← Product Code [BSZ](/productcode/BSZ) · K171292

# A7 Anesthesia System (K171292)

_Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · BSZ · Jan 12, 2018 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K171292

## Device Facts

- **Applicant:** Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
- **Product Code:** [BSZ](/productcode/BSZ.md)
- **Decision Date:** Jan 12, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5160
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The A7 Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The A7 is intended for use by licensed clinicians, for patients requiring anesthesia within a health care facility, and can be used in adult and pediatric (including neonate, infant, child and adolescent) populations.

## Device Story

Continuous flow inhalation gas anesthesia system; delivers anesthetic vapor; provides automatic/manual ventilation. Inputs: O2, N2O, anesthetic agents (Desflurane, Isoflurane, Enflurane, Sevoflurane, Halothane). Monitors: CO2, O2, N2O, agent concentration. Operation: Licensed clinicians in health care facilities. Features: Fresh Flow Optimizer, lung recruitment function, heated breathing circuit, spirometry loops. Output: Anesthetic delivery, ventilation support, real-time gas/pressure monitoring. Clinical decision-making: Clinicians use real-time data to adjust anesthesia depth and ventilation parameters. Benefits: Precise anesthetic delivery and respiratory support for diverse patient populations.

## Clinical Evidence

Bench testing only. Testing included software unit/integration/system testing, performance testing of gas modules, fresh flow optimizer, suction, heating, power, thermal, and cleaning/disinfection. Biocompatibility testing (VOCs, particulates, cytotoxicity, sensitization, irritation, E&L, inorganic gases) and human factors validation were performed. Device conforms to AAMI/ANSI ES60601-1, IEC 60601-1-2, ISO 80601-2-13, ISO 80601-2-55, ASTM F1101-90, and ISO 10079-3.

## Technological Characteristics

Continuous flow gas anesthesia system. Materials: Airway pressure gauge housing material changed. Sensing: O2, CO2, N2O, and anesthetic agent sensors. Connectivity: Data integration with Passport 12M/17M monitors. Control: DSP hardware platform for ventilator control. Modes: VCV, PCV, PCV-VG, SIMV-VC, SIMV-PC, PS, CPAP/PS, APRV. Standards: ISO 80601-2-13, ISO 80601-2-55, IEC 60601-1-2.

## Regulatory Identification

A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).

## Predicate Devices

- A7 Anesthesia System ([K151954](/device/K151954.md))

## Reference Devices

- A5 Anesthesia Delivery System ([K123211](/device/K123211.md))
- Evita XL Ventilator ([K083050](/device/K083050.md))
- GE Datex-Ohmeda Aisys ([K110213](/device/K110213.md))
- GE Datex-Ohmeda Aisys CS2 ([K132530](/device/K132530.md))
- Passport 12M/17M patient monitor ([K170876](/device/K170876.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.

January 12, 2018

Shenzhen Mindray Bio-Medical Electronics Co., Ltd. Yanhong Bai Manager Regulatory Affairs, Technical Regulation Department Mindray Building, Keji 12th Road South Hi-tech Industrial Park. Shenzhen, 518057 China

Re: K171292

Trade/Device Name: A7 Anesthesia System Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine for Anesthesia or Analgesia Regulatory Class: Class II Product Code: BSZ, CCK, NHO, CBS, CBQ, CBR, CCL, NHQ, NHP, KDP Dated: December 8, 2017 Received: December 11, 2017

Dear Yanhong Bai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/SafetyReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Tara A. Ryan -S
DN: c=US, o=U.S. Government, ou=HHS,
Tara A. Ryan -s ou=FDA, ou People, cn=Tara A. Ryan-5,
0.9.2342.19200300.100.1.1=1300030749
Date: 2018.01.12 11:26:05-05'00'

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K171292

Device Name A7 Anesthesia System

Indications for Use (Describe)

The A7 Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation.

The A7 is intended for use by licensed clinicians, for patients requiring anesthesia within a health care facility, and can be used in adult and pediatric (including neonate, infant, child and adolescent) populations.

