X-PSI Knee System

K171269 · Orthosoft Inc (D/B/A Zimmer Cas) · JWH · Dec 28, 2017 · Orthopedic

Device Facts

Record IDK171269
Device NameX-PSI Knee System
ApplicantOrthosoft Inc (D/B/A Zimmer Cas)
Product CodeJWH · Orthopedic
Decision DateDec 28, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3560
Device ClassClass 2

Indications for Use

The X-PSI Knee System is indicated as an orthopedic instrument system to assist in the positioning of knee replacement components. It involves surgical planning software used pre-operatively to plan the surgical placement of the components on the basis of provided patient radiological images and 3-D reconstructed bones with identifiable placement anatomical landmarks, and surgical instrument components that include patient specific or customized guides fabricated on the basis of the surgical plan to precisely reference the placement of the implant components intra-operatively per the surgical plan. The X-PSI Knee system is indicated for patients without severe bone deformities, such as a HKA greater than 15° or deformities due to prior fracture of the distal femur or proximal tibia. The X-PSI Knee System is to be used with the following fixed bearing knee replacement systems in accordance with their indications and contraindications: NexGen® CR, NexGen CR-Flex Gender, NexGen LPS, NexGen LPS-Flex, NexGen LPS-Flex Gender, Persona® CR, Persona PS, Vanguard® CR and Vanguard PS. The patient specific guide components are intended for single-use only.

Device Story

System uses patient X-ray images to construct 3D bone models; software suite enables pre-operative surgical planning and design of patient-specific surgical guides (jigs). Guides are 3D-printed via Selective Laser Sintering (SLS) to fit patient-specific bony/articular topography. Intra-operatively, surgeon uses guides to reference placement of femoral and tibial implant components; guides set resection depth, varus/valgus, and slope. System includes 3D bone models for visual reference and reusable surgical instrumentation (stylus, cut blocks). Used in clinical settings by orthopedic surgeons. Output guides assist in precise bone cuts and implant alignment, potentially improving surgical accuracy and patient outcomes. Compatible with specific NexGen, Persona, and Vanguard knee systems.

Clinical Evidence

Non-significant risk clinical study evaluated HKA alignment accuracy by comparing resulting frontal HKA angle to planned HKA angle. Results demonstrated satisfactory performance. Additional bench testing included software system tests, guide mechanical resistance (drop, debris, aging, sterilization), and full system validation using cadaver specimens and bone models to verify cut accuracy (frontal/sagittal angles, rotation, resection depth).

Technological Characteristics

System utilizes 3D-printed (Selective Laser Sintering) patient-specific guides and bone models. Inputs are X-ray DICOM images. Software suite performs triage, 3D reconstruction, landmark acquisition, and surgical planning. Guides are single-use; reusable instrumentation includes femoral stylus and cut blocks. Fixation pins are manual surgical instruments. Compatible with fixed-bearing knee replacement systems.

