The CONTACT LENS CASE ROUND GLASSES is intended for the storage of soft (hydrophilic), rigid gas permeable, and/or hard contact lenses. Used for storage during chemical disinfection only. Not for use during heat disinfection.
Device Story
Contact Lens Case Round Glasses is a storage container for contact lenses; features two adjoining wells with snap lids and silicone seals; constructed from PC, TPE, and ABS plastics with stainless steel hinges. Used by patients for chemical disinfection storage; lenses are immersed in solution within compartments marked L (left) and R (right). Leak-proof design prevents solution loss. Device provides a safe, biocompatible environment for lens storage; does not perform disinfection itself but supports chemical disinfection protocols.
Clinical Evidence
No clinical data. Bench testing only: leakage testing (48 units, 100% pass) and biocompatibility testing (cytotoxicity, eye irritation, systemic toxicity) per ISO 10993 standards.
Technological Characteristics
Materials: PC (Covestro 2805), TPE (Zhongsuwang), ABS (Chi Mei 757), stainless steel hinges. Dimensions: 76mm x 29mm x 17mm. Capacity: 4.4ml per compartment. Design: Two-well container with snap lids and silicone seals. Biocompatibility: ISO 10993-5, 10993-10, 10993-11 compliant.
Indications for Use
Indicated for storage of soft (hydrophilic), rigid gas permeable, and hard contact lenses during chemical disinfection. Not for use with heat disinfection.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, resembling a bird-like shape.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 27, 2017
FiftyTwoWavs Ltd Jos Reinders Official Correspondent Unit 4, 7th Floor, Flat B, Grand Ind. Bldg, 159-165 Wo Yi Hop Rd, Kwai Chung, Hong Kong S.A.R
Re: K171163 Trade/Device Name: Contact Lens Case Round Glasses Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (Hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LRX Dated: April 4, 2017 Received: May 2, 2017
Dear Jos Reinders:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Denise L. Hampton -S
for Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K171163
Device Name
Contact Lens Case Round Glasses
#### Indications for Use (Describe)
The CONTACT LENS CASE ROUND GLASSES is intended for the storage of soft (hydrophilic), rigid gas permeable, and/or hard contact lenses. Used for storage during chemical disinfection only. Not for use during heat disinfection.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | |
| | X Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510 (k) summary
This summary of 510k safety and effectiveness information is being submitted in accordance with the requirements of 21CFR 807.92
| Submitter & Foreign<br>Manufacture<br>ldentification | FiftyTwoWays Ltd<br>Unit 4, 7th Floor, Flat B, Grand Industrial Building<br>159-165 Wo Yi Hop Road, Kwai Chung<br>HONG KONG<br>tel: +852 27352527<br>E-mail: jos@fiftytwoways.com<br>Contact: Jos Jacob Reinders | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Date of Initial Summary: | 04.04.2017 | |
| Device Name:<br>Proprietary Name:<br>Trade Name:<br>Common Name:<br>Classification Name:<br>Device Classification:<br>Regulation Number:<br>Panel:<br>Product Code: | Contact Lens Case Round Glasses<br>Contact Lens Case Round Glasses<br>Contact Lens Case Round Glasses<br>Contact Lens Case<br>Soft (hydrophilic) contact lens care products<br>Class II<br>21 CFR 886.5928<br>Ophthalmic<br>LRX | |
| Predicate Device<br>Information: | K130930<br>OWL Contact Lens Case, manufactured by Ningbo Lishunda<br>Electronics Co Ltd in Ningbo, CHINA | |
| Device Description: | The Contact Lens Case Round Glasses is leak-proof and made from durable<br>plastic with secure, snap lids. Note: can be used with hard and soft contact<br>lenses; for chemical disinfection only; rinse case thoroughly before initial<br>use; replace contact case every 2 months.<br>The secure snap lids are connected to the body by stainless steel hinges and<br>securely close by use of the silicon seal within the plastic<br>Each compartment has a capacity of 4,4ml and diameter of 26,5mm and an<br>overall dimension of 76mm x 29mm x 17mm<br>The Case body is in White Color and the Lids in Black Color with transparent<br>windows.<br>Contact lenses can be fully immersed in each compartment, and each<br>compartment fit all standard lenses currently being sold in the market. The<br>side of each compartment is marked with L (left) or R (right) | |