Type of Use (Select one or both, as applicable)

| <div> <span style="font-size:12px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)    </div> |
|----------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:12px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)    </div>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) SUMMARY

In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the A7 Anesthesia System is provided below:

| Device Common Name:                                 | Gas-Machine, Anesthesia                                                                                                                                                                                                                                                                      |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Proprietary Name:                            | A7 Anesthesia System                                                                                                                                                                                                                                                                         |
| Submitter:                                          | SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>Shenzhen 518057, P.R. China<br>Tel: +86 755 81885635<br>Fax: +86 755 26582680                                                                       |
| Contact:                                            | Yanhong Bai<br>Manager Regulatory Affairs<br>SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>518057, P.R. China<br>Tel: +86 755 81885635<br>Fax: +86 755 26582680<br>E-mail: baiyanhong@mindray.com |
| Date Prepared:                                      | December 6th, 2017                                                                                                                                                                                                                                                                           |
| Panel:                                              | Anesthesiology                                                                                                                                                                                                                                                                               |
| Classification Regulation and Product Code:         |                                                                                                                                                                                                                                                                                              |
| Primary:                                            |                                                                                                                                                                                                                                                                                              |
| 868.5160- BSZ Anesthesia Gas Machine Class II       |                                                                                                                                                                                                                                                                                              |
| Secondary:                                          |                                                                                                                                                                                                                                                                                              |
| 868.1400 – CCK - Carbon Dioxide Gas Analyzer        |                                                                                                                                                                                                                                                                                              |
| 868.1500 – NHO/CBQ/NHQ/NHP - Enflurane gas analyzer |                                                                                                                                                                                                                                                                                              |
| 868.1620 - CBS - Halothane Gas Analyzer             |                                                                                                                                                                                                                                                                                              |
| 868.1700 – CBR - Nitrous Oxide Gas Analyzer         |                                                                                                                                                                                                                                                                                              |
| 868.1720 – CCL- Oxygen Gas Analyzer                 |                                                                                                                                                                                                                                                                                              |
| 880.6740 – KDP- Vacuum Regulator                    |                                                                                                                                                                                                                                                                                              |

{4}------------------------------------------------

| 510(k) number | Trade or proprietor or model<br>name | Manufacturer                                           |
|---------------|--------------------------------------|--------------------------------------------------------|
| K151954       | A7 Anesthesia System                 | Shenzhen Mindray Bio-<br>Medical Electronics Co., Ltd. |

#### Predicate Device:

#### Reference Devices:

| 510(k) number | Trade or proprietor or model<br>name | Manufacturer                                           |
|---------------|--------------------------------------|--------------------------------------------------------|
| K123211       | A5 Anesthesia Delivery System        | Shenzhen Mindray Bio-<br>Medical Electronics Co., Ltd. |
| K083050       | Evita XL Ventilator                  | Dräger Medical AG & Co. KG                             |
| K110213       | GE Datex-OhmedaAisys                 | Datex-Ohmeda Inc.                                      |
| K132530       | GE Datex-OhmedaAisys CS2             | Datex-Ohmeda Inc.                                      |

### Indications for Use:

The A7 Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation.

The A7 is intended for use by licensed clinicians, for patients requiring anesthesia within a health care facility, and can be used in adult and pediatric (including neonate, infant, child and adolescent) populations.

## Device Description:

The A7 Anesthesia System is a continuous flow inhalation gas anesthesia system that delivers anesthetic vapor and provides for automatic and manual modes of ventilation. The A7 incorporates 02, CO2, N2O and Agent concentration monitoring (Desflurane, Isoflurane, Enflurane, Sevoflurane and Halothane).

### Nonclinical testing and Performance:

The following areas have been tested:

- . Software
	- o Unit testing
	- Integration testing o
	- o System testing
- Performance ●
	- o AG Module
	- Fresh Flow Optimizer O
	- Sample Gas Return and CO2-absorbent O

{5}------------------------------------------------

- Suction o
- Heating Module O
- Power Supply O
- Thermal O
- Cleaning and Disinfection o
- Waveform Comparison o
- Biocompatibility ●
	- Volatile Organic Compounds o
	- Particulate Testing O
	- Cytotoxicity O
	- Sensitization O
	- Irritation / intracutaneous reactivity O
	- Extractables and leachables (E&L) testing O
	- Inorganic gases (03, 02, CO2, NO2) testing O
- Human Factors Validation Testing ●
- . Testing as per consensus standards:
	- o AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, ,C1:2009/(R)2012 and A2:2010/(R)2012 Medical electrical equipment - Part 1: General Requirements for basic safety and essential performance
	- IEC 60601-1-2 Edition 3: 2007 Medical electrical equipment Part 1-2: General o requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
	- ISO 80601-2-13 First edition 2011-08-11 Medical electrical equipment Part 2- O 13: Particular requirements for basic safety and essential performance of an anesthetic workstation
	- ISO 80601-2-55 First edition 2011-12-15 Medical electrical equipment -Part 2-55: o Particular requirements for the basic safety and essential performance of respiratory gas monitors
	- ASTM F1101-90 (Reapproved 2003) Standard Specification for Ventilators O Intended for Use During Anesthesia
	- ISO 10079-3:2014 Medical suction equipment Part 3:Suction equipment o powered from a vacuum or positive pressure gas source
	- AIM 7351731 medical electrical equipment and system electromagnetic O immunity test for exposure to radio frequency identification readers - an aim standard. (General II (ES/EMC))

{6}------------------------------------------------

The functional and system level testing showed that the device continues to meet specifications and the performance of the device is equivalent to the predicate.

