← Product Code [MAI](/productcode/MAI) · K171114

# HEALIX ADVANCE KNOTLESS Anchor (K171114)

_Medos International SARL · MAI · Nov 27, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K171114

## Device Facts

- **Applicant:** Medos International SARL
- **Product Code:** [MAI](/productcode/MAI.md)
- **Decision Date:** Nov 27, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The HEALIX ADVANCE KNOTLESS Anchors are indicated for use in the following procedures for reattachment of soft tissue to bone: Shoulder: Rotator Cuff, Biceps Tenodesis, Deltoid Repair Knee: Posterior Oblique, Medial Collateral Ligament (MCL), Lateral Ligament (LCL), Iliotibial (IT) Band Tenodesis Foot/Ankle*: Achilles Tendon Repair *HEALIX ADVANCE KNOTLESS PEEK Anchors only

## Device Story

One-piece, cannulated, threaded suture anchor; secures soft tissue to bone. Provided pre-loaded on disposable inserter driver. Available in 4.75mm, 5.5mm, and 6.5mm sizes. Used by surgeons in orthopedic procedures. Implanted into bone to anchor sutures; facilitates tissue reattachment. Benefits include knotless fixation, reducing surgical complexity. Sterile, single-use device.

## Clinical Evidence

Bench testing only. Performance testing included in-vitro anchor pull-out, time-zero anchor pull-out, and torque testing. Safety evaluations addressed biological, sterility, packaging, and shelf-life. Device confirmed non-pyrogenic per ANSI/AAMI ST-72:2011, USP, and EP standards using bacterial endotoxin testing.

## Technological Characteristics

Materials: Biocryl Rapide (BR) and polyetheretherketone (PEEK). Design: One-piece, cannulated, threaded anchor. Sizes: 4.75mm, 5.5mm, 6.5mm. Delivery: Disposable inserter driver. Sterilization: Sterile, single-use.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- HEALIX ADVANCE KNOTLESS BR Anchor ([K130917](/device/K130917.md), [K131974](/device/K131974.md))
- HEALIX ADVANCE KNOTLESS PEEK Anchor ([K130539](/device/K130539.md), [K131683](/device/K131683.md))
- Arthrex PushLock Anchors ([K101679](/device/K101679.md), [K082810](/device/K082810.md), [K071177](/device/K071177.md), [K061863](/device/K061863.md), [K051219](/device/K051219.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

November 27, 2017

Medos International SARL % Ms. Julie Vafides Manager, Regulatory Affairs DePuy Mitek, a Johnson & Johnson Company 325 Paramount Drive Raynham, Massachusetts 02767

Re: K171114

Trade/Device Name: HEALIX ADVANCE KNOTLESS™ Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI, MBI Dated: October 19, 2017 Received: October 23, 2017

Dear Ms. Vafides:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known)

K171114

Device Name

HEALIX ADVANCE™ KNOTLESS Anchor

Indications for Use (Describe)

The HEALIX ADVANCE KNOTLESS Anchors are indicated for use in the following procedures for reattachment of soft tissue to bone:

Shoulder: Rotator Cuff, Biceps Tenodesis, Deltoid Repair

Knee: Posterior Oblique, Medial Collateral Ligament (MCL), Lateral Ligament (LCL), Iliotibial (IT) Band Tenodesis

- Foot/Ankle*: Achilles Tendon Repair
*HEALIX ADVANCE KNOTLESS PEEK Anchors only

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D)    | Over-The-Counter Use (21 CFR 801 Subpart C) |

