← Product Code [KMF](/productcode/KMF) · K171148

# Aleo BME Liquid Bandage (K171148)

_Aleo Bme, Inc. · KMF · Jan 12, 2018 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K171148

## Device Facts

- **Applicant:** Aleo Bme, Inc.
- **Product Code:** [KMF](/productcode/KMF.md)
- **Decision Date:** Jan 12, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5090
- **Device Class:** Class 1
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

Over the counter: Aleo BME Liquid Bandage is intended to cover and protect the skin from outside dirt and microbial penetration in minor cuts, scrapes, burns, skin irritations and abrasions. Prescription Use: Aleo BME Liquid Bandage is intended to cover and protect the skin from outside dirt and microbial penetration for minor cuts, scrapes, burns, irritations and abrasions, as well as closed surgical incisions and excisions.

## Device Story

Aleo BME Liquid Bandage is a topical skin protectant; applied as liquid to form a rapid, waterproof, breathable, transparent polymeric film. Acts as physical barrier against external dirt and microbial penetration. Used in clinical or home settings; applied by clinicians or patients. Protects minor wounds and closed surgical sites; promotes barrier integrity during healing.

## Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993 (cytotoxicity, irritation, sensitization, acute systemic toxicity, subacute/subchronic toxicity, implantation, material-mediated pyrogenicity). Physical, mechanical, and preservative effectiveness testing confirmed design specifications met.

## Technological Characteristics

Polymeric liquid film-forming agent. Non-sterile. Provided in 15 mL glass vial. Biocompatibility per ISO 10993-5, 10993-6, 10993-10, 10993-11.

## Regulatory Identification

A liquid bandage is a sterile device that is a liquid, semiliquid, or powder and liquid combination used to cover an opening in the skin or as a dressing for burns. The device is also used as a topical skin protectant.

## Predicate Devices

- 3M NexcareTM Liquid Bandage ([K053409](/device/K053409.md))
- KeriCure's Natural SealTM Liquid Bandage ([K131384](/device/K131384.md))
- NewSkin® Liquid Bandage (Pre-amendment)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

January 12, 2018

Aleo BME Inc. % LeAnn Latham Consultant M Squared Associates. Inc. 575 Eighth Avenue, Suite 1212 New York, NY 10018

Re: K171148

Trade/Device Name: Aleo BME Liquid Bandage Regulation Number: 21 CFR 880.5090 Regulation Name: Liquid Bandage Regulatory Class: Class I Product Code: KMF Dated: December 13, 2017 Received: December 13, 2017

Dear Ms. Latham:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

1000-1050.

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K171148

Device Name

Aleo BME Liquid Bandage

Indications for Use (Describe)

Over the counter: Aleo BME Liquid Bandage is intended to cover and protect the skin from outside dirt and microbial penetration in minor cuts, scrapes, burns, skin irritations and abrasions.

Prescription Use: Aleo BME Liquid Bandage is intended to cover and protect the skin from outside dirt and microbial penetration for minor cuts, scrapes, burns, irritations and abrasions, as well as closed surgical incisions and excisions.

| Type of Use (Select one or both, as applicable)                                  |                                                                                 |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span>      Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>      Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary

## Aleo BME Liquid Bandage

| Sponsor:                         | Aleo BME, Inc.<br>200 Innovation Blvd<br>Ste. 210A<br>State College, PA 16803<br>Phone: (814) 954 7381<br>Fax: (814) 826 4447                                                                                                                                                                                                                                       |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                  | LeAnn Latham<br>M Squared Associates, Inc.<br>E-mail: leann.latham@hotmail.com<br>Phone: (703) 562 9800                                                                                                                                                                                                                                                             |
| Date Prepared:                   | January 9, 2018                                                                                                                                                                                                                                                                                                                                                     |
| Proprietary Name:                | Aleo BME Liquid Bandage                                                                                                                                                                                                                                                                                                                                             |
| Common Name                      | Bandage, Liquid                                                                                                                                                                                                                                                                                                                                                     |
| Regulatory Class                 | Class I, 510(k)                                                                                                                                                                                                                                                                                                                                                     |
| Regulation Name,<br>Number       | Liquid bandage, 21 CFR 880.5090                                                                                                                                                                                                                                                                                                                                     |
| Device Product Code<br>and Panel | KMF, General & Plastic Surgery                                                                                                                                                                                                                                                                                                                                      |
| Predicate Devices                | 3M NexcareTM Liquid Bandage, K053409<br>KeriCure's Natural SealTM Liquid Bandage, K131384<br>NewSkin® Liquid Bandage, Pre-amendment                                                                                                                                                                                                                                 |
| Device Description               | ALEO liquid bandage is a tough elastic skin protectant for covering minor<br>skin cuts, scrapes, abrasions and cracks. When applied, it rapidly forms<br>water-proof, breathable and transparent protection that forms an<br>effective barrier to prevent microbial penetration from the external<br>environment while the polymeric film remains intact.           |
| Intended Use                     | Over the counter: Aleo BME Liquid Bandage is intended to cover and<br>protect the skin from outside dirt and microbial penetration in minor<br>cuts, scrapes, burns, skin irritations and abrasions.                                                                                                                                                                |
|                                  | Prescription Use: Aleo BME Liquid Bandage is intended to cover and<br>protect the skin from outside dirt and microbial penetration for minor<br>cuts, scrapes, burns, irritations and abrasions, as well as closed surgical<br>incisions and excisions.                                                                                                             |
| Biocompatibility<br>Testing      | Aleo BME Liquid Bandage passed biocompatibility testing requirements<br>according to ISO 10993 for Cytotoxicity (ISO 10993-5), Irritation (ISO<br>10993-10), sensitization (ISO 10993-10), acute systemic toxicity (ISO<br>10993-11), subacute/subchronic toxicity (ISO 10993-11), implantation<br>(ISO 10993-6) and material mediated pyrogenicity (ISO 10993-11). |
| Shelf Life                       | Aleo BME Liquid Bandage is provided non-sterile.  Product formulation<br>inhibits microbial growth for preservation of the liquid bandage while in<br>the 15 mL glass vial and on the shelf.  Shelf life studies have been<br>performed and the results indicate that the Aleo BME Liquid Bandages is<br>safe and effective for use for the labeled shelf life.     |
| Performance Testing              | Physical, mechanical and preservative effectiveness testing results<br>confirm that the Aleo BME Liquid Bandage meets the product design<br>specifications. Where applicable, performance testing of the Aleo BME<br>Liquid Bandage was compared to the predicate devices.                                                                                          |
| Substantial<br>Equivalence       | Characterization and performance tests of Aleo BME Liquid Bandage<br>have confirmed that it is substantially equivalent in design, function and<br>intended use to the predicate devices and confirmed that there are no<br>significant differences between the proposed and predicate devices that<br>raise new questions of safety or efficacy.                   |

{4}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/device/K171148](https://fda.innolitics.com/device/K171148)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