Indications for Use

Indicated for patients undergoing knee replacement surgery who do not have severe bone deformities (HKA > 15°) or deformities resulting from prior distal femur or proximal tibia fractures.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 28, 2017 Orthosoft Inc. (d/b/a Zimmer CAS) Paul Hardy Senior Specialist, Regulatory Affairs 75 Queen Street. Suite 3300 Montreal, H3C 2N6 Canada Re: K171269 Trade/Device Name: X-PSI Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, OOG, MBH, LLZ Dated: November 17, 2017 Received: November 20, 2017 Dear Paul Hardy: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K171269 Device Name X-PSI Knee System #### Indications for Use (Describe) The X-PSI Knee System is indicated as an orthopedic instrument system to assist in the positioning of knee replacement components. It involves surgical planning software used pre-operatively to plan the surgical placement of the components on the basis of provided patient radiological images and 3-D reconstructed bones with identifiable placement anatomical landmarks, and surgical instrument components that include patient specific or customized guides fabricated on the basis of the surgical plan to precisely reference the placement of the implant components intra-operatively per the surgical plan. The X-PSI Knee system is indicated for patients without severe bone deformities, such as a HKA greater than 15° or deformities due to prior fracture of the distal femur or proximal tibia. The X-PSI Knee System is to be used with the following fixed bearing knee replacement systems in accordance with their indications and contraindications: NexGen® CR, NexGen CR-Flex Gender, NexGen LPS, NexGen LPS-Flex, NexGen LPS-Flex Gender, Persona® CR, Persona PS, Vanguard® CR and Vanguard PS. The patient specific guide components are intended for single-use only. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| | <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Zimmer Biomet • X-PSI Knee System 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)'s, issued on August 12, 2005. | Sponsor: | Orthosoft, Inc. d/b/a Zimmer CAS<br>75 Queen St., Suite 3300<br>Montreal, Quebec, CANADA H3C 2N6<br>Establishment Registration Number: 9617840 | |-----------------|------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Paul Hardy<br>Zimmer Biomet<br>574-372-6799 | | Date: | December 27, 2017 | | Subject Device: | Trade Name: Zimmer Biomet® X-PSI Knee System<br>Common Name: Surgical Knee Guide | | Product Codes: | OOG, JWH, MBH, LLZ | ### Classification Name: - . Knee Joint Patello/Femorotibial Polymer/Metal/Polymer Semi-Constrained Cemented Prosthesis / 21 CFR § 888.3560 / Product Code JWH - . Knee Joint Patello/Femorotibial Polymer/Metal/Polymer Semi-Constrained Cemented Prosthesis / 21 CFR § 888.3560 / Product Code OOG - Prosthesis. Knee, Patello/Femorotibial. Semi-Constrained, uncemented. Porous. Coated, Polymer/Metal / MBI-, Knee Joint Patello/Femorotibial Metal/Polymer Porous-Coated Uncemented Prosthesis / 21 CFR § 888.3565/ Product Code MBH - Picture Archiving and Communications System/ 21 CFR § 892.2050 / Product Code LLZ # Predicate Device(s): | Manufacturer | Device | 510(k)<br>Number | Type | |---------------|------------------------|------------------|-----------| | Zimmer CAS | CAS PSI<br>Knee System | K131409 | Predicate | | ONEFITMedical | Knee EOS | K161828 | Reference | {4}------------------------------------------------ #### Device Description: The present Zimmer Biomet® X-PSI Knee System is an instrumentation system that includes customized surgical guides to mate each patient's bony and articular surface topographies to reference the location and orientation of the implant system's instruments which in turns sets the position and alignment of the femoral and tibial implant components. It involves the following: - A CAS X-PSI Knee Software Suite used in preparation for the surgery to sequentially construct 3-D surface models of each patient's knee joint bony structures and articular surfaces from the patient's X-Ray images, plan the location and orientation of the knee replacement implant components upon the patient's model, and create the corresponding specification models for the patient specific surgical guides (PSI Guides) with surfaces and elements to uniquely fit each patient topographical features and set or reference the placement of the implant system components per the plan, - . The Zimmer Biomet ® X-PSI Guides (also called jigs) that are manufactured per the above models and plan, for intra-operative and single use, which include one to set the placement of the distal femoral cut guides which set the resection depth, the varus/valgus and the flexion of the distal cut and one to set the placement of the tibial cut guides which set the resection depth, the varus/valgus, and the posterior slope of the proximal cut. - . Each patient's Zimmer Biomet ®3-D Bone Models (femur and tibia components) that are fabricated and provided along with the PSI Guides for use intra-operatively to provide the surgeon with an intra-operative visual reference of the planned location of the PSI Guides in order to help guide their locations on the patient's actual joint, - Zimmer Biomet X-PSI Reusable Surgical Instrumentations are provided