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

## 510(k) SUMMARY

## HEALIX ADVANCE™ KNOTLESS Anchor

| Submitter's<br>Name and<br>Address                                                | Medos International SARL<br>Chemin-Blanc 38, Case Postale<br>Le Locle<br>CH 2400<br>Switzerland                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                          |                                                         |
|-----------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
|                                                                                   | Date Prepared: April 13, 2017                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                          |                                                         |
| Contact Person                                                                    | Julie Vafides<br>Manager, Regulatory Affairs<br>DePuy Mitek, Inc.<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767, USA                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                          | Telephone: 508-977-6645<br>e-mail: jvafides@its.jnj.com |
| Name of<br>Medical Device                                                         | Proprietary Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | a) HEALIX ADVANCE™ KNOTLESS BR Anchor<br>b) HEALIX ADVANCE™ KNOTLESS PEEK Anchor                                                                                                         |                                                         |
|                                                                                   | Classification Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | a) Single/multiple component metallic bone fixation<br>appliances and accessories<br>b) Smooth or threaded metallic bone fixation fasteners                                              |                                                         |
|                                                                                   | Common Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Suture Anchor                                                                                                                                                                            |                                                         |
| Substantial<br>Equivalence                                                        | The HEALIX ADVANCE™ KNOTLESS Anchor is substantially equivalent to:<br>K130917, K131974 HEALIX ADVANCE™ KNOTLESS BR Anchor K130539, K131683 HEALIX ADVANCE™ KNOTLESS PEEK Anchor K101679, K082810, K071177, K061863, K051219 Arthrex PushLock Anchors                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                          |                                                         |
| Device<br>Classification                                                          | > HEALIX ADVANCE™ KNOTLESS BR Anchor is classified as:<br>Single/multiple component metallic bone fixation appliances and accessories,<br>classified as Class II, product code MAI, regulated under 21 CFR 888.3030.<br>> HEALIX ADVANCE™ KNOTLESS PEEK Anchor is classified as:<br>Smooth or threaded metallic bone fixation fasteners, classified as Class II, product                                                                                                                                                                                     |                                                                                                                                                                                          |                                                         |
| Device<br>Description                                                             | The HEALIX ADVANCE™ KNOTLESS Anchor is a one piece implantable<br>cannulated, threaded anchor designed to secure soft tissue to bone. The anchor is<br>provided loaded on a disposable inserter driver device. The proposed anchor is offered<br>in three sizes, 4.75 mm, 5.5 mm and 6.5mm. The HEALIX ADVANCE™<br>KNOTLESS Anchors are molded from Biocryl Rapide® (BR) and<br>polyetheretherketone (PEEK) materials. The HEALIX ADVANCE™ KNOTLESS<br>Anchor is provided sterile and is for single use only.                                                |                                                                                                                                                                                          |                                                         |
| Technological<br>Characteristics                                                  | The suture anchor design, anchor and suture materials, and finished good assembly<br>design remain the same when compared to the predicate HEALIX ADVANCE™<br>KNOTLESS Anchors (BR: K130917, K131974; PEEK: K130539, K131683). Minor<br>modifications have been made to the distal tip geometry of the BR Anchors.<br>Additional minor changes have been made to the disposable inserter to increase suture<br>management usability.                                                                                                                         |                                                                                                                                                                                          |                                                         |
| Indications for<br>Use                                                            | The HEALIX ADVANCE KNOTLESS Anchors are indicated for use in the following<br>procedures for reattachment of soft tissue to bone:                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                          |                                                         |
|                                                                                   | Shoulder:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Rotator Cuff Repair, Biceps Tenodesis, Deltoid Repair                                                                                                                                    |                                                         |
|                                                                                   | Knee:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Posterior Oblique, Medial Collateral Ligament (MCL), Lateral<br>Collateral Ligament (LCL), Iliotibial (IT) Band Tenodesis                                                                |                                                         |
|                                                                                   | Foot/Ankle*:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Achilles Tendon Repair                                                                                                                                                                   |                                                         |
|                                                                                   | *HEALIX ADVANCE™ KNOTLESS PEEK Anchors only                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                          |                                                         |
| Non-clinical<br>Testing                                                           | Non-clinical testing has been performed on the proposed device and / or its predicates.<br>Performance testing included in-vitro anchor pull out, time zero anchor pull out and<br>torque testing. Safety evaluations were conducted to address biological, sterility,<br>packaging and shelf-life. Note: The proposed device has been determined to be non-<br>pyrogenic per the requirements set forth in ANSI/AAMI ST-72:2011, United States<br>Pharmacopeia (USP), and European Pharmacopeia (EP) using the bacterial endotoxin<br>testing (BET) method. |                                                                                                                                                                                          |                                                         |
| Safety and<br>Performance                                                         | Results of safety and performance testing have demonstrated that the proposed devices<br>are suitable for their intended use.                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                          |                                                         |
|                                                                                   | Based on similarities in the indications for use, technological characteristics, and<br>performance in comparison to the predicate devices, the proposed HEALIX<br>ADVANCE™ KNOTLESS Anchor has shown to be substantially equivalent to the<br>predicate device under the Federal Food, Drug and Cosmetic Act.                                                                                                                                                                                                                                               |                                                                                                                                                                                          |                                                         |
| Proposed Device                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                          |                                                         |
| HEALIX<br>ADVANCE™<br>KNOTLESS Anchor                                             | The HEALIX ADVANCE KNOTLESS Anchors are indicated for use in the following<br>procedures for reattachment of soft tissue to bone:                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                          |                                                         |
|                                                                                   | Shoulder:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Rotator Cuff Repair, Biceps Tenodesis, Deltoid Repair                                                                                                                                    |                                                         |
|                                                                                   | Knee:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Posterior Oblique, Medial Collateral Ligament (MCL), Lateral<br>Collateral Ligament (LCL), Iliotibial (IT) Band Tenodesis                                                                |                                                         |
|                                                                                   | Foot/Ankle*:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Achilles Tendon Repair                                                                                                                                                                   |                                                         |
|                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | * HEALIX ADVANCE™ KNOTLESS PEEK Anchors only                                                                                                                                             |                                                         |
| Predicate Devices                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                          |                                                         |
| HEALIX<br>ADVANCE™<br>KNOTLESS Anchors<br>(K130917, K131974,<br>K130539, K131683) | Shoulder:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Rotator Cuff Repair, Biceps Tenodesis                                                                                                                                                    |                                                         |
| Arthrex PushLock™<br>(K101679, K082810,<br>K071177, K061863,<br>K051219)          | Shoulder:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps<br>Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or<br>Capsulolabral Reconstruction    |                                                         |
|                                                                                   | Foot/Ankle:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair,<br>Metatarsal Ligament Repair/Tendon Repair, Hallux Valgus Reconstruction, Mid-Foot<br>Reconstruction, Bunionectomy |                                                         |
|                                                                                   | Knee:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar<br>Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis                           |                                                         |
|                                                                                   | Hand/Wrist:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral<br>Ligament Reconstruction                                                                                              |                                                         |
|                                                                                   | Elbow:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral<br>Ligament Reconstruction, Lateral Epicondylitis Repair                                                     |                                                         |
|                                                                                   | Hip:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Capsular Repair, Acetabular Labral Repair                                                                                                                                                |                                                         |

{5}------------------------------------------------

{6}------------------------------------------------

# INDICATIONS FOR USE

---

**Source:** [https://fda.innolitics.com/device/K171114](https://fda.innolitics.com/device/K171114)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
