K171186 · Cerebrotech Medical Systems · QAF · Dec 8, 2017 · Cardiovascular
Device Facts
Record ID
K171186
Device Name
Cerebrotech CMS-5000
Applicant
Cerebrotech Medical Systems
Product Code
QAF · Cardiovascular
Decision Date
Dec 8, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Indications for Use
The CMS-5000 is a bioimpedance spectroscopy device for use on adult human patients utilizing impedance ratios that are displayed as a B-DEX ratio as an aid in the assessment of fluid volume differences between the cerebral hemispheres in patients undergoing neurologic assessment. The CMS-5000 is an adjunct to standard methodologies routinely used in the clinical evaluation of intracranial fluid distribution and is not intended to be a sole determinant.
Device Story
CMS-5000 is a bioimpedance spectroscopy device using Volumetric Integral Phase-Shift Spectroscopy (VIPS) to monitor intracranial fluid distribution. It utilizes low-power radio frequency (RF) signals transmitted between an emitter antenna and a detector antenna placed on opposite sides of the head. The device measures the frequency response of the phase angle shift to calculate a B-DEX ratio, representing the asymmetry in fluid volume between cerebral hemispheres. Used as an adjunct to standard clinical neurologic evaluation, the device provides data to clinicians to assist in assessment; it is not a sole diagnostic determinant. The system includes a console and antenna modules. It is intended for use in clinical settings on adult patients.
Clinical Evidence
No clinical data. Performance validated via bench testing, including human head phantoms, IEC 60601 testing, and studies on 79 healthy adult volunteers to establish normative bioimpedance and person-to-person variability. Additional bench studies demonstrated measurement stability and the ability to detect induced bioimpedance asymmetry.
Indicated for adult human patients undergoing neurologic assessment to aid in the assessment of fluid volume differences between cerebral hemispheres.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
Cerebrotech Medical Systems % Pamela Buckman Consultant Pamela M. Buckman, Msn 2800 Pleasant Hill Rd. #175 Pleasant Hill, California 94523 December 8, 2017
Re: K171186
Trade/Device Name: Cerebrotech CMS-5000 Regulation Number: 21 CFR 870.2770 Regulation Name: Impedance Plethysmograph Regulatory Class: Class II Product Code: OAF Dated: November 7, 2017 Received: November 8, 2017
Dear Ms. Buckman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{2}------------------------------------------------
# Page 2 - Pamela Buckman
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael J. Hoffmann -S
Carlos L. Peña. PhD. MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171186
Device Name Cerebrotech CMS-5000
#### Indications for Use (Describe)
The CMS-5000 is a bioimpedance spectroscopy device for use on adult human patients utilizing impedance ratios that are displayed as a B-DEX ratio as an aid in the assessment of fluid volume differences between the cerebral hemispheres in patients undergoing neurologic assessment.
The CMS-5000 is an adjunct to standard methodologies routinely used in the clinical evaluation of intracranial fluid distribution and is not intended to be a sole determinant.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY
# K171186
#### General Information 1.
| Submission | |
|------------|-----------------------------------|
| Sponsor: | Cerebrotech Medical Systems, Inc. |
| | 1048 Serpentine Lane, Suite 301 |
| | Pleasanton, CA 94566 |
| | T 510-501-6571 |
| | F 925 399 5827 |
| | Contact Person: Dawnel Scott |
# Submission Contact:
Pamela M. Buckman, MSN 2800 Pleasant Hill Rd., Suite 175 Pleasant Hill, CA 94523 T 925 980 7007 F 925 705 7381
Date Prepared: December 8, 2017
#### 2. Device Identification
| Device Name: | Cerebrotech™ CMS-5000™ |
|-------------------------------|--------------------------|
| Classification Name: | Impedance Plethysmograph |
| Classification<br>Regulation: | 21 CFR Part 870.2770 |
| Classification Panel: | Neurology |
| Product Code: | QAF |
| Device Class: | Class II |
{5}------------------------------------------------
## 3. Predicate Device
K130338, Impedimed L-Dex U400, Impedance Plethysmograph
- 4. Device Description
The CMS-5000 is a bioimpedance spectroscopy device used to monitor patients who have conditions that predispose them to fluid volume changes. The device uses Volumetric Integral Phase-Shift Spectroscopy (VIPS) technology. Changes in the volume of fluids in the human skull can be detected by monitoring changes in electrical properties. VIPS technology measures the frequency response of the phase angle shift between an emitter antenna and a detector antenna, when placed on opposite sides of the head. The device is intended to be used as an aid to clinical assessment in conjunction with current standards of care.
