AR_N Type Implant System

K171297 · Biotem Co., Ltd. · DZE · Jan 31, 2018 · Dental

Device Facts

Record IDK171297
Device NameAR_N Type Implant System
ApplicantBiotem Co., Ltd.
Product CodeDZE · Dental
Decision DateJan 31, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Indications for Use

AR N Type Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. AR N Type Implant System is for two stage surgical procedures. It is intended for delayed load.

Device Story

AR N Type Implant System consists of endosseous dental implants and associated abutments; used to replace missing teeth in mandible or maxilla. Implants are surgically placed in jaw bone; osseointegrate to provide prosthetic anchorage. System includes fixtures (CP Ti Gr 4) with RBM surface treatment and various abutments (cover screws, healing, angled, cemented, solid). Used by dental professionals in clinical settings. Output is a stable foundation for dental restorations; restores chewing function. Benefits include tooth replacement and improved oral function.

Clinical Evidence

Bench testing only. Includes sterilization validation (ISO 11137, ISO 11737), steam sterilization validation (ANSI/AAMI ST79), surface chemical/SEM analysis, comparative fatigue testing (ISO 14801:2007), and biocompatibility (cytotoxicity ISO 10993-5, sensitization ISO 10993-10).

Technological Characteristics

Material: CP Ti Gr 4 (ASTM F67). Surface: RBM (Resorbable Blast Media). Connection: Internal Hexagon. Sterilization: Gamma (fixtures), Steam (abutments). Dimensions: Fixture diameters 3.7-5.1mm, lengths 7.5-15.0mm. Abutment angulations 0°, 15°, 25°.