## Substantial Equivalence:

### Comparison of Indications:

Both the predicate device and the subject A7 Anesthesia System are gas anesthesia machine used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The indications for use of the subject device (A7 Anesthesia System) have been modified to add the definition of the pediatric population subgroup. All of the pediatric population subgroup was cleared in the predicate device A7 Anesthesia System (K151954). In conclusion, the minor changes to the indications for use do not change the intended use of the A7 Anesthesia System.

{7}------------------------------------------------

# Comparison of Technological Characteristics:

The table below compares the key technological feature of the subject device to the predicate device (A7 Anesthesia System (K151954)).

# Device Comparison Table

|                                                                         | Technical<br>Characteristics                                                                                          | Subject Device<br>A7 Anesthesia System<br>(subject device)                                                            | Predicate Device<br>A7 Anesthesia System<br>(K151954) |
|-------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
|                                                                         | Vaporizers                                                                                                            | Two or Three, variable bypass                                                                                         | Two or three, variable bypass                         |
|                                                                         | Agent - Sevoflurane                                                                                                   | Yes                                                                                                                   | Yes                                                   |
|                                                                         | Agent - Isoflurane                                                                                                    | Yes                                                                                                                   | Yes                                                   |
|                                                                         | Agent - Desflurane                                                                                                    | Yes                                                                                                                   | Yes                                                   |
|                                                                         | Agent - Halothane                                                                                                     | Yes                                                                                                                   | Yes                                                   |
|                                                                         | Agent - Enflurane                                                                                                     | Yes                                                                                                                   | Yes                                                   |
|                                                                         | Automatic Ventilator                                                                                                  | Yes                                                                                                                   | Yes                                                   |
|                                                                         | Bellows                                                                                                               | Yes                                                                                                                   | Yes                                                   |
|                                                                         | Bellows Volume                                                                                                        | 1500mL                                                                                                                | 1500mL                                                |
| Ventilation Modes                                                       |                                                                                                                       |                                                                                                                       |                                                       |
|                                                                         | VCV                                                                                                                   | Yes                                                                                                                   | Yes                                                   |
|                                                                         | PCV                                                                                                                   | Yes                                                                                                                   | Yes                                                   |
|                                                                         | PCV – VG                                                                                                              | Yes                                                                                                                   | Yes                                                   |
|                                                                         | SIMV - VC                                                                                                             | Yes                                                                                                                   | Yes                                                   |
|                                                                         | SIMV - PC                                                                                                             | Yes                                                                                                                   | Yes                                                   |
|                                                                         | PS                                                                                                                    | Yes                                                                                                                   | Yes                                                   |
| Specifications                                                          |                                                                                                                       |                                                                                                                       |                                                       |
|                                                                         | Tidal Volume                                                                                                          | Yes                                                                                                                   | Yes                                                   |
|                                                                         | Range, ml                                                                                                             | 20 to 1500                                                                                                            | 20 to 1500                                            |
|                                                                         | Minute Volume                                                                                                         | Yes                                                                                                                   | Yes                                                   |
|                                                                         | Rate, bpm                                                                                                             | 4 to 100 bpm                                                                                                          | 4 to 100 bpm                                          |
|                                                                         | I:E Ratio                                                                                                             | 4:1 to 1:8 with 0.5 increment                                                                                         | 4:1 to 1:8 with 0.5 increment                         |
|                                                                         | Inspiratory Pause                                                                                                     | Off, 5 to 60% of insp. Period                                                                                         | Off, 5 to 60% of insp. Period                         |
|                                                                         | Air Flow Range                                                                                                        | 0 to 15 L/min                                                                                                         | 0 to 15 L/min                                         |
| Fresh Gas                                                               | N2O Flow Range                                                                                                        | 0 to 12 L/min                                                                                                         | 0 to 12 L/min                                         |
|                                                                         | O2 Flow Range                                                                                                         | 0 to 15 L/min                                                                                                         | 0 to 15 L/min                                         |
| Technical<br>Characteristics                                            | Subject Device<br>A7 Anesthesia System<br>(subject device)                                                            | Predicate Device<br>A7 Anesthesia System<br>(K151954)                                                                 |                                                       |
| Individual Gas Flow<br>Accuracy                                         | ±50 ml/min or ±5% of setting<br>value, whichever is greater                                                           | ±50 ml/min or ±5% of setting<br>value, whichever is greater                                                           |                                                       |
| Pressure Limit, cm H₂O                                                  | 0 to 100cm H₂O                                                                                                        | 0 to 100cm H₂O                                                                                                        |                                                       |