- 5. Indications for Use
The CMS-5000 is a bioimpedance spectroscopy device for use on adult human patients utilizing impedance ratios that are displayed as a B-DEX ratio as an aid in the assessment of fluid volume differences between the cerebral hemispheres in patients undergoing neurologic assessment. The CMS-5000 is an adjunct to standard methodologies routinely used in the clinical evaluation of intracranial fluid distribution and is not intended to be a sole determinant.
- 6. Comparison of Technological Characteristics
The technological characteristics of the CMS-5000 are substantially equivalent to those of the predicate device.
{6}------------------------------------------------
| Technological Characteristics Comparison Table | | |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|
| Characteristic | Impedimed L-Dex U400 | Cerebrotech Fluids Monitor CMS-5000 |
| Monitoring Mode | Periodic. Baseline, then repeated as required | Periodic. Baseline, then repeated as required |
| Operating Principle | Multi-frequency Bioimpedance Spectroscopy | Multi-frequency Bioimpedance Spectroscopy |
| Embedded Software | Yes | Yes |
| Type of Energy | Electric Current | Low-Power Radio Frequency |
| Frequency Range | 30-310 MHz | 4-1000 KHz |
| Displayed parameter | Bioimpedance ratio | Bioimpedance ratio |
| Device Configuration | Non-invasive, antenna modules and system controller | Non-invasive; adhesive contact electrode modules, system controller |
| Power Source | Rechargeable Battery | Rechargeable Battery |
| Performance | Device detects and monitors bioimpedance ratios | Device detects and monitors bioimpedance ratios |
| Voltage | 100-240 | 100-240 |
| Frequency | 50-60 Hz | 50-60 Hz |
| Current | 1A Apple MagSafe 2 Power Adapter (for recharging Console)<br>0.2A Cell-Con plug (Scanner Battery Recharger) | Not available |
# 7. Substantial Equivalence
The compared systems utilize the same multi-frequency bioimpedance spectroscopy technology to measure tissue fluid differences between contralateral anatomies. Bioimpedance has been used for decades to measure tissue fluids in many anatomical regions, including limbs, lungs, brain, and large portions of the body for body composition analysis. While the type of energy used for monitoring differs between the CMS-5000 (Low-Power RF) and L-Dex U400 (Current), the principles of comparison are similar. The differences between the CMS-5000 and the L-Dex U400 identified above introduce no additional risks.
{7}------------------------------------------------
- 8. Safety and Effectiveness information
The CMS-5000 and the L-Dex U400 use the same technical principles to achieve the same goal: to provide information about tissue fluid differences derived from bioimpedance ratios. Both are intended to supply adjunct information and do not replace standard diagnostic techniques, which may include physical examination, medical imaging, and professional judgment. The systems are therefore substantially equivalent.
- 9. Non-Clinical Performance
The performance of the CMS-5000 was validated in a series of bench tests, including the use of standardized, validated bioimpedance human head phantoms. In addition, all required standardized testing was completed, including IEC 60601 tests. The results of all performance, verification and validation testing showed no new issues of safety or efficacy.
#### Summary of Performance Data
The bioimpedance index, or B-DEX, is a proprietary index derived from the electrical signals received by the CMS-5000 across the range of device frequencies. B-DEX asymmetry is the percent difference in B-DEX between the two hemispheres (right minus left, divided by the average). A series of studies were conducted to establish device performance.
| Study | Description |
|--------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Normative human<br>bioimpedance measurements | 79 healthy adult volunteers established normative<br>human bioimpedance and person-to-person<br>variability. |
| Detection of bioimpedance<br>asymmetry | Human head phantom with a septum dividing the<br>right and left halves was used to measure<br>bioimpedance asymmetry between the two<br>hemispheres. |
| Mathematical model | A mathematical model validated the accuracy of<br>the bioimpedance asymmetry readings. |
| Measurement of<br>bioimpedance asymmetry<br>during head tilt | Demonstrated the ability to detect a bioimpedance<br>asymmetry induced by gravity. |
| Measurement Stability<br>(human) | Demonstrated day-to-day measurement stability in<br>human subjects. |
| Measurement Stability<br>(human head phantom) | Demonstrated day-to-day measurement stability of<br>the device, independent of human day-to-day<br>variability. |
{8}------------------------------------------------
### 10. Clinical Performance
There was no clinical testing required to support this medical device as the indications for use and technology are equivalent to those of the predicate devices.
# 11.Statement of Substantial Equivalence
A device is substantially equivalent when the subject device has the same intended use and the same technological characteristics as the previously cleared predicate device(s). The CMS-5000 component change from patient contacting electrodes in the predicate device to a non-contact antenna technology does not alter either of these characteristics and does not raise any new questions of safety or efficacy as compared to the predicate.
## 12. Conclusion
The Cerebrotech CMS-5000 was found to be substantially equivalent to the predicate device. It shares the same indications for use, similar design features as well as functional features and thus is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.