Indications for Use

Indicated for partially or fully edentulous adult patients requiring single or multiple-unit dental restorations, including cemented, screw-retained, or overdenture restorations, and fixed bridgework support. For two-stage surgical procedures with delayed loading.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. January 31, 2018 Biotem Co., Ltd. c/o Joyce Bang Consultant Provision Consulting Group Inc. 14071 Peyton Dr. #967 Chino Hills, California 91709 Re: K171297 Trade/Device Name: AR N Type Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: November 2, 2017 Received: January 4, 2018 Dear Joyce Bang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Mary S. Runner -S For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K171297 Device Name AR N Type Implant System #### Indications for Use (Describe) AR N Type Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. AR N Type Implant System is for two stage surgical procedures. It is intended for delayed load. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------| | <span style="font-size:100%">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 16. 510(K) SUMMARY #### Submitter: Hong Koo Yeo Biotem Co., Ltd. 57, Sasang-ro , 401 beon-gil , Sasang-gu, Busan , Korea Tel : +82-70-8220-4934 Fax : +82-51-908-8257 #### Official Correspondent/US Agent: Jovce Bang Provision Consulting Group, Inc. 1370 Valley Vista Suite 200 Phone: +1-909-550-0131 EXT.1131 Date Prepared: January 26, 2018 #### Device Information: Device Name: AR_N Type Implant System Classification Name: Endosseous Dental Implant Common Name: Endosseous Dental Implant Classification: Class II Product Code: DZE Subsequent Product Code: NHA Regulation number: 21 CFR 872.3640 #### Predicate devices • CMI IS Implant System (K113554) manufactured by Neobiotech Implant Co., Ltd. #### Indication for use AR N Type Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. AR_N Type Implant System is for two stage surgical procedures. It is intended for delayed loading. #### Device Description The AR_N Type Implant System is a dental implant system made of CP TiGr 4 per ASTM F67, intended to be surgically placed in the bone of the upper or lower jaw arches for loading after a conventional healing period. The implants may be used to replace one or more missing teeth. The surface of the implants has been treated with RBM (Resorbable Blast Media). The AR_N Type Implant System is offered in the following sizes. | Platform | Body Diameter | Total Length (mm) | |----------|---------------|----------------------------------| | Ø 3.7 | Ø 3.5 | 7.5, 8.5, 10.0, 11.5, 13.0, 15.0 | | Ø 4.2 | Ø 4.0 | 7.5, 8.5, 10.0, 11.5, 13.0, 15.0 | | Ø 4.6 | Ø 4.5 | 7.5, 8.5, 10.0, 11.5, 13.0, 15.0 | | Ø 5.1 | Ø 5.0 | 7.5, 8.5, 10.0, 11.5, 13.0, 15.0 | {4}------------------------------------------------ ## Abutment Each product line includes corresponding abutments in multiple designs (Cover Screws, Healing Abutments, Angled Abutments, Cemented Abutments (Hex, Non-hex), and Solid Abutments) for restorations. Small diameter implants and angled abutments are not recommended for the posterior region. They are made of Titanium Grade 4 / ASTM F67, and provided as non-sterile. | Abutments | Connection | Angulation (°) | Sizes | | |-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|-------------------|-----------------------| | | | | Cuff Heights (mm) | Platform diameter (Ø) | | Angled<br>abutment | Assembled with Fixture and<br>Angled Abutment (Hex) by<br>inserting Abutment body with set<br>screw. The connection type is<br>internal Hexagon shape. | 15, 25 | 2, 4 | 4.5~5.8 | | Cemented<br>Abutment<br>(Hex) | Assembled with fixture and<br>cemented abutment (Hex) by<br>inserting abutment body with set<br>screw. The connection type is<br>Internal Hexagon shape | 0 | 1~6 | 4.5~6.5 | | Cemented<br>Abutment<br>(Non-Hex) | Assembled with Fixture and<br>Cemented Abutment (Non-Hex)<br>by inserting Abutment body with<br>set screw. The connection type is<br>internal Non-Hexagon shape. | 0 | 1~6 | 4.5~6.5 | | Healing<br>Abutment | Assembled with Fixture and<br>Healing Abutment by inserting the<br>screw part. And the connection is<br>internal Hexagon shape | 0 | 1~5 | 4.5~6.5 | | Solid<br>Abutment | Assembled with fixture by<br>inserting set screws. The<br>connection type is internal<br>Hexagon shape. | 0 | 1~6 | 4.5~6.5 | | Cover Screw | Assembled with Fixture by<br>inserting screw part.<br>And the connection is internal<br>Hexagon shape. | 0 | 2.7 | 2.8 -3.3 | | Set screw | Connection with abutment with<br>screw | 0 | 10 (length) | 2.1, 2.5 | ## Substantial Equivalence Comparison Chart | | Subject Device | Predicate Device | |---------------|--------------------------|-----------------------| | 510(k) Number | K171297 | K113554 | | Device Name | AR_N Type Implant System | CMI IS Implant System | | Manufacturer | Biotem Co., Ltd. | Neobiotech Co., Ltd. | {5}------------------------------------------------ | | | Principles of Operation | Subject device is an endosseous<br>implants and are integrated in the<br>bone (osseointegration). They are<br>intended to be used for surgical<br>placement in the upper or lower jaw<br>to provide a means for prosthetic<br>attachment for anchoring or<br>supporting tooth replacements to<br>restore chewing function. | Predicate device is an endosseous<br>implants and are integrated in the<br>bone (osseointegration). They are<br>intended to be used for surgical<br>placement in the upper or