| PEEP, cm H₂O                                                            | Off, 3 to 30, 1 cm H₂O increment                                                                                      | Off, 3 to 30, 1 cm H₂O increment                                                                                      |                                                       |
| System Checks                                                           | Auto at start                                                                                                         | Auto at start                                                                                                         |                                                       |
| Airway Pressure  Measured at                                            | Inspiratory                                                                                                           | Inspiratory                                                                                                           |                                                       |
| High/Low Airway Pressure Alarm                                          | Yes                                                                                                                   | Yes                                                                                                                   |                                                       |
| Pressure Limiting Alarm                                                 | Yes                                                                                                                   | Yes                                                                                                                   |                                                       |
| Sub Atmospheric Pressure Alarm                                          | Yes                                                                                                                   | Yes                                                                                                                   |                                                       |
| Continuous Press Alarm                                                  | Yes                                                                                                                   | Yes                                                                                                                   |                                                       |
| Apnea >2 Minute Alarm                                                   | Yes                                                                                                                   | Yes                                                                                                                   |                                                       |
| Apnea Alarm                                                             | Yes                                                                                                                   | Yes                                                                                                                   |                                                       |
| High/Low Minute Volume Alarm                                            | Yes                                                                                                                   | Yes                                                                                                                   |                                                       |
| High/Low O₂ Concentration Alarm                                         | Yes                                                                                                                   | Yes                                                                                                                   |                                                       |
| Heated Breathing Circuit                                                | Yes                                                                                                                   | Yes                                                                                                                   |                                                       |
| Spirometry: Pressure-Volume and<br>Flow-Volume loops                    | Yes                                                                                                                   | Yes                                                                                                                   |                                                       |
| Anesthetic Gas Module Sampling Rate                                     | Adult/pediatric: 120, 150,<br>200mL/min<br>Neonate: 70, 90, 120mL/min                                                 | High volume: 120, 150,<br>200mL/min<br>Low volume: 70, 90, 120mL/min                                                  |                                                       |
| Anesthetic Gas Module Sampling<br>Delay Time:                           | <4 seconds                                                                                                            | <4 seconds                                                                                                            |                                                       |
| Anesthetic Gas Module  Refresh Rate:                                    | 1 second                                                                                                              | 1 second                                                                                                              |                                                       |
| Anesthetic Gas Module  Warm-up<br>Time:                                 | 45 seconds to warm-up status<br>10 minutes to ready-to-measure<br>status                                              | 45 seconds to warm-up status<br>10 minutes to ready-to-measure<br>status                                              |                                                       |
| Anesthetic Gas Module   Accuracy<br>CO₂:                                | 0 to 1%: +/-.1%<br>1 to 5%: +/-.2%<br>5 to 7%: +/-.3%<br>7 to10%: +/-.5%<br>>10%: unspecified                         | 0 to 1%: +/-.1%<br>1 to 5%: +/-.2%<br>5 to 7%: +/-.3%<br>7 to10%: +/-.5%<br>>10%: unspecified                         |                                                       |
| Anesthetic Gas Module   Accuracy<br>N₂O:                                | 0 to 20%: +/-2%<br>20 to 100%: +/-3%                                                                                  | 0 to 20%: +/-2%<br>20 to 100%: +/-3%                                                                                  |                                                       |
| Technical<br>Characteristics                                            | Subject Device<br>A7 Anesthesia System<br>(subject device)                                                            | Predicate Device<br>A7 Anesthesia System<br>(K151954)                                                                 |                                                       |
| Anesthetic Gas Module Accuracy<br>Desflurane:                           | 0 to 1%: +/-.15%<br>1 to 5%: +/-.2%<br>5 to 10%: +/-.4%<br>10 to 15%: +/-.6%<br>15 to 18%: +/-1%<br>>18%: unspecified | 0 to 1%: +/-.15%<br>1 to 5%: +/-.2%<br>5 to 10%: +/-.4%<br>10 to 15%: +/-.6%<br>15 to 18%: +/-1%<br>>18%: unspecified |                                                       |
| Anesthetic Gas Module   Accuracy<br>Sevoflurane:                        | 0 to 1%: +/-.15%<br>1 to 5%: +/-.2%<br>5 to 8%: +/-.4%<br>>8%: unspecified                                            | 0 to 1%: +/-.15%<br>1 to 5%: +/-.2%<br>5 to 8%: +/-.4%<br>>8%: unspecified                                            |                                                       |
| Anesthetic Gas Module   Accuracy<br>Enflurane/Isoflurane/<br>Halothane: | 0 to 1%: +/-.15%<br>1 to 5%: +/-.2%<br>>5%: unspecified                                                               | 0 to 1%: +/-.15%<br>1 to 5%: +/-.2%<br>>5%: unspecified                                                               |                                                       |
| Anesthetic Gas Module Accuracy O2:                                      | 0 to 25%: +/-1%<br>25 to 80%: +/-2%<br>80 to 100%: +/-3%                                                              | 0 to 25%: +/-1%<br>25 to 80%: +/-2%<br>80 to 100%: +/-3%                                                              |                                                       |
| Anesthetic Gas Module   Accuracy<br>awRR:                               | 2 to 60rpm: +/-1rpm<br>>60rpm: unspecified                                                                            | 2 to 60rpm: +/-1rpm<br>>60rpm: unspecified                                                                            |                                                       |
| Anesthetic Gas Module<br>Measurement<br>Rise Time:                      | CO2: <250ms<br>N2O: <250ms<br>O2: ≤500ms<br>Hal/Iso/Sev/Des: ≤300ms<br>Enf: <350ms                                    | CO2: <250ms<br>N2O: <250ms<br>O2: ≤500ms<br>Hal/Iso/Sev/Des: ≤300ms<br>Enf: <350ms                                    |                                                       |
| Measurement Range CO2                                                   | 0 to 30%                                                                                                              | 0 to 30%                                                                                                              |                                                       |
| Measurement Range N2O                                                   | 0 to 100%                                                                                                             | 0 to 100%                                                                                                             |                                                       |
| Measurement Range Des                                                   | 0 to 30%                                                                                                              | 0 to 30%                                                                                                              |                                                       |
| Measurement Range Sev                                                   | 0 to 30%                                                                                                              | 0 to 30%                                                                                                              |                                                       |
| Measurement Range Enf/Iso/Hal                                           | 0 to 30%                                                                                                              | 0 to 30%                                                                                                              |                                                       |
| Measurement Range O2                                                    | 0 to 100%                                                                                                             | 0 to 100%                                                                                                             |                                                       |

