomatocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT54 | 79 | 1 | 0 | 0 | 0 | 99%  |
|  ReT55 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT56 | 80 | 0 | 0 | 0 | 0 | 100%  |

Acanthocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT54 | 0 | 80 | 0 | 0 | 1 | 100%  |
|  ReT55 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT56 | 80 | 0 | 0 | 0 | 0 | 100%  |

{18}

19

# Echinocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 1 | 21 | 52 | 6 | 2 | 65%  |
|  ReT52 | 1 | 69 | 10 | 0 | 1 | 86%  |
|  ReT53 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT54 | 43 | 37 | 0 | 0 | 0 | 54%  |
|  ReT55 | 79 | 1 | 0 | 0 | 0 | 99%  |
|  ReT56 | 80 | 0 | 0 | 0 | 0 | 100%  |

# Howell-Jolly bodies

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 0 | 0 | 80 | 0 | 2 | 100%  |
|  ReT52 | 3 | 62 | 15 | 0 | 1 | 78%  |
|  ReT53 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT54 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT55 | 1 | 75 | 4 | 0 | 1 | 94%  |
|  ReT56 | 78 | 2 | 0 | 0 | 0 | 98%  |

# Pappenheimer bodies

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT54 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT55 | 0 | 0 | 0 | 80 | 3 | 100%  |
|  ReT56 | 80 | 0 | 0 | 0 | 0 | 100%  |

# Basophilic stippling

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT52 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT53 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT54 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT55 | 79 | 0 | 1 | 0 | 0 | 99%  |
|  ReT56 | 80 | 0 | 0 | 0 | 0 | 100%  |

# Parasites

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReT51 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT52 | 0 | 79 | 1 | 0 | 1 | 99%  |
|  ReT53 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT54 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT55 | 80 | 0 | 0 | 0 | 0 | 100%  |
|  ReT56 | 80 | 0 | 0 | 0 | 0 | 100%  |

{19}

20

# Reproducibility

The reproducibility study was performed at three sites with samples collected from routine workflow that included normal and elevated levels for each of the 21 RBC morphological characteristics. From each sample, three slides were prepared. The slides were then run at each site two times a day for 5 days. Seven blood samples were used for the DM96 study and five samples were used for the DM1200 study. The grading reproducibility was calculated by determining the relation between the occurrence of true grade and the total number of runs. The results met the predefined acceptance criteria.

## Summary of the reproducibility data for each of the 21 RBC morphological characteristics obtained with the Advanced RBC Application using CellaVision DM96 (all 3 sites)

### Polychromatic cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 4 | 86 | 0 | 0 | 1 | 96%  |
|  ReR02 | 1 | 33 | 55 | 1 | 2 | 61%  |
|  ReR03 | 83 | 6 | 1 | 0 | 0 | 92%  |
|  ReR04 | 0 | 0 | 90 | 0 | 2 | 100%  |
|  ReR05 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR06 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR07 | 66 | 24 | 0 | 0 | 0 | 73%  |

### Hypochromatic Cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR02 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR03 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR04 | 0 | 23 | 67 | 0 | 2 | 74%  |
|  ReR05 | 0 | 1 | 11 | 78 | 3 | 87%  |
|  ReR06 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR07 | 90 | 0 | 0 | 0 | 0 | 100%  |

### Anisocytosis

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 0 | 90 | 0 | 0 | 1 | 100%  |
|  ReR02 | 0 | 38 | 52 | 0 | 2 | 58%  |
|  ReR03 | 44 | 46 | 0 | 0 | 1 | 51%  |
|  ReR04 | 0 | 0 | 0 | 90 | 3 | 100%  |
|  ReR05 | 0 | 10 | 80 | 0 | 2 | 89%  |
|  ReR06 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR07 | 0 | 90 | 0 | 0 | 1 | 100%  |

{20}

Microcytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 0 | 90 | 0 | 0 | 1 | 100%  |
|  ReR02 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR03 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR04 | 0 | 90 | 0 | 0 | 1 | 100%  |
|  ReR05 | 0 | 20 | 70 | 0 | 2 | 78%  |
|  ReR06 | 84 | 6 | 0 | 0 | 0 | 93%  |
|  ReR07 | 90 | 0 | 0 | 0 | 0 | 100%  |

Macrocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR02 | 0 | 0 | 0 | 90 | 3 | 100%  |
|  ReR03 | 0 | 0 | 0 | 90 | 3 | 100%  |
|  ReR04 | 0 | 5 | 85 | 0 | 2 | 94%  |
|  ReR05 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR06 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR07 | 0 | 90 | 0 | 0 | 1 | 100%  |

Poikilocytosis

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 0 | 0 | 90 | 0 | 2 | 100%  |
|  ReR02 | 0 | 59 | 31 | 0 | 1 | 66%  |
|  ReR03 | 23 | 67 | 0 | 0 | 1 | 74%  |
|  ReR04 | 0 | 0 | 66 | 24 | 2 | 73%  |
|  ReR05 | 14 | 76 | 0 | 0 | 1 | 84%  |
|  ReR06 | 0 | 90 | 0 | 0 | 1 | 100%  |
|  ReR07 | 0 | 2 | 87 | 1 | 2 | 97%  |

Target cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR02 | 0 | 90 | 0 | 0 | 1 | 100%  |
|  ReR03 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR04 | 24 | 66 | 0 | 0 | 1 | 73%  |
|  ReR05 | 89 | 1 | 0 | 0 | 0 | 99%  |
|  ReR06 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR07 | 90 | 0 | 0 | 0 | 0 | 100%  |

{21}

Schistocytes®

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 0 | 78 | 12 | 0 | 1 | 87%  |
|  ReR02 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR03 | 87 | 3 | 0 | 0 | 0 | 97%  |
|  ReR04 | 3 | 87 | 0 | 0 | 1 | 97%  |
|  ReR05 | 0 | 83 | 7 | 0 | 1 | 92%  |
|  ReR06 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR07 | 85 | 5 | 0 | 0 | 0 | 94%  |

Helmet cells®

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR02 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR03 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR04 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR05 | 87 | 3 | 0 | 0 | 0 | 97%  |
|  ReR06 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR07 | 90 | 0 | 0 | 0 | 0 | 100%  |

Schistocytes+Helmet cells®

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 0 | 77 | 13 | 0 | 1 | 86%  |
|  ReR02 | 82 | 8 | 0 | 0 | 0 | 91%  |
|  ReR03 | 46 | 44 | 0 | 0 | 0 | 51%  |
|  ReR04 | 30 | 59 | 1 | 0 | 1 | 66%  |
|  ReR05 | 0 | 60 | 30 | 0 | 1 | 67%  |
|  ReR06 | 89 | 1 | 0 | 0 | 0 | 99%  |
|  ReR07 | 64 | 26 | 0 | 0 | 0 | 71%  |

®Helmet cells can be relevant if Schistocytes are present in the sample; therefore, the results of all three are presented

Sickle cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR02 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR03 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR04 | 0 | 0 | 1 | 89 | 3 | 99%  |
|  ReR05 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR06 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR07 | 90 | 0 | 0 | 0 | 0 | 100%  |

{22}

Spherocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR02 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR03 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR04 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR05 | 82 | 8 | 0 | 0 | 0 | 91%  |
|  ReR06 | 89 | 1 | 0 | 0 | 0 | 99%  |
|  ReR07 | 1 | 89 | 0 | 0 | 1 | 99%  |

Elliptocytes*

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 0 | 30 | 29 | 31 | 2 | 32%  |
|  ReR02 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR03 | 81 | 9 | 0 | 0 | 0 | 90%  |
|  ReR04 | 50 | 40 | 0 | 0 | 0 | 56%  |
|  ReR05 | 6 | 54 | 30 | 0 | 1 | 60%  |
|  ReR06 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR07 | 90 | 0 | 0 | 0 | 0 | 100%  |