both sterile . and non-sterile and are reusable for intra-operative use, which include femoral stylus, and femoral and tibial cut block instrumentations to allow setting the resection level and performing the bone cuts as defined by the PSI Guides. - Fixation Pins are accessories for use of the guides, these accessories are Class I devices under the classification "Orthopedic manual surgical instrument (21 CFR § 888.4540)". The Zimmer Biomet® X-PSI Knee System is compatible for use with the following class II Zimmer Biomet Nexgen, Persona and Vanguard total knee replacement implant systems: - . NexGen®family: NexGen CR, NexGen CR-Flex, NexGen CR-Flex Gender, NexGen LPS, NexGen LPS-Flex, NexGen LPS-Flex Gender - . Persona® family: Persona CR, Persona PS - Vanguard® family: Vanguard CR, Vanguard PS Finally, the following accessories are used for the acquisition of the x-ray images: - X-Ray Marker 3D Zimmer Biomet ® X-PSI (re-usable) ● - X-Ray Calibration Straps, Short and Long (single-use) {5}------------------------------------------------ These accessories are class I device under the classifications "Radiologic quality assurance instrument (21 CFR § 892.1940)" and "Radiographic film marking system (21 CFR § 892.1640)". ### Indications for Use: The Zimmer Biomet® X-PSI Knee System is indicated as an orthopedic instrument system to assist in the positioning of knee replacement components. It involves surgical planning software used pre-operatively to plan the surgical placement of the components on the basis of provided patient radiological images and 3-D reconstructed bones with identifiable placement anatomical landmarks, and surgical instrument components that include patient specific or customized guides fabricated on the basis of the surgical plan to precisely reference the placement of the implant components intra-operatively per the surgical plan. The X-PSI Knee system is indicated for patients without severe bone deformities, such as a HKA greater than 15° or deformities due to prior fracture of the distal femur or proximal tibia. The Zimmer Biomet® X-PSI Knee System is to be used with the following fixed bearing knee replacement systems in accordance with their indications and contraindications: NexGen® CR, NexGen CR-Flex, NexGen CR-Flex Gender, NexGen LPS, NexGen LPS-Flex, NexGen LPSFlex Gender, Persona® CR, Persona PS, Vanguard® CR and Vanguard PS. The patient specific guide components are intended for single-use only. ### Summary of Technological Characteristics: - While the CAS PSI Knee System uses MRI DICOM as an input, the Zimmer Biomet® XPSI ● Knee System utilizes x-ray DICOM as an input as does the Knee EOS System. - . Both the CAS PSI Knee System and the Zimmer Biomet • X-PSI Knee System utilize the same software suites for triage, reconstruction, landmark acquisition, planning, guide creation, and bone model creation. - . The Guides are substantially equivalent, however the Zimmer Biomet®X-PSI Knee System guides utilize a different method for attachment to the knee joint than that used in the CAS PSI Knee System. - The guides and bone models are manufactured using a 3-D printing method, Selective Laser Sintering (SLS). # Summary of Performance Data: # (Nonclinical and/or Clinical) Four different types of non-clinical and/or clinical tests were conducted to verify and validate the performance of the system and assess that no new safety and efficacy issues were raised in the device. - Software System Tests: They were performed to ensure that no hazardous anomalies were present in the system software components. They consisted of testing software features and functionalities in correspondence to software design requirements. - Guide Mechanical Resistance Tests: Tests were performed to verify the mechanical . {6}------------------------------------------------ performance of the guides including resistance to use or drop breakage, debris generation, aging stability, sterilization, and extreme shipping conditions. - . Full System Validation Tests: Full use simulations tests using cadaver specimens and/or bone models were performed by multiple surgeons in multiple settings to verify and validate the overall system performance in terms of system usage and instrument ergonomics. The results demonstrated that the X-PSI Knee System is substantially equivalent in cut accuracy with regards to femoral and tibial frontal and sagittal angles, as well as femoral rotation and resection depth. - Non-Significant Risk Clinical Study: A clinical study was performed to evaluate ● proper positioning of the Zimmer Biomet® X-PSI Knee guides. This study measured the resulting frontal Hip-Knee-Ankle (HKA) alignment angle versus the surgically planned HKA alignment angle. The results demonstrated satisfactory performance per the intended use. # Substantial Equivalence Conclusion: In summary, the proposed Zimmer Biomet® X-PSI Knee System is substantially equivalent to the CAS PSI Knee System predicate, as well as the kneeEOS reference in regards to X-Ray imaging and reconstruction with the same intended use and similar technology. Design verification and validation tests, including performance and clinical tests, demonstrated the substantial equivalence of the system.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...