lower jaw<br>to provide a means for prosthetic<br>attachment for anchoring or<br>supporting tooth replacements to<br>restore chewing function. | |---------------------|--------|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use | | | AR_N Type Implant System is<br>indicated for use in partially or fully<br>edentulous mandibles and maxillae,<br>in support of single or multiple-unit<br>restorations including; cemented<br>retained, screw retained, or<br>overdenture restorations, and<br>terminal or intermediate abutment<br>support for fixed bridgework. AR_N<br>Type Implant System is for two stage<br>surgical procedures. It is intended for<br>delayed load. | The CMI Implant IS System is<br>indicated for use in partially or fully<br>edentulous mandibles and maxillae, in<br>support of single or multiple-unit<br>restorations including; cemented<br>retained, screw retained, or<br>overdenture restorations, and<br>terminal or intermediate abutment<br>support for fixed bridgework. IS<br>System is dedicated for two stage<br>surgical procedures and for<br>immediate loading when there is<br>good primary stability and an<br>appropriate occlusal load. Also,<br>implants with diameters larger than 5<br>min are indicated for molar regions. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Design | Internal Hex Submerged<br>Macro thread | Internal Hex Submerged<br>Macro thread | | | | Material | CP Ti Gr 4<br>ASTM F67 | CP Ti Gr 4<br>ASTM F67 | | | | Fixture Sterilization | Gamma sterilization | Gamma sterilization | | | | Fixture Diameter (mm) | Ø3.7 - 5.1 | Ø3.5 - 8.0 | | | | Fixture Length (mm) | 7.5 - 15.0 | 7.0 - 15.0 | | | | Surface treatment | RBM | RBM | | | | Diameters | Ø4.5 - 6.5mm | Ø3.5 - 6.5 | | | | Lengths | 4.0 - 9.0mm | 4.0-9.0mm | | Abutment | Angled | | 15°, 25° | 15°,25° | | Abutment material | | | CP Ti Gr 4<br>ASTM F67 | CP Ti Gr 4<br>ASTM F67 | | | | | | | | | | Abutment sterilization | Non-sterile | Non-sterile | | Product Code | | | DZE, NHA | DZE, NHA | {6}------------------------------------------------ | | Subject abutments<br>Biotem Product (AR N type) | | | Predicate abutments<br>NEO BIOTECH Product (IS-II type) | | | |-----------------------------------|-------------------------------------------------|----------------------|----------------|---------------------------------------------------------|----------------------|----------------| | | Platform<br>diameter (Ø) | Cuff heights<br>(mm) | Angulation (°) | Platform<br>diameter (Ø) | Cuff heights<br>(mm) | Angulation (°) | | Healing<br>abutment | 4.5~6.5 | 1~5 | 0 | 4.0~6.0 | 2~7 | 0 | | Cemented<br>abutment<br>(hex) | 4.5~6.5 | 1~6 | 0 | 4.5~6.5 | 1~4 | 0 | | Cemented<br>abutment<br>(non-hex) | 4.5~6.5 | 1~6 | 0 | 4.5~6.5 | 1~4 | 0 | | Solid<br>abutment | 4.5~6.5 | 1~6 | 00 | 4.5~6.5 | 1~7 | 0 | | Angled<br>abutment | 4.5~5.8 | 2,4 | 15, 25 | 4.5~5.7 | 2,3,4 | 15, 25 | ## Basis for Substantial Equivalence The AR N Type Implant System has a substantially equivalent intended use as the identified predicate. Both are used for mandible and maxilla endosseous dental implant and accessories. The AR N Type Implant System is similar in fundamental scientific technology to the predicate device in that they all have been designed, manufactured and tested in compliance with FDA's Class II special controls guidance document root-form endosseous dental implants and endosseous dental implant abutments. They all share same internal hexagon submerged connection system. The subject and predicate devices are both bone-level implants that share same tapered body design and cutting edge. The subject and predicate devices are similar in size, same surface treatment and materials. ## Non-Clinical Test data The subject device was tested to evaluate its performance as below. - Sterilization Validation testing for sterile devices (fixtures) has been performed in accordance with ISO 11137, SIO 11737-1& ISO 11737-2 for gamma sterilization - Steam Sterilization validation for non-sterile devices (abutments) has been performed in accordance with ANSI/AAMI ST79. - Surface Characteristics Test Report Chemical and SEM image analyses have been performed to verify that there is no residual after RBM treatment on the fixtures. - Comparative Fatigue testing of the subject device and predicate device has been performed in accordance with ISO 14801:2007. - Cytotoxicity Test performed according to ISO 10993-5:2009 - · Sensitization test performed according to ISO 10993-10:2010 Those tests have been performed to evaluate the substantial equivalence in the surface characteristics compared to the predicate device. The result of the above tests have met the standard, and proved the substantial equivalence with the predicate device. Non-clinical testing consisted of performance of testing in accordance with the FDA guidance "Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments." The results of the non-clinical testing demonstrate that the subject device is substantially equivalent to the predicate device. ## Conclusions Overall, the AR N Type Implant System has the following similarities to the predicate devices: {7}------------------------------------------------ *has the same intended use, *uses the same operating principle, *incorporates the same basic design, *incorporates the same material and the surface treatment. In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, we conclude that the AR_N Type Implant System is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...