Ovalocytes*

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 30 | 35 | 18 | 7 | 1 | 39%  |
|  ReR02 | 30 | 60 | 0 | 0 | 1 | 67%  |
|  ReR03 | 2 | 32 | 56 | 0 | 2 | 62%  |
|  ReR04 | 22 | 41 | 27 | 0 | 1 | 46%  |
|  ReR05 | 35 | 55 | 0 | 0 | 1 | 61%  |
|  ReR06 | 89 | 1 | 0 | 0 | 0 | 99%  |
|  ReR07 | 81 | 9 | 0 | 0 | 0 | 90%  |

Elliptocytes+Ovalocytes*

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 0 | 0 | 37 | 53 | 3 | 59%  |
|  ReR02 | 30 | 60 | 0 | 0 | 1 | 67%  |
|  ReR03 | 0 | 32 | 28 | 30 | 1 | 31%  |
|  ReR04 | 31 | 59 | 0 | 0 | 1 | 66%  |
|  ReR05 | 0 | 60 | 30 | 0 | 1 | 67%  |
|  ReR06 | 89 | 1 | 0 | 0 | 0 | 99%  |
|  ReR07 | 67 | 23 | 0 | 0 | 0 | 74%  |

*The results for Elliptocytes and Ovalocytes are presented separately but were also evaluated together due to the fact that the names of these morphologies are used interchangeably in the daily practice.

{23}

Tear drop cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 0 | 0 | 0 | 90 | 3 | 100%  |
|  ReR02 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR03 | 81 | 9 | 0 | 0 | 0 | 90%  |
|  ReR04 | 81 | 9 | 0 | 0 | 0 | 90%  |
|  ReR05 | 6 | 50 | 34 | 0 | 1 | 56%  |
|  ReR06 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR07 | 90 | 0 | 0 | 0 | 0 | 100%  |

Stomatocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR02 | 85 | 5 | 0 | 0 | 0 | 94%  |
|  ReR03 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR04 | 70 | 20 | 0 | 0 | 0 | 78%  |
|  ReR05 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR06 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR07 | 90 | 0 | 0 | 0 | 0 | 100%  |

Acanthocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 86 | 4 | 0 | 0 | 0 | 96%  |
|  ReR02 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR03 | 89 | 1 | 0 | 0 | 0 | 99%  |
|  ReR04 | 89 | 1 | 0 | 0 | 0 | 99%  |
|  ReR05 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR06 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR07 | 0 | 86 | 4 | 0 | 1 | 96%  |

Echinocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 59 | 31 | 0 | 0 | 0 | 66%  |
|  ReR02 | 59 | 31 | 0 | 0 | 0 | 66%  |
|  ReR03 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR04 | 0 | 50 | 40 | 0 | 1 | 56%  |
|  ReR05 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR06 | 0 | 57 | 33 | 0 | 1 | 63%  |
|  ReR07 | 0 | 14 | 76 | 0 | 2 | 84%  |

{24}

Howell-Jolly bodies

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 88 | 2 | 0 | 0 | 0 | 98%  |
|  ReR02 | 49 | 37 | 4 | 0 | 0 | 54%  |
|  ReR03 | 51 | 32 | 7 | 0 | 0 | 57%  |
|  ReR04 | 0 | 0 | 33 | 57 | 3 | 63%  |
|  ReR05 | 69 | 16 | 5 | 0 | 0 | 77%  |
|  ReR06 | 31 | 38 | 21 | 0 | 1 | 42%  |
|  ReR07 | 5 | 45 | 40 | 0 | 1 | 50%  |

Pappenheimer bodies

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR02 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR03 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR04 | 5 | 25 | 29 | 31 | 2 | 32%  |
|  ReR05 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR06 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR07 | 85 | 5 | 0 | 0 | 0 | 94%  |

Basophilic stippling

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR02 | 86 | 4 | 0 | 0 | 0 | 96%  |
|  ReR03 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR04 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR05 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR06 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR07 | 90 | 0 | 0 | 0 | 0 | 100%  |

Parasites

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR01 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR02 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR03 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR04 | 89 | 1 | 0 | 0 | 0 | 99%  |
|  ReR05 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR06 | 0 | 90 | 0 | 0 | 1 | 100%  |
|  ReR07 | 90 | 0 | 0 | 0 | 0 | 100%  |