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

There are some differences between the subject A7 Anesthesia System and the predicate device A7 Anesthesia System (K151954). Per the FDA Guidance, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" issued July 28, 2014, the following devices are provided to support substantial equivalence.

| Differences                                                                                                                                                                 | Reference Device |                                                          |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|----------------------------------------------------------|
| Add Three Ventilation<br>Modes                                                                                                                                              | SIMV-VG          | Evita XL Ventilator (K083050)                            |
|                                                                                                                                                                             | CPAP/PS          | Aisys anesthesia system (K110213)                        |
|                                                                                                                                                                             | APRV             | Evita XL Ventilator (K083050)                            |
| Lung Recruitment Ventilation Function                                                                                                                                       |                  | GE Datex-Ohmeda Aisys CS2<br>Anesthesia System (K132530) |
| Added O2 Sensor                                                                                                                                                             |                  | A5 Anesthesia Delivery System<br>(K123211)               |
| Redesign the Existing Ventilator Control Board Using<br>DSP Hardware Platform                                                                                               |                  | A7 Anesthesia System (K151954)                           |
| Change about the Anesthetic Gas Module and<br>Accessories                                                                                                                   |                  | A7 Anesthesia System (K151954)                           |
| Get Data from Anesthetic Gas Module plugged into<br>the Passport 12M/17M patient monitor (K170876) to<br>support optimizer function and agent usage calculation<br>function |                  | A7 Anesthesia System (K151954)                           |
| Increase the Maximum Inspiratory Flow                                                                                                                                       |                  | Evita XL Ventilator (K083050)                            |
| Change the Material of the Airway Pressure Gauge<br>Housing                                                                                                                 |                  | A7 Anesthesia System (K151954)                           |

### Substantial Equivalence Conclusion:

Based on the detailed comparison of specifications for each of the modifications to the previously cleared A7 Anesthesia System (K151954) and cited reference devices, the performance testing results and conformance with applicable standards show that the A7 Anesthesia System can be found substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K171292](https://fda.innolitics.com/device/K171292)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