| | Materials Used:<br>Hinge (Part 1):<br>Glasses Frame (Part 2):<br>Glasses (Part 3):<br>Anti Water Layer (Part 4):<br>Cups (Part 5):<br>Holder/Backside (Part 6):<br>Colorants: | Stainless Steel<br>PC (Covestro 2805)<br>PC (Covestro 2805)<br>TPE (Zhongsuwang)<br>ABS (Chi Mei 757)<br>PC (Covestro 2805)<br>Critsal Tiona |
| Indications for Use: | The CONTACT LENS CASE ROUND GLASSES is intended for the storage<br>of soft (hydrophilic), rigid gas permeable, and/or hard contact lenses.<br>Used for storage during chemical disinfection only.<br>Not for use during heat disinfection. | |
| Clinical Test | No Clinical Test Were Performed | |
| Anti Leakage: | Units of the Contact Lens Case Round Glasses have undergone<br>leakage testing. 48 products were filled to 2/3 with liquid. Each set<br>was turned upside down for 15 minutes and the test repeated for 4<br>times. None of the tested lens cases showed any leakage and all of<br>the Leakage Test Passed Successfully | |
| Comparison to Predicate Devices: | The Contact Lens Case Round Glasses is compared with the following predicate device in terms of intended use, design, material, specifications and performace<br>* K130930 (Owl Contact Lens Case) | |
{4}------------------------------------------------
{5}------------------------------------------------
The following table shows similarities and differences of use, design, and materials between our device and the predicate device
| Table 1: Comparison of<br>IFU and technological<br>characteristics | | | |
|--------------------------------------------------------------------|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Indication for<br>Use | Our Device | Predicate Device<br>(K130930) |
| | | The CONTACT LENS CASE<br>ROUND GLASSES is<br>intended for the storage<br>of soft (hydrophilic), rigid<br>gas permeable, and/or<br>hard contact lenses. Used<br>for storage during<br>chemical disinfection<br>only. Not for use during<br>heat disinfection. | The Owl Contact Lens Case<br>is intended for the storage<br>of soft (hydrophilic), rigid<br>gas permeable, and/or<br>hard contact lens cases.<br>Used for storage during<br>chemical disinfection only.<br>Not for use during heat<br>disinfection |
| | Materials: | PC (Covestro 2805), TPE<br>(Zhongsuwang), ABS (Chi<br>Mei 757). Critsal Tiona | Acrylonitrile-Butadiene<br>Styrene Copolymer |
| | Basic Design | Two adjoining wells with<br>snap lids | Two adjoining wells with<br>screw down lids |
| | Size | 76x29x17mm | 711x38,5x17mm |
| | Volume | 4.4ml | 5.5ml |
| | Colors | White. Black | Black. White. Blue |
Our design of Contact Lens Case Round Glasses are the same with the predicate device in indications for use and adding a fun design to the item. Although there are differences in the outlook they serve the same indication for use. Our device and the predicate device have some minor difference in material and color which are described in above table 1 and in the text below.
# Table 2: Comparison of Biocompatibility
| | Our Device | Predicate Device<br>(K130930) |
|--------------------------------------|----------------------|-------------------------------|
| Cytotoxicity<br>(ISO-10993-5) | No cytotoxicity | No cytotoxicity |
| Eye Irritation<br>(ISO 10993-10) | No Eye Irritation | No Eye Irritation |
| Systemic Toxicity (ISO-<br>10993-11) | No systemic Toxicity | No systemic Toxicity |
{6}------------------------------------------------
## Our device and the predicate device differ in the following areas:
Our device is made from PC Bayer, ABS and TPE plastics and has two colors on the outside (black and white) and has snap lids, while the predicate device is made from Acrylonitrile-Butadiene-Styrene Copolymer and has three colors and screw lids.
The above mentioned differences do not affect the effectiveness and safety of our device. Our device has been tested for cytotoxicity, eye irritation and systematic cytotoxicity. All results show that the material is not cytotoxic, causes no eye irritation and has no systematic cytotoxic.
Above mentioned test results were submitted in the application to demonstrate that the device is safe and effective and meets all requirements
### Substantial Equivalence Conclusions:
Based on above comparisons our device (Contact Lens Case Round Glasses) is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.