{25}

# Summary of the reproducibility data for each of the 21 RBC morphological characteristics obtained with the Advanced RBC Application using CellaVision DM1200 (all 3 sites)

Polychromatic cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 84 | 6 | 0 | 0 | 0 | 93%  |
|  ReR12 | 11 | 79 | 0 | 0 | 1 | 88%  |
|  ReR13 | 0 | 0 | 56 | 34 | 2 | 62%  |
|  ReR14 | 89 | 1 | 0 | 0 | 0 | 99%  |
|  ReR15 | 7 | 72 | 11 | 0 | 1 | 80%  |

Hypochromatic Cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 81 | 9 | 0 | 0 | 0 | 90%  |
|  ReR12 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR13 | 0 | 85 | 5 | 0 | 1 | 94%  |
|  ReR14 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR15 | 90 | 0 | 0 | 0 | 0 | 100%  |

Anisocytosis

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR12 | 0 | 90 | 0 | 0 | 1 | 100%  |
|  ReR13 | 0 | 0 | 0 | 90 | 3 | 100%  |
|  ReR14 | 6 | 84 | 0 | 0 | 1 | 93%  |
|  ReR15 | 0 | 22 | 68 | 0 | 2 | 76%  |

Microcytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR12 | 0 | 90 | 0 | 0 | 1 | 100%  |
|  ReR13 | 25 | 65 | 0 | 0 | 1 | 72%  |
|  ReR14 | 88 | 2 | 0 | 0 | 0 | 98%  |
|  ReR15 | 90 | 0 | 0 | 0 | 0 | 100%  |

Macrocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 59 | 31 | 0 | 0 | 0 | 66%  |
|  ReR12 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR13 | 0 | 1 | 89 | 0 | 2 | 99%  |
|  ReR14 | 1 | 89 | 0 | 0 | 1 | 99%  |
|  ReR15 | 0 | 0 | 0 | 90 | 3 | 100%  |

{26}

Poikilocytosis

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR12 | 0 | 0 | 83 | 7 | 2 | 92%  |
|  ReR13 | 0 | 0 | 49 | 41 | 2 | 54%  |
|  ReR14 | 0 | 22 | 65 | 3 | 2 | 72%  |
|  ReR15 | 0 | 90 | 0 | 0 | 1 | 100%  |

Target cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR12 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR13 | 17 | 73 | 0 | 0 | 1 | 81%  |
|  ReR14 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR15 | 0 | 90 | 0 | 0 | 1 | 100%  |

Schistocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR12 | 0 | 63 | 27 | 0 | 1 | 70%  |
|  ReR13 | 14 | 45 | 12 | 19 | 1 | 50%  |
|  ReR14 | 72 | 18 | 0 | 0 | 0 | 80%  |
|  ReR15 | 87 | 3 | 0 | 0 | 0 | 97%  |

Helmet cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 89 | 1 | 0 | 0 | 0 | 99%  |
|  ReR12 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR13 | 88 | 2 | 0 | 0 | 0 | 98%  |
|  ReR14 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR15 | 88 | 2 | 0 | 0 | 0 | 98%  |

Schistocytes+Helmet cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 87 | 3 | 0 | 0 | 0 | 97%  |
|  ReR12 | 0 | 61 | 29 | 0 | 1 | 68%  |
|  ReR13 | 0 | 58 | 11 | 21 | 1 | 64%  |
|  ReR14 | 49 | 41 | 0 | 0 | 0 | 54%  |
|  ReR15 | 71 | 19 | 0 | 0 | 0 | 79%  |

$^{\text{※}}$ Helmet cells can be relevant if Schistocytes are present in the sample; therefore, the results of all three are presented

{27}

Sickle cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR12 | 88 | 2 | 0 | 0 | 0 | 98%  |
|  ReR13 | 0 | 0 | 1 | 89 | 3 | 99%  |
|  ReR14 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR15 | 90 | 0 | 0 | 0 | 0 | 100%  |

Spherocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR12 | 82 | 8 | 0 | 0 | 0 | 91%  |
|  ReR13 | 85 | 5 | 0 | 0 | 0 | 94%  |
|  ReR14 | 0 | 86 | 4 | 0 | 1 | 96%  |
|  ReR15 | 90 | 0 | 0 | 0 | 0 | 100%  |

Elliptocytes*

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR12 | 0 | 90 | 0 | 0 | 1 | 100%  |
|  ReR13 | 39 | 51 | 0 | 0 | 1 | 57%  |
|  ReR14 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR15 | 90 | 0 | 0 | 0 | 0 | 100%  |

Ovalocytes*

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 89 | 1 | 0 | 0 | 0 | 99%  |
|  ReR12 | 0 | 0 | 81 | 9 | 2 | 90%  |
|  ReR13 | 11 | 79 | 0 | 0 | 1 | 88%  |
|  ReR14 | 65 | 25 | 0 | 0 | 0 | 72%  |
|  ReR15 | 62 | 28 | 0 | 0 | 0 | 69%  |

Elliptocytes+Ovalocytes*

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 86 | 4 | 0 | 0 | 0 | 96%  |
|  ReR12 | 0 | 0 | 31 | 59 | 3 | 66%  |
|  ReR13 | 0 | 83 | 7 | 0 | 1 | 92%  |
|  ReR14 | 51 | 39 | 0 | 0 | 0 | 57%  |
|  ReR15 | 57 | 33 | 0 | 0 | 0 | 63%  |

*The results for Elliptocytes and Ovalocytes are presented separately but were also evaluated together due to the fact that the names of these morphologies are used interchangeably in the daily practice.

{28}

Tear drop cells

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR12 | 0 | 0 | 0 | 90 | 3 | 100%  |
|  ReR13 | 57 | 24 | 9 | 0 | 0 | 63%  |
|  ReR14 | 89 | 1 | 0 | 0 | 0 | 99%  |
|  ReR15 | 87 | 3 | 0 | 0 | 0 | 97%  |

Stomatocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 89 | 1 | 0 | 0 | 0 | 99%  |
|  ReR12 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR13 | 84 | 6 | 0 | 0 | 0 | 93%  |
|  ReR14 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR15 | 73 | 17 | 0 | 0 | 0 | 81%  |

Acanthocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR12 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR13 | 87 | 3 | 0 | 0 | 0 | 97%  |
|  ReR14 | 0 | 87 | 3 | 0 | 1 | 97%  |
|  ReR15 | 90 | 0 | 0 | 0 | 0 | 100%  |

Echinocytes

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR12 | 68 | 22 | 0 | 0 | 0 | 76%  |
|  ReR13 | 0 | 33 | 57 | 0 | 2 | 63%  |
|  ReR14 | 0 | 47 | 43 | 0 | 1 | 52%  |
|  ReR15 | 90 | 0 | 0 | 0 | 0 | 100%  |

Howell-Jolly bodies

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR12 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR13 | 0 | 2 | 70 | 18 | 2 | 78%  |
|  ReR14 | 67 | 22 | 1 | 0 | 0 | 74%  |
|  ReR15 | 87 | 3 | 0 | 0 | 0 | 97%  |

{29}

Pappenheimer bodies

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR12 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR13 | 0 | 0 | 2 | 88 | 3 | 98%  |
|  ReR14 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR15 | 89 | 1 | 0 | 0 | 0 | 99%  |

Basophilic stippling

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR12 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR13 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR14 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR15 | 45 | 45 | 0 | 0 | 0 | 50%  |

Parasites

|  Sample ID | Grade |   |   |   | True grade | Agreement  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3  |   |   |
|  ReR11 | 0 | 0 | 40 | 50 | 3 | 56%  |
|  ReR12 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR13 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR14 | 90 | 0 | 0 | 0 | 0 | 100%  |
|  ReR15 | 90 | 0 | 0 | 0 | 0 | 100%  |

c. Linearity:

Not applicable

d. Carryover:

Not applicable

e. Interfering Substances:

Not applicable

2. Other Supportive Instrument Performance Data Not Covered Above:

Not applicable

K. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

{30}

L. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision

31

---

**Source:** [https://fda.innolitics.com/device/K171315](https://fda.innolitics.com/device/